FOCUS BLOOD GLUCOSE MONITORING SYSTEM, FOCUS PRO BLOOD GLUCOSE MONITORING SYSTEM

K131013 · Nova Biomedical Corporation · NBW · Oct 17, 2013 · Clinical Chemistry

Device Facts

Record IDK131013
Device NameFOCUS BLOOD GLUCOSE MONITORING SYSTEM, FOCUS PRO BLOOD GLUCOSE MONITORING SYSTEM
ApplicantNova Biomedical Corporation
Product CodeNBW · Clinical Chemistry
Decision DateOct 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The FOCUS Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood obtained from the fingertip or forearm. Intended to be used by a single-patient and should not be shared. Intended for self-testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. Alternative site testing on the forearm can be used only during steady-state blood glucose conditions. Not intended for the diagnosis of or screening for diabetes, and not intended for use on neonates. FOCUS Blood Glucose Test Strips are used only with FOCUS Meters to quantitatively measure whole blood glucose in fresh, human capillary whole blood taken from the fingertip or forearm. FOCUS Control Solution is for use with FOCUS and FOCUS PRO Meters and Test Strips as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. The FOCUS PRO Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in venous and arterial whole blood and in fresh capillary whole blood obtained from the fingertip or forearm. Intended for testing outside the body (in vitro diagnostic use) for multi-patient use in a professional healthcare setting as an aid to monitor the effectiveness of a diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. Alternative site testing on the forearm should be used only during steady-state blood glucose conditions. Not intended for the diagnosis of or screening for diabetes, and not intended for use on neonates. FOCUS PRO Blood Glucose Test Strips are used only with FOCUS PRO Meters for the quantitative measurement of glucose in venous and arterial whole blood and in fresh capillary whole blood taken from the fingertip or forearm..

Device Story

Handheld blood glucose monitoring system; measures glucose in capillary, venous, or arterial whole blood via electrochemical sensing (GDH-FAD enzyme). User inserts test strip into meter; applies blood sample via capillary action; meter performs analysis in 4 seconds. Output displayed on LCD; results plasma-calibrated. FOCUS version for single-patient self-testing; FOCUS PRO version for multi-patient professional use. Meter stores up to 400 results; provides audible/visual alerts; includes ketone alert (>240 mg/dL). Powered by 3V lithium battery. Control solutions verify system performance; meter automatically detects control solution via impedance measurement and excludes results from patient averages. Benefits include rapid, quantitative glucose monitoring to aid diabetes control.

Clinical Evidence

Bench testing only. No new clinical performance studies were required for the modifications. Analytical performance (precision, linearity, specificity, hematocrit range, altitude, and sample volume) was established in previous submissions (k112638, k122435). Verification studies included a Glucose Control Auto Detection Study (n=360, 100% correct identification) and a Ketone Test Alert Study. Infection control robustness was validated for 10,950 cleaning/disinfection cycles.

Technological Characteristics

Electrochemical glucose sensor using GDH-FAD enzyme. Handheld meter powered by 3V CR2032 lithium battery. No-coding system. Measures glucose in 4 seconds using 0.4µL sample. Connectivity: standalone. Storage: 400 results. Features: test strip ejector, ketone alert, automatic control solution detection via impedance.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood (fingertip/forearm) for single-patient self-testing (FOCUS) or venous/arterial/capillary whole blood for multi-patient professional use (FOCUS PRO) as an aid in diabetes control monitoring. Contraindicated for neonates, diabetes screening/diagnosis, and critically ill/shock/dehydrated/hyper-osmolar patients.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k131013 B. Purpose for Submission: Modification of previously cleared device to include an optional ketone test alert and auto control detection. C. Measurand: FOCUS Blood Glucose Monitoring System: Glucose in fresh capillary whole blood obtained from the fingertip or forearm FOCUS Pro Blood Glucose Monitoring System: Glucose in venous and arterial whole blood and in fresh capillary whole blood obtained from the fingertip or forearm. D. Type of Test: Quantitative amperometric assay (GDH-FAD) E. Applicant: Nova Biomedical Corporation F. Proprietary and Established Names: FOCUS Blood Glucose Monitoring System FOCUS Pro Blood Glucose Monitoring System G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NBW | Class II | 21 CFR § 862.1345, Glucose Test System | Clinical Chemistry (75) | | LFR | Class II | 21 CFR § 862.1345, Glucose Test System | Clinical Chemistry (75) | | JJX | Class I, reserved | 21 CFR § 862.1660, Quality control material | Clinical Chemistry (75) | {1} H. Intended Use: 1. Intended use(s): See Indications for Use below. 2. Indication(s) for use: The FOCUS Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood obtained from the fingertip or forearm. Intended to be used by a single-patient and should not be shared. Intended for self-testing outside the body by people with diabetes mellitus as an aid to monitor the effectiveness of diabetes control. Alternative site testing on the forearm can be used only during steady-state blood glucose conditions. Not intended for the diagnosis of or screening for diabetes, and not intended for use on neonates. FOCUS Blood Glucose Test Strips are used only with FOCUS Meters to quantitatively measure whole blood glucose in fresh, human capillary whole blood taken from the fingertip or forearm. FOCUS Control Solution is for use with FOCUS and FOCUS PRO Meters and Test Strips as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. The FOCUS PRO Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in venous and arterial whole blood and in fresh capillary whole blood obtained from the fingertip or forearm. Intended for testing outside the body (in vitro diagnostic use) for multi-patient use in a professional healthcare setting as an aid to monitor the effectiveness of a diabetes control program. This system should only be used with single-use, auto-disabling lancing devices. Alternative site testing on the forearm should be used only during steady-state blood glucose conditions. Not intended for the diagnosis of or screening for diabetes, and not intended for use on neonates. FOCUS PRO Blood Glucose Test Strips are used only with FOCUS PRO Meters for the quantitative measurement of glucose in venous and arterial whole blood and in fresh capillary whole blood taken from the fingertip or forearm.. 3. Special conditions for use statement(s): FOCUS Blood Glucose Monitoring System FOCUS Pro Blood Glucose Monitoring System The FOCUS Blood Glucose Monitoring System - For Over-the-Counter use - Not for neonatal use {2} - Not for screening or diagnosis of diabetes mellitus - Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar patients - Alternative site testing (AST) can be used only during steady-state blood glucose conditions - AST results should not be used for continuous glucose monitor calibration or for use in insulin dose calculations - Single-patient use only system; should not be shared ## The FOCUS PRO Blood Glucose Monitoring System - For Prescription Use - Not for neonatal use - Not for screening or diagnosis of diabetes mellitus - This system has not been validated in the critically ill - Alternative site testing (AST) can be used only during steady-state blood glucose conditions - AST results should not be used for continuous glucose monitor calibration or for use in the insulin dose calculations - This system should only be used with single-use, auto-disabling lancing devices ## 4. Special instrument requirements: - FOCUS Blood Glucose Meter - FOCUS Pro Blood Glucose Meter ## I. Device Description: This is a modification of a previously cleared device to include housing/material construction changes and software modification to include optional ketone test alert and auto control detection. The FOCUS Blood Glucose Monitoring System consists of the FOUCS Blood Glucose meter (sold separately), 10 FOCUS Test Strips (sold separately), FOCUS Control Solution Level 1 (sold separately), Lancing Device for single patient use only and lancets, clear cap (sold separately), log book for recording test results carry case, and User Manual. The FOCUS PRO Blood Glucose Monitoring System consists of the FOUCS PRO Blood Glucose meter (sold separately), 10 PRO FOCUS Test Strips (sold separately), FOCUS Control Solution Level 1(sold separately), and User Manual. Each test strip contains the following reagent compositions: flavin adenine dinucleotide-glucose dehydrogenase (FAD-GDH from Aspergillus Oryzae); mediator ; and other non-reactive ingredients. Control Solution level 1 is included in the kits, and control solution levels 2 and 3 are sold {3} separately. Each level control solution vial contains 3.0 ml aqueous solution. # J. Substantial Equivalence Information: 1. Predicate device name(s): Nova One Blood Glucose Monitoring System 2. Predicate 510(k) number(s): k122435 5. Comparison with predicate: | Item | Candidate device FOCUS Blood Glucose Monitoring System (Self-Monitoring) | NOVA One Blood Glucose Monitoring System (Predicate Device, k122435) | | --- | --- | --- | | Similarities | | | | Indication For Use | It is intended to be used for quantitative measurement of glucose in fresh capillary whole blood as an aid to monitor the effectiveness of diabetes control in people with diabetes. | Same | | Assay Method | Colorimetric Electrochemical Sensor | Same | | Test Range | 20 – 600 mg/dL | Same | | Sample volume | 0.40μl | same | | Test time | 4 seconds | same | | Sample type | Glucose in venous and arterial whole blood and in fresh capillary whole blood obtained from the fingertip or forearm. | same | | Hematocrit | 25% to 60% | Same | | Coding | None | Same | | | | | | Differences | | | | Ketone Alert | Yes, when glucose value is above 240 mg/dL. Feature can be turned on/off by the user during set up. | N/A | | Automatic Control Detection | Yes, identifies control sample as a control. | N/A | {4} | Item | Candidate Device FOCUS PRO Blood Glucose Monitoring System (Professional Monitoring) | NOVA One Blood Glucose Monitoring System (Professional Monitoring) (Predicate Device, k122435) | | --- | --- | --- | | Similarities | | | | Indication For Use | It is intended to be used for the quantitative measurement of glucose in venous, arterial and fresh capillary whole blood from the finger and forearm as an aid to monitor the effectiveness of diabetes control program. | Same | | Assay Method | Colorimetric Electrochemical Sensor | Same | | Test Range | 20 – 600 mg/dL | Same | | Sample volume | 0.40μl | same | | Test time | 4 seconds | same | | Sample type | Glucose in venous and arterial whole blood and in fresh capillary whole blood obtained from the fingertip or forearm. | same | | Hematocrit | 25% to 60% | Same | | Coding | None | Same | | Differences | | | | Ketone Alert | Yes, when glucose value is above 240 mg/dL. Feature can be turned on/off by the user during set up. | N/A | | Automatic Control Detection | Yes, identifies control sample as a control. | N/A | | Item | Candidate device FOCUS Control Solutions | NOVA One Glucose Control Solutions (Predicate Device, k122435) | | --- | --- | --- | | Similarities | | | | Indication For Use | It is intended to be used for use as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. | Same | | Levels | 1,2, and 3 | same | | Matrix | aqueous | same | {5} 6 K. Standard/Guidance Document Referenced (if applicable): IEC 60601: Medical electrical equipment-Part 1-2: general requirements for safety- collateral standard: Electromagnetic compatibility – Requirements and tests. EN55011: Limits and methods of measurement of radio disturbance characteristics of industrial, scientific, and medical RF equipment. IEC 61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General Requirements. IEC 61010-2-101: Safety requirements for electrical equipment for measurement, control and laboratory use – Particular requirements for In Vitro Diagnostic (IVD) Medical Equipment. L. Test Principle: The FOCUS Blood Glucose Monitor and FOCUS PRO Blood Glucose Monitor measures glucose electrochemically. The glucose biosensor is capable of recognizing the glucose present in whole blood or control solutions by virtue of the glucose specificity of the enzyme glucose dehydrogenase (GDH) present on the glucose test strip. The electrons liberated by this reaction are transferred via a co-factor and mediator (FAD) to the meter where they are read as a small electrical current. The current is integrated over the analysis time to generate charge which is directly proportional to the level of the glucose in the applied sample. M. Performance Characteristics (if/when applicable): This submission is seeking clearance for modifications to a previously cleared device to include an optional ketone test alert and auto control detection. The sponsor uses the same analytical claim for the current test system and refers to the predicate 510(k) for the respective studies. No new performance studies were necessary for this device modification. The two systems in the submission only differ in their intended use (single-use vs. multiple-patient use) and the name; therefore, the performance data applies to both systems. 1. Analytical performance: a. Precision/Reproducibility: Established in k112638. b. Linearity/assay reportable range: The measuring range of the device is 20-600 mg/dL based on studies conducted in k112638. c. Traceability, Stability, Expected values (controls, calibrators, or methods): {6} Traceability Traceable to NIST SRM 91, dry D-glucose. Stability Glucose test strip: The candidate blood glucose test strips were previously cleared in k112638. However the brand name of the glucose strip has been updated to FOCUS Test Strips and FOCUS Pro Test Strips. The stability claims are 24 months when stored between temperatures of 15 to 30 °C between 10- 90% relative humidity and open-vial stability of 90 days when strips are stored at 15 to 30 °C between 10- 90% relative humidity as determined in k112638. Control Solution: The control solutions were previously cleared and stability and value assignment was previously established in k112638. The only difference is that they have been renamed to FOCUS Control Solutions (Levels 1, 2, 3). d. Detection limit: The reportable range is 20-600 mg/dL based on linearity study referenced in Section M.1.b. above. e. Analytical specificity: Established in k112638. The sponsor has provided the following statements in the labeling: - Cholesterol up to 500 mg/dL or triglyceride up to 750 mg/dL does not significantly affect results. - Therapeutic levels of n-acetylcysteine and elevated uric acid may affect results. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with YSI: The system accuracy on finger stick and forearm samples was evaluated in k112638. The system accuracy on venous and arterial samples was evaluated in k122435. 7 {7} b. Matrix comparison: Established in k122435. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): A lay user finger and forearm study was evaluated in k112638. 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: In the labeling the sponsor provides the following expected values and reference citation: The normal fasting adult blood glucose value for a person without diabetes is <100 mg/dL. One to two hours after meals normal blood glucose levels should be less than 140 mg/dL. American Diabetes Association, Diagnosis and Classification of Diabetes Mellitus. Diabetes Care, Volume 36, Supplement 1, January 2013. N. Instrument Name: FOCUS Blood Glucose Meter FOCUS PRO Blood Glucose Meter O. System Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. {8} Does the applicant’s device contain the ability to transmit data to a computer, webserver or mobile device? Yes ________ or No ______ X Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? Yes ________ or No ______ X 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ______ X ______ or No ______ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. Controls samples are identified as controls. 4. Specimen Sampling and Handling: FOCUS Blood Glucose Monitor is intended to be used with capillary whole blood from the fingertip, which is directly applied to the test strip. FOCUS PRO Blood Glucose Monitor is intended to be used with capillary whole blood from the finger and forearm, as well as venous and arterial whole blood collected in sodium or lithium heparin tubes. The whole blood sample is applied directly to the test strip by capillary action. 5. Calibration: The meter is a non-coding meter, therefore no coding is required by the user. 6. Quality Control: The sponsor has three levels of control solutions (level 1 control solution is included with the kits). When a test strip is inserted into the meter, each control can be measured by following the instructions for “Running Control Solution” provided in the user’s manual. Auto Control Detection automatically detects control solution when used in testing. The ionic strength of the control solution provides a current flow unique to just control solution. An acceptable range for each control level is printed on the test strip vial label. If the test results fall outside the range printed on the test strip vial, the user is instructed to contact Customer Service (available 24 hours a day, 7 days a week). 9 {9} P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: 1. Infection control Infection Control Studies: The device systems are intended for single-patient use (FOCUS Blood Glucose Monitoring System) or multiple-patient use (FOCUS Pro Blood Glucose Monitoring System). Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, PDI Super Sani-Cloth Wipes (EPA Registration # 9480-4). Robustness studies were performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 10,950 cleanings and 10,950 disinfection steps with the 10,950 wipes. The robustness studies were designed to simulate 3 years of use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures. 2. Temperature and humidity operating conditions The sponsor claims operating temperature from 57°F-104°F (14-40°C) and relative humidity range from 10% to 93%, as established in k112638. 3. EMC testing EMC testing was evaluated and certified by The Compliance Management Group. A test certificate was issued to Nova on January 30, 2013. 4. Altitude Study The sponsor claims that an altitude up to 10,000 feet does not affect the test results of the proposed device as established in k112638. 5. Hematocrit study The sponsor claims a hematocrit range of 25 to 60% as established in k112638. 6. Usability Study The sponsor performed a Usability Study using 101 subjects to evaluate the FOCUS blood glucose monitoring system with respect to ergonomics, ease-of-use, and instructions for use. 7. Readability Evaluation of FOCUS Instructions The sponsor performed a readability assessment of the labeling and states that the owner's guide, strip insert and control insert are written at the 8th grade level or below based on Flesch-Kincaid Readability Assessment. 8. Glucose Control Auto Detection Study The sponsor performed a Glucose Control Auto Detection Study to verify the FOCUS blood glucose monitoring system correctly performs automatic glucose control detection 10 {10} by identifying glucose control samples as a control. The sponsor evaluated three levels of glucose control solutions and 6 levels of venous whole blood (n=360) using three lots of test strips and six FOCUS/FOCUS PRO meters and correctly identified 100% of the control solutions. 9. Ketone Test Alert Study The Ketone Test Alert (optional) is intended to alert users when blood glucose results are greater than 240 mg/dL so that users may check their ketone levels using an alternate method if it is part of their treatment plan. The sponsor performed a Ketone Test Alert Study to verify the FOCUS blood glucose monitoring system correctly performs the Ketone Test Alert on samples > 240 mg/dL. 10. Sample Volume Study The sponsor claims that a minimum of 0.40μl of sample volume are required for this system, as established in k112638. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 11
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...