← Product Code [PDJ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/PDJ) · K140654

# Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c Controls (K140654)

_Abbott Laboratories · PDJ · May 22, 2015 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/PDJ/K140654

## Device Facts

- **Applicant:** Abbott Laboratories
- **Product Code:** [PDJ](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/PDJ.md)
- **Decision Date:** May 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1373
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Hemoglobin A1c assay is used in clinical laboratories for the quantitative in vitro measurement of percent hemoglobin A1c (NGSP) or HbA1c fraction mmol/mol (IFCC) in human whole blood and hemolysate on the ARCHITECT c 4000 System. Hemoglobin A1c measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. The Hemoglobin A1c Calibrators are for use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 4000 System. The Hemoglobin A1c Controls are used for the estimation of test precision and the detection of systematic analytical deviations of the Hemoglobin A1c assay on the ARCHITECT c 4000 System.

## Device Story

The Hemoglobin A1c assay is an in vitro diagnostic test for use on the ARCHITECT c 4000 System. It processes human whole blood or hemolysate samples to quantify HbA1c and total hemoglobin (THb). The system automatically lyses whole blood or uses manual hemolysate preparation. The assay uses an enzymatic method: protease cleaves the N-terminal fructosyl dipeptide of the HbA1c beta-chain; fructosyl peptide oxidase reacts with the peptide to produce hydrogen peroxide, which is measured. Total hemoglobin is measured via absorbance after oxidation to methemoglobin azide. The system calculates the %HbA1c or mmol/mol fraction from the HbA1c/THb ratio. Results are displayed to clinicians to aid in diabetes diagnosis and long-term glucose monitoring. The device is intended for professional use in clinical laboratory settings.

## Clinical Evidence

Bench testing only. Studies included 20-day precision (CLSI EP5-A2), LoB/LoD (CLSI EP17-A), interference testing (CLSI EP7-A2), linearity (CLSI EP6-A), and method comparison (CLSI EP9-A2-IR) against an NGSP secondary reference laboratory method (Tosoh HPLC). Method comparison (n=120) showed Deming regression slopes of 1.00-1.01 and r-values of 0.995-0.996. Total error was calculated at 5.0%, 6.5%, 8.0%, and 12.0% HbA1c levels, with results demonstrating acceptable performance.

## Technological Characteristics

Enzymatic assay; reagents include protease (bacterial), peroxidase (horseradish), and fructosyl-peptide-oxidase (E. coli, recombinant). Analyte detection via absorbance measurement. Platform: ARCHITECT c 4000 clinical chemistry analyzer. Standardization: traceable to IFCC reference calibrators; NGSP certified. Calibration/Controls: lyophilized human whole blood matrix. Connectivity: integrated with ARCHITECT c 4000 system.

## Regulatory Identification

A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.

## Special Controls

(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following:
(i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method.
(iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.

## Predicate Devices

- Hemoglobin A1c ([K130255](/device/K130255.md))
- Hemoglobin A1c Calibrators ([K130255](/device/K130255.md))
- Hemoglobin A1c Controls ([K130255](/device/K130255.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k140654

B. Purpose for Submission:
New Device

C. Measurand:
Glycosylated Hemoglobin (HbA1c)

D. Type of Test:
Quantitative, enzymatic

E. Applicant:
Abbott Laboratories

F. Proprietary and Established Names:
Hemoglobin A1c Assay
Hemoglobin A1c Calibrators
Hemoglobin A1c Controls

G. Regulatory Information:

|  Regulatory Description | Classification | Regulation | Product Code | Panel  |
| --- | --- | --- | --- | --- |
|  Hemoglobin A1c Test System | II | 21 CFR 862.1373 | PDJ | Chemistry, 75  |
|  Glycosylated Hemoglobin Assay | II | 21 CFR 864.7470 | LCP | Hematology, 81  |
|  Calibrator | II | 21 CFR 862.1150 | JIT | Chemistry, 75  |
|  Quality Control Material | II | 21 CFR 862.1660 | JJX | Chemistry, 75  |

H. Intended Use:

1. Intended use(s):

{1}

See indications for use below.

2. Indication(s) for use:

The Hemoglobin A1c assay is used in clinical laboratories for the quantitative *in vitro* measurement of percent hemoglobin A1c (NGSP) or HbA1c fraction mmol/mol (IFCC) in human whole blood and hemolysate on the ARCHITECT c 4000 System.

Hemoglobin A1c measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.

The Hemoglobin A1c Calibrators are for use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 4000 System.

The Hemoglobin A1c Controls are used for the estimation of test precision and the detection of systematic analytical deviations of the Hemoglobin A1c assay on the ARCHITECT c 4000 System.

3. Special conditions for use statement(s):

- For prescription use only.
- This device has significant negative interference with fetal hemoglobin (HbF). HbA1c results are invalid for patients with abnormal amounts of HbF including those with known Hereditary Persistence of Fetal Hemoglobin.
- Glycated HbF is not detected by the assay as it does not contain the β-chain that characterizes HbA1c. However, HbF is measured in the total Hb assay and as a consequence, specimens containing high amounts of HbF (&gt;5%) may result in lower than expected mmol/mol HbA1c values (IFCC) and % HbA1c values (DCCT/NGSP).
- The Hemoglobin A1c assay should not be used to diagnose diabetes during pregnancy. Hemoglobin A1c reflects the average blood glucose levels over the preceding 3 months (i.e., the average life span of a red blood cell) and therefore may be falsely low during pregnancy or any other condition associated with recent onset of hyperglycemia and/or decreased red blood cell survival.
- The Hemoglobin A1c assay should not be used to diagnose or monitor diabetes in patients with the following conditions:
- hemoglobinopathies except as demonstrated to produce acceptable performance (e.g., sickle cell trait).
- abnormal red blood cell turnover. (e.g., anemias from hemolysis and iron deficiency).
- malignancies, and severe chronic hepatic and renal disease.

2

{2}

- In cases of rapidly evolving Type 1 diabetes, the increase of HbA1c values might be delayed compared to the acute increase in glucose concentrations. In these conditions, diabetes mellitus must be diagnosed based on plasma glucose concentrations and/or the typical clinical symptoms.
- This test should not replace glucose testing for patients with Type 1 diabetes, pediatric patients or pregnant women.

4. Special instrument requirements:

For use on the ARCHITECT c 4000 System

I. Device Description:

The Hemoglobin A1c Reagent Kit contains two working reagents, Reagent 1 and Reagent 2 and a hemolyzing reagent (Diluent) with the following constituents:

|  Reagent | Reactive Ingredients | Concentration  |
| --- | --- | --- |
|  Reagent 1 | 10-(carboxymethylaminocarbonyl)-3,7-bis(dimethylamino)phenothiazine sodium salt | 0.000817%  |
|   |  Protease (bacterial) | < 1 mU/dL  |
|  Reagent 2 | Peroxidase (horseradish) | 5 to 15 kU/dL  |
|   |  Fructosyl-peptide-oxidase (E. coli, recombinant) | 300 to 900 U/dL  |
|  Diluent | Sodium nitrite | > 0.05 to < 0.3%  |

Inactive Ingredients: Reagent 1 contains sodium azide as a stabilizer and preservative. Reagent 1 and Diluent contain ProClin 300 as a preservative. Reagent 2 contains ofloxacin as a preservative.

The Hemoglobin A1c assay consists of two application types: The Whole Blood application used an automated on-board pretreatment with hemolyzing reagent, (Diluent). The Hemolysate application consists of a manual pretreatment step which is performed using the hemolyzing reagent before the sample is placed on the analyzer.

The Hemoglobin A1c Calibrators

The Hemoglobin A1c Calibrators are for use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 4000 System.

The Hemoglobin A1c Calibrator Kit contains two levels of calibrator material, Calibrator 1 (Cal 1) and Calibrator 2 (Cal 2). A1c Calibrators (lyophilized) contain hemoglobin and glycated hemoglobin from human whole blood. Prior to lyophilization, the calibrator matrix is an MES-buffered solution. Preservative: Ofloxacin.

{3}

The value-assigned A1c Calibrator values are within the following hemoglobin A1c ranges:
Calibrator 1: 4.59% to 6.02% HbA1c
Calibrator 2: 10.52% to 13.37% HbA1c
Actual analyte concentrations for each lot of calibrators are listed in the Hemoglobin A1c Calibrator Value Sheet, packaged with the calibrator.

## The Hemoglobin A1C Controls

The Hemoglobin A1c Control Kit contains: a low control (Control L) and a high control (Control H) A1c Controls (lyophilized) contain hemoglobin and glycated hemoglobin from human whole blood. Prior to lyophilization, the control matrix is an MES-buffered solution. Preservative: Ofloxacin.

The value-assigned A1c Control values are within the following hemoglobin A1c ranges:
Control L: 4.59% to 6.02% HbA1c
Control H: 9.42% to 11.07% HbA1c.

Actual analyte concentrations for each lot of controls are listed in the Hemoglobin A1c Control Value Sheet, packaged with the controls.

All human source materials were tested by FDA approved methods and found to be negative for the presence of HBs Ag and antibody to HIV1/HIV2, and HCV

## J. Substantial Equivalence Information:

Predicate device name(s):

|  Predicate Device Name | Predicate Device 510(k) Number  |
| --- | --- |
|  Hemoglobin A1c | K130255  |
|  Hemoglobin A1c Calibrators | K130255  |
|  Hemoglobin A1c Controls | K130255  |

Comparison with predicate:

|  Reagents
Similarities and Differences  |   |   |
| --- | --- | --- |
|  Characteristics | Candidate Device
Hemoglobin A1c Assay | Predicate Device
Hemoglobin A1c Assay (k130255)  |
|  Intended Use and Indications for Use | Same | For the quantitative in vitro measurement of percent hemoglobin A1c (NGSP) or HbA1c fraction mmol/mol (IFCC) in human whole blood and hemolysate.  |

{4}

|  Reagents Similarities and Differences  |   |   |
| --- | --- | --- |
|  Characteristics | Candidate Device Hemoglobin A1c Assay | Predicate Device Hemoglobin A1c Assay (k130255)  |
|  Platform | ARCHITECT c 4000 Systems (clinical chemistry analyzer) | ARCHITECT c 8000 System (clinical chemistry analyzer)  |
|  Methodology | Same | Enzymatic  |
|  Specimen Type | Same | Whole blood and Hemolysate: Dipotassium EDTA Lithium Heparin Sodium Heparin Sodium Fluoride/Disodium EDTA Tripotassium EDTA  |
|  Measuring Interval | Same | 4.0 to 14.0 %HbA1c (DCCT/NGSP) 20.22 to 129.51 mmol/mol HbA1c (IFCC)  |
|  Calibrators Similarities and Differences  |   |   |
| --- | --- | --- |
|  Characteristics | Candidate Device Hemoglobin A1c Calibrators | Predicate Device Hemoglobin A1c Calibrators (k130255)  |
|  Intended use | For use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 4000 Systems. | For use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 8000 System.  |
|  Platform | ARCHITECT c 4000 Systems (clinical chemistry analyzer) | ARCHITECT c 8000 System (clinical chemistry analyzer)  |
|  Levels | Same | 2 levels (Calibrator 1 and 2) Each lot of calibrators is value-assigned and values are reported in both NGSP and IFCC units. Actual analyte concentrations for each lot of calibrators are listed in the Hemoglobin A1c Calibrator Value Sheet, packaged with the calibrator. The concentration of glycated hemoglobin (HbA1c) and total hemoglobin (THb) is provided for each lot.  |

{5}

|  Calibrators Similarities and Differences  |   |   |
| --- | --- | --- |
|  Characteristics | Candidate Device Hemoglobin A1c Calibrators | Predicate Device Hemoglobin A1c Calibrators (k130255)  |
|  Standardization/ Traceability | Same | Calibrators are prepared gravimetrically, lyophilized, and then value assigned using secondary calibrators that are traceable to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method.  |
|  Controls Similarities and Differences  |   |   |
| --- | --- | --- |
|  Characteristics | Candidate Device Hemoglobin A1c Controls | Predicate Device Hemoglobin A1c Controls (k130255)  |
|  Intended use | For the estimation of test precision and the detection of systematic analytical deviations of the Hemoglobin A1c assay on the ARCHITECT c 4000 Systems. | For the estimation of test precision and the detection of systematic analytical deviations of the Hemoglobin A1c assay on the ARCHITECT c 8000 System.  |
|  Platform | ARCHITECT c 4000 Systems (clinical chemistry analyzer) | ARCHITECT c 8000 System (clinical chemistry analyzer)  |
|  Levels | Same | 2 levels (Low and High Control) Lyophilized Assignment of values is specific for each lot.  |
|  Material | Same | Controls are prepared using hemoglobin and glycated hemoglobin from human whole blood. Prior to lyophilization, the matrix used is MES buffered solution.  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI EP9-A2-IR: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition

CLSI EP6-A: Evaluation of the Linearity of Quantitative Measuring Procedures: A Statistical Approach; Approved Guideline

{6}

CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition

CLSI EP7-A2: Interference Testing in Clinical Chemistry, Approved Guideline – Second Edition

CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantification; Approved Guideline

L. Test Principle:

The Hemoglobin A1c assay consists of two separate concentration measurements: glycated hemoglobin (HbA1c) and total hemoglobin (THb). The two concentrations are used to determine the percent HbA1c (NGSP units) or the hemoglobin fraction in mmol/mol (IFCC units). The individual concentration values of HbA1c and THb generated by the Hemoglobin A1c assay are used only for calculating the percent hemoglobin A1c or HbA1c fraction, and must not be used individually for diagnostic purposes. The anticoagulated whole blood specimen is lysed automatically on the system for the Whole Blood application or may be lysed manually using the Hemoglobin A1c Diluent for the Hemolysate application.

Glycated Hemoglobin (HbA1c)

The Hemoglobin A1c assay utilizes an enzymatic method that specifically measures N-terminal fructosyl dipeptides of the β-chain of HbA1c. In the pretreatment process, the erythrocytes are lysed and the hemoglobin is transformed to methemoglobin by reaction with sodium nitrite. With the addition of Reagent 1 to the sample, the glycosylated N-terminal dipeptide (fructosyl-VH) of the β-chain of hemoglobin is cleaved by the action of protease. The hemoglobin is transformed to stable methemoglobin azide by the action of sodium azide and the concentration of the hemoglobin is determined by measuring absorbance. Addition of Reagent 2 starts a reaction and fructosyl peptideoxidase (FPOX) is allowed to react with fructosyl-VH. The HbA1c concentration is measured by determining the resultant hydrogen peroxide.

Total Hemoglobin (THb)

The hemoglobin is oxidized to stable methemoglobin azide by the action of sodium nitrite and sodium azide and the concentration of the hemoglobin is determined by measuring absorbance (sample + R1).

Hemoglobin A1c Calculations

The final result is expressed as %HbA₁c (NGSP) or mmol/mol HbA₁c (IFCC) and is automatically calculated by the system from the HbA₁c/THb ratio as follows:

mmol/mol HbA1c IFCC:
HbA1c (mmol/mol) = (HbA₁c/THb) × 1000

%HbA1c DCCT/NGSP:
HbA₁c (%) = IFCC × 0.09148 + 2.152

7

{7}

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

# Within-Laboratory Precision (20-Day)

A 20-day precision study was conducted to evaluate the precision performance of the Hemoglobin A1c assay based on guidance from the CLSI document EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods. Testing was performed using 3 lots of Hemoglobin A1c Reagents, 3 lots of Hemoglobin A1c Calibrators, 1 lot of Hemoglobin A1c Controls (Low and High), and 1 lot of commercially available human whole blood controls (Control Levels 1, 2, and 3) on 3 ARCHITECT c 4000 analyzers. Three levels of controls made from reconstituted lyophilized anticoagulated human whole blood and three levels of pooled human anticoagulated venous whole blood (panels) were tested a minimum of 2 replicates, twice per day, for a total of 20 testing days. Total number of measurements per sample was 720 (=3 instruments x 3 lots x 20 days x 2 runs x 2 replicates). Results are shown below in NGSP and IFCC units. The HbA1c sample abbreviations are as follows: LC= Hemoglobin A1c Low Control (target value  $5.2\%$ ), HC= Hemoglobin A1c High Control (target value  $10.0\%$ ), CL3= commercially available control level 3 (target value  $14.6\%$ ); PS1 = human whole blood near the lower limit of the measuring interval (target range  $4.0 - 4.6\%$ ), PS2=human whole blood near the medical decision point (target range  $6.0 - 7.0\%$ ) PS3= human whole blood above the medical decision point (target range  $8.0 - 10.0\%$ ).

# Hemoglobin A1c Reproducibility

# Hemolysate Application NGSP

ARCHITECT c 4000 analyzer # 1

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 5.2% | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.1 | 0.02 | 0.3  |
|  HC 10.0% | 0.01 | 0.1 | 0.01 | 0.1 | 0.02 | 0.2 | 0.03 | 0.3 | 0.04 | 0.4  |
|  CL3 14.5% | 0.04 | 0.2 | 0.02 | 0.2 | 0.04 | 0.3 | 0.13 | 0.9 | 0.14 | 1.0  |
|  PS1 4.5% | 0.01 | 0.2 | 0.01 | 0.3 | 0.01 | 0.2 | 0.02 | 0.5 | 0.3 | 0.7  |
|  PS2 6.5% | 0.01 | 0.2 | 0.02 | 0.2 | 0.02 | 0.3 | 0.04 | 0.6 | 0.05 | 0.7  |
|  PS3 9.0% | 0.02 | 0.2 | 0.01 | 0.1 | 0.02 | 0.2 | 0.06 | 0.7 | 0.07 | 0.8  |

{8}

ARCHITECT c 4000 analyzer #2

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA_{1c} | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 5.3% | 0.01 | 0.2 | 0.01 | 0.3 | 0.01 | 0.2 | 0.02 | 0.4 | 0.03 | 0.6  |
|  HC 10.1% | 0.01 | 0.1 | 0.02 | 0.2 | 0.02 | 0.2 | 0.06 | 0.5 | 0.06 | 0.6  |
|  CL3 14.4% | 0.05 | 0.4 | 0.03 | 0.2 | 0.00 | 0.0 | 0.02 | 0.1 | 0.06 | 0.4  |
|  PS1 4.5% | 0.01 | 0.3 | 0.01 | 0.3 | 0.02 | 0.4 | 0.04 | 0.8 | 0.05 | 1.0  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.2 | 0.02 | 0.4 | 0.05 | 0.8 | 0.06 | 0.9  |
|  PS3 9.0% | 0.02 | 0.2 | 0.02 | 0.2 | 0.02 | 0.2 | 0.07 | 0.7 | 0.07 | 0.8  |

ARCHITECT c 4000 analyzer # 3

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA_{1c} | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 5.3% | 0.01 | 0.2 | 0.01 | 0.2 | 0.00 | 0.0 | 0.03 | 0.7 | 0.04 | 0.7  |
|  HC 10.1% | 0.01 | 0.1 | 0.01 | 0.1 | 0.01 | 0.1 | 0.07 | 0.7 | 0.07 | 0.7  |
|  CL3 14.3% | 0.04 | 0.3 | 0.04 | 0.3 | 0.01 | 0.1 | 0.10 | 0.7 | 0.11 | 0.8  |
|  PS1 4.5% | 0.01 | 0.3 | 0.01 | 0.2 | 0.01 | 0.1 | 0.05 | 1.0 | 0.5 | 1.1  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.1 | 0.02 | 0.2 | 0.05 | 0.8 | 0.06 | 0.9  |
|  PS3 8.9% | 0.02 | 0.2 | 0.01 | 0.1 | 0.02 | 0.2 | 0.06 | 0.6 | 0.07 | 0.7  |

{9}

ARCHITECT c 4000 analyzers (combined)

|  Reproducibility - Hemolysate Application, NGSP  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Between Instrument |   | Total  |   |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 5.3% | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.2 | 0.02 | 0.5 | 0.02 | 0.3 | 0.03 | 0.6  |
|  HC 10.1% | 0.01 | 0.1 | 0.01 | 0.1 | 0.02 | 0.2 | 0.05 | 0.5 | 0.03 | 0.3 | 0.7 | 0.6  |
|  CL3 14.4% | 0.04 | 0.3 | 0.04 | 0.3 | 0.02 | 0.1 | 0.09 | 0.7 | 0.11 | 0.8 | 0.16 | 1.1  |
|  PS1, 4.5% | 0.01 | 0.3 | 0.01 | 0.3 | 0.01 | 0.3 | 0.04 | 0.8 | 0.01 | 0.3 | 0.04 | 1.0  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.2 | 0.02 | 0.3 | 0.05 | 0.7 | 0.01 | 0.2 | 0.06 | 0.9  |
|  PS3 9.0% | 0.02 | 0.2 | 0.01 | 0.2 | 0.02 | 0.2 | 0.06 | 0.7 | 0.02 | 0.2 | 0.07 | 0.8  |

Whole Blood Application NGSP

ARCHITECT c 4000 analyzer #1

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 4.9% | 0.01 | 0.3 | 0.01 | 0.3 | 0.02 | 0.3 | 0.00 | 0.1 | 0.02 | 0.5  |
|  HC 9.6% | 0.02 | 0.2 | 0.02 | 0.2 | 0.01 | 0.1 | 0.01 | 0.1 | 0.03 | 0.3  |
|  CL3 14.6% | 0.03 | 0.2 | 0.05 | 0.4 | 0.03 | 0.2 | 0.03 | 0.2 | 0.07 | 0.5  |
|  PS1 4.4% | 0.01 | 0.2 | 0.01 | 0.3 | 0.01 | 0.2 | 0.03 | 0.7 | 0.04 | 0.9  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.2 | 0.04 | 0.6 | 0.05 | 0.7  |
|  PS3 9.0% | 0.01 | 0.1 | 0.01 | 0.1 | 0.02 | 0.2 | 0.05 | 0.6 | 0.06 | 0.6  |

{10}

ARCHITECT c 4000 analyzer #2

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 4.9% | 0.01 | 0.3 | 0.02 | 0.5 | 0.02 | 0.5 | 0.00 | 0.0 | 0.03 | 0.7  |
|  HC 9.6% | 0.02 | 0.2 | 0.02 | 0.2 | 0.02 | 0.2 | 0.01 | 0.1 | 0.04 | 0.4  |
|  CL3 14.5% | 0.03 | 0.2 | 0.06 | 0.4 | 0.04 | 0.3 | 0.04 | 0.3 | 0.08 | 0.6  |
|  PS1 4.5% | 0.01 | 0.2 | 0.02 | 0.4 | 0.02 | 0.4 | 0.03 | 0.7 | 0.04 | 0.9  |
|  PS2 6.5% | 0.01 | 0.1 | 0.02 | 0.3 | 0.02 | 0.3 | 0.04 | 0.5 | 0.04 | 0.7  |
|  PS3 9.0% | 0.01 | 0.1 | 0.01 | 0.1 | 0.02 | 0.2 | 0.05 | 0.5 | 0.05 | 0.6  |

ARCHITECT c 4000 analyzer #3

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 4.9% | 0.01 | 0.2 | 0.02 | 0.4 | 0.00 | 0.0 | 0.02 | 0.4 | 0.03 | 0.6  |
|  HC 9.6% | 0.02 | 0.2 | 0.02 | 0.2 | 0.01 | 0.1 | 0.03 | 0.3 | 0.04 | 0.4  |
|  CL3 14.5% | 0.02 | 0.2 | 0.06 | 0.4 | 0.03 | 0.2 | 0.05 | 0.4 | 0.09 | 0.6  |
|  PS1 4.4% | 0.01 | 0.2 | 0.01 | 0.3 | 0.01 | 0.2 | 0.04 | 0.9 | 0.04 | 1.0  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.2 | 0.04 | 0.7 | 0.05 | 0.8  |
|  PS3 9.0% | 0.01 | 0.2 | 0.02 | 0.2 | 0.01 | 0.2 | 0.05 | 0.6 | 0.06 | 0.6  |

{11}

ARCHITECT c 4000 analyzers (combined)

|  Reproducibility - Whole Blood Application, NGSP  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Between Instrument |   | Total  |   |
|  % HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 4.9% | 0.01 | 0.3 | 0.02 | 0.4 | 0.02 | 0.3 | 0.01 | 0.2 | 0.01 | 0.3 | 0.03 | 0.7  |
|  HC 9.6% | 0.02 | 0.2 | 0.02 | 0.2 | 0.02 | 0.2 | 0.02 | 0.2 | 0.01 | 0.1 | 0.04 | 0.4  |
|  CL3 14.5% | 0.03 | 0.2 | 0.06 | 0.4 | 0.03 | 0.2 | 0.04 | 0.3 | 0.03 | 0.2 | 0.09 | 0.6  |
|  PS1 4.4% | 0.01 | 0.2 | 0.01 | 0.3 | 0.01 | 0.3 | 0.04 | 0.8 | 0.01 | 0.2 | 0.04 | 1.0  |
|  PS2 6.5% | 0.01 | 0.2 | 0.01 | 0.2 | 0.01 | 0.2 | 0.04 | 0.6 | 0.01 | 0.2 | 0.05 | 0.7  |
|  PS3 9.0% | 0.01 | 0.1 | 0.01 | 0.2 | 0.02 | 0.2 | 0.05 | 0.5 | 0.01 | 0.1 | 0.06 | 0.6  |

Hemolysate Application IFCC

ARCHITECT c 4000 analyzer #1

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 33.86 | 0.083 | 0.2 | 0.113 | 0.3 | 0.095 | 0.3 | 0.058 | 0.2 | 0.179 | 0.5  |
|  HC 86.31 | 0.134 | 0.2 | 0.112 | 0.1 | 0.188 | 0.2 | 0.308 | 0.4 | 0.401 | 0.5  |
|  CL3 135.02 | 0.385 | 0.3 | 0.250 | 0.2 | 0.471 | 0.3 | 1.413 | 1.0 | 1.558 | 1.2  |
|  PS1 25.26 | 0.107 | 0.4 | 0.139 | 0.6 | 0.105 | 0.4 | 0.262 | 1.0 | 0.333 | 1.3  |
|  PS2 47.16 | 0.118 | 0.3 | 0.168 | 0.4 | 0.199 | 0.4 | 0.414 | 0.9 | 0.503 | 1.1  |
|  PS3 74.52 | 0.200 | 0.3 | 0.126 | 0.2 | 0.218 | 0.3 | 0.662 | 0.9 | 0.736 | 1.0  |

{12}

ARCHITECT c 4000 analyzer #2

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 34.03 | 0.106 | 0.3 | 0.164 | 0.5 | 0.133 | 0.4 | 0.250 | 0.7 | 0.344 | 1.0  |
|  HC 86.86 | 0.149 | 0.2 | 0.167 | 0.2 | 0.174 | 0.2 | 0.605 | 0.7 | 0.688 | 0.8  |
|  CL3 133.68 | 0.569 | 0.4 | 0.369 | 0.3 | 0.000 | 0.0 | 0.185 | 0.1 | 0.703 | 0.5  |
|  PS1 25.36 | 0.147 | 0.6 | 0.161 | 0.6 | 0.182 | 0.7 | 0.404 | 1.6 | 0.494 | 1.9  |
|  PS2 47.27 | 0.127 | 0.3 | 0.152 | 0.3 | 0.256 | 0.5 | 0.543 | 1.1 | 0.633 | 1.3  |
|  PS3 74.56 | 0.191 | 0.3 | 0.185 | 0.2 | 0.277 | 0.2 | 0.724 | 1.0 | 0.792 | 1.1  |

ARCHITECT c 4000 analyzer #3

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 34.01 | 0.099 | 0.3 | 0.104 | 0.3 | 0.021 | 0.1 | 0.378 | 1.1 | 0.405 | 1.2  |
|  HC 86.67 | 0.118 | 0.1 | 0.142 | 0.2 | 0.124 | 0.1 | 0.732 | 0.8 | 0.765 | 0.9  |
|  CL3 133.30 | 0.392 | 0.3 | 0.432 | 0.3 | 0.133 | 0.1 | 1.079 | 0.8 | 1.233 | 0.9  |
|  PS1 25.15 | 0.131 | 0.5 | 0.089 | 0.4 | 0.056 | 0.2 | 0.509 | 2.0 | 0.536 | 2.1  |
|  PS2 47.02 | 0.137 | 0.3 | 0.105 | 0.2 | 0.174 | 0.4 | 0.574 | 1.2 | 0.624 | 1.3  |
|  PS3 74.24 | 0.218 | 0.3 | 0.124 | 0.2 | 0.230 | 0.3 | 0.632 | 0.9 | 0.718 | 1.0  |

{13}

ARCHITECT c 4000 analyzers (combined)

|  Reproducibility - Hemolysate Application, IFCC  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Between Instrument |   | Total  |   |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 33.97 | 0.097 | 0.3 | 0.130 | 0.4 | 0.095 | 0.3 | 0.261 | 0.8 | 0.169 | 0.5 | 0.363 | 1.1  |
|  HC 86.61 | 0.135 | 0.2 | 0.142 | 0.2 | 0.164 | 0.2 | 0.577 | 0.7 | 0.333 | 0.4 | 0.713 | 0.8  |
|  CL3 134.00 | 0.457 | 0.3 | 0.405 | 0.3 | 0.207 | 0.2 | 1.032 | 0.8 | 1.236 | 0.9 | 1.735 | 1.3  |
|  PS1, 25.26 | 0.130 | 0.5 | 0.133 | 0.5 | 0.125 | 0.5 | 0.405 | 1.6 | 0.153 | 0.6 | 0.487 | 1.9  |
|  PS2 47.15 | 0.127 | 0.3 | 0.144 | 0.3 | 0.213 | 0.5 | 0.515 | 1.1 | 0.146 | 0.3 | 0.608 | 1.3  |
|  PS3 74.44 | 0.203 | 0.3 | 0.148 | 0.2 | 0.210 | 0.3 | 0.674 | 0.9 | 0.173 | 0.2 | 0.769 | 1.0  |

Whole Blood Application IFCC
ARCHITECT c 4000 analyzer #1

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 30.50 | 0.143 | 0.5 | 0.133 | 0.4 | 0.176 | 0.6 | 0.034 | 0.1 | 0.265 | 0.9  |
|  HC 81.84 | 0.194 | 0.2 | 0.219 | 0.3 | 0.133 | 0.2 | 0.147 | 0.2 | 0.353 | 0.4  |
|  CL3 135.83 | 0.326 | 0.2 | 0.590 | 0.4 | 0.334 | 0.2 | 0.329 | 0.2 | 0.821 | 0.6  |
|  PS1 24.61 | 0.116 | 0.5 | 0.153 | 0.6 | 0.100 | 0.4 | 0.353 | 1.4 | 0.414 | 1.7  |
|  PS2 46.61 | 0.117 | 0.2 | 0.111 | 0.2 | 0.145 | 0.3 | 0.453 | 1.0 | 0.502 | 1.1  |
|  PS3 74.29 | 0.129 | 0.2 | 0.147 | 0.2 | 0.226 | 0.3 | 0.546 | 0.7 | 0.623 | 0.8  |

{14}

ARCHITECT c 4000 analyzer #2

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 30.38 | 0.136 | 0.4 | 0.253 | 0.8 | 0.249 | 0.8 | 0.029 | 0.1 | 0.381 | 1.3  |
|  HC 81.75 | 0.171 | 0.2 | 0.222 | 0.3 | 0.237 | 0.3 | 0.151 | 0.2 | 0.397 | 0.5  |
|  CL3 135.22 | 0.321 | 0.2 | 0.636 | 0.5 | 0.435 | 0.3 | 0.397 | 0.3 | 0.925 | 0.7  |
|  PS1 25.14 | 0.086 | 0.3 | 0.192 | 0.8 | 0.204 | 0.8 | 0.335 | 1.3 | 0.445 | 1.8  |
|  PS2 47.23 | 0.106 | 0.2 | 0.175 | 0.4 | 0.204 | 0.4 | 0.385 | 0.8 | 0.481 | 1.0  |
|  PS3 74.70 | 0.126 | 0.2 | 0.152 | 0.2 | 0.189 | 0.3 | 0.499 | 0.7 | 0.569 | 0.8  |

ARCHITECT c 4000 analyzer #3

|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 30.32 | 0.125 | 0.4 | 0.199 | 0.7 | 0.000 | 0.0 | 0.220 | 0.7 | 0. | 0.  |
|  HC 81.70 | 0.191 | 0.2 | 0.220 | 0.3 | 0.139 | 0.2 | 0.338 | 0.4 | 0. | 0.  |
|  CL3 135.42 | 0.258 | 0.2 | 0.708 | 0.5 | 0.271 | 0.2 | 0.564 | 0.4 | 0. | 0.  |
|  PS1 24.96 | 0.100 | 0.4 | 0.127 | 0.5 | 0.087 | 0.3 | 0.453 | 1.8 | 0. | 0.  |
|  PS2 47.03 | 0.123 | 0.3 | 0.164 | 0.3 | 0.113 | 0.2 | 0.476 | 1.0 | 0. | 0.  |
|  PS3 74.46 | 0.157 | 0.2 | 0.164 | 0.2 | 0.150 | 0.2 | 0.543 | 0.7 | 0. | 0.  |

{15}

ARCHITECT c 4000 analyzers (combined)

|  Reproducibility - Whole Blood Application, IFCC  |   |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean | Repeatability |   | Between Run |   | Between Day |   | Between Lot |   | Between Instrument |   | Total  |   |
|  mmol/mol HbA1c | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  LC 30.40 | 0.135 | 0.4 | 0.204 | 0.7 | 0.173 | 0.6 | 0.130 | 0.4 | 0.136 | 0.4 | 0.354 | 1.2  |
|  HC 81.76 | 0.186 | 0.2 | 0.221 | 0.3 | 0.176 | 0.2 | 0.230 | 0.3 | 0.096 | 0.1 | 0.420 | 0.5  |
|  CL3 135.49 | 0.303 | 0.2 | 0.646 | 0.5 | 0.353 | 0.3 | 0.442 | 0.3 | 0.299 | 0.2 | 0.959 | 0.7  |
|  PS1, 25.06 | 0.101 | 0.4 | 0.159 | 0.6 | 0.140 | 0.6 | 0.384 | 1.5 | 0.110 | 0.4 | 0.463 | 1.8  |
|  PS2 47.10 | 0.116 | 0.2 | 0.153 | 0.3 | 0.158 | 0.3 | 0.440 | 0.9 | 0.125 | 0.3 | 0.520 | 1.1  |
|  PS3 74.58 | 0.138 | 0.2 | 0.154 | 0.2 | 0.190 | 0.3 | 0.530 | 0.7 | 0.121 | 0.2 | 0.612 | 0.8  |

# b. Linearity/assay reportable range:

A linearity study was performed based on guidance from the CLSI document EP6-A, Evaluation of the Linearity of Quantitative Measuring Procedures: A Statistical Approach.

Commercially available whole-blood HbA1c linearity sets, comprised of Levels 1, 2, 3, and 4, were obtained. Five additional samples were prepared by combining the 4 levels of the commercially available linearity sets in specific ratios. The 9 samples were tested using the Abbott Hemoglobin  $\mathrm{A_{1c}}$  assay on the ARCHITECT c 4000 analyzer. The mean observed  $\% \mathrm{HbA1c}$  value was determined for each intermediate dilution and plotted versus the relative analyte concentration. The linear regression equation is summarized below:

NGSP:

|  Slope | Intercept | R2 | Sample range tested  |
| --- | --- | --- | --- |
|  0.9709 | -0.40 | 0.999 | 3.1-19.0 %HbA1c  |

IFCC:

|  Slope | Intercept | R2 | Sample range tested  |
| --- | --- | --- | --- |
|  0.9831 | -6.17 | 0.999 | 11-184 mmol/mol HbA1c  |

{16}

The linearity study supports the claimed assay measuring range of 4.0 to 14.0% HbA1c (20.22 to 129.51 mmol/mol HbA1c (IFCC)).

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:

The Architect HbA1c assay standardization is traceable to the International Federation of Clinical Chemistry (IFCC) reference calibrators. The Architect HbA1c assay is NGSP certified. The NGSP certification expires in one year. See NGSP website for current certification at http://www.ngsp.org.

The derived result of (%) from the NGSP correlation is calculated from the individual quantitative results for total hemoglobin and Hemoglobin A1c (HbA1c). The IFCC units of mmol/mol are calculated using the Master Equation IFCC=(NGSP-2.15/0.092). Two different units are provided to the users: NGSP equivalent units (%) and IFCC equivalent units (mmol/mol).

Value Assignment:

The ARCHITECT HbA1c calibrators are aligned to IFCC reference calibrators through internal value assignment in which calibrator values must meet the sponsor's pre-determined acceptance criteria within a set specification, determined by the manufacturer.

Each lot of calibrators is value-assigned. The concentration of glycated hemoglobin (HbA1c) and total hemoglobin (THb) is provided for each lot. Calibrators are prepared gravimetrically, lyophilized, and then value assigned using secondary calibrators that are traceable to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method.

The value-assigned A1c Calibrator values are within the following hemoglobin A1c ranges:

Calibrator 1: 4.59% to 6.02% HbA1c

Calibrator 2: 10.52% to 13.37% HbA1c

The Hemoglobin A1C Controls, low and high, are value assigned using the secondary calibrators. The values obtained must meet the sponsor's pre-determined acceptance criteria. The value-assigned A1c Control values are within the following hemoglobin A1c ranges:

Control L: 4.59% to 6.02% HbA1c

Control H: 9.42% to 11.07% HbA1c.

Stability:

Shelf-life claims: Un-opened calibrators and controls can be stored at 2-8°C until the expiration date or for 12 months. The storage and stability study protocols and acceptance criteria provided by the sponsor were reviewed and found acceptable to

17

{17}

support the claimed conditions.

Open-vial claims: The recommended storage condition for in-use calibrators and controls is 2-8 °C and is stable for 10 days post-reconstitution. Labeling indicates not to freeze the control and calibrator materials. The storage and stability study protocols and acceptance criteria provided by the sponsor were reviewed and found acceptable to support the claimed conditions.

On-board stability for the ARCHITECT HbA1c was established by real time studies on the ARCHITECT c4000 analyzer and demonstrated on-board reagent stability of 50 days. The unopened ARCHITECT HbA1c reagent is stable until the expiration date printed on the label when stored at 2 to 8°C. The storage and stability study protocols and acceptance criteria provided by the sponsor were reviewed and found acceptable to support the claimed conditions.

d. Detection limit:

The claimed measuring range of 4.0-14.0% for the ARCHITECT HbA1c assay is based on linearity. See 1b. above.

e. Analytical specificity:

i.) An interference study was performed based on CLSI EP7-A2 guideline to assess common or known substances that could interfere with the HbA1c assay. The potential interferents listed below were spiked into human EDTA whole blood samples with different levels of % HbA1c (approximately 6.5 and &gt; 8.0% HbA1c). Each sample was tested in a minimum of 12 replicates. The % HbA1c values of the spiked samples were compared to the reference samples containing no interferent. Significant interference was defined as greater or equal to ±5% difference from the expected concentration. The interference study results are summarized in the following table:

|  Potential Interferent | Highest Tested Concentration at which no significant interference (≥±5%) was observed  |
| --- | --- |
|  Bilirubin, conjugated | 20.0 mg/dL  |
|  Bilirubin, unconjugated | 15.0 mg/dL  |
|  Total Protein | 22g/dL  |
|  Triglycerides | 3000 mg/dL  |
|  Rheumatoid Factor | 200 IU/mL  |
|  Ascorbic Acid | 3.0 mg/dL  |
|  Urea | 667 mg/dL  |
|  Glucose | 1000 mg/dL  |
|  Vitamin E | 8.6 mg/dL  |

ii.) A hemoglobin variant interference

The hemoglobin variant interference study was performed according to guidance

{18}

from the CLSI EP7-A2. Interference effects were assessed by comparing the Hemoglobin A1c values to a comparative method for samples containing potentially interfering hemoglobin variants. The samples were tested in duplicate using one lot of HbA1c Reagents on one ARCHITECT c 4000 analyzer. The numbers and concentrations hemoglobin variants tested, and the range of  $\%$  HbA1c concentrations in which they were tested are shown below:

|  Hemoglobin Variant | n | Range in % Abnormal Variant | Range in %HbA1c Concentration  |
| --- | --- | --- | --- |
|  HbC | 43 | 27% - 43% | 4.7 – 13.6  |
|  HbD | 40 | 34% - 43% | 4.7 – 8.8  |
|  HbE | 50 | 14% - 32% | 4.7 – 11.8  |
|  HbS | 31 | 18% - 42% | 4.6 – 13.5  |
|  HbA2 | 24 | 4.1% - 6.1% | 4.6 – 11.7  |
|  HbF | 28 | 3% - 33.0% | 4.2 – 7.9  |

The results for the hemoglobin variant study are summarized below:

|  Hemoglobin Variant | Relative % Difference from Reference Concentration at Low and High Concentrations  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  ~ 6.0 %HbA1c (5.5 to 6.5 %HbA1c) |   | ~ 9.0 %HbA1c (7.5 to 10.5 %HbA1c)a  |   |
|   |  Relative % Difference | Rangeb | Relative % Difference | Rangeb  |
|  HbC | -3.1 | -6.9 to 3.3 | -0.5 | -4.2 to 2.7  |
|  HbD | 0.6 | -3.4 to 3.2 | 0.2 | -1.3 to 2.6  |
|  HbE | 1.0 | -3.3 to 7.8 | 2.5 | -2.1 to 6.3  |
|  HbS | -0.8 | -3.6 to 3.3 | -0.5 | -3.8 to 2.2  |
|  HbA2 | 0.7 | 0.0 to 1.7 | 2.9 | 1.4 to 4.5  |
|  HbF | Bias exceeds -5% when the amount of HbF in the sample exceeds 5%c  |   |   |   |

a. The HbA2 results at  $\sim 9.0\%$  HbA1c consisted of samples between 7.2 to  $11.2\%$  HbA1c.
b. The range is defined as the minimum and maximum relative % difference at each concentration level (~6.0 and ~9.0 %HbA1c).
c. A negative % difference with HbF is proportional in magnitude to the % HbF present in the sample. For example, when the amount of HbF in the sample was  $20.4\%$ , the % difference was  $-20.0\%$  on the ARCHITECT c 4000 System.

{19}

The sponsor claims that no significant interference was observed for the HbC, HbD, HbE, HbS, and HbA2 variants at the concentrations tested above.

In addition, the device labeling contains the following prominent boxed warning in the package insert labeling:

WARNING: The Hemoglobin A1c assay has significant interference with the fetal hemoglobin (HbF). Hemoglobin A1c results are invalid for patients with abnormal amounts of HbF, including those with known Hereditary Persistence of Fetal Hemoglobin.

iii) A drug interference study was performed using a low level whole blood sample with a concentration value of between 6.0 and 7.0% HbA1c and a high whole blood sample with a concentration value of ≥ 8.0% HbA1c. Test samples were prepared by spiking each drug at the interferent concentrations into the low and high samples and comparing to control samples containing no drug. The test and control samples were tested in a minimum of 12 replicates using 1 lot of HbA1c reagents and calibrators. The sponsor defined significant interference as greater or equal to ±5% difference from the expected concentration.

The sponsor states that the HbA1c assay is not susceptible to interference effects from the following drugs at the listed concentration levels evaluated.

|  Potential Drug Interferent | Highest Level Tested Showing No Significant Interference.  |
| --- | --- |
|  Acarbose | 50 mg/dL  |
|  Acetaminophen | 20 mg/dL  |
|  Acetylsalicylate | 50.8 mg/dL  |
|  Atorvastatin | 600 μg Eq/L  |
|  Captopril | 0.5 mg/dL  |
|  Chlorpropamide | 74.7 mg/dL  |
|  Cyanate | 50 mg/dL  |
|  Furosemide | 6.0 mg/dL  |
|  Gemfibrozil | 7.5 mg/dL  |
|  Ibuprofen | 50 mg/dL  |
|  Insulin | 450 micro units/mL  |
|  Losartan | 5 mg/dL  |
|  Metformin | 5.1 mg/dL  |
|  Nicotinic Acid | 61 mg/dL  |
|  Propranolol | 0.2 mg/dL  |
|  Repaglinide | 60 ng/mL  |

{20}

iv.) A hemoglobin derivative study was performed using a low level whole blood sample with a concentration between 6.0 and 7.0% HbA1c and a high whole blood sample with a concentration of ≥ 8.0% HbA1c. The potentially interfering hemoglobin derivatives were prepared by adding the following substance at the levels shown below.

- Acetylated Hemoglobin ≥ 50 mg/dL of ASA (aspirin)
- Carbamylated Hemoglobin ≥ 10 mmol/L of Cyanate
- Labile Hemoglobin ≥ 1000 mg/dL of Glucose

The test samples were prepared by spiking the substances at the interferent concentration listed above into the low and high samples. The test and reference samples were tested in a minimum of 12 replicates. Significant interference was defined as ± 5% when comparing test samples containing the potentially interfering hemoglobin derivatives to the corresponding reference samples.

The sponsor states that the HbA1c assay is not susceptible to interference effects from the hemoglobin derivatives at the interference levels evaluated.

f. Assay cut-off:

Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:

A method comparison study using human EDTA whole blood specimens (n=128) was performed based on CLSI EP9-A2, Method Comparison and Bias Estimation Using Patient Samples guidelines. Each sample was analyzed internally in replicates of two, using two lots each of HbA1c reagents and HbA1c calibrators on two ARCHITECT c 4000 analyzers over a minimum of 5 days. The samples were also tested in duplicate with an NGSP secondary reference laboratory using a FDA cleared HPLC-based method, the Tosoh G8. The total number of comparison results obtained was 256. The sample range tested was 4.1 to 13.2% HbA1c. The method comparison was performed using both the hemolysate and whole blood application. The distribution of samples spanned the measuring interval with a concentration of samples around the clinical decision points as follows:

21

{21}

|  %HbA1c Level* | Number of Samples  |
| --- | --- |
|  < 5.0% | 5  |
|  5.0 – 5.9% | 16  |
|  6.0 – 6.4% | 32  |
|  6.5 – 7.0% | 32  |
|  7.1 – 8.0% | 22  |
|  8.1 – 9.0% | 11  |
|  > 9.0% | 10  |
|  Total | 128  |

Deming (weighted) and Passing-Bablok regression analyses were performed for the ARCHITECT HbA1c versus the reference method.

Summary of the first replicate versus first replicate regression analysis results are presented in the tables below for the hemolysate and whole blood applications:

Summary of Method Comparison Results for NGSP Whole blood Application:

|   | y-Intercept | 95% CI | Slope | 95% CI  |
| --- | --- | --- | --- | --- |
|  Deming | -0.2 | -0.4 to 0.0 | 1.01 | 0.99 to 1.04  |
|  Passing-Bablok | -0.1 | -0.3 to -0.1 | 1.00 | 1.00 to 1.03  |

![img-0.jpeg](img-0.jpeg)
Whole Blood Scatter Plot with Deming Regression (NGSP):

{22}

![img-1.jpeg](img-1.jpeg)
Whole Blood Scatter Plot with Passing-Bablok Regression (NGSP):

Bias and % Bias Summary for Whole Blood (NGSP)

|  Regression Method | %HbA1c | Bias | % Bias  |
| --- | --- | --- | --- |
|  Deming | 5.0 | -0.1 | -3.0  |
|   |  6.5 | -0.1 | -2.0  |
|   |  8.0 | -0.1 | -1.4  |
|   |  12.0 | -0.1 | -0.6  |
|  Passing-Bablok | 5.0 | -0.1 | -2.0  |
|   |  6.5 | -0.1 | -1.5  |
|   |  8.0 | -0.1 | -1.3  |
|   |  12.0 | -0.1 | -0.8  |

Summary of Method Comparison Results for NGSP Hemolysate Application:

|   | y-Intercept | 95% CI | Slope | 95% CI  |
| --- | --- | --- | --- | --- |
|  Deming | -0.20 | -0.4 to 0.0 | 1.01 | 0.99 to 1.04  |
|  Passing-Bablok | -0.01 | -0.1 to -0.1 | 1.00 | 1.00 to 1.00  |

{23}

![img-2.jpeg](img-2.jpeg)
Hemolysate Scatter Plot with Deming Regression

![img-3.jpeg](img-3.jpeg)
Hemolysate Scatter Plot with Passing-Bablok Regression (NGSP):

{24}

Hemoglobin A₁c Method Comparison (Correlation) Bias and % Bias Summary – Hemolysate (NGSP)

|  Regression Method | %HbA₁c | Bias | % Bias  |
| --- | --- | --- | --- |
|  Deming | 5.0 | -0.1 | -2.4  |
|   |  6.5 | -0.1 | -1.7  |
|   |  8.0 | -0.1 | -1.3  |
|   |  12.0 | -0.1 | -0.7  |
|  Passing-Bablok | 5.0 | -0.1 | -2.0  |
|   |  6.5 | -0.1 | -1.5  |
|   |  8.0 | -0.1 | -1.2  |
|   |  12.0 | -0.1 | -0.8  |

Hemoglobin A₁c Method Comparison (Correlation) Bias and % Bias Summary - Hemolysate (IFCC)

|  Regression Method | HbA1c mmol/mol | Bias | % Bias  |
| --- | --- | --- | --- |
|  Deming | 31.13 | -0.94 | -3.0  |
|   |  47.53 | -1.14 | -2.4  |
|   |  63.93 | -1.33 | -2.1  |
|   |  107.65 | -1.85 | -1.7  |
|  Passing-Bablok | 31.13 | -1.35 | -4.3  |
|   |  47.53 | -1.18 | -2.5  |
|   |  63.93 | -1.02 | -1.6  |
|   |  107.65 | -0.57 | -0.5  |

{25}

Bias and % Bias Summary - Whole Blood (IFCC)

|  Regression Method | HbA1c mmol/mol | Bias | % Bias  |
| --- | --- | --- | --- |
|  Deming | 31.13 | -1.29 | -4.1  |
|   |  47.53 | -1.34 | -2.8  |
|   |  63.93 | -1.39 | -2.2  |
|   |  107.65 | -15.4 | -1.4  |
|  Passing-Bablok | 31.13 | -1.59 | -5.1  |
|   |  47.53 | -1.27 | -2.7  |
|   |  63.93 | -0.95 | -1.5  |
|   |  107.65 | -0.10 | -0.1  |

Total Error Near the Cutoff

Using the results of bias estimation (%Bias) in the method comparison study and precision estimates in the reproducibility study, the Total Error (TE) at the following concentrations (5%, 6.5%, 8% and 12%) was calculated as follows:  $\% \mathrm{TE} = |\% \mathrm{Bias}| + 1.96 * \% \mathrm{CV} * (1 + \% \mathrm{Bias})$ . The results are presented in the tables below.

Hemoglobin  $\mathbf{A}_{1c}$  Method Comparison (Correlation)  $\%$  Total Error Summary - Hemolysate (NGSP)

|  %HbA1c | Average % Bias | % CV | % TE  |
| --- | --- | --- | --- |
|  5 | -2.2 | 0.6 | 3.4  |
|  6.5 | -1.6 | 0.9 | 3.3  |
|  8 | -1.3 | 0.8 | 2.8  |
|  12.0 | -0.8 | 0.9 | 2.5  |

% Total Error Summary - Whole Blood (NGSP)

|  %HbA1c | Average % Bias | % CV | % TE  |
| --- | --- | --- | --- |
|  5 | -2.5 | 0.6 | 3.6  |
|  6.5 | -1.8 | 0.7 | 3.1  |
|  8 | -1.3 | 0.6 | 2.5  |
|  12.0 | -0.7 | 0.6 | 1.9  |

b. Matrix comparison:

A matrix study was performed to determine the suitability of different anticoagulant collection tube types for use in the ARCHITECT HbA1c assay. Specimens with concentration values spanning the measuring interval of the assay (4.0 to  $14.0\%$  HbA1c)

{26}

were collected from a minimum of 43 different donors in the control tube type (Dipotassium EDTA, plastic) and in the following tube types under evaluation:

- Lithium heparin
- Sodium heparin
- Sodium Fluoride/Disodium EDTA
- Tripotassium EDTA

The blood collection tubes collected from one individual constituted one sample set. Each sample was tested in a minimum of 2 replicates using one lot of HbA1c reagents and HbA1c calibrators on one ARCHITECT c4000 analyzer.

The regression results are as follows:

|  Dipotassium-EDTA Control vs Tube Types  |   |   |   |
| --- | --- | --- | --- |
|  Tube type | Linear fit  |   |   |
|   |  Slope (95% CI) | Intercept (95% CI) | r value  |
|  Sodium Fluoride/Sodium EDTA | 1.00 | 0.0 | 1.00  |
|   |  1.00 to 1.00 | 0.0 to 0.0  |   |
|  Lithium Heparin | 1.02 | 0.0 | 1.00  |
|   |  1.01 to 1.03 | -0.1 to 0.0  |   |
|  Tripotassium EDTA | 1.00 | 0.0 | 1.00  |
|   |  1.00 to 1.00 | 0.0 to 0.0  |   |
|  Sodium Heparin | 1.02 | 0.0 | 1.00  |
|   |  1.01 to 1.03 | -0.1 to 0.0  |   |

The data support the use of the following blood collection tubes with the HbA1c assay:

- Dipotassium EDTA (control tube)
- Lithium heparin
- Sodium heparin
- Sodium Fluoride/Disodium EDTA
- Tripotassium EDTA

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

{27}

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

The American Diabetes Association (ADA) recommendations¹ are summarized in the following table.

|  HbA1c Value | Glycemic Goal  |
| --- | --- |
|  < 8% HbA1c (64 mmol/mol) | Less stringent  |
|  < 7% HbA1c (53 mmol/mol) | General (Non-Pregnant Adults)  |
|  < 6.5% HbA1c (48 mol/mol) | More stringent  |

HbA1c values above 6.5% HbA1c (48mmol/mol) are an indication of hyperglycemia during the preceding 2 to 3 months or longer. According to the recommendations of the ADA, AbA1c values above 6.5% HbA1c (48 mmol/mol) are suitable for the diagnosis of diabetes mellitus. Patients with HbA1c values in the range of 5.7-6.4%HbA1c (39-46 mmol/mol) may be at risk of developing diabetes.²,³

¹American Diabetes Association. Position Statement: Standards of medical care in diabetes - 2012. Diabetes Care 2012;35 (Suppl 1):S11–S63.

²American Diabetes Association Workgroup Report: International Expert Committee report on the role of the A1C assay in the diagnosis of diabetes. Diabetes Care 2009;32(7):1327–1334.

³American Diabetes Association. Position Statement: Diagnosis and classification of diabetes mellitus. Diabetes Care 2010;33 (Suppl 1):S62–S69.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/PDJ/K140654](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/PDJ/K140654)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
