← Product Code [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW) · K090629

# EASY STEP AND DIACHEX+ PRO BLOOD GLUCOSE MONITORING SYSTEMS (K090629)

_Tyson Bioresearch, Inc. · NBW · Jun 4, 2009 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K090629

## Device Facts

- **Applicant:** Tyson Bioresearch, Inc.
- **Product Code:** [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW.md)
- **Decision Date:** Jun 4, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The Easy Step / DIACHEX PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.

## Device Story

Blood glucose monitoring system; measures glucose in capillary whole blood samples. Modifications include: addition of backlight; graphic indicator for pre/post-meal results; aesthetic changes; removal of hypo/hyperglycemic alarms; conversion to no-coding technology; automatic PC link mode activation upon USB connection. Used by patients at home for diabetes management. Output displayed on meter screen; used by patients to monitor glucose levels and adjust diabetes control programs. Data transferable to PC via USB.

## Clinical Evidence

Clinical accuracy assessed in 129 participants (male/female) with glucose range 33.2–577 mg/dL and hematocrit 34%–53%. Results compared to YSI laboratory method. ISO 15197 compliance achieved: 98.4%–99.2% of fingertip results within ±15 mg/dL (<75 mg/dL) or ±20% (≥75 mg/dL). Bench testing included precision, linearity, and interference studies.

## Technological Characteristics

Electrochemical biosensor; glucose oxidase enzyme; potassium ferricyanide mediator. Measures current generated by electron release. Range: 20–600 mg/dL. Connectivity: USB for PC link. Form factor: handheld meter with LCD. Power: battery-operated. Sterilization: not specified.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Diachex Determine and Diachex Infinity blood glucose monitoring system (k073492)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification

ODE Review Memorandum (Decision Making Document is Attached)

To: THE FILE

RE: DOCUMENT NUMBER k090629

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):

1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k073492 Diachex Determine and Diachex Infinity blood glucose monitoring system.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for the addition of a backlight feature, graphic indicator for pre/post meal results, changes to the appearance of the meters, removal of hypoglycemic and hyperglycemic alarm functionalities, and conversion to no-coding. Also, the meters now enter the PC link mode automatically when the USB cable is plugged in, compared to the predicate in which the user had to manually enter the PC link mode.

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics.

5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:

i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.

6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K090629](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K090629)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
