PRESAGE ST2 ASSAY

K111452 · Critical Care Diagnostics, Inc. (Dba Critical Diag · OYG · Dec 9, 2011 · Clinical Chemistry

Device Facts

Record IDK111452
Device NamePRESAGE ST2 ASSAY
ApplicantCritical Care Diagnostics, Inc. (Dba Critical Diag
Product CodeOYG · Clinical Chemistry
Decision DateDec 9, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1117
Device ClassClass 2

Indications for Use

The Critical Diagnostics Presage® ST2 Assay kit is an in vitro diagnostic device that quantitatively measures ST2 in serum or plasma by enzyme-linked immunosorbant assay (ELISA) in a microtiter plate format. The Presage® ST2 Assay is indicated to be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure. The Presage® ST2 Assay Kit Controls, Level 1 and Level 2, are designed to be used for monitoring the performance of test procedures on the Critical Diagnostics Presage® ST2 Assay kit.

Device Story

The Presage® ST2 Assay is an in vitro diagnostic ELISA kit used to quantitatively measure ST2 protein levels in human serum or plasma. The assay utilizes a microtiter plate format to perform the measurement. It is intended for use by laboratory professionals in a clinical setting. The output is a quantitative concentration of ST2, which healthcare providers use alongside standard clinical evaluations to assess the prognosis of patients with chronic heart failure. The device includes Level 1 and Level 2 controls for monitoring test performance.

Clinical Evidence

No clinical data provided in the summary document.

Technological Characteristics

Quantitative sandwich monoclonal ELISA in 96-well microtiter plate format. Components: mouse monoclonal anti-human ST2 capture/tracer antibodies, recombinant human ST2 calibrators, Streptavidin-HRP conjugate, TMB substrate. Detection: spectrophotometry at 450 nm. Measuring range: 3.1-200.0 ng/mL. Requires microtiter plate shaker (750 rpm) and 450 nm reader. Stability: 12 months at 2-8°C. No significant interference from common endogenous substances (bilirubin, hemoglobin, lipids, protein) or 49 therapeutic drugs.

Indications for Use

Indicated for use in patients diagnosed with chronic heart failure as an aid in assessing prognosis. Not for diagnosis of heart failure. Prescription use only.

Regulatory Classification

Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111452 B. Purpose for Submission: New device C. Measurand: Soluble ST2 D. Type of Test: Quantitative sandwich monoclonal ELISA E. Applicant: Critical Diagnostics F. Proprietary and Established Names: 1. Presage ST2 Assay Kit 2. Presage ST2 Assay Kit Controls G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | OYG, ST2 Assay | Class II | 21 CFR 862.1117, Test, Natriuretic Peptide | Clinical Chemistry (75) | | JJX, Quality Control Material | Class I, reserved | 21 CFR 862.1660 Quality Control Material | Clinical Chemistry (75) | H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: {1} The Critical Diagnostics Presage® ST2 Assay kit is an in vitro diagnostic device that quantitatively measures ST2 in serum or plasma by enzyme-linked immunosorbant assay (ELISA) in a microtiter plate format. The Presage® ST2 Assay is indicated to be used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure. The Presage® ST2 Assay Kit Controls, Level 1 and Level 2, are designed to be used for monitoring the performance of test procedures on the Critical Diagnostics Presage® ST2 Assay kit. 3. Special conditions for use statement(s): For prescription use only. The Presage ST2 Assay is intended to be used only in patients with chronic heart failure and should not be used for diagnosis of heart failure. 4. Special instrument requirements: Microtiter plate shaker capable of shaking plate at 750 rpm. 96-well microplate reader capable of reading at 450 nm. I. Device Description: 1. The Presage® ST2 Assay kit is provided in microplate configuration (96 tests). The content of the kit includes: | Item | Qty | Description | | --- | --- | --- | | Plate | 1 plate | Ready-to-use microtiter plate coated with mouse monoclonal anti-human ST2 monoclonal antibodies (96 wells in 12x8 strips). | | Standard | 2 vials | Recombinant human ST2 Standard Calibrator (lyophilized) (8 ng/vial which corresponds to a range of standards of 3.1 to 200.0 ng/ml when used with specimens diluted 1:50) | | Standard Diluent | | ST2 Standard Diluent (fetal bovine serum, tissue culture grade) (13 ml) | | Biotinylated Antibody | 1 bottle | Anti ST2 Biotinylated Antibody Reagent in phosphate buffered saline (13 ml) | | Sample Diluent | 1 bottle | ST2 Sample Diluent (50 ml) (MOPS buffered) | | Tracer Concentrate | 1 bottle | 100X Streptavidin-HRP Conjugate Concentrate (100X) (0.2 ml) | | Tracer Diluent | 1 bottle | Streptavidin-HRP Conjugate Diluent (MOPS buffer) (13 ml) | | Wash Concentrate | 1 bottle | 20X Wash Buffer (potassium phosphate buffer with NaCl and Tween 20) (50 ml) | | TMB Reagent | 1 bottle | Tetramethylbenzidine Reagent (11 ml) | | Stop Solution | 1 bottle | Stop Solution of diluted HCl (11 ml) | {2} 2. Presage ST2 Assay Controls are packaged separately and contain: Two levels of controls are provided in a sealed, lyophilized vial format. The base matrix of the controls is delipidized human serum. The lower control (C1) is formulated with a target ST2 concentration between 18.8 ng/ml and 35.0 ng/ml. The higher control (C2) is formulated with a target ST2 concentration between 65.3 ng/ml and 105.0 ng/ml. The sponsor stated in the labeling that all donor units of serum used in the preparation of Controls were tested by FDA-approved method and found negative for HIV (HIV I/II antibody), HBV (HBsAg) and HCV (antibody). J. Substantial Equivalence Information: 1. Predicate device name(s): BG Medicine Galectin-3 Assay 2. Predicate 510(k) number(s): k093758 3. Comparison with predicate: | Characteristic | New device Presage ST2 Assay | Predicate Device Galectin-3 (k093758) | | --- | --- | --- | | Intended Use | For the in vitro quantitative determination of soluble ST2 protein in human serum and plasma. | Quantitative measurement of galectin-3 in serum or EDTA-plasma. | | Indication for Use | The assay is used in conjunction with clinical evaluation as an aid in assessing the prognosis of patients diagnosed with chronic heart failure. | same | | Assay Type | Sandwich ELISA | same | | Analyte | ST2 | Galectin-3 | | Antibody | Mouse monoclonal antibody pair | Rat and mouse monoclonal antibody pair | | Detection Method | Spectrophotometry Absorbance at 450 nm | same | | Reference standards | Critical Diagnostics Presage ST2 Assay Control materials | BGM Galectin-3 Quality Control materials | | Reportable Range | 3.1-200.0 ng/ml | 1.4 - 94.8 ng/ml | {3} | Characteristic | New device Presage ST2 Assay | Predicate Device Galectin-3 (k093758) | | --- | --- | --- | | Analytical Sensitivity | 1.8 ng/ml (Limit of Detection) 2.4 ng/ml (Limit of Quantitation) | 1.13 ng/ml (Limit of Detection) 1.32 ng/ml (Limit of Quantitation) | | Sample Volume | 0.020 ml | 0.030 ml | | Sample Type | Serum or plasma (EDTA or heparin plasma) | EDTA plasma and serum | | Hook Effect | No high dose effect up to 200.0 ng/ml | No high dose effect up to 500 ng/ml | | Calibration Interval | Calibration performed with each analysis | same | # K. Standard/Guidance Document Referenced (if applicable): - CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline - CLSI EP7-A: Interference Testing in Clinical Chemistry; Approved Guideline - CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline - CLSI EP25-P: Evaluation of Stability of In Vitro Diagnostic Products; Proposed Guideline - CLSI C28-A3: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline-Third Edition # L. Test Principle: ST2 (for growth STimulation expressed gene 2; also known as IL1RL1, or Interleukin-1 Receptor-Like 1) is a member of the interleukin-1 receptor family. The ST2 protein has two isoforms directly implicated in the progression of cardiac disease: soluble ST2 and a cell membrane-bound isoform, ST2L. When soluble ST2 levels are low, ST2's ligand, IL-33, is available to bind to ST2L and has a cardioprotective effect resulting in preserved cardiac function. However, when soluble ST2 levels are high, soluble ST2 competitively binds to IL-33, making it less likely to bind to ST2L and thereby making IL-33 unavailable for cardioprotective signaling. As a result, the heart is subjected to greater stress in the presence of high levels of soluble ST2, leading to cellular death and tissue fibrosis, reduced cardiac function, and increasing the rate of disease progression (Kakkar &amp; Lee 2008; Seki et al. 2009). The Critical Diagnostics Presage ST2 Assay is a quantitative sandwich monoclonal ELISA in {4} a 96-well microtiter plate format for measurement of soluble ST2 in serum or plasma (EDTA or heparin). A mouse monoclonal anti-human ST2 antibody is coated onto the surface of the microtiter plate wells and acts as the capture antibody to bind ST2 molecules in solution. A second mouse monoclonal anti-human ST2 antibody is biotinylated and functions as the tracer antibody for detecting ST2 molecules that have bound to the capture antibody. After binding of Streptavidin-HRP to the tracer antibody, a substrate is added which yields a blue color that absorbs at 450 nm. The absorbance is proportional to the ST2 levels in the specimens. The test results of the specimens are read from the standard curve. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: **Study Protocol:** Assay precision was evaluated according to CLSI EP5-A2 Guidelines at two CLIA-certified clinical laboratories as well as the sponsor's internal lab, for a total of 3 sites. EDTA plasma pools at 3 different ST2 concentrations, one in the normal reference range (low), one near the 35 ng/ml cutpoint (medium) and one at an elevated concentration (high) were prepared from plasma purchased from a commercial vender. A set of 20 aliquots of each pool was prepared and shipped to each testing site so that all 3 sites analyzed the same test materials and all used assay materials from the same lot of kits. All 3 sites followed the same test protocol performing 2 runs per day with 4 replicates of each test sample per run for 20 consecutive days. **Result Summary:** Results of within-run and total precision were summarized in the below tables. **Precision Analysis Results from Site 1** | Test Pool | N | Mean ST2 (ng/ml) | Within Run | | Between Runs | | Between Days | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Low | 152 | 20.2 | 0.95 | 4.7% | 1.40 | 7.0% | 0.98 | 4.9% | 1.96 | 9.7% | | Medium | 160 | 32.1 | 1.73 | 5.4% | 1.94 | 6.0% | 0.95 | 3.0% | 2.76 | 8.6% | | High | 160 | 76.1 | 3.63 | 4.8% | 4.84 | 6.4% | 1.71 | 2.2% | 6.05 | 8.0% | **Precision Analysis Results from Site 2** | Test Pool | N | Mean ST2 (ng/ml) | Within Run | | Between Runs | | Between Days | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Low | 160 | 21.6 | 0.73 | 3.4% | 0.40 | 1.8% | 0.72 | 3.3% | 1.10 | 5.1% | | Medium | 160 | 36.7 | 0.98 | 2.7% | 0.91 | 2.5% | 0.83 | 2.3% | 1.57 | 4.3% | | High | 160 | 87.6 | 1.72 | 2.0% | 0.95 | 1.1% | 3.08 | 3.5% | 3.66 | 4.2% | **Precision Analysis Results from Site 3** {5} 6 | Test Pool | N | Mean ST2 (ng/ml) | Within Run | | Between Runs | | Between Days | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Low | 152 | 20.0 | 1.44 | 7.2% | 1.81 | 9.0% | 0.64 | 3.2% | 2.40 | 12.0% | | Medium | 152 | 30.2 | 2.36 | 7.8% | 1.64 | 5.4% | 1.83 | 6.1% | 3.40 | 11.3% | | High | 144 | 79.0 | 3.34 | 4.2% | 5.71 | 7.2% | 3.84 | 4.9% | 7.65 | 9.7% | b. Linearity/assay reportable range: Study Protocol: Linearity was evaluated following CLSI guideline EP6-A. A direct dilution series of 11 test samples was prepared by mixing the high patient plasma pool and low patient plasma pool obtained from a commercial source at different ratios. The high patient pool was prepared at a ST2 concentration slightly greater than the 200.0 ng/ml upper limit of the standard curve and the low concentration pool had a mean concentration of 3.2 ng/ml, a value just above the proposed lower range of 3.1 ng/mL. Each test sample was measured in replicates of 4. Result Summary: Data from the linearity analysis demonstrated that the tested replicates were highly consistent with CV values of &lt;5% across the entire concentration range and linear with coefficient of R² = 0.999, liner regression of Y=0.99 X-0.97 Conclusion: The result support the sponsor claimed measuring range of 3.1 to 200.0 ng/ml, which is the full dynamic range of the assay calibration curve. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability and value assignment: Calibrators Calibrators (Standard included in each assay kit, provided lyophilized at 400 ng/vial) are prepared using purified recombinant ST2 protein. Concentration is determined by Bradford protein assay and confirmed using the Presage ST2 Assay. The calibrator is normalized to a "gold standard" stock of the initial preparation of recombinant ST2 protein that was used as the antigen to generate hybridomas of the ST2 antibody. Controls Controls are produced at 2 target concentrations using recombinant ST2 protein in delipidized human serum matrix. The lower control (C1) is formulated with a target ST2 concentration between 18.8 ng/ml and 30.0 ng/ml and corresponds to a normal concentration range. The higher control (C2) is formulated with a target ST2 concentration between 65.0 ng/ml and 85.0 ng/ml and corresponds to the concentration range of patients with severe cardiac disease and worse prognosis. Lot-specific ranges are assigned by testing a minimum of 3 vials per lot, each vial in duplicate, over 3 days. The average of all control values for {6} each level is calculated and reported on the Certificate of Analysis along with the ± 10% performance range. ## Stability: The Presage ST2 Assay Kit and Presage ST2 Controls were evaluated for storage stability at various temperatures and for successively longer periods of time, up to 12 months. Stability study protocol and the acceptance criteria have been reviewed and found to be acceptable. - Presage ST2 Assay Kit Shelf-life Stability: Real-time testing supports a shelf life of 12 months when stored at 2-8°. However, exposure to elevated temperatures, 37°C or above, for even a brief time (1 day) or a prolonged period (≥4 months) of exposure to room temperature results in compromised performance of the product as evaluated by the characteristics of the standard curve, specifically the decrease in standard curve slope over time. Consequently, the sponsor stated in the labeling that care should be taken to ensure that the product is always maintained at refrigerated temperatures until ready to use, including the use of cold packs during shipment. Open-Vial Stability: Not applicable. - Presage ST2 Control Shelf-life Stability: Testing has demonstrated a closed bottle shelf-life of 1 year, stored at 2-8°C. Open-Vial Stability: Testing has demonstrated an open bottle stability of 7 days after reconstitution, capped at 2-8°C - Test Sample Stability of samples stored at different conditions was evaluated using 12 plasma samples collected from the ER department of a hospital. Baseline ST2 plasma concentrations were determined immediately after blood collection, and at the same time the plasma samples were aliquoted into 1.5 ml plastic tubes which were stored at the specified temperatures for various amount of time. % Recovery from the baseline were calculated for each sample and plotted against time of storage. The sponsor's acceptance criterion for analyte stability was that the mean % recovery of the 12 plasma samples is &gt;=90%. Analyte freeze/thaw stability was evaluated at 0, 5, 10 and 15 cycles using 5 7 {7} commercial EDTA plasma samples. Linear regression analysis showed that none of the samples tested resulted in a slope that significantly deviated from the baseline over the course of 15 freeze/thaw cycles and had an average CV of $4\%$ . This analysis confirms that ST2 is not significantly affected by sample freeze/thaw cycles. Based on the testing results, the sponsor made the following recommendation in the labeling: | Storage conditions | Sample stability | | --- | --- | | -20°C and -80°C | 18 months | | 4°C | 7 days | | 20°C | 48 hours | The sponsor also stated that "ST2 in EDTA-plasma is not significantly affected by sample freeze/thaw cycles and has been shown to be stable for 15 freeze-thaw cycles." # d. Detection limit: - Analytical sensitivity was determined per CLSI EP17-A. For LoB determination, calibrator diluent (fetal bovine serum), was measured over 4 days, with 16 replicates per day giving 64 determinations in total. LoB was calculated as $\mathrm{LoB} = \mu_{\mathrm{B}} + 1.645\sigma_{\mathrm{B}}$ where $\mu \mathrm{B}$ and $\sigma_{\mathrm{B}}$ are the mean and standard deviation values respectively. For LoD determination, 60 replicates of 4 different low ST2 concentration human plasma samples ( were measured in sets of 15 replicates each over 4 consecutive days. LoD was calculated as $\mathrm{LoD} = \mathrm{LoB} + 1.645\sigma_{\mathrm{S}}$ The same sample measurements used for the LoD determination were used to estimate bias and imprecision. LoQ was calculated as $\mathrm{LoQ} = \mathrm{bias} + 2\mathrm{xSDs}$ The resulting values from this analysis are summarized in the following table. | Parameter | ST2 Value (ng/ml) | | --- | --- | | Limit of Blank (LoB) | 0.5 | | Limit of Detection (LoD) | 1.8 | | Limit of Quantitation (LoQ) | 2.4 | - The sponsor claimed that the assay has a measuring range from 3.1 to $200\mathrm{ng / mL}$ . # e. Analytical specificity: Interference {8} Interference testing was performed on 5 most common endogenous substances i.e. total protein (BSA); triglycerides; hemoglobin; cholesterol; and bilirubin as well as 49 common therapeutic substances. ## Endogenous Substances Three human EDTA plasma pools with low (~15 ng/ml); medium (~25 ng/ml); and high (~100 ng/ml) concentrations of ST2 were used in this study. Potentially interfering test substances were added at low and high levels to each plasma pool and assayed in sets of 8 replicates. The sponsor defines no significant interference as &lt;10% difference between the spiked and the control (solvent only) samples. No significant interference was observed on any of the substances/concentrations tested as summarized in the below table. | Endogenous Substances | Concentration tested (mg/dL) | | | --- | --- | --- | | Total Protein (BSA) | 1500 | 6000 | | Hemoglobin | 100 | 200 | | Bilirubin | 10 | 30 | | Cholesterol (total) | 150 | 500 | | Triglycerides (total) | 150 | 3000 | Based on these results, the sponsor concluded in the labeling that no significant interference was observed from: Bilirubin &lt; 30 mg/dl, Hemoglobin &lt; 200 mg/dl, Triglycerides &lt; 3000 mg/dl, Cholesterol &lt; 500 mg/dl, Total protein (BSA) &lt; 6000 mg/dl. ## Therapeutic Substances A plasma sample pooled from 5 normal individuals was used in the testing of 49 common therapeutic substances. Each substance was tested at two concentrations per CLSI EP7-A2 recommendations or at 1x and 5x normal therapeutic levels if the substances are not described in CLSI EP7-A2. The mean of 8 replicate measurements was used to calculate the deviation between the spiked and the solvent only plasma samples. The sponsor defines no significant interference as &lt;10% difference between the spiked and the control samples. Among the 49 commonly used drugs tested, 6 substances had nominal &gt;10% deviations from the controls and are tested further using plasma pools with ST concentrations at low (~15 ng/ml), medium (~35 ng/ml) and high (~90 ng/ml). None of the compounds tested showed an effect of &gt;=10% change from the control in any of the plasma pools. Based on these results, the sponsor concluded in the labeling that no significant interference was observed from the commonly used therapeutic substances as listed in the below Table. | Acetaminophen | Cinnarizine | Heparin sodium | Propranolol hydrochloride | | --- | --- | --- | --- | {9} 10 | Acetylsalicyclic acid | Cyclosporin A | Hydralazine hydrochloride | Quinidine | | --- | --- | --- | --- | | Allopurinol | Digitoxin | Hydrochlorothiazide | Simvastatin | | Ambroxol | Digoxin | Indomethacin | Spironolactone | | Amiodarone | Diltiazem | Lisinopril | Sulfamethoxazole | | Amlodipine besylate | Dopamine | Nicotine | Theophylline | | Ampicillin | Dipyridamole | Nifedipine | L-Thyroxine | | Ascorbic Acid | Enalapril Maleate | Nitrofurantoin | Trimethoprim | | Atenolol | Eplerenone | Oxytetracycline | Verapamil hydrochloride | | Bivalirudin | Erythromycin | Phenytoin | Warfarin | | Caffeine | Fluvastatin | Pravastatin Sodium | | | Captopril | Furosemide | Probenecid | | | Chloramphenicol | Glyburide | Procainamide hydrochloride | | - Cross-Reactivity **Study Protocol:** ST2 is recognized to be structurally and genetically related to the interleukin-1 receptor class of molecules but has not been shown to exhibit significant sequence homology (Tominaga et al. 1989). To confirm that the ST2 assay does not cross react with proteins from this class of receptors, a representative set of three (3) molecules was tested, including: 1) human interleukin-1 soluble receptor type I (IL-1 sR-1); 2) recombinant human IL-Iα; 3) recombinant human IL-Iβ". Each molecule was tested at equimolar and 5x molar concentrations of ST2. Evaluations at each test concentration were performed in replicates of 8. **Results and Conclusion:** Based on the results summarized in the below table, the sponsor concluded that none of the three molecules tested, at concentrations 1x and 5x the measured ST2 concentration, exhibited measurable cross reactivity. A summary of the cross reactivity test is presented in the below table. | | substance concentration (pg/ml) | Ave ST2 (ng/ml) | SD | CV | % deviation | | --- | --- | --- | --- | --- | --- | | solvent (PBS) | 0 | 18.9 | 0.9 | 4.6% | | | IL-1 sR-1 | 18,700 | 19.3 | 0.7 | 3.4% | 2.0% | | | 98,500 | 18.5 | 0.6 | 3.0% | 2.1% | | IL-1 alpha | 6,120 | 19.0 | 0.4 | 2.1% | 0.4% | | | 30,600 | 19.1 | 0.7 | 3.6% | 0.6% | | IL-1 beta | 5,780 | 19.7 | 0.5 | 2.7% | 3.7% | | | 28,900 | 19.8 | 0.5 | 2.6% | 4.3% | f. Assay cut-off: {10} See Clinical Cutoff in 3 c below. ## 2. Comparison studies: a. Method comparison with predicate device: See clinical studies in 3 c below b. Matrix comparison: **Study Protocol:** The Presage ST2 assay has been validated for use with plasma or serum samples. Matched sets of native samples from 40 donors were obtained from healthy donors and hospital emergency department patients. The reported ST2 values for each sample and for each matrix were obtained from single measurements. **Results of linear regressions:** | Matrix Y | Matrix X | Slope (95% CI) | Intercept (95% CI) | R² | N | ST2 range (ng/mL) | | --- | --- | --- | --- | --- | --- | --- | | Lithium heparin plasma | Serum | 0.96 (0.94 to 0.98) | 0.32 (-0.58 to 1.2) | 0.9965 | 40 | 3.1-200.0 | | K3-EDTA plasma | Serum | 1.01 (0.99-1.03) | -0.27 (-1.2 to 0.6) | 0.9969 | | | ## 3. Clinical studies: a. Clinical Sensitivity: See 3 (c) below b. Clinical specificity: See 3 (c) below c. Other clinical supportive data (when a. and b. are not applicable): The ST2 analysis cutpoint of 35 ng/mL was selected by choosing a Presage ST2 Assay concentration value above the 90th and below the 95th percentile of the Reference Group (See section M [5] Expected values/Reference range below for further details). The prognostic performance of ST2 at 35 ng/mL cutoff was evaluated in the following study: **Description of study cohort:** {11} The intended use of the Presage® ST2 Assay as an aid in assessing the prognosis of patients diagnosed with chronic heart failure was established in an analysis of the HF-ACTION (Heart Failure: A Controlled Trial Investigating Outcomes of Exercise TraiNing) study. HF-ACTION was a multicenter, randomized, controlled, blinded trial of 2,331 patients with chronic heart failure that compared patients in the exercise training group vs. patients in the usual care group. Patients were randomized from April 2003 through February 2007 within the United States, Canada, and France. A complete description of the design of HF-ACTION has been published previously (Whellan et al. 2007). Plasma samples from a total of 912 patients from this cohort were available for biomarker analysis. Sensitivity analysis was performed comparing the set of 912 HF-ACTION subjects having evaluable ST2 values with all other HF-ACTION participants, and it was found that the clinical validation results based on the evaluable set of subjects were robust and representative of the larger study population. # Results: A Kaplan-Meier analysis using the cutpoint of $35\mathrm{ng / ml}$ in the HF-ACTION Study is presented in Figure 1. Mortality risk is higher in patients with ST2 $&gt;35$ ng/ml (log rank $\mathfrak{p} &lt; 0.001$ ). ST2 Value $&gt;35\mathrm{ng / ml}$ As a Predictor of All Cause Mortality in the HF-ACTION study. ![img-0.jpeg](img-0.jpeg) Figure 1. Kaplan-Meier Analysis {12} &gt; ST2 is also prognostic for clinical outcomes in addition to all-cause mortality. Figures 2 through 4 summarize a quantitative analysis of ST2 at $35\mathrm{ng / ml}$ as a predictor of all-cause mortality, all-cause hospitalization, cardiovascular mortality, and cardiovascular hospitalization, respectively. ![img-1.jpeg](img-1.jpeg) Figure 2: ST2 Relationship to Time of All Cause Mortality Event ![img-2.jpeg](img-2.jpeg) Figure 3: ST2 Relationship to Time to All Cause Hospitalization Event {13} ![img-3.jpeg](img-3.jpeg) Figure 4: ST2 Relationship to Time to CVD Mortality Event ![img-4.jpeg](img-4.jpeg) Figure 5: ST2 Relationship to Time to CVD Hospitalization Event As illustrated in Figures 2-5 absolute event rates in patient with ST2 $&gt;35\mathrm{ng / ml}$ for the endpoints of all-cause mortality, all-cause hospitalization, CVD mortality, and CVD hospitalization are higher than for those patients with ST2 $\leq 35\mathrm{ng / ml}$ . These results are summarized in the below Table. Duration of follow-up in this study is 4 {14} years. | Outcome | ST2 ≤35 ng/ml | ST2 >35 ng/ml | | --- | --- | --- | | All-Cause Mortality within 1 year | 2.7% | 12.8% | | All-Cause Mortality within study | 15.2% | 33.5% | | All-Cause Hospitalization within study | 65.0% | 79.9% | | CVD Mortality within study | 12.2% | 28.5% | | CVD Hospitalization within study | 53.9% | 72.21% | * Study follow up period is 4 years. # Multivariate Covariate Risk-Adjusted Analysis Using Cox proportional hazards models of ST2 as a dichotomous marker at $35\mathrm{ng / ml}$ shows that ST2 is a statistically significant predictor of all clinical endpoints assessed in this study i.e. all-cause mortality within 1 year, all-cause mortality within study, all-cause hospitalization within study, death due to CVD within study, and hospitalization due to CVD within study. Results from Multivariate Covariate Risk-Adjusted Analysis are summarized in the below table. Variables in this adjusted model include age, gender, NYHA class, eGFR, LVEF, diabetes status, hypertension status and smoking status. Importantly the prognostic utility of the Presage ST2 Assay is not adversely influenced by the common confounders; age, gender and renal function (eGFR). Table. Summary of Risk-Adjusted Cox Proportional Hazards Prediction Models for ST2 at ${35}\mathrm{\;{ng}}/\mathrm{{ml}}$ | Outcome | HR (95% CI) | P | | --- | --- | --- | | All-Cause Mortality within 1 year | 3.20 (1.66-6.19) | <0.001 | | All-Cause Mortality within study | 1.76 (1.26-2.46) | <0.001 | | All-Cause Hospitalization within study | 1.35 (1.10-1.65) | <0.043 | | CVD Mortality within study | 1.87 (1.29-2.69) | <0.001 | | CVD Hospitalization within study | 1.33 (1.06-1.66) | 0.0128 | * Study follow up period is 4 years. Variables adjusted include age, gender, NYHA class, eGFR, LVEF, diabetes status, hypertension status and smoking status. # $\succ$ Use of ST2 in Conjunction with Natriuretic Peptides ST2 and natriuretic peptides, such as NT-proBNP, are measures of separate and distinct biological processes. ST2 and natriuretic peptides provide independent and complementary prognostic information. Table below summarizes an analysis of ST2, at $35\mathrm{ng / ml}$ , and NT-proBNP, at the study median value of $852~\mathrm{pg / ml}$ , for assessment of risk of mortality within the study follow-up period. {15} Table. Additive Value of ST2 and NT-proBNP | Category | Death Rate | HR (95% CI) | p | | --- | --- | --- | --- | | ST2 ≤35 ng/ml NT-proBNP ≤Median | 9.0% | 1 | NA | | ST2 ≤35 ng/ml NT-proBNP >Median | 23.3% | 2.87 (1.9 – 4.32) | <0.001 | | ST2 >35 ng/ml NT-proBNP ≤Median | 22.2% | 2.70 (1.25 – 5.84) | 0.0115 | | ST2 >35 ng/ml NT-proBNP >Median | 38.9% | 5.59 (3.61 – 8.66) | <0.001 | # 4. Clinical cut-off: The ST2 analysis cutpoint of $35\mathrm{ng / mL}$ was selected by choosing a Presage ST2 Assay concentration value above the 90th and below the 95th percentile of the Reference Group (See section M [5] Expected values/Reference range below for further details). # 5. Expected values/Reference range: Soluble ST2 concentrations representative of the general population were determined from 490 individuals (245 women and 245 men) in a reference group of donors from two sources selected for this purpose. Donor group one was comprised of 240 individuals from a reference laboratory. Donor group two was comprised of 250 individuals recruited by a commercial vender. All donors were self declared to be healthy with no history of heart disease. No other inclusion or exclusion criteria were applied. Demographic information collected on these donors is summarized in the Tables below. In summary this population included apparently healthy individuals, selected without regard to diabetes, hypertension, pulmonary disease, and renal insufficiency status. Reference Group: Males and Females | Presage® ST2 Assay (ng/ml) | Reference Group - All | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Metric | All | <45 | 45-54 | 55-64 | 65-74 | >=75 | <75 | | Mean | 20.9 | 21.7 | 18.8 | 20.7 | 20.4 | 22.8 | 20.8 | | SD | 9.3 | 9.7 | 8.4 | 8.3 | 9.3 | 10.2 | 9.3 | | Median | 18.8 | 20.0 | 17.4 | 18.8 | 18.9 | 19.3 | 18.8 | | 90th percentile | 34.2 | 35.4 | 30.0 | 32.5 | 33.9 | 37.4 | 34.2 | | % > 35 ng/ml | 9.0 | 10.8 | 6.5 | 6.2 | 6.8 | 12.5 | 8.8 | | N | 490 | 249 | 93 | 65 | 59 | 24 | 466 | {16} Reference Group: Males | Presage® ST2 Assay (ng/ml) | Reference Group - Males Only | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Metric | All | <45 | 45-54 | 55-64 | 65-74 | >=75 | <75 | | Mean | 24.9 | 25.9 | 22.8 | 23.6 | 23.9 | 28.6 | 24.8 | | SD | 10.1 | 10.4 | 9.9 | 9.1 | 10.1 | 9.1 | 10.1 | | Median | 23.6 | 24.5 | 20.7 | 22.1 | 21.9 | 29.5 | 23.4 | | 90th percentile | 37.1 | 37.7 | 35.5 | 34.8 | 35.5 | 39.4 | 36.9 | | % > 35 ng/ml | 16.3 | 19.4 | 14.0 | 9.4 | 13.3 | 18.2 | 16.2 | | N | 245 | 129 | 43 | 32 | 30 | 11 | 234 | Reference Group: Females | Presage® ST2 Assay (ng/ml) | Reference Group - Females Only | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Metric | All | <45 | 45-54 | 55-64 | 65-74 | >=75 | <75 | | Mean | 16.9 | 17.1 | 15.4 | 17.8 | 16.8 | 18.0 | 16.8 | | SD | 6.2 | 6.3 | 4.7 | 6.4 | 6.8 | 8.7 | 6.1 | | Median | 16.2 | 16.0 | 14.9 | 17.4 | 17.5 | 15.6 | 16.4 | | 90th percentile | 23.7 | 23.7 | 21.8 | 24.4 | 22.8 | 20.2 | 23.7 | | % > 35 ng/ml | 1.6 | 1.7 | - | 3.0 | - | 7.7 | 1.3 | | N | 245 | 120 | 50 | 33 | 29 | 13 | 232 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...