BASESENS BLOOD GLUCOSE MONITORING SYSTEM

K093475 · Hygeia, Inc. · NBW · Feb 24, 2010 · Clinical Chemistry

Device Facts

Record IDK093475
Device NameBASESENS BLOOD GLUCOSE MONITORING SYSTEM
ApplicantHygeia, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateFeb 24, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The BaseSens I Blood Glucose Monitoring System is used for the quantitative measurement of glucose levels in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. The BaseSens I Blood Glucose Monitoring System should be used only for testing outside the body (in vitro diagnostic use only). The BaseSens I Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. Testing sites include the fingertip along with alternate sites testing (AST) on the forearm, palm, thigh and calf. AST in this system can be used only during steady-state blood glucose conditions. BaseSens I Test Strip is used with the BaseSens I Blood Glucose Meter for quantitatively measuring glucose in capillary whole blood. The BaseSens I Test Strip is intended for self-testing outside the body (in vitro diagnostic use only). The BaseSens I Test Strips are not intended for the diagnosis of or screening for diabetes mellitus, and are not intended for use on neonates. Testing sites include the fingertip along with alternate sites testing (AST) on the forearm, palm, thigh and calf. AST in this system can be used only during steady-state blood glucose conditions. BaseSens I Control A&B Solutions are red liquid to check that both the meters and test strips are working together properly. It contains a known range of glucose as written on the bottle.

Device Story

BaseSens I Blood Glucose Monitoring System is a rebranded i-Sens CareSens N system. Device measures glucose in fresh capillary whole blood samples using test strips. Principle of operation: amperometric method; glucose oxidase chemical reaction; instrument measures current proportional to glucose concentration. Used by patients at home or clinicians in clinical settings. Output: quantitative glucose level displayed on meter. System includes meter, test strips, and control solutions. Provides aid in diabetes management; not for diagnosis or screening. Features automatic code identification, 250-test memory, and post-meal flagging.

Clinical Evidence

No new clinical data provided. Performance characteristics, including precision, linearity, analytical specificity, and method comparison, were established in the predicate submissions (k080923 and k083468).

Technological Characteristics

Amperometric measurement principle using glucose oxidase enzyme and carbon electrodes. Test time: 5 seconds. Sample volume: 0.5 µL. Test range: 20-600 mg/dL. Hematocrit range: 20-60%. Features automatic code identification, 3-button interface, and memory for 250 tests. Powered by battery (implied).

Indications for Use

Indicated for quantitative measurement of glucose in capillary whole blood to monitor diabetes management effectiveness. For home or clinical use. Not for neonates, diagnosis, or screening of diabetes. Testing sites: fingertip, forearm, palm, thigh, calf (AST only during steady-state).

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k093475 B. Purpose for Submission: Rebranding of a cleared device C. Measurand: Whole blood glucose D. Type of Test: Whole blood glucose concentration through a quantitative amperometric assay (Glucose Oxidase) E. Applicant: Hygieia, Inc. F. Proprietary and Established Names: BaseSens I Blood Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1345, Glucose Test System 21 CFR § 862.1660, Quality Control Material (assayed and unassayed) 2. Classification: Class II (assay) and Class I, reserved (controls) 3. Product code: NBW, Blood Glucose Test System, Over-the-Counter CGA, Glucose Oxidase, Glucose JJX, Single (specified) analyte controls (assayed and unassayed) 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See indication for use below. 2. Indication(s) for use: The BaseSens I Blood Glucose Monitoring System is used for the quantitative measurement of glucose levels in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. The BaseSens I Blood Glucose Monitoring System should be used only for testing outside the body (in vitro diagnostic use only). The BaseSens I Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes {1} mellitus, and is not intended for use on neonates. Testing sites include the fingertip along with alternate sites testing (AST) on the forearm, palm, thigh and calf. AST in this system can be used only during steady-state blood glucose conditions. BaseSens I Test Strip is used with the BaseSens I Blood Glucose Meter for quantitatively measuring glucose in capillary whole blood. The BaseSens I Test Strip is intended for self-testing outside the body (in vitro diagnostic use only). The BaseSens I Test Strips are not intended for the diagnosis of or screening for diabetes mellitus, and are not intended for use on neonates. Testing sites include the fingertip along with alternate sites testing (AST) on the forearm, palm, thigh and calf. AST in this system can be used only during steady-state blood glucose conditions. BaseSens I Control A&B Solutions are red liquid to check that both the meters and test strips are working together properly. It contains a known range of glucose as written on the bottle. 3. Special conditions for use statement(s): Not intended for diagnosis or screening of diabetes mellitus - Not intended for use on neonates - For in vitro diagnostic use only - Not for use on critically ill patients, patients in shock, dehydrated patients or hyperosmolar patients - AST in this system can be used only during steady-state blood glucose conditions 4. Special instrument requirements: BaseSens I Blood Glucose Meter I. Device Description: The BaseSens I Blood Glucose Monitoring System (BGMS) is comprised of the BaseSens I Blood Glucose Meter, the BaseSens I Blood Glucose Test Strips (glucose oxidase) and the BaseSens I Control Solutions (2 levels). The BaseSens I BGMS is identical to the CareSens N BGMS (k083468). The BaseSens I System is a rebranded CareSens N System. J. Substantial Equivalence Information: 1. Predicate device name(s): i-Sens CareSens BGMS CareSens N BGMS 2. Predicate K number(s): k080923 and k083468 respectively {2} 3. Comparison with predicate: | Item | Proposed Device | CareSens N BGMS (k083468) | CareSens BGMS (k080923) | | --- | --- | --- | --- | | Similarities | | | | | Intended Use | The BaseSens I BGMS is used for the quantitative measurement of glucose levels in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. The BaseSens I BGMS should be used only for testing outside the body (in vitro diagnostic use only). The BaseSens I BGMS is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. Testing sites include the fingertip along with alternate sites testing (AST) on the forearm, palm, thigh and calf. AST in this system can be used only during steady-state blood glucose conditions. | same | same | | Enzyme | Amperometric method | same | same | | Measurement Principle | Glucose oxidase (A. Niger) | same | same | | Test Principle | Glucose oxidase chemical reaction. The instrument measures the extent of current caused by the presence of glucose in sample. | same | same | | Sample | Fresh capillary whole blood | same | same | | Electrode | Carbon | same | same | | Calibration | Plasma-equivalent | same | same | | Test time (seconds) | 5 | same | same | | Sample volume (μl) | 0.5 | same | same | | Memory | 250 | same | same | | Test range (mg/dL) | 20 to 600 | same | same | | Hematocrit range (%) | 20 to 60 (below 400 mg/dL) | same | same | | Glucose units | mg/dL or mmol/L (default is mg/dL) | same | same | | Alternate site testing | Yes | same | same | | Operating humidity | 10 to 90 % | same | same | | Differences | | | | | Coding | Automatic code identification | same | Manual input | | Self-diagnosis of code identification function | Yes | same | No | {3} | 3 time set alarms and 2 hour post meal alarm | Yes | same | No | | --- | --- | --- | --- | | Post-meal flagging | Yes | same | No | | Number of buttons | 3 | same | 2 (CareSens II) 1 (CareSens POP) | K. Standard/Guidance Document Referenced (if applicable): - ISO 15197:2003, In Vitro Diagnostic Test Systems—Requirements for Blood Glucose Test Systems for Self Managing Diabetes Mellitus. - IEC/EN 61601-1-2, (Second Edition, 2001), Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests. (General) L. Test Principle: The BaseSens I Blood Glucose Monitoring System uses electrochemical methodologies. The system quantitatively measures blood glucose levels using an amperometric method, which involves detecting the current produced from glucose oxidation. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Established in k080923 and k083468 b. Linearity/assay reportable range: Established in k080923 and k083468 c. Traceability, Stability, Expected values (controls, calibrators, or methods): Established in k080923 and k083468 d. Detection limit: Established in k080923 and k083468 e. Analytical specificity: Established in k080923 and k083468 f. Assay cut-off: Not applicable {4} 5 2. Comparison studies: a. Method comparison with predicate device: Established in k080923 and k083468 b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected blood glucose levels for people without diabetes (referenced from Diagnosis of Diabetes, NIH Publication No. 05-4642, January 2005.) | Time | Range (mg/dL) | Range (mmol/L) | | --- | --- | --- | | Fasting | 70 to 110 | 3.9 to 6.1 | | One hour after meals | less than 160 | less than 8.9 | N. Instrument Name: BaseSens I Blood Glucose Meter O. Systems Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?: Yes ☐ or No ☑ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?: Yes ☐ or No ☑ {5} 2. Software: Reviewed under k080923 and k083468 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger, which can be applied directly to the test strip. 5. Calibration: The device has automatic code identification. No further calibration is required. 6. Quality Control: The sponsor has two levels of controls available for the system (not included in the kit but available for purchase). When a test strip is inserted into the meter, each control can be measured by following the instructions for “Checking the System” provided in the Owner’s Booklet of the system. An acceptable range for each control level is printed on the test strip vial label. The user is instructed to contact the Customer Help Line during the operational times or a healthcare provider outside the operational times if the control results fall outside these ranges. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above: Not applicable Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...