FORA D40 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM

K101926 · Taidoc Technology Corporation · NBW · Dec 20, 2011 · Clinical Chemistry

Device Facts

Record IDK101926
Device NameFORA D40 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM
ApplicantTaidoc Technology Corporation
Product CodeNBW · Clinical Chemistry
Decision DateDec 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended to be used by a single person and should not be shared. The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady – state times (when glucose is not changing rapidly). The FORA D40d Test Strips are for use with the FORA D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm. This system contains speaking function, but is not intended for use by the visually impaired. The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use auto-disabling lancing devices. The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, and upper arm should be done only during steady – state times (when glucose is not changing rapidly). The FORA Wisdom D40d Test Strips are for use with the FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm. This system contains speaking function, but is not intended for use by the visually impaired.

Device Story

System combines glucose meter and oscillometric blood pressure monitor. Glucose measurement: electrochemical amperometric assay using glucose oxidase; requires 0.5µL capillary blood; provides results in 5 seconds. Blood pressure measurement: non-invasive oscillometry via arm cuff; provides systolic/diastolic pressure and pulse rate. Used at home (single-patient) or professional settings (multiple-patient). Features include speaking function, USB/Bluetooth data transmission, and automatic temperature compensation via thermistor. Healthcare providers/patients use output to monitor diabetes control and blood pressure trends. Benefits include integrated monitoring, ease of use, and data connectivity for clinical review.

Clinical Evidence

Consumer study performed for blood glucose accuracy at fingertip and alternative sites (palm, forearm, upper arm) compared to YSI-2300 reference. Blood pressure accuracy validated against reference method per ANSI/AAMI SP10 requirements. Device met all ISO 15197 and ANSI/AAMI SP10 performance criteria.

Technological Characteristics

System includes monitor, test strips, control solution, and inflatable cuff. Glucose sensing via electrochemical test strips. Blood pressure via oscillometric method. Complies with IEC/EN 60601-1, IEC/EN 60601-1-2, EN 61326-1, IEC/EN 61010-1, and IEC/EN 61010-2-101. Includes audio output functionality.

Indications for Use

Indicated for quantitative measurement of capillary whole blood glucose and non-invasive blood pressure/pulse rate in adults. Intended for diabetes management aid. Contraindicated for neonates, diabetes diagnosis/screening, patients with dehydration, hypotension, shock, hyperglycemic-hyperosmolar state, or common arrhythmia. Alternative site testing restricted to steady-state conditions.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k101926 B. Purpose for Submission: New device C. Measurand: Capillary whole blood glucose and blood pressure D. Type of Test: Whole blood glucose concentration through a quantitative amperometric assay (Glucose Oxidase) E. Applicant: TaiDoc Technology Corporation F. Proprietary and Established Names: There are two proprietary names for the same device intended for: 1. Single patient use: FORA D40d Blood Glucose plus Blood Pressure Monitoring System 2. Multiple patient use: FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1345 Glucose Test System 21 CFR 870.1130 Noninvasive blood pressure measurement system 2. Classification: Class II (assay) 3. Product code: NBW, Blood Glucose Test System, Over-the-Counter CGA, Glucose Oxidase DXN, System, Measurement, Blood-Pressure, Non-Invasive 4. Panel: Clinical Chemistry (75) and Cardiovascular H. Intended Use: 1. Intended use(s): See indication for use below. {1} 2. Indication(s) for use: 1. FORA D40d Blood Glucose plus Blood Pressure Monitoring System Indications for Use: The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended to be used by a single person and should not be shared. The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady – state times (when glucose is not changing rapidly). The FORA D40d Test Strips are for use with the FORA D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm. This system contains speaking function, but is not intended for use by the visually impaired. 2. FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System Indications for Use: The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use auto-disabling lancing devices. The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, and upper arm should be done only during steady – state times (when glucose is not changing rapidly). The FORA Wisdom D40d Test Strips are for use with the FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. 2 {2} This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm. This system contains speaking function, but is not intended for use by the visually impaired. 3. Special conditions for use statement(s): - For in vitro diagnostic use only - Over-the-Counter and Prescription Use The following limitations apply to the glucose measurement function: - Not intended for use on neonates - Not for the diagnosis of or screening for diabetes mellitus - Not for use on patients who are dehydrated, hypotensive, in shock, or for individuals in hyperglycemic-hyperosmolar state, with or without ketosis. - Critically ill patients should not be tested with a blood glucose meter. - Allows testing on the fingertip, palm, forearm, upper arm, calf, or thigh - Alternative site testing (AST) can be used only during steady-state blood glucose conditions. AST should ONLY be used in the following intervals: - In a pre-meal or fasting state (more than 2 hours since the last meal) - Two hours or more after taking insulin - Two hours or more after exercise - Not intended for use by the visually impaired The following limitations apply to the blood pressure measurement function: - Allows testing only on the arm - Not for use in the presence of common arrhythmia 4. Special instrument requirements: 1. FORA D40d Blood Glucose plus Blood Pressure Meter 2. FORA Wisdom D40d Blood Glucose plus Blood Pressure Meter I. Device Description: This system consists of a glucose meter, arm cuff, user manual, protective wallet, quick start user guide, daily log book, warranty card, and batteries. Control solutions (cleared in k093724) are not included in the kit with the meter and are sold separately. J. Substantial Equivalence Information: 1. Predicate device name(s): TD-3252 Blood Glucose plus Blood Pressure Monitoring System 2. Predicate k number(s): k091555 {3} 4 3. Comparison with predicate: | Item | FORA D40d/FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems | TD-3252 Blood Glucose plus Blood Pressure (k091555) | | --- | --- | --- | | Blood Glucose Test Principle: Similarities and Differences | | | | Intended use | The FORA D40d/ Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: palm, forearm and upper-arm. It is intended for use by people with diabetes mellitus for a single patient at home as an aid in monitoring the effectiveness of diabetes control. The alternative site testing can be used only during steady-state blood glucose conditions. It is not intended for the diagnosis of or screening for diabetes mellitus, hypertension, or for testing on neonates. The systems are also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm. The systems offer cable and wireless communication function which is able to transmit the test result. It contains speaking function but is not intended for use by the | Same | {4} | | visually impaired. | | | --- | --- | --- | | Enzyme | Glucose Oxidase | Same | | Test sample | Whole blood from fingertip, palm, forearm, and upper-arm | Whole blood from fingertip, palm, forearm, upper-arm, calf and thigh | | Measuring range | 20 – 600 mg/dL | Same | | PC connection/Transmission function | USB and Bluetooth | RS232 and Bluetooth | | Reaction time | 5 seconds | 7 seconds | | Code function | No coding | No coding | | Storage/Transportation condition | 39°F – 104°F (4°C - 40°C), below 85% R.H. | Same | | Operation condition | 10°C (50°F) to 40°C (104°F), below 85% R.H. | Same | | Temperature compensation | Automatic compensation with built-in thermister | Same | | Speaking function | Yes | Yes | | Strip ejector | Yes | Yes | | Earphone port | Yes | Yes | | Memory | 864 measurements | 400 measurements | | Measurement mode | General, AC, PC and QC (quality control) | Same | | Special message | Lo/Hi/Ketone | Same | | Blood Pressure Test Principle: Similarities and Differences | | | | Detection method | Oscillometry – non-invasive arm measurement | Same | | Inflation & Deflation | Automatic control | Same | | Maximum inflation pressure | 300 mmHg | 280 mmHg | | Systolic Measurement Range | 55 mmHg – 255 mmHg | 50 mmHg – 250 mmHg | | Diastolic Measurement Range | 25 mmHg – 195 mmHg | 30 mmHg – 180 mmHg | | Pulse rate | 40 -199 per minute | Same | | Measurement mode | Single measurement and triple measurements for obtaining an average value | Single measurement | | Accuracy | Blood pressure: ±3 mmHg or ±2% of reading; Pulse rate: ±4% of reading | Same | # K. Standard/Guidance Document Referenced (if applicable): - ISO 14971:2007. Medical devices-Application of risk management to medical {5} devices. - ISO 15197. In vitro diagnostic test systems. Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. - EN 60601-1-1. Medical electrical equipment, Part 1-1. General requirements for safety. Safety requirements for medical electrical systems. - EN 60601-1-2:2001 (A1:2006). Medical electrical equipment, Part 1-2. General requirements for basic safety and essential performance. Electromagnetic Compatibility. - EN 61326-1:2006. Electrical equipment for measurement, control, and laboratory use. EMC Requirements. General requirements. - IEC/EN 61010-2-101:2002. Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 2-101. Particular requirements for in vitro diagnostic (IVD) medical equipment. - ISO15197:2003 ISO 14971:2000 - Medical devices – Application of risk management L. Test Principle: The FORA D40d/FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring uses electrochemical methodologies. This system quantitatively measures blood glucose levels using an amperometric method, which involves detecting the current produced from glucose oxidation. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The sponsor performed precision studies in accordance with the ISO 15197 and CLSI EP-5A. Venous whole blood adjusted to 5 glucose levels with hematocrit ~45% was used for within-day precision studies. Each concentration was tested 10 times each on 10 meters, using 3 test strip lots, for a total of 300 tests per blood glucose level. Results are summarized below: Within Day precision: | | Level 1 | | | Level 2 | | | Level 3 | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Lot | Average (mg/dL) | SD (mg/dL) | %CV | Average (mg/dL) | SD (mg/dL) | %CV | Average (mg/dL) | SD (mg/dL) | %CV | | 1 | 42.8 | 1.86 | 4.36 | 80.3 | 3.13 | 3.90 | 146.0 | 4.20 | 2.88 | | 2 | 42.6 | 1.56 | 3.67 | 77.7 | 2.99 | 3.85 | 146.8 | 3.54 | 2.41 | | 3 | 42.7 | 1.85 | 4.33 | 78.8 | 3.18 | 4.03 | 148.3 | 2.64 | 1.78 | {6} 7 | | Level 4 | | | Level 5 | | | | --- | --- | --- | --- | --- | --- | --- | | Lot | Average (mg/dL) | SD (mg/dL) | %CV | Average (mg/dL) | SD (mg/dL) | %CV | | 1 | 203.5 | 4.47 | 2.20 | 307.1 | 6.93 | 2.26 | | 2 | 204.4 | 5.75 | 2.81 | 302.7 | 6.59 | 2.18 | | 3 | 203.7 | 3.82 | 1.88 | 307.0 | 7.13 | 2.32 | In addition to the study above, the sponsor also evaluated day-to-day precision using control samples with 3 different levels of glucose. Three lots of test strips and 10 meters were used in the study, with 1 test performed on each meter per day for 10 days, for a total of 100 tests per control level. Results for each test strip lot are summarized below: Between Day precision: | | Level 1 | | | Level 2 | | | Level 3 | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Lot | Average (mg/dL) | SD (mg/dL) | %CV | Average (mg/dL) | SD (mg/dL) | %CV | Average (mg/dL) | SD (mg/dL) | %CV | | 1 | 83.7 | 2.06 | 2.46 | 145.0 | 3.83 | 2.64 | 334.9 | 4.82 | 1.44 | | 2 | 82.5 | 2.75 | 3.33 | 145.4 | 2.98 | 2.05 | 334.9 | 5.76 | 1.72 | | 3 | 82.1 | 2.45 | 2.98 | 144.9 | 2.56 | 1.77 | 334.8 | 5.66 | 1.69 | b. Linearity/assay reportable range: The sponsor performed linearity studies using whole blood samples with 11 different glucose concentration ranges (from $\sim 10$ to $\sim 780\mathrm{mg / dL}$ ). Testing was performed using the FORA D40d/ FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems and the YSI-2300 glucose analyzer (reference method). Three lots of test strips were used in the study. Linear regression analysis for each test strip lot is summarized below: Lot 1 $y = 0.9621x + 5.6476, r^2 = 0.9975$ Lot 2 $y = 0.9514x + 7.1043, r^2 = 0.9973$ Lot 3 $y = 0.9605x + 6.0252, r^2 = 0.9971$ The measuring range of the FORA D40d/ FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is $20 - 600\mathrm{mg / dL}$ c. Traceability, Stability, Expected values (controls, calibrators, or methods): {7} The method comparison was performed using the proposed device and YSI 2300 glucose analyzer. Controls: The 3-level control solutions were originally cleared in k093724. However, values of these Control Solutions specific for use with the FORA D40d/ FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System, were established by repeat analysis using 25 FORA D40d/ FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring devices and 25 corresponding test strips. The mean and standard deviation are used to establish the acceptable range for the strips. Calibration: no calibration is required by the user. ## Stability: Storage Stability Study: The storage claim for this device is: 90 days for open vial of test strips and controls stored at 39 – 104 °F (4 – 40 °C) and 24 months for unopened vials of strips and controls stored at 39 – 104 °F. The study protocols (accelerated testing and real time studies), summary of results and acceptance criteria were reviewed and found to be adequate. ## d. Detection limit: The measuring range of the devices is 20-600 mg/dL. This range was validated via the linearity study (see section M.1.b.). ## e. Analytical specificity: The sponsor performed interference studies with spiked venous blood samples at three glucose concentrations (~70, 120 and 300 mg/dL) that were prepared and divided into a test (dosed) pool and a control pool. The potential interferants were added to the sample, tested at least at two concentrations (therapeutic/normal and above therapeutic/toxic) and each sample was analyzed in duplicate using 4 FORA D40d/ FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems. The table below lists all substances tested at concentrations with insignificant (&lt;10%) interference: | Substance | Concentration with <10% interference (mg/dL) | | --- | --- | | Acetaminophen | 6.25 | | Acetylsalicilic acid | 50 | | Acyclovir | 3.1 | | Allopurinol | 5 | | Amitriptyline | 0.27 | | Amoxicillin | 12.5 | | Ampicilin | 5 | | Ascorbic acid | 5 | {8} | Aspirin | 60 | | --- | --- | | Atenolol | 10 | | Bicarbonate | 40 mM | | Bile acids | 6 | | Bilirubin | 20 | | Caffeine | 10 | | Calcium | 5 mM | | Chloride | 140 mM | | Cholesterol | 500 | | Clonidine | 2 | | Creatinine | 5 | | Digoxin | 0.16 | | Diphenhydramine | 1 | | Dopamine | 1.25 | | Enalapril | 0.15 | | K2EDTA | 180 | | K3EDTA | 175.5 | | Erythromycin | 20 | | Estrone | 0.1 | | Ephedrine HCl | 60 | | Famotidine | 0.13 | | Fluoxetine | 0.8 | | Fructose | 1000 | | Furosemide | 2 | | Galactose | 250 | | Gentisic acid | 2 | | Glutathione | 23 | | Glyburide | 1.07 | | Hemoglobin | 100 | | Heparin (Li) | 1700 U/dL | | Heparin (Na) | 1700 U/dL | | Ibuprofen | 55 | | Lactose | 1000 | | Lactitole | 1000 | | L-dopa | 1.4 | | Lidocaine | 6 | | Magnesium | 5 mM | | Maltitole | 1000 | | Maltose | 1000 | | Mannitol | 1000 | | Mannose | 250 | | Metaproterenol | 1.81 | | Methyl-dopa | 1.875 | | Metoprolol | 0.3 | | Naproxen | 100 | 9 {9} | Nifedipine | 0.17 | | --- | --- | | Nortriptyline | 0.15 | | Penicillin | 12 | | pH | 6.7 – 9.8 | | Phenytoin | 10 | | Piroxicam | 5 | | Potassium | 10 mM | | Pralidoxime Iodide | 5 | | Sodium | 200 mM | | Sorbitol | 1000 | | Sulfamethoxazole | 120 | | Sulfate | 5 mM | | Terfenadine | 0.45 | | Tetracycline | 4 | | Theophylline | 25 | | Tolazamide | 12.5 | | Tolbutamide | 64 | | Total protein | 12000 | | Triglycerides | 2000 | | Urea | 600 | | Uric acid | 10 | | Vancomycin | 25 | | Verapamil | 0.45 | | Vitamin E | 20 | | Warfarin | 2 | | Xylitol | 1000 | | Xylose | 6.25 | f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The sponsor conducted a combined accuracy study and consumer study, with trained operators and 150 samples (136 study subjects/natural patient samples + 14 altered samples) comparing accuracy of the candidate device to the YSI 2300 analyzer. The study subjects ranged in age, education, and were about equally divided between males and females. The blood samples with glucose concentrations ranging from 22 to 581 mg/dL were tested. Samples with glucose concentrations less than 40 and greater than 400 mg/dL were either glycolyzed or spiked and tested by trained operators as well as lay-users. Distribution of blood glucose concentration across the samples tested was in accordance with the ISO15197. The studies met the ISO 15197 accuracy {10} criteria where ninety-five percent (95%) of the individual glucose results shall fall within $\pm 15\mathrm{mg/dL}$ of the results at glucose concentrations $&lt; 75\mathrm{mg/dL}$ and within $\pm 20\%$ at glucose concentrations $\geq 75\mathrm{mg/dL}$. The results relative to YSI are summarized in the tables below: **Trained operators vs. YSI** **Linear Regression Analysis:** Finger: $y = 1.004x + 3.8033$, $r^2 = 0.9849$ Palm: $y = 0.9969x + 1.8077$, $r^2 = 0.9801$ Forearm: $y = 1.004x + 3.6933$, $r^2 = 0.9867$ Upper arm: $y = 0.9835x + 6.2784$, $r^2 = 0.9831$ For samples $&lt; 75\mathrm{mg/dL}$ | | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | | --- | --- | --- | --- | | Fingertip | 20/26 (77%) | 25/26 (96%) | 26/26 (100%) | | Palm | 15/26 (58%) | 25/26 (96%) | 26/26 (100%) | | Forearm | 15/24 (54%) | 23/24 (96%) | 24/24 (100%) | | Upper arm | 16/25 (64%) | 23/25 (96%) | 25/25 (100%) | For samples $\geq 75\mathrm{mg/dL}$ | | Within±5% | Within±10% | Within±15% | Within±20% | | --- | --- | --- | --- | --- | | Fingertip | 64/124 (52%) | 100/124 (81%) | 120/124 (97%) | 124/124 (100%) | | Palm | 57/124 (44%) | 92/124 (73%) | 120/124 (97%) | 124/124 (100%) | | Forearm | 55/126 (44%) | 92/126 (73%) | 119/126 (94%) | 126/126 (100%) | | Upper arm | 48/125 (38%) | 87/125 (70%) | 121/125 (97%) | 125/125 (100%) | **Lay-users vs. YSI** **Linear Regression Analysis:** Finger: $y = 1.0622x + 0.55771$, $r^2 = 0.9789$ Palm: $y = 1.0427x - 4.9914$, $r^2 = 0.9722$ Forearm: $y = 1.0154x + 1.393$, $r^2 = 0.9747$ Upper arm: $y = 1.0306x - 1.956$, $r^2 = 0.9711$ | For samples < 75 mg/dL | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | | --- | --- | --- | --- | | Fingertip | 13/18 (72%) | 17/18 (94%) | 18/18 (100%) | {11} | Palm | 9/18 (50%) | 17/18 (94%) | 18/18 (100%) | | --- | --- | --- | --- | | Forearm | 11/16 (69%) | 15/16 (94%) | 16/16 (100%) | | Upper arm | 8/17 (47%) | 16/17 (94%) | 17/17 (100%) | For samples $\geq 75\mathrm{mg / dL}$ | | Within±5% | Within±10% | Within±15% | Within±20% | | --- | --- | --- | --- | --- | | Fingertip | 50/118 (42%) | 82/118 (69%) | 112/118 (95%) | 118/118 (100%) | | Palm | 38/118 (32%) | 78/118 (66%) | 113/118 (96%) | 118/118 (100%) | | Forearm | 46/120 (38%) | 83/120 (69%) | 114/120 (95%) | 120/120 (100%) | | Upper arm | 38/119 (32%) | 80/119 (67%) | 113/119 (95%) | 119/119 (100%) | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected blood glucose levels for people without diabetes (referenced from the American Diabetes Association, Clinical Practice Recommendations (2010). Diabetes Care, Vol. 33, Supplement 1, p. S1-S100. Time Fasting and before meals Two hours after meals Range (mg/dL) Less than 100 mg/dL Less than 140 mg/dL Range (mmol/L) 5.6mmol/L 7.8 mmol/dL N. Instrument Name: FORA D40d (TD-3216A) Blood Glucose Meter FORA Wisdom D40d (TD-3216B) Blood Glucose Meter {12} 13 O. Systems Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?: Yes ☐ X (reviewed under k070941) or No ☐ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?: Yes ☐ X or No ☐ 2. Software: FDA reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X or No ☐ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger and other alternate sites (palm, forearm, and upper-arm only). Since the whole blood sample is applied directly to the test strip, there are no special handling or storage issues. 5. Calibration: No calibration is required by the user. The meter accommodates auto-coding, in that each strip is designed and manufactured to code the meter appropriately when the strip is inserted. 6. Quality Control: The sponsor has three levels of controls that are not provided with the device, but can be purchased separately. When a test strip is inserted into the meter, each control can be measured by following the instructions for “Performing a Control Solution Test” provided in the User’s Manuals for the meters. An acceptable range for each control level is printed on the test strip vial label. If the test results fall outside the range printed on the test strip vial, the user is instructed to contact the Customer Care Line at 1-866-469-2632 for customer support. The Customer Care service is available 24 hours a day, 7 days a week, 365 days a year. {13} P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: 1. These devices are intended for single-patient use (FORA D40d) and multiple-patient use (FORA Wisdom D40d). Disinfection studies (viral efficacy studies) were performed by an outside service to evaluate the virus elimination effectiveness of disinfecting wipes in preventing the spread of bloodborne pathogens, particularly hepatitis B virus (HBV). Specifically, Micro-Kill+ cleaner disinfectant towels (with EPA registration # 59894-10-37549) were validated demonstrating complete inactivation of live virus. The sponsor does not market a lancing device for use with the device. The sponsor also performed robustness studies and demonstrated that there was no change in performance or in the external materials of the meters after 10,000 cleaning and disinfection cycles designed to simulate 5 years (5 cleaning and disinfection cycles per day) of single-patient device use or 3 years of multiple-patient use with cleaning and disinfecting after each patient up to 9 uses per day. In addition, the sponsor provided a risk analysis specific to the potential harm due to infection. Labeling has been reviewed for adequate instructions in validated cleaning and disinfection procedures. 2. Hematocrit Study: A study to evaluate the effect of hematocrit was conducted on samples with 9 glucose concentrations spanning the range of 20-600 mg/dL at 9 hematocrit levels (20, 25, 30, 35, 40, 45, 50, 55, and 60%). Each glucose level/hematocrit combination was tested in duplicates of 6 using one test strip lot. The results from this study were within ±15% individual bias, supporting the claimed hematocrit range of 20-60%. Results of samples at each hematocrit level were compared to samples with the same glucose concentration at normal (40%) hematocrit as well as to the corresponding YSI value. All individual results for each meter type were ±15% of the YSI, supporting the claimed hematocrit range of 20-60%. 3. Altitude Study: An altitude study was performed in an altitude simulation chamber with whole blood spiked samples spanning the range of ~20 to 600 mg/dL. The device-specific measurements at level 0 ("sea level") were compared to the "high altitude" measurements and to the YSI measurements as well. The study protocol and data were provided, reviewed and found to be adequate. Altitude up to 11500 feet (3500 m) does not affect test results of the FORA D40d/FORA Wisdom D40d Monitoring Systems. 4. Temperature and humidity studies: Temperature and humidity studies were conducted to demonstrate that the FORA D40d/FORA Wisdom D40d Blood Glucose Monitoring Systems can be used at temperatures of 4 to 40°C and at a relative humidity below 85%, and stored at temperatures of 4 to 40°C with a relative humidity below 85%. The FORA D40d test strips are stable for 24 months when stored unopened and 90 days when stored opened at 4 to 40°C with a relative humidity below 85%. The study protocol and data of individual measurements compared to the YSI across the entire measuring range were provided and found to be adequate. 14 {14} 5. Open-vial-in-use test strip studies were performed at the combined extremes of 10±2°C (46.4-56.3°F) and RH: 10 %, 10±2°C RH:85%, 40±2°C (100.4- 107.6°F) RH:10 %, and 40±2°C RH:85% with venous blood samples with concentrations across the measuring range and compared to the YSI. The study protocol, data and ±10% bias of individual measurements across the entire measuring range when compared to YSI were provided and found to be adequate. 6. A study to evaluate the effect of different sample volumes (0.3, 0.4, 0.5, 0.6, and 1.0μl) was conducted to demonstrate that the 0.5 μl volume provides accurate reading. The results from this study for the sample volume of 0.5 μl were within ±10% individual bias across the entire measuring range when FORA D40d/FORA Wisdom D40d measurements were compared to the YSI-2300. 7. Lay-User Study for USB and Bluetooth mediated Data Transmission: This study was performed as a part of the lay-user study described above in 2.a. The 50 study participants, following instructions in User's manual, were able to transmit results from the meter to the PC application (cleared under k070941) via either the USB or Bluetooth features. The study participants ranged in age, education, and were about equally divided between males and females. Those study participants also completed a questionnaire in response to whether the data transmission feature is easy to use. The sponsor concluded that the users' responses indicated that data transmission function was easy to operate by following the User's manuals for both the USB and the Bluetooth mediated data transmission. 8. Lay user questionnaire: 136 lay users evaluated the ease of use of the device and the presentation of the labeling. All users thought that the FORA D40d and FORA Wisdom D40d Monitoring Systems were easy to use and most answer "easy" and "very easy" that the user manuals were written to make it easy to use. 9. The sponsor provided a readability study and obtained Flesch-Kincaid grade level scores of 8 or lower for the User's Manual and test strip insert. 10. Additional electromagnetic interference study was performed to verify these new products. The results of this study complied with standards listed in the Standard/Guidance Document Referenced section above. Electromagnetic Compatibility (EMC) testing was performed/passed and a certificate to Taidoc Technology Corporation was provided. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 15
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...