← Product Code [NDY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NDY) · K063647

# CYSTATIN C ANTISERUM, CALIBRATOR, CONTROL AND CONTROL HIGH (K063647)

_Thermo Electron OY · NDY · Mar 12, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NDY/K063647

## Device Facts

- **Applicant:** Thermo Electron OY
- **Product Code:** [NDY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NDY.md)
- **Decision Date:** Mar 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1225
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Cystatin C is intended for quantitative in-vitro diagnostic determination of Cystatin C in human serum or Li-heparin plasma and EDTA plasma by turbidimetry using T60 Clinical Chemistry Analyzers. Cystatin C measurements in serum and plasma are used as an aid in the diagnosis and treatment of renal diseases. Cystatin C Calibrator is intended for in vitro diagnostic use on T60 analyzer. Cystatin C Calibrator is used as a calibrator for quantification of Cystatin C in serum and plasma by immunoturbidimetry using methods defined by Thermo Electron Oy. Cystatin C control is intended for in vitro diagnostic use on T60 analyzer. Cystatin C control is used as a quality control to monitor precision of the Cystatin C test using method defined by Thermo Electron Oy. Cystatin C Control High is intended for in vitro diagnostic use on T60 analyzer. Cystatin C Control High is used as a quality control serum to monitor precision of the Cystatin C test using methods defined by Thermo Electron Oy.

## Device Story

Device performs quantitative in-vitro diagnostic determination of Cystatin C in human serum or plasma; utilizes particle-enhanced immunoturbidimetric assay on T60 Clinical Chemistry Analyzers. Input: patient serum/plasma samples; reagents consist of microparticles coated with rabbit anti-human Cystatin C. Principle: immunoprecipitation measured at 540 nm; increase in absorbance proportional to Cystatin C concentration. Output: quantitative Cystatin C concentration (mg/L). Used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing and monitoring renal disease. Includes calibrators and bi-level controls (low/high) for quality monitoring.

## Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A (total CV 0.8-4.2%). Linearity assessed per CLSI EP6-A (range 0.577-5.80 mg/L). Method comparison (n=54) against predicate showed high correlation (r=0.999, y=1.06x-0.1). Interference studies confirmed no significant interference from hemoglobin, bilirubin, rheumatoid factor, triglycerides, or Intralipid. No clinical studies performed.

## Technological Characteristics

Particle-enhanced immunoturbidimetric assay. Reagents: microparticles coated with rabbit anti-human Cystatin C, MPOS buffered saline, PBS, NaN3. Detection: 540 nm absorbance. Form factor: liquid reagents, calibrators, and controls. Connectivity: T60 clinical chemistry analyzer integration. Sterilization: N/A (in vitro diagnostic reagents).

## Regulatory Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

## Predicate Devices

- Dako Cystatin C reagent ([K041627](/device/K041627.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k063647

B. Purpose for Submission:
New Device

C. Measurand:
Cystatin C

D. Type of Test:
Quantitative

E. Applicant:
Thermo Fisher Oy

F. Proprietary and Established Names:
Cystatin C
Cystatin C Calibrator
Cystatin C Control and Cystatin C Control High

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NDY | II | 21 CFR 862.1225 | Clinical Chemistry  |
|  JIT | II | 21 CFR 862.1150 | Clinical Toxicology  |
|  JJX | I | 21 CFR 862.1660 | Clinical Toxicology  |

H. Intended Use:

1. Intended use(s):
See indications for use statements below.

{1}

2. Indication(s) for use:

Cystatin C is intended for quantitative in-vitro diagnostic determination of Cystatin C in human serum or Li-heparin plasma and EDTA plasma by turbidimetry using T60 Clinical Chemistry Analyzers.

Cystatin C measurements in serum and plasma are used as an aid in the diagnosis and treatment of renal diseases.

Cystatin C Calibrator is intended for in vitro diagnostic use on T60 analyzer. Cystatin C Calibrator is used as a calibrator for quantification of Cystatin C in serum and plasma by immunoturbidimetry using methods defined by Thermo Electron Oy.

Cystatin C control is intended for in vitro diagnostic use on T60 analyzer. Cystatin C control is used as a quality control to monitor precision of the Cystatin C test using method defined by Thermo Electron Oy.

Cystatin C Control High is intended for in vitro diagnostic use on T60 analyzer. Cystatin C Control High is used as a quality control serum to monitor precision of the Cystatin C test using methods defined by Thermo Electron Oy.

3. Special conditions for use statement(s):

For prescription use only.

4. Special instrument requirements:

T60 clinical chemistry analyzers.

I. Device Description:

The Cystatin C has reagents (2 vials), buffer (2 vials) and diluent (1 vial). The Cystatin C reagent is comprised of microparticles coated with rabbit anti-human Cystatin C. Cystatin C Buffer is a solution of polymers in MPOS buffered saline and the diluent is PBS and NaN³.

The Cystatin C Calibrator contains 2 vials of delipidated human serum liquid pools enriched with recombinant human Cystatin C produced in E. coli.

The Cystatin C controls contains 3 vials of delipidated human serum liquid pools enriched with recombinant human Cystatin C produced in E. coli. The Cystatin C controls is a bi-level control with levels of approximately 1.06 and 4.27.

This product has been tested, and found negative for HBsAg and for antibodies to HIV 1/2 and HCV. The tests used were cleared for in vitro use in the EU (Directive 98/79/EC, Annex II).

2

{2}

J. Substantial Equivalence Information:

1. Predicate device name(s):
Cystatin C Immunoparticles
Cystatin C Calibrator
Cystatin C Control Set

2. Predicate 510(k) number(s):
k041627

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Indications for Use | In vitro diagnostic determination of Cystatin C in human serum or Li-heparin plasma and EDTA plasma. Cystatin C measurements in serum and plasma are used as an aid in the diagnosis and treatment of renal diseases. | Intended for the quantitative determination of Cystatin C in human serum, heparinized plasma and EDTA plasma by turbidimetry and nephelometry. Cystatin C measurements are used as an aid in diagnosis and treatment of renal diseases.  |
|  Method | Particle enhanced immunoturbidimetric | Particle enhanced immunoturbidimetric.  |
|  Matrix | Human serum, Li-Heparin Plasma and EDTA Plasma | Human Serum, Heparinized Plasma and EDTA Plasma  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Instrument | T60 clinical chemistry analyzers | Hitachi 911, 917 and Cobas Mira Plus and IMMAGE  |
|  Measuring Range | 0.44 – 7.0 mg/L | 0.4 – 7.5 mg/L  |

{3}

4

K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A Evaluation of Precision Performance of Clinical Chemistry Devices;
CLSI EP6-P2 Evaluation of the Linearity of Quantitative Analytical Methods;
CLSI EP9-A Method Comparison and Bias Estimation Using Patient Samples;

L. Test Principle:

The Cystatin C assay uses a method based on measurement of immunoprecipitation at 540 nm. Microparticles coated with anti-human Cystatin C are added to buffered samples. The increase in absorbance caused by immunoprecipitation is recorded after a fixed timing. The amount of immunoprecipitate is proportional to the Cystatin C concentration in the solution.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Within run, between run and total precision was conducted according to CLSI EP5-A. Four samples (2 controls levels and 2 serum pools) were run on the T60 in duplicate, twice a day for twenty days. The results are summarized below.

|  Precision  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | Low Control mean 1.03 mg/L |   | High Control mean 4.59 mg/L |   | Low Pool mean 0.70 mg/L |   | High Pool mean 1.49 mg/L  |   |
|   | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  Within run | 0.027 | 2.7 | 0.054 | 1.2 | 0.010 | 1.4 | 0.039 | 2.6  |
|  Between run | 0.032 | 3.1 | 0.038 | 0.8 | 0.011 | 1.5 | 0.006 | 0.4  |
|  Total | 0.044 | 4.2 | 0.074 | 1.6 | 0.016 | 2.3 | 0.038 | 2.6  |

b. Linearity/assay reportable range:

Linearity was assessed according to CLSI EP6-A. Serum samples diluted with saline were run in triplicate. Linear regression results (expected vs. observed values) are summarized in the chart below.

|   | n | Range mg/L | Slope mg/L | Intercept mg/L  |
| --- | --- | --- | --- | --- |
|  Cystatin C | 10 | 0.577- 5.80 | 1.02 | -0.018  |

{4}

An antigen excess study (hook effect) was conducted through a dilution study of de-lipidated human serum enriched with human recombinant Cystatin C. The measured signal was plotted against the concentration of Cystatin C for each dilution. The antigen excess was defined as the concentration range between the highest standard point and the critical concentration where the signal becomes lower than the signal of the highest standard point. There was no hook effect/antigen excess found up to a Cystatin C concentration of 37.4 mg/L.

The sponsor recommends dilution of samples above the assay range up to 15 mg/L. They conducted a study to assess the linearity of the assay above 7 mg/L with human serum samples spiked with recombinant Cystatin C. Linear regression results are summarized in the chart below.

|   | n | Range mg/L | Slope mg/L | Intercept mg/L  |
| --- | --- | --- | --- | --- |
|  Cystatin C | 10 | 1.26- 15.91 | 1.045 | -0.360  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The Cystatin C calibrator is traceable to a pure recombinant human Cystatin C reference preparation. The Cystatin C concentration was established by a dry mass determination.

Stability

An accelerated stability study was conducted on two lots of the reagents at 30° C for 65 days. Real time reagent stability studies are ongoing.

Real time stability studies were conducted for the Cystatin C calibrators and controls (low and high). Three lots of each were stored at 4-8°C and were tested with two levels of frozen serum pools at 3 month intervals. The Cystatin C test was calibrated with the calibrator and controls and the serum pools were measured. The sponsors' recovery acceptance criteria were +/- 15% or 0.1 mg/L. The results support the sponsor's open vial and closed vial stability of 12 months.

Value Assignment

A Cystatin C primary calibrator is prepared gravimetrically. The assay calibrator values are assigned by multiple runs on assays calibrated using the primary calibrator. The Cystatin C controls were value assigned from multiple runs on assays calibrated by the Cystatin C calibrators.

{5}

d. Detection limit:

The detection limit was determined as the lowest concentration that gave a signal higher than the signal of the blank plus 3 SD. The results support the claimed detection limit of  $0.18\mathrm{mg / L}$ .

e. Analytical specificity:

Interference studies were conducted with 3 serum pools with low, medium and high Cystatin C concentrations. Varying concentrations of interferents were added and samples were analyzed in triplicate. The sponsor defined "no interference" as deviations less than or equal to  $+/- 10\%$  of initial value or  $+/- 0.1 \mathrm{mg} / \mathrm{L}$  of initial value.

|  Interfering substance | Cystatin C (mg/L)  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Low Cystatin C | Dev. (%) | Medium Cystatin C | Dev. (%) | High Cystatin C | Dev. (%)  |
|  Hemoglobin | 0 mg/dL | 0.68 | - | 1.66 | - | 5.63 | -  |
|   |  500 mg/dL | 0.66 | -3.9 | 1.60 | -3.8 | 5.65 | 0.4  |
|   |  1000 mg/dL | 0.62 | -9.3 | 1.58 | -4.6 | 5.70 | 1.4  |
|  Bilirubin conjugated | 0 mg/dL | 0.73 | - | 1.44 | - | 5.88 | -  |
|   |  30 mg/dL | 0.74 | 0.9 | 1.44 | 0.5 | 6.00 | 2.0  |
|   |  60 mg/dL | 0.73 | 0.0 | 1.40 | -2.6 | 5.94 | 1.0  |
|  Bilirubin nonconjugated | 0 mg/dL | 0.72 | - | 1.38 | - | 5.91 | -  |
|   |  30 mg/dL | 0.67 | -7.0 | 1.34 | -2.9 | 5.85 | -1.0  |
|   |  60 mg/dL | 0.69 | -3.7 | 1.33 | -3.6 | 5.88 | -0.5  |
|  Rheumatoid factor | 0 IU/mL | 0.70 | - | 1.20 | - | 5.81 | -  |
|   |  600 IU/mL | 0.70 | 0.0 | 1.20 | 0.3 | 5.74 | -1.1  |
|   |  1200 IU/mL | 0.77 | 10.0 | 1.24 | 3.3 | 5.85 | 0.7  |
|  Triglyceride | 0 mg/dL | 0.64 | - | 1.40 | - | 5.69 | -  |
|   |  750 mg/dL | 0.63 | -0.5 | 1.26 | -9.8 | 5.70 | 0.1  |
|   |  1500 mg/dL | 0.60 | -5.2 | 1.37 | -1.7 | 5.61 | -1.4  |
|  Intralipid® | 0 mg/dL | 0.71 | - | 1.17 | - | 5.99 | -  |
|   |  500 mg/dL | 0.67 | -5.2 | 1.14 | -2.6 | 5.68 | -5.2  |
|   |  800 mg/dL | 0.68 | -4.7 | 1.12 | -4.3 | 5.66 | -5.6  |

The sponsor claims no interference for:

Hemoglobin up to  $1000\mathrm{mg / dL}$

Bilirubin conjugated and nonconjugated up to  $58.5\mathrm{mg / dL}$

Rheumatoid factor up to 1200 IU/mL

Triglyceride up to  $1500\mathrm{mg / dL}$

Intralipid® up to  $800\mathrm{mg / dL}$

f. Assay cut-off:

Not applicable.

{6}

7

2. Comparison studies:

a. Method comparison with predicate device:

The sponsor conducted a method comparison study according to CLSI EP9-A. Fifty-four serum samples were analyzed with Cystatin C test on the T60 and the predicate Dako Cystatin C reagents on the Hitachi 917. The samples were run in singlets on a single day with a new calibration for each run. The results were recalculated using EP Evaluator version 7 method comparison CLSI EP 9 module and the Deming regression equation was determined.

|   | n | Range mg/L | Equation mg/L | Correlation Coefficient  |
| --- | --- | --- | --- | --- |
|  Cystatin C | 54 | 0.25-6.17 | y = 1.06x - 0.1 | .999  |

b. Matrix comparison:

The sponsor conducted a comparison study of serum samples paired with lithium heparin and EDTA plasma samples. Thirty paired serum samples, lithium heparin and EDTA plasma samples were analyzed in duplicate using deming regression. Six of the thirty paired samples were spiked with recombinant Cystatin C.

|  n | Serum mean (mg/L) | EDTA mean (mg/L) | % Deviation | Range (mg/L) | Regression Equation | r  |
| --- | --- | --- | --- | --- | --- | --- |
|  30 | 1.55 | 1.59 | 1.9 | 0.68 to 7.03 | Y= 1.03x + 0 | 0.998  |
|  n | Serum mean (mg/L) | Lithium Heparin mean (mg/L) | % Deviation | Range (mg/L) | Regression Equation | r  |
|  30 | 1.55 | 1.43 | -6.4 | 0.68 to 6.75 | Y=0.88x + 0.06 | 0.998  |

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

{7}

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

From a study of apparently healthy individuals, the sponsor reports the following expected values:

For individuals 1-50 years: 0.55-1.15 mg/L

For individuals &gt; 50 years: 0.63 – 1.44 mg/L

The sponsor recommends that these values should serve as a guide only and that each laboratory should verify the range or derive a reference interval for its own population.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports substantial equivalence decision.

8

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NDY/K063647](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NDY/K063647)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
