← Product Code [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW) · K091296

# POLYTEL GMA, MODELS PWR-08-67, PWR-08-07, PWR-08-09 (K091296)

_Polymap Wireless · NBW · Jul 17, 2009 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K091296

## Device Facts

- **Applicant:** Polymap Wireless
- **Product Code:** [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW.md)
- **Decision Date:** Jul 17, 2009
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Polytel GMA (models PWR-08-06, PWR-08-07, and PWR-08-09) are remote communications link devices intended to be used to wirelessly transmit glucose meter readings from a compatible Blood Glucose Monitor to a compatible cellular phone, such as the Nokia N73 or access point, such as the Polytel® APT. The receiving device, in turn, sends the readings to the healthcare provider in another location. The device does not send any real-time alarms. Clinical judgment and experience are required to check and interpret the information delivered.

## Device Story

Polytel GMA is a battery-powered telemedicine accessory; connects to blood glucose meters via phono jack; transmits data wirelessly using Bluetooth v2.0 (2.402-2.480 GHz) to compatible access points (e.g., Nokia N73, Polytel APT). Device acts as a communication bridge; sends glucose readings to remote healthcare providers. Does not provide real-time alarms; requires clinical interpretation of transmitted data. Used in home or clinical settings to facilitate remote monitoring of glucose levels.

## Clinical Evidence

No clinical data. Bench testing only. User study previously performed under k070559; no new user study required as modifications do not impact user functions. Reading level analysis for manuals confirmed below 8th grade level.

## Technological Characteristics

Battery-powered; Bluetooth v2.0 wireless transmission; frequency band 2.402 to 2.480 GHz; phono jack interface for glucose meter connection.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- eSan, Ltd. Think Plus (t+) diabetes management system (k061328)
- Polymap Wireless LLC Polymap PWR-08-03 (k070559)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
k091296

B. Purpose for Submission:
This submission is for a modification to a previously cleared device (k070559). This glucose meter accessory is intended to transmit data to compatible access points such as a personal computer or cellular phone with Bluetooth capability.

C. Manufacturer and Instrument Name:
Manufacturer: Polymap Wireless, LLC.
Instrument Names: Polytel GMA models PWR-08-06, PWR-08-07, and PWR-08-09

D. Type of Test or Tests Performed:
Not applicable.

E. System Descriptions:

1. Device Description:
The Polytel PWR-08-06 GMA is a battery-powered, plug-in accessory to the Lifescan OneTouch Ultra, OneTouch Ultra 2, OneTouch UltraMini, and OneTouch Select Blood Glucose Monitors. The Polytel PWR-08-07 GMA is a battery-powered, plug-in accessory to the Bayer Contour Blood Glucose Monitors. The Polytel PWR-08-09 GMA is a battery-powered, plug-in accessory to the Abbott FreeStyle Freedom Lite and FreeStyle Lite Blood Glucose Monitors. Each of these devices acts as a remote unit to transmit data from the meter to a compatible access point, such as a Nokia Model N73 cell phone. The GMA plugs into a port on the glucose meter. In the case of the PWR-08-06, when a glucose reading is taken and the strip withdrawn from the meter, unit turns on automatically, accepts the data, and transmits it wirelessly to the compatible Access Point. The Access Point then sends the encrypted data via cell or land telephone lines to a computer server. Due to the electrical design of the Bayer and Abbott glucose meters, both the PWR-08-07 and PWR-08-09 require the user pressing a button to initiate data transmission.

The access point can be a compatible cell phone, personal computer, or other device having Bluetooth capability. This allows glucose readings to be monitored from a remote location. The glucose readings transmitted by the Polytel glucose accessory are intended to be used for monitoring and historical trending. Clinical judgment and experience are required to check and interpret the information delivered. The unit does not transmit any alarms.

The device hardware platform is a Cambridge Silicon Radio BlueCore4-Ext radio chip with 1MB of flash memory.

{1}

The GMA will be sold both through approved groups who can offer service support and OTC. It will be advertised to physicians, in a variety of publications, and on the Polymap website which will direct interested users to contact one of the approved support groups that Polymap has identified. Users may contact Polymap directly, or go to the Polymap website. A support technician from one of the approved groups will contact the user, answer questions about the GMA, validate that it works properly with the user's particular access point, and validate accurate data transmission to the monitoring site.

2. Principles of Operation:
Not applicable.

3. Modes of Operation:
Transmits data over a wireless Bluetooth link from a Blood Glucose Monitor to a compatible server.

4. Specimen Identification:
Glucose meter generated – unique unit ID

5. Specimen Sampling and Handling:
Not applicable.

6. Calibration:
Not applicable.

7. Quality Control:
Not applicable.

8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐

F. Regulatory Information:

1. Regulation section:
21 CFR § 862.1345

2. Classification:
Class II

3. Product code:
NBW

4. Panel:
Chemistry (75)

{2}

G. Intended Use:

1. Indication(s) for Use:
The Polytel GMA (models PWR-08-06, PWR-08-07, and PWR-08-09) are remote communications link devices intended to be used to wirelessly transmit glucose meter readings from a compatible Blood Glucose Monitor to a compatible cellular phone, such as the Nokia N73 or access point, such as the Polytel APT. The receiving device, in turn, sends the readings to the healthcare provider in another location. The device does not send any real-time alarms. Clinical judgment and experience are required to check and interpret the information delivered.

2. Special Conditions for Use Statement(s):
For prescription and over-the-counter use.

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:
k061328 – eSan, Ltd. Think Plus (t+) diabetes management system
k070559 – Polymap Wireless LLC Polymap PWR-08-03

2. Comparison with Predicate Device:

|  Similarities  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | eSan Predicate | PWR-08-03 Predicate  |
|  Users | Patient & Healthcare provider | Patient & Healthcare provider | Patient & Healthcare provider  |
|  Site of use | In the home | In the home | In the home  |
|  System description | Transmitting device that sends data from a glucose meter to an access point | Transmitting device that sends data from a glucose meter to an access point | Transmitting device that sends data from a glucose meter to an access point  |
|  Connection | Wired connection between glucometer and device. Wireless connection between glucometer and access point | Wired connection between glucometer and device. Wireless connection between glucometer and access point | Wired connection between glucometer and device. Wireless connection between glucometer and access point  |
|  Monitoring Devices | Blood Glucose Meter | Blood Glucose Meter | Blood Glucose Meter  |
|  Power | Battery | Battery | Battery  |
|  Alarms | None | None | None  |

{3}

|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Item | Device | eSan Predicate | PWR-08-03 Predicate  |
|  Intended Use/indications | Transmit data between a monitoring device and enabled cell phone or other compatible device | Transmit data between monitoring devices and enabled cell phone | Transmit data between a monitoring device and enabled cell phone or other compatible device  |
|  Prescription/OTC | Prescription/OTC | OTC | Prescription Sold through approved groups who set up the system  |
|  Data Transmission | Residential telephone lines or cellular connections | Cellular Connections | Residential telephone lines or cellular connections  |
|  Patient interactions | PWR-08-06: None PWR-08-07, PWR-08-09: press button to start transmission | Press button to start the data transmission | None  |
|  Wireless link | Class II Bluetooth radio | Class II Bluetooth radio | Class I Bluetooth radio  |
|  Storage | Fits inside original zippered case | Not addressed | Fits inside original zippered case  |
|  Range | 10 meters | 10 Meters | 100 Meters  |

I. Special Control/Guidance Document Referenced (if applicable):
- IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety; Safety Requirements for Medical Electrical Systems, 2004
- IEC 6060 1-1-2, Medical Electrical Equipment - Part 1-2: General Requirements for Safety; Electromagnetic Compatibility -- Requirements and Tests, 2005
- ISO 14971:2000, Medical devices - Application of risk management to medical devices

J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy: Not applicable.
b. Precision/Reproducibility: Not applicable.

{4}

c. Linearity: Not applicable.
d. Carryover: Not applicable.
e. Interfering Substances: Not applicable.

2. Other Supportive Instrument Performance Data Not Covered Above:

A user study was performed and found to be acceptable under k070559. A new user study is not required under this submission because the modifications made to the device are not expected to affect user functions.

Reading level analyses were performed for the user manuals. All are below an 8th grade level.

K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K091296](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K091296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
