ONETOUCH VERIO SYNC BLOOD GLUCOSE MONITORING SYSTEM
K120708 · Cilag GmbH International · NBW · Feb 7, 2013 · Clinical Chemistry
Device Facts
Record ID
K120708
Device Name
ONETOUCH VERIO SYNC BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Cilag GmbH International
Product Code
NBW · Clinical Chemistry
Decision Date
Feb 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OneTouch Verio Sync Blood Glucose Monitoring System (BGMS) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The system is intended to be used by a single person and should not be shared. The OneTouch Verio Sync BGMS is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch Verio Sync BGMS should not be used for the diagnosis of or screening of diabetes or for neonatal use. The OneTouch Verio Test Strips are for use with the OneTouch Verio Sync Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Device Story
System measures glucose in fresh capillary whole blood via amperometric electrochemical reaction in test strips. Input: blood sample on test strip; output: glucose concentration displayed on meter. Device includes meter, test strips, control solutions, lancing device, and lancets. Used at home by patients with diabetes. Features wireless communication with OneTouch Reveal Diabetes Management Application and rechargeable battery. Output allows patients to monitor diabetes control effectiveness. Differences from predicate include ergonomic design, simplified UI, wireless connectivity, and rechargeable battery; no changes to core electrochemical sensing technology.
Clinical Evidence
System accuracy evaluated via method comparison against YSI 2300 reference using 100 participants (glucose range 40-434 mg/dL). User performance study conducted with 273 untrained lay users (glucose range 49-506 mg/dL). Results showed 100% of samples within ±20% of reference for concentrations ≥75 mg/dL and 100% within ±15 mg/dL for concentrations <75 mg/dL. Precision studies (repeatability and intermediate) demonstrated CVs between 1.49% and 3.04%. Bench testing confirmed performance across hematocrit (20-60%), altitude (up to 10,000 ft), and temperature/humidity ranges.
Technological Characteristics
Amperometric measurement using GDH-FAD enzyme dry reagent assay. Plasma-calibrated. Dimensions: 3.9 x 1.6 x 0.5 inches. Connectivity: Bluetooth wireless to mobile devices. Power: AC adaptor/USB. Operating range: 50–111°F, 10–90% humidity. Hematocrit range: 20–60%. Standards: CLSI EP06-A, EP07-A, ISO 15197:2003, IEC 61010-1, IEC 61326-1.
Indications for Use
Indicated for quantitative measurement of glucose in fresh capillary whole blood from fingertips in people with diabetes for home monitoring of diabetes control. Not for neonatal use, diabetes diagnosis/screening, or use on critically ill, shock, dehydrated, hypotensive, or hyper-osmolar patients.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch Verio Blood Glucose Monitoring System (k093745)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k120708
B. Purpose for Submission:
New Device
C. Measurand:
Capillary Whole Blood Glucose
D. Type of Test:
Quantitative, Amperometric method, Glucose dehydrogenase (FAD)
E. Applicant:
Cilag Gmbh International
F. Proprietary and Established Names:
OneTouch Verio Sync Blood Glucose Monitoring System
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LFR – Glucose dehydrogenase, glucose | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
| NBW – system, test, blood glucose, over the counter | Class II | 21 CFR § 862.1345 | Clinical Chemistry (75) |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The OneTouch Verio Sync Blood Glucose Monitoring System (BGMS) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The system is intended to be used by a single person and should not be shared.
The OneTouch Verio Sync BGMS is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch Verio Sync BGMS should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The OneTouch Verio Test Strips are for use with the OneTouch Verio Sync Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
3. Special conditions for use statement(s):
For over-the-counter use.
Not for neonatal use, nor for screening for or diagnosis of diabetes mellitus.
Not for use on critically ill patients, patients in shock, dehydrated patients, hypotensive patients or hyper-osmolar patients.
For single patient use only and should not be shared.
4. Special instrument requirements:
OneTouch Verio Sync Blood Glucose Meter
I. Device Description:
The OneTouch® Verio™ Sync Blood Glucose Monitoring System consists of the following components:
- OneTouch Verio Sync Meter
- OneTouch Verio Level 3 Control Solution (available separately)
- OneTouch Verio Level 4 Control Solution (available separately)
{2}
- OneTouch Verio Test Strips (available separately)
- OneTouch Delica Lancing Device
- OneTouch Delica Sterile Lancets
- Carrying Case
- AC Adaptor (wall charger)
- USB Cable
- OneTouch® Verio Sync Product Labeling
Accessories available separately for the OneTouch® Verio Sync Blood Glucose Monitoring System include the OneTouch Reveal Diabetes Management Application (cleared under k120558) and the OneTouch® Diabetes Management Software (DMS - cleared under k 984527).
# J. Substantial Equivalence Information:
1. Predicate device name(s):
OneTouch Verio Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k093745
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate (k093745) |
| | | |
| Indications for Use | Same | For the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples. |
| Operating principle / methodology | Same | Electrochemical reaction with GDH-FAD Glucose Dehydrogenase |
| Minimum sample volume | Same | 0.4 μL |
| Test time | Same | 5 seconds |
| Measuring range | Same | 20 – 600 mg/dL |
| Calibration | Same | No coding required |
| Hematocrit range | Same | 20 – 60% |
| Operating humidity range | Same | 10 – 90% non-condensing |
| Units | Same | mg/dL, not user |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate (k093745) |
| | | changeable |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate (k093745) |
| Sampling sites | Fingertip only | Fingertip, forearm, and palm |
| Physical dimensions | 3.9 x 1.6 x 0.5 inches, 1.7 oz. | 2.9 x 2.2 x 0.8 inches, 1.9 oz |
| Operating Temperature Range | 50 – 111° F 10 – 44 ° C | 43 – 111° F 6 – 44 ° C |
| Buttons | 1 button: Turns meter off and on, changes the illumination level and toggles device between bluetooth and airplane mode | 4 buttons: Up/down/OK/Back on face of the device |
| Control solutions and target levels | Level 3: 120 mg/dL | Mid: 120 mg/dL |
| | Level 4: 350 mg/dL | High: 350 mg/dL |
| Result averaging | None. Averages are calculated only by the OneTouch Reveal Diabetes Management Application | 7, 14, and 30 day averages |
| Date / time setting | The date and time are set by the OneTouch Reveal Diabetes Management Application. | Date and time are user set |
| Wireless data transmission capability | Bluetooth wireless transmission to the paired Apple partner device. | None |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach.
{4}
CLSI EP07-A: Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition.
ISO 15197:2003 In Vitro Diagnostic Test Systems- Requirements For In Vitro Whole Blood Glucose
## L. Test Principle:
The OneTouch Verio Blood Glucose Monitoring System employs flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry as the standard dry reagent assay for glucose in whole blood. This enzyme assay, with a redox chemical "mediator" reaction, is used to generate an electrical current proportional to the glucose concentration in the blood sample. The system is designed as an amperometric measurement device using current generated from the redox reaction as the measurable response.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Repeatability**
Venous blood from one donor with nominal hematocrit (37 - 45%) collected with lithium heparin anticoagulant, was adjusted with glucose to five glucose levels (30-50, 51-110, 111-150, 151-250, 251-400) across the claimed range and tested on three lots of production strips on 30 meters (10 meters per test strip lot). Ten replicates were tested per meter, test strip lot and glucose concentration. The testing for each group of 10 meters was executed by one operator over one day. Results are summarized below:
Test Strip Lot 1
| Glucose conc. (mg/dL) | 30-50 | 51-110 | 111-150 | 151-250 | 251-400 |
| --- | --- | --- | --- | --- | --- |
| n | 100 | 100 | 100 | 100 | 100 |
| Mean (mg/dL) | 44.51 | 102.42 | 133.41 | 203.8 | 354.04 |
| Std Dev (mg/dL) | 0.82 | 1.98 | 2.49 | 4.18 | 7.56 |
| CV (%) | 1.84 | 1.93 | 1.87 | 2.05 | 2.14 |
Test Strip Lot 2
| Glucose conc. (mg/dL) | 30-50 | 51-110 | 111-150 | 151-250 | 251-400 |
| --- | --- | --- | --- | --- | --- |
| n | 100 | 100 | 100 | 100 | 100 |
| Mean (mg/dL) | 35.92 | 104.48 | 132.17 | 201.68 | 333.16 |
| Std Dev (mg/dL) | 1.02 | 3.17 | 3.68 | 4.5 | 10.12 |
| CV (%) | 2.84 | 3.03 | 2.78 | 2.23 | 3.04 |
{5}
Test Strip Lot 3
| Glucose conc.(mg/dL) | 30-50 | 51-110 | 111-150 | 151-250 | 251-400 |
| --- | --- | --- | --- | --- | --- |
| n | 100 | 100 | 100 | 100 | 100 |
| Mean (mg/dL) | 34.99 | 97.49 | 136.54 | 193.8 | 349.54 |
| Std Dev (mg/dL) | 0.75 | 2.37 | 2.62 | 5.17 | 9.33 |
| CV (%) | 2.14 | 2.43 | 1.92 | 2.67 | 2.67 |
# Intermediate Precision
In this study, three test strip lots were tested with control solutions at three different concentrations. The test was performed with 30 meters over 10 days, and 20 replicates per day.
Test Strip Lot 1
| Glucose concentration range (mg/dL) | Control Level 1
30-50 | Control Level 2
96-144 | Control Level 3
280-420 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 37.76 | 116.51 | 335.27 |
| n | 200 | 200 | 200 |
| Std Dev (mg/dL) | 0.82 | 1.74 | 5.61 |
| CV (%) | 2.17 | 1.49 | 1.67 |
Test Strip Lot 2
| Glucose concentration range (mg/dL) | Control Level 1
30-50 | Control Level 2
96-144 | Control Level 3
280-420 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 38.65 | 116.91 | 335.75 |
| n | 200 | 200 | 200 |
| Std Dev (mg/dL) | 0.88 | 2.10 | 6.35 |
| CV (%) | 2.28 | 1.80 | 1.89 |
Test Strip Lot 3
| Glucose concentration range (mg/dL) | Control Level 1
30-50 | Control Level 2
96-144 | Control Level 3
280-420 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 39.49 | 118.75 | 336.70 |
| n | 200 | 200 | 200 |
| Std Dev (mg/dL) | 0.93 | 2.07 | 6.73 |
| CV (%) | 2.36 | 1.74 | 2.00 |
b. Linearity/assay reportable range:
{6}
In this study, venous blood from nine donors (with nominal hematocrits of 37-45%) was adjusted to eight target glucose levels (20, 100, 200, 300, 400, 500, 600 and 700 mg/dL) and tested with 3 OneTouch Verio Test Strip lots on eight OneTouch Verio Sync Blood Glucose Meters. Each sample was also assayed with the YSI reference method. Note: The target glucose of 700 mg/dL exceeds that of the claimed OneTouch Verio Sync Blood Glucose Monitoring System glucose concentration range which is 20 - 600 mg/dL. The meters were modified to display a glucose result greater than 600 mg/dL. Linear regression of the data produced the following:
| Strip Lot | Intercept (with 95% CI) | Slope (with 95% CI) | r² |
| --- | --- | --- | --- |
| 1 | 2.8 (2.1 – 3.5) | 0.969 (0.965 – 0.972) | 0.996 |
| 2 | 4.6 (3.9 – 5.3) | 0.958 (0.954 – 0.961) | 0.996 |
| 3 | 5.2 (4.4 – 5.9) | 0.973 (0.969 – 0.976) | 0.996 |
The results of the study support the sponsor’s claimed glucose measurement range of 20 – 600 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The OneTouch Verio Sync Blood Glucose Monitoring System is traceable to the NIST SRM 917b reference material.
Test Strip Stability
Closed vial (shelf life) stability was assessed in real-time studies. The protocols and acceptance criteria were reviewed and found to be acceptable. The testing supported the claimed shelf life of 22 months when stored at 41 - 86° F (5 - 30° C).
Opened vial (in-use) stability was assessed in real-time studies. The protocols and acceptance criteria were reviewed and found to be acceptable. The testing supported the open vial stability of six months when stored at 41 - 86° F (5 - 30° C).
Control Solution Stability:
Control shelf-life stability (closed vial) was assessed in real-time studies. The protocols and acceptance criteria were reviewed and found to be acceptable. The testing supported the claimed shelf life of 24 months when stored at 41 - 86° F (5 - 30° C).
{7}
Control in-use stability (open vial) was assessed in real-time studies. The protocols and acceptance criteria were reviewed and found to be acceptable. The testing supported the open vial stability of six months when stored at 41 - 86° F (5 - 30° C).
d. Detection limit:
The measuring range of the OneTouch Verio Sync Blood Glucose Monitoring System is 20 – 600 mg/dL. This range was validated by the linearity study (M.1.b).
e. Analytical specificity:
The sponsor performed interference studies in accordance with CLSI EP7-A. Testing was performed in parallel (control samples vs. test samples) to minimize the effects of glucose metabolism.
Venous blood was drawn into Lithium heparin anticoagulant tubes from healthy volunteers. The glucose levels tested were 65 and 240 mg/dL. Samples were allowed to glycolyze or were spiked, as needed, to achieve the 65 and 240 mg/dL glucose levels. A low and high concentration of each potential interferent was then tested at each glucose level. The following substances were found not to interfere at the concentrations listed:
| Endogenous Substances | No interference up to (mg/dL in plasma): |
| --- | --- |
| Ascorbic Acid | 10.2 |
| Bilirubin | 32.5 |
| Cholesterol | 837 |
| Creatinine | 8.5 |
| Glutathione | 12.8 |
| Triglycerides | 3331 |
| Urea | 424 |
| Uric Acid | 8 |
| Lactose | 35.8 |
| Galactose | 105 |
| Maltose | 349 |
| Exogenous Substances | No interference up to (mg/dL in plasma): |
| Acetaminophen | 8 |
| Dopamine | 0.18 |
| Ephedrine | 0.35 |
| Gentisic Acid | 1.8 |
| Ibuprofen | 86.0 |
| Levo dopa | 1.7 |
{8}
The sponsor has the following limitations in their labeling:
| Tolazamide | Concentrations greater than 10.8mg/dL may cause falsely low results |
| --- | --- |
| Uric Acid | Concentrations greater than 8mg/dL may cause falsely low results |
f. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison with predicate device:
# System Accuracy Study
To assess system accuracy, results from the OneTouch Verio Sync Blood Glucose Monitoring System were compared to a reference method (YSI 2300). Capillary fingerstick samples from 100 participants with glucose concentrations ranging from $40 - 434\mathrm{mg / dL}$ were analyzed with the candidate device and with YSI. Three samples were collected from each fingerstick. The first was used for the Verio Sync measurement, the second for the YSI measurement, and the third for a hematocrit measurement. Two participant samples were allowed to glycolyze to obtain a glucose concentration $< 50\mathrm{mg / dL}$ and three participant samples were spiked to obtain a glucose concentration $>400\mathrm{mg / dL}$ . The results relative to the reference method are summarized in the tables below:
Concentrations $< 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 1 | 15/19 (79%) | 18/19 (95%) | 19/19 (100%) |
| 2 | 14/19 (74%) | 18/19 (95%) | 19/19 (100%) |
| 3 | 11/19 (58%) | 19/19 (100%) | 19/19 (100%) |
| All | 40/57 (70%) | 55/57 (96%) | 57/57 (100%) |
{9}
Concentrations $\geq 75\ \mathrm{mg/dL}$
| Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- | --- |
| 1 | 60/81 (74%) | 78/81 (96%) | 81/81 (100%) | 81/81 (100%) |
| 2 | 59/81 (73%) | 79/81 (98%) | 81/81 (100%) | 81/81 (100%) |
| 3 | 55/81 (68%) | 78/81 (96%) | 80/81 (99%) | 81/81 (100%) |
| All | 174/243 (72%) | 235/243 (97%) | 242/243 (99.59%) | 243/243 (100%) |
Linear Regressions
| Strip Lot | n | Intercept (with 95% CI) | Slope (with 95% CI) | r² |
| --- | --- | --- | --- | --- |
| 1 | 100 | 1.16 (-2.11 – 4.43) | 1.00 (0.99 – 1.02) | 0.993 |
| 2 | 100 | 1.96 (-1.31 – 5.23) | 0.99 (0.97 – 1.01) | 0.993 |
| 3 | 100 | 4.7 (0.77 – 8.66) | 0.99 (0.97 – 1.01) | 0.990 |
b. Matrix comparison:
Not Applicable. Only fresh capillary whole blood samples from the fingertip may be used.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
User Performance Study:
To assess the performance of the Verio Sync Blood Glucose Monitoring System in the hands of the intended users, the sponsor performed a study with 273 untrained lay user participants. Each participant collected and analyzed their own fingerstick sample. A second fingerstick sample was then collected and analyzed on the YSI reference method. The study was performed at three clinical sites and the gender breakdown was 53% male and 47% female. Three lots of test strips were used and single measurements collected by the users were compared to the reference method. None of the samples were altered before analysis. The range of glucose values for the samples as measured by the reference method was 49 - 506 mg/dL. Results are summarized below:
{10}
Concentrations < 75 mg/dL
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- |
| 5/23 (22%) | 15/23 (65%) | 22/23 (96%) |
Concentrations ≥ 75 mg/dL
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- |
| 141/250 (56%) | 219/250 (88%) | 242/250 (97%) | 248/250 (99%) |
Linear Regression
| n | Intercept (with 95% CI) | Slope (with 95% CI) | r² |
| --- | --- | --- | --- |
| 273 | 8.84 (5.64 – 12.0) | 0.96 (0.94 – 0.98) | 0.98 |
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
Expected blood glucose levels for non-pregnant people without diabetes:
| Time | Range (mg/dL) |
| --- | --- |
| Fasting | Less than 100 |
| 2 hours after meals | Less than 140 |
American Diabetes Association, Position Statement, Diagnosis and Classification of Diabetes Mellitus, Diabetes Care 35:S4-S10, 2012.
N. Instrument Name:
OneTouch Verio Sync Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings. The minimum sample volume is 0.4 uL.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
☐ Yes ☐ No
Does the applicant’s device transmit data to a computer, webserver, or mobile device
{11}
using wireless transmission?
☐ Yes ☐ No
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
☐ Yes ☐ No
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The device is intended to be used with capillary whole blood from the finger only. The whole blood sample is applied directly to the test strip by capillary action.
5. Calibration:
There is no calibration required for the OneTouch Verio BGMS by the user. The meter is plasma-calibrated.
6. Quality Control:
Glucose control solutions at two different concentrations are supplied with the device. The meter has an algorithm to automatically recognize when a control solution is being analyzed. Control solution test results are stored in the meter, but are not sent to the Apple device. Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Hematocrit Study
Three OneTouch Verio Test Strip lots were tested on 12 Verio Sync meters at each of the following five hematocrit levels: 19, 30, 44, 50 and 61%. At each hematocrit level, each lot was tested with three donor venous blood samples at each of five glucose levels: 30, 66, 118, 233, 455, and 563 mg/dL. A single replicate on each of the 12 meters was obtained for each combination of test
12
{12}
strip/glucose level/hematocrit and donor. This resulted in n = 180 data points per strip lot at each hematocrit level (5 glucose levels x 3 donors x 12 meters). Samples were also tested by the YSI reference method. The results demonstrated that the OneTouch Verio Blood Glucose Monitoring System produces accurate results over the claimed hematocrit range of 20 – 60%.
2. Sample Volume Study
The sponsor performed a sample volume study to support the claimed minimum sample volume requirement for the OneTouch Verio Sync Blood Glucose Monitoring System (0.4 μL) using blood samples at two glucose concentrations (65 and 450 mg/dL). The system displays an error code when insufficient sample is detected. Results support the claimed sample volume of 0.4 μL.
3. Altitude Study
In this study, three (3) test strip lots were tested on sixteen (16) Verio Sync meters using venous blood from three (3) donors (with nominal hematocrit (37-45%)) and adjusting them to 70, 240 and 450 mg/dL. The samples were tested at three altitude levels (3000, 6000 and 10,000 feet), and at sea level (0 feet) as a control. This resulted in n = 288 data points per strip lot at each altitude level (3 donors x 3 glucose levels x 16 meters x 2 replicates). Each blood sample was also tested by the YSI reference method. The bias relative to the reference method was acceptable to support the claim that altitudes up to 10,000 feet do not significantly affect the glucose results.
4. Temperature and Humidity Studies
The sponsor performed temperature and humidity studies using venous blood samples at target glucose concentrations of 65, 240 and 450 mg/dL to evaluate temperatures ranging from 50 – 111°F (10 – 44°C) and relative humidity from 10 – 90 %. Combinations of the claimed temperature and humidity operating conditions were evaluated by comparing meter results to the YSI reference method. The bias relative to the reference method was acceptable to support the claim that temperatures from 50 – 111°F (10 – 44°C) and relative humidity from 10 – 90% do not significantly affect the glucose results.
5. Infection Control Studies
This device is intended for single patient use.
Clorox Germicidal wipes (EPA registration #67619-2) were validated demonstrating complete inactivation of live virus using materials from the meter and lancing device. The sponsor also demonstrated (separately for each product) that there was no change in performance or in the external materials after 310 cleaning and disinfection cycles for the meter and lancing device (one cycle
13
{13}
includes one cleaning wipe plus one disinfecting wipe). This simulated 3 years of use for the meter and lancing device. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
6. EMC testing
Electromagnetic Compatibility (EMC) Testing was evaluated and certified by Elliott Laboratories, Inc. and a letter of attestation dated September 6, 2011 was submitted stating that the device met all of the requirements of the following standards:
IEC 60068-2-64:1993
IEC 61010-1 :2001
IEC 61326-1 :2005, IEC 61326-2-6:2005 and IEC 61010-2-101
7. Readability Assessment
The sponsor provided a readability assessment of the Owner's Booklet and Test Strip Insert, The Flesch-Kincaid analysis produced a grade level of 8.3 and 7.9 respectively.
8. Customer service is available 7 days a week from 8:00 am to 10:00 pm EST by calling 1-888-567-3003. The labeling instructs users to contact their healthcare professional if they cannot reach customer service.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.