Finetest Lite Blood Glucose Monitoring System

K203562 · Osang Healthcare Co. , Ltd. · NBW · Jul 13, 2021 · Clinical Chemistry

Device Facts

Record IDK203562
Device NameFinetest Lite Blood Glucose Monitoring System
ApplicantOsang Healthcare Co. , Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateJul 13, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The Finetest Lite Blood Glucose Monitoring System is intended for single patient home use.

Device Story

System comprises blood glucose meter, test strips, control solution, and lancing device. Input: fresh capillary whole blood sample applied to test strip. Principle: electrochemical measurement; test strip generates electrical current proportional to glucose concentration; meter measures current to calculate glucose level. Used at home by patients with diabetes for self-monitoring. Output: digital glucose concentration displayed on meter screen. Healthcare providers use results to assess diabetes management effectiveness. Benefits: enables convenient, frequent monitoring of glycemic control.

Clinical Evidence

No new clinical data provided. Substantial equivalence based on design control activities, risk analysis, and verification testing of physical modifications. Disinfection efficacy and robustness studies (350 cycles) previously established in K103021.

Technological Characteristics

Electrochemical glucose measurement system. Components: meter, test strips, control solution, lancing device. Connectivity: not specified. Sterilization: not specified. Software: embedded firmware for signal processing and data storage.

Indications for Use

Indicated for single-patient home use for the quantitative measurement of glucose in capillary whole blood from the fingertip, palm, or forearm.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K203562 B Applicant Osang Healthcare Co., Ltd. C Proprietary and Established Names Finetest Lite Blood Glucose Monitoring System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable. 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.): GluNEO Lite Glucose Monitoring System, Gluneo Lite Professional Blood Glucose Monitoring System, k132966. 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device has not changed. This change was for: - Change in device dimensions and weight (from 81 X 52 X 16 mm to 83 X 56 X 18 ±1 mm; 43 g to 46 g with battery), addition feet to the bottom of the meter and change in certain meter material colors. - Change in the name of the device from GluNEO Lite Glucose Monitoring System to Finetest Lite Blood Glucose Monitoring System. - Increased the number of glucose results that can be stored in the meter memory from 365 to 500 results. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. The Finetest Lite Blood Glucose Monitoring System is intended for single patient home use. Disinfection efficacy studies were previously performed (k103021) on the exterior meter materials demonstrating complete inactivation of hepatitis B virus with CaviWipes Disinfecting Towelettes (EPA registration number 46781-8). The sponsor also conducted robustness studies and demonstrated that there was no change in performance or in the external materials of the meter after 350 cleaning and disinfection cycles designed to simulate 3 years of single-patient use. Labeling was reviewed for adequate instruction for the validated cleaning and disinfection procedures. K203562 - Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...