LIPIDPRO LIPID PROFILE AND GLUCOSE MEASURING SYSTEM, LIPIDPRO PROFESSIONAL LIPID PROFILE AND GLUCOSE MEASURING SYSTEM.
Device Facts
| Record ID | K130295 |
|---|---|
| Device Name | LIPIDPRO LIPID PROFILE AND GLUCOSE MEASURING SYSTEM, LIPIDPRO PROFESSIONAL LIPID PROFILE AND GLUCOSE MEASURING SYSTEM. |
| Applicant | Infopia Co, Ltd. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Jun 21, 2013 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Indications for Use
The LipidPro® Lipid Profile and Glucose Measuring System is intended for self testing (in home) and for testing outside the body (in vitro diagnostic use only). The LipidPro® Lipid Profile and Glucose Measuring System, which consists of a meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG) and glucose in capillary whole blood. The LipidPro® Lipid Profile and Glucose Measuring System is intended to be used by a single patient and should not be shared. The LipidPro® Total Cholesterol Test Strip, LipidPro® HDL-cholesterol Test Strip, LipidPro® Triglyceride Test Strip, Lipid Profile Test Strip, and LividPro® Glucose Test Strip are for use with the LipidPro® Lipid Profile and Glucose Meter. The LipidPro® Total Cholesterol Test Strip and the LipidPro® HDL-cholesterol Test Strip are used to measure total cholesterol (TC) and high density lipoprotein cholesterol (HDL-C) in capillary whole blood respectively and the measurements obtained are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid, and lipoprotein metabolism disorders. The LipidPro® Total Cholesterol Test Strip and LipidPro® HDLcholesterol Test Strip measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. The LipidPro® Triglyceride Test Strip is used to measure triglyceride (TG) in capillary whole bloodandthe obtained measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL-cholesterol values are reported only when triglycerides are ≤ 350 mg/dL; when triglycerides are > 350 mg/dL. calculated LDL-cholesterol are not reported. The LipidPro® Lipid Profile Test Strip is used to measure TC, HDL-C, and TG in capillary whole blood at the same time. The LipidPro® Glucose Test Strip is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand. upper arm, forearm, caff and/or thigh by diabetic patients as an aid in the management of diabetes, Glucose measurement is not to be used for the diagnosis of or screening for diabetes and/or for neonatal use. Alternate site testing should be done during steady-state times when glucose is not changing rapidly. LipidPro® Total Cholesterol Control Solution, HDL-cholesterol Control Solution, Triglycerides Control Solution, and Glucose Control Solution are used to test the precision of the LipidPro® Lipid Profile and Glucose Measuring System and to detect systematic analytic deviations that may arise from reagent or analytical variation and they are for testing outside the body (in vitro diagnostic use only). LipidPro® Control Solutions are intended for use in home. The LipidPro® Professional Lipid Profile and Glucose Measuring Systemis intended for multiple patient use in professional health care settings and for testing outside the body (in vitro diagnostic use only). The LipidPro® Professional Lipid Profile and Glucose Measuring System which consists of meter and test strips, measures total cholesterol (TC), high density lipoprotein cholesterol (HDL-C), triglyceride (TG), and glucose in capillary whole blood. The LipidPro® Professional Lipid Profile and Glucose Measuring System should only be used with auto-disabling, single-use lancing device. The LipidPro® Professional Total Cholesterol Test Strip, LipidPro® Professional HDL-cholesterol Test Strip, LipidPro® Professional Triglyceride Test Strip, LipidPro® Professional Lipid Profile Test Strip, and LipidPro® Professional Glucose Test Strip are for use with the LipidPro® Professional Lipid Profile and Glucose Meter. The LipidPro® Professional Total Cholesterol Test strip and the LipidPro® Professional HDL-cholesterol Test Strip are used to measure total cholesterol (TC) and high density lipoprotein cholesterol (HDL-C) in capillary whole blood respectively and the measurements obtained are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid, and lipoprotein metabolism disorders. The LipidPro® Professional Total Cholesterol Test strip and ' LipidPro® Professional HDL-cholesterol Test Strip measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. The LipidPro® Professional Triglyceride Test Strip is used to measure triglyceride (TG) in capillary whole blood and the obtained measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders. Calculated LDL-cholesterol values are reported only when triglycerides are ≤ 350 mg/dL; when triglycerides are > 350 mg/dL, calculated LDLcholesterol are not reported. The LipidPro® Professional Lipid Profile Test Strip is used to measure TC, HDL-C, and TG in capillary whole blood at the same time. The LipidPro® Professional Glucose Test Strip is for the quantitative measurement of the concentration of glucose in capillary whole blood that can be taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh by diabetic patients as an aid in the management of diabetes. Glucose measurement is not to be used for the diagnosis of or screening for diabetes and/or for neonatal use. Alternate site testing should be done during steady-state times when glucose is not changing rapidly. The LipidPro® Professional Total Cholesterol Control Solution, HDL-cholesterol Control Solution, Triglycerides Control Solution, and Glucose Control Solution are used to test the precision of the LipidPro® Professional Lipid Profile and Glucose Measuring System and to detect systematic analytic deviations that may arise from reagent or analytical variation and they are for testing outside the body (in vitro diagnostic use only). The LipidPro® Professional Control Solutions are intended for use in hospital and professional health care settings.
Device Story
System comprises meter, test strips, and control solutions; measures TC, HDL-C, TG, and glucose in capillary whole blood. Used in home (patient self-use) or professional healthcare settings (physician/nurse/technician). Input: capillary blood sample on test strip. Processing: electrochemical measurement of analytes; calculates LDL-C if TG ≤ 350 mg/dL. Output: quantitative concentration values displayed on meter; data transfer to PC/printer. Benefits: aids management of lipid disorders and diabetes. Professional version requires auto-disabling lancing device. Disinfection validated for multi-patient use.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed performance, functionality, and reliability against pre-determined pass/fail criteria. Disinfection study validated CaviWipes (EPA 46781-8) for meter and lancing device, demonstrating no performance degradation after 10,980 cycles (simulated 3 years of use).
Technological Characteristics
In vitro diagnostic meter for lipid/glucose measurement. Features include test strips, check strips, and control solutions. Materials include modified control solution bottles and redesigned strip covers/capillary rods. Connectivity includes PC data transfer. Software includes error messaging and temperature unit selection. Sterilization/disinfection validated for multi-patient use via CaviWipes (EPA Reg. No. 46781-8).
Indications for Use
Indicated for the quantitative measurement of lipids and glucose in blood. Intended for either single-patient use (LipidPro System) or multiple-patient use (LipidPro Professional System).
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- LipidPro™ System (K090405)