COBAS C TINA-QUANT CYSTATIN C GEN.2 ASSAY, C.F.A.S. CYSTATIN C CALIBRATOR, CYSTATIN C GEN.2 CONTROL SET

K141143 · Roche Diagnostics · NDY · Jul 17, 2014 · Clinical Chemistry

Device Facts

Record IDK141143
Device NameCOBAS C TINA-QUANT CYSTATIN C GEN.2 ASSAY, C.F.A.S. CYSTATIN C CALIBRATOR, CYSTATIN C GEN.2 CONTROL SET
ApplicantRoche Diagnostics
Product CodeNDY · Clinical Chemistry
Decision DateJul 17, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Indications for Use

The Tina-quant Cystatin C Gen.2 is an in vitro test for the quantitative determination of cystatin C in human serum and plasma on Roche/Hitachi cobas c systems. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Calibrator: The C.f.a.s. (Calibrator for automated systems) Cystatin C is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. Control: The Cystatin C Control Set Gen.2 is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

Device Story

Tina-quant Cystatin C Gen. 2 is an in vitro diagnostic reagent for quantitative measurement of cystatin C in human serum and plasma. Used on Roche/Hitachi cobas c systems; operated by laboratory professionals. Principle of operation: latex-enhanced immunoturbidimetric assay; human cystatin C agglutinates with latex particles coated with rabbit anti-cystatin C antibodies; aggregate measured turbidimetrically at 546 nm. Input: patient serum or plasma samples. Output: quantitative cystatin C concentration (mg/L). Results assist clinicians in diagnosing and monitoring renal disease. System includes C.f.a.s. Cystatin C calibrator and 3-level Cystatin C Control Set Gen.2 for quality control. Reagents are provided in cassettes with R1 (polymers in MOPS-buffered saline) and R2 (latex particles in glycine buffer).

Clinical Evidence

Bench testing only. Precision/reproducibility (N=80) showed CVs < 2.2%. Linearity established 0.4–6.8 mg/L. LoD/LoQ determined to be 0.4 mg/L. Method comparison (N=103) against predicate showed high correlation (r=0.988). Matrix comparison (serum vs. plasma) showed high agreement. No clinical studies performed.

Technological Characteristics

Latex-enhanced immunoturbidimetric assay. Reagents: polymers in MOPS-buffered saline (R1); latex particles coated with rabbit anti-cystatin C antibodies in glycine buffer (R2). Calibrator/Controls: pooled delipidated human serum enriched with recombinant human cystatin C. Optical measurement at 546 nm. Platform: cobas c 501 analyzer. Standards: CLSI EP5-A2, EP6-A, EP17-A2.

Indications for Use

Indicated for the quantitative determination of cystatin C in human serum and plasma to aid in the diagnosis and treatment of renal diseases. For prescription use only on Roche/Hitachi cobas c systems.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k141143 B. Purpose for Submission: New device C. Measurand: Cystatin C D. Type of Test: Quantitative turbidimetric method E. Applicant: Roche Diagnostics F. Proprietary and Established Names: Tina-quant Cystatin C Gen. 2 Assay C.f.a.s (Calibrator for automated systems) Cystatin C Cystatin C Control Set Gen. 2 G. Regulatory Information: 1. Regulation section: 21 CFR 862.1225, Creatinine Test System 21 CFR 862.1150, Calibrator 21 CFR 862.1660, Quality Control Material 2. Classification: Class II, Class I, reserved 3. Product code: NDY: Test, Cystatin C {1} JIT: Calibrator, Secondary JJX: Single (Specified) Analyte Controls 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See Indications for Use below 2. Indication(s) for use: The Tina-quant Cystatin C Gen.2 is an in vitro test for the quantitative determination of cystatin C in human serum and plasma on Roche/Hitachi cobas c systems. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Calibrator: The C.f.a.s. (Calibrator for automated systems) Cystatin C is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. Control: The Cystatin C Control Set Gen.2 is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: For use on Roche/Hitachi cobas c 501 system I. Device Description: The Tina-quant Cystatin C Gen. 2 reagent provides quantitative measurement of the cystatin C that is present in human serum and plasma. Assay: Reagents are packaged in a cassette with two bottles labeled with their instrument positioning, R1 (Reagent 1) and R2 (Reagent 2). 2 {2} R1 contains a solution of polymers in MOPS-buffered saline; preservative and stabilizers. R2 is latex particles in glycine buffer coated with anti-Cystatin C antibodies (rabbit); preservative and stabilizers. ## Calibrator: C.f.a.s. Cystatin C is a liquid, ready-to-use calibrator based on pooled delipidated human serum enriched with recombinant human Cystatin C produced in E. Coli. Single level calibrators with lot specific values are diluted on board the analyzer to create a 6-point calibration curve. ## Control: Cystatin C Control Set contains 3 controls based on pooled delipidated human serum enriched with human recombinant Cystatin C produced in E. Coli. All human source materials used in the preparation of this device were tested and found to be non-reactive for the presence of HBsAg, anti-HIV and HCV by FDA approved methods. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Diazyme Cystatin C Assay Calibrator f.a.s. Cystatin C and Cystatin C Control Set 2. Predicate 510(k) number(s): k093680 k080811 3. Comparison with predicate: | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate reagents/device | Predicate (k093680) | | Indications for Use | The Tina-Quant Cystatin C Gen.2 is an in vitro test for the quantitative determination of Cystatin C in human serum and plasma. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. | Same. | | Sample Types | Serum and plasma | Same | | Permissible Anticoagulants | Li-heparin plasma K2-EDTA, K3-EDTA plasma | Li-heparin plasma K2-EDTA plasma | | Instrument Platform | cobas c 501 analyzer | Hitachi 917 analyzer | {3} | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate reagents/device | Predicate (k093680) | | Measuring range | 0.40 – 6.80 mg/L | 0.2 – 8.0 mg/L | | Similarities and Differences | | | | Item | Candidate calibrator | Predicate (k080811) | | Analyte | Cystatin C | Same | | Form | Liquid | Same | | Instrument Platform | cobas c 501 analyzer | Hitachi 917 analyzer | | Differences | | | | --- | --- | --- | | Item | Candidate controls | Predicate (k080811) | | Analyte | Cystatin C | Same | | Form | Liquid | Same | | Levels | Three (Low Level: 0.90 - 1.30 mg/L; Mid Level: 1.20 – 2.00 mg/L; and High Level: 4.00 – 5.00 mg/L) | Two (Low Level: 0.90 - 1.30 mg/L and High Level: 4.00 – 5.00 mg/L) | | Instrument Platform | cobas c 501 analyzer | Hitachi 917 analyzer | # K. Standard/Guidance Document Referenced (if applicable): CLSI EP5-A2: Evaluation of Precision of Clinical Chemistry Devices, Approved Guideline-2nd Edition CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. # L. Test Principle: The Tina-quant Cystatin C Gen. 2 Assay is based on a latex enhanced immunoturbidimetric assay. Cystatin C in the sample binds to the specific anti-Cystatin C antibody (rabbit), which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically at $546~\mathrm{nm}$ and is proportional to the amount of Cystatin C in the sample. # M. Performance Characteristics (if/when applicable): Performance testing was performed on the cobas c501 analyzer {4} # 1. Analytical performance: # a. Precision/Reproducibility: The sponsor performed precision studies in accordance with the CLSI EP5-A2 guideline. Three control samples containing $1.0\mathrm{mg / L}$ , $1.8\mathrm{mg / L}$ , and $4.1\mathrm{mg / L}$ and three human serum pools containing $0.56\mathrm{mg / L}$ , $2.80\mathrm{mg / L}$ and $6.39\mathrm{mg / L}$ Cystatin C were analyzed in four replicates, twice a day, for 21 days $(N = 80)$ using one lot of reagents on one, cobas c 501 analyzer. Results are summarized below: | Samples | Mean (mg/L) | Within-run precision | | Between-run precision | | | --- | --- | --- | --- | --- | --- | | | | SD (mg/L) | CV (%) | SD (mg/L) | CV (%) | | Control 1 | 1.00 | 0.02 | 1.7 | 0.02 | 2.2 | | Control 2 | 1.84 | 0.02 | 0.9 | 0.03 | 1.4 | | Control 3 | 4.12 | 0.03 | 0.7 | 0.06 | 1.4 | | Human serum 1 | 0.560 | 0.010 | 1.8 | 0.011 | 2.0 | | Human serum 2 | 2.80 | 0.02 | 0.6 | 0.04 | 1.3 | | Human serum 3 | 6.39 | 0.04 | 0.6 | 0.07 | 1.1 | # b. Linearity/assay reportable range: The sponsor performed linearity studies in accordance with the CLSI EP6-A guideline. 21 levels of serum based (in the range from 0.1 to $7.6\mathrm{mg / L}$ ) and 13 levels of plasma based (in the range from 0.1 to $8.9\mathrm{mg / L}$ ) samples were tested in triplicates on the cobas c501 analyzer. Results are summarized below: Linear regression equation for serum: $y = 1.001x - 0.0057r = 1.000$ Linear regression equation for plasma: $y = 1.006x - 0.0178r^2 = 1.000$ The linearity data and the LoQ data support the sponsor's claim that the measuring range of this device is 0.4 to $6.8\mathrm{mg / L}$ . Auto-rerun validation study: An auto-rerun study was performed to evaluate the accuracy of the analyzer's auto dilution function to obtain results for samples with analyte concentrations above the measuring range. Three samples were used for the study (7.26 mg/L; 8.75 mg/L and 9.87 mg/L). Each sample was manually diluted with 1:1.5 ratio. The results of the auto-rerun study support the claim that samples having higher concentrations can be run via the rerun function which uses a 1:1.5 dilution. {5} High-dose hook effect: The potential for a high-dose hook effect was evaluated for this assay. A serum sample containing 27.7 mg/L of Cystatin C was diluted up to the lower end of the measuring range. The samples were analyzed in triplicate on one cobas c 501 analyzer. Based on the results, the sponsor concluded that Cystatin C levels up to 20 mg/L did not exhibit any high dose effect. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrators: The calibrator is traceable to the European Reference Material ERM-DA471/IFCC. The calibrators were previously cleared in k080811 for use on the Hitachi 917 analyzer. Roche has verified the value assignment on the cobas c 501 analyzer (see Value assignment below). Controls: The typical assigned values for the Cystatin C Control Set Gen.2 are: Low Level: 0.90 - 1.30 mg/L Mid Level: 1.20 - 2.00 mg/L High Level: 4.00 - 5.00 mg/L The Low and High levels were previously cleared in k080811 for use on the Hitachi 917 analyzer. The addition of a Mid-level control and the cobas c 501 analyzer are the only differences between this control set and the one that was cleared in k080811. Value assignment: In order to transfer the values previously obtained for the calibrator and the Low and High level controls to the cobas c501, Roche verified the value assignment obtained in k080811 and for the new Mid level control previously assigned on the Hitachi 917 analyzer by running the calibrator and control material (all three levels) in triplicate, using 2 lots in 3 runs on the cobas c 501 analyzer and checked for the % recovery. The % recovery was within +/-2 % bias for the calibrator and +/-3% bias for the control material. Stability: Calibrators and Controls: Closed vial stability claim of 24 months for the calibrator and 10 months for the controls at 2-8°C was supported by a real-time stability study performed on the cobas c 501 analyzer. The open vial stability claim of 30 days at 2-8°C was established in k080811. Open vial stability of 30 days at 2-8°C for the Mid Level Control was established on the cobac c 501 analyzer. Reagents: Closed reagent stability was assessed on the cobas c 501 analyzer at day 0, 5 months, 8 months and 17 months. The real-time stability testing is still on-going and the results support the sponsor's current shelf-life claim {6} of 15 months at 2-8°C. On-board reagent stability was assessed on the cobas c 501 analyzer and the results support the sponsor's open vial claims of 30 days when stored on-board or at 2-8°C. d. Detection limit: The sponsor determined the limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) according to the CLSI EP-17-A guideline. The limit of the blank (LoB) was determined by analyzing a blank sample with 60 replicates daily for 3 days on the cobas c 501 analyzer. The LoB was calculated as the 95th percentile which is the average of the 57th and 58th highest values for the true blanks and was determined to be 0.3 mg/L. To determine the LoD five low patient serum samples (0.08; 0.19, 0.77, 1.22 and 1.74 mg/L) were tested in 6 runs daily for 3 days on the cobas c 501 analyzer. The results demonstrated that the LoD is 0.4 mg/L. To determine the LoQ 5 diluted patient serum samples (0.645, 0.338, 0.439, 0.575 and 0.576 mg/L) were tested in 2 replicates, for 5 days on the cobas c 501 analyzer. The results demonstrated that the LoQ is 0.4 mg/L at 13.3% CV. The claimed measuring range of the device is 0.4 to 6.8 mg/L. e. Analytical specificity: Three to seven levels of each interferent were tested in triplicate with patient serum samples representing low (~1 mg/L) and high (~6 mg/L) Cystatin C levels (controls). The level of interference was considered not to be significant by the sponsor if there was no more than 10% difference between the result in the presence of the interferent and the control result. The table below lists the substances tested and the concentrations at which no significant interference was observed: Endogenous substances: | Tested substance | Concentration with ≤10% interference | | --- | --- | | Bilirubin - conjugated | 60 I Index* | | Bilirubin - unconjugated | 60 I Index | | Hemoglobin | 1000 H Index* | | Lipemia | 1000 L Index* | | Rheumatoid Factor | 1200 IU/mL | * I Index value – corresponds approximately to 1 mg/dL of bilirubin (17 μmol/L) * H Index value – corresponds approximately to 1 mg/dL of hemoglobin (0.625 μmol/L) * L index value – corresponds approximately to 1000 mg/dL Intralipid solution {7} Exogenous substances: | Tested substance | Concentration with ≤10% interference | | --- | --- | | Acetaminophen | 200 mg/L | | Acetylcystein | 150 mg/L | | Acetylsalicylic Acid | 1000 mg/L | | Ampicillin | 1000 mg/L | | Ascorbic acid | 300 mg/L | | Cefoxitin | 5 mg/L | | Cyclosporine | 5 mg/L | | Doxycyclin | 50 mg/L | | Heparin | 5000 U | | Ibuprofen | 500 mg/L | | Levodopa | 20 mg/L | | Methyldopa | 20 mg/L | | Metronidazole | 200 mg/L | | Phenylbutazone | 400 mg/L | | Rifampicin | 60 mg/L | | Theophylline | 100 mg/L | f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: A total of 103 human serum samples in the range from 0.420 mg/mL to 6.21 mg/L were tested on the candidate device and compared with the predicate device. The results were calculated using Passing/Bablok and Linear regressions shown below: Passing/Bablok (P/B): y = 0.997x - 0.064 r = 0.937 Linear regression: y = 1.031x - 0.153 r = 0.988 b. Matrix comparison: The effect of the presence of anticoagulants on analyte recovery is determined by matrix comparison. A total of 57 tubes were collected per anticoagulant type and filled completely, in addition 9 tubes (Gel Separation tubes, K₂EDTA), 8 tubes (Li-Heparin) and 6 tubes (K₃EDTA) were half-filled covering the ranges specified in the table: {8} | Anticoagulants | Sample concentration range tested (mg/L) | | --- | --- | | Li-Heparin (full) | 0.560 - 6.63 | | Li-Heparin (half) | 0.650 - 5.04 | | K_{2}-EDTA (full) | 0.550 - 6.67 | | K_{2}-EDTA (half) | 0.630 - 5.05 | | K_{3}-EDTA (full) | 0.580 - 6.72 | | K_{3}-EDTA (half) | 0.620 - 4.89 | | Gel Separation Tube | 0.510 - 6.55 | Plasma results were compared to serum results and percent recovery was determined. In addition, method comparisons with plasma vs. serum were calculated with the following results: Serum vs. Li-heparin Passing/Bablok: $y = 1.010x + 0.020$ , $r = 1.000$ Serum vs. K2-EDTA Passing/Bablok: $y = 1.020x - 0.010$ , $r = 1.000$ Serum vs. K3-EDTA Passing/Bablok: $y = 1.030x + 0.000$ , $r = 1.000$ Serum vs. Gel Separation Passing/Bablok: $y = 1.000x - 0.010$ , $r = 1.000$ 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The sponsor performed reference range studies using samples collected from 273 healthy study subjects. The samples were evenly distributed across gender and age between ages 21 and 77. The analysis of the data with the $2.5\%$ and the $97.5\%$ percentile gave a Cystatin C range from $0.61 \, \mathrm{mg/L}$ to $0.95 \, \mathrm{mg/L}$ . {9} The sponsor recommends that “each laboratory should investigate the transferability of the expected values to its own patient population and if necessary determine its own reference ranges”. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...