CARESENS N BLOOD GLUCOSE MONITORING SYSTEM, CARESENS N METER, MODEL GM5051A OR GM5051B OR GM5051C (COLOR)
Device Facts
| Record ID | K083468 |
|---|---|
| Device Name | CARESENS N BLOOD GLUCOSE MONITORING SYSTEM, CARESENS N METER, MODEL GM5051A OR GM5051B OR GM5051C (COLOR) |
| Applicant | I-Sens, Inc. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Feb 10, 2009 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
Indications for Use
The CareSens N Meter is used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens N Blood Glucose System should be used only for testing outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis or screening of diabetes. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf. CareSens N Test Strip is used with the CareSens N Blood Glucose Meter for quantitatively measuring glucose in capillary whole blood. The CareSens N Test Strip is intended for selftesting outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis of diabetes without the guidance of healthcare professional. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf. CareSens Control A&B Solutions are a red liquid to check that both the meters and test strips are working together properly. It contains a known range of glucose as written on the bottle.
Device Story
CareSens N BGMS is an in vitro diagnostic system measuring glucose concentration in capillary whole blood. Input: capillary blood sample applied to CareSens N test strip. Principle: amperometric method using glucose oxidase enzyme; meter detects current proportional to glucose concentration. Key modification: automatic code identification via infrared sensors detecting patterns on test strips, eliminating manual coding. Features: 3-button interface, time-set alarms, 2-hour post-meal alarm, post-meal flagging. Used at home or point-of-care by patients or clinicians. Output: digital glucose reading on meter display. Benefits: simplifies user workflow, reduces errors from manual coding, aids diabetes management monitoring.
Clinical Evidence
Bench testing only; verification and validation activities performed per design control requirements (21 CFR 820.30) to confirm modifications met predetermined acceptance criteria.
Technological Characteristics
Blood glucose monitoring system; auto-code test strip calibration; three-button user interface; reminder alarm functionality; post-meal flagging capability. Fundamental scientific technology consistent with predicate K080923.
Indications for Use
Indicated for quantitative glucose measurement in capillary whole blood to monitor diabetes management effectiveness. For use by patients (home) or clinicians (clinical settings). Testing sites include fingertip, forearm, palm, thigh, and calf. Not for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.