← Product Code [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW) · K080538

# IDEAL LIFE POD (MODEL ILP 0001) (K080538)

_Ideal Life, Inc. · NBW · Jul 3, 2008 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K080538

## Device Facts

- **Applicant:** Ideal Life, Inc.
- **Product Code:** [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW.md)
- **Decision Date:** Jul 3, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The IDEAL LIFE Pod™ receives data wirelessly from IDEAL LIFE devices to transmit over the Internet or common telephone lines from the user's home. The Pod is an optional accessory to IDEAL LIFE devices, including the GlucoManager™, the BP-Manager™ and/or the Body Manager™. The Pod is intended to aid people at home and health care professionals to review and evaluate historical blood glucose, weight and blood pressure test results, to support effective health care management. The IDEAL LIFE Pod makes no interpretation, evaluation, medical judgment or recommendations for treatment. This device is not intended as a substitute for medical care.

## Device Story

Wireless modem accessory; receives physiological data (blood glucose, blood pressure, weight) from compatible IDEAL LIFE devices via Bluetooth; transmits data to secure website via telephone line (RJ11) or Internet (RJ45/Ethernet). Used in home settings by patients; data reviewed by patients and healthcare professionals via secure Internet access. Facilitates remote monitoring and historical data tracking; does not provide medical interpretation or treatment advice. Improves health management by enabling remote data transmission for clinical review.

## Clinical Evidence

No clinical data; bench testing only. Usability study conducted with 20 subjects (ages 18-70+) to evaluate setup and transmission success via RJ11 and RJ45 interfaces. All subjects successfully transmitted preloaded data and viewed results on the website. User feedback was positive with average usability scores below 2 on a 1-10 scale (1 being most favorable).

## Technological Characteristics

Wireless modem; Bluetooth (2400-2483.5 MHz); RJ11 and RJ45 connectivity. Power: AC adapter. Dimensions: 120mm x 105mm x 30mm; Weight: 142g. Operating temp: 10°C to 40°C. Standards: ISO 15197, ANSI/AAMI/ISO 14971, ANSI C63.4.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- In Touch Diabetes Management Software ([K984527](/device/K984527.md))
- IDEAL LIFE Pod, IDEAL LIFE INC. ([K060504](/device/K060504.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k080538

B. Purpose for Submission:
New submission for the addition of whole blood glucose data acquisition and transmission to previously cleared blood pressure data acquisition and transmission device (k060504).

C. Measurand:
Whole blood glucose

D. Type of Test:
The IDEAL LIFE Pod is a wireless modem accessory that wirelessly receives data from IDEAL LIFE devices and transmits the data to the IDEAL LIFE website. The IDEAL LIFE database is updated securely with the user's information, which is available for review via secure Internet access. Using the Pod, an individual can more easily view and track their blood pressure (cleared under k060504), blood glucose, and/or body weight (510(k) exempt) information.

E. Applicant:
Ideal Life Inc.

F. Proprietary and Established Names:
Ideal Life POD (Model ILP 0001)

G. Regulatory Information:

1. Regulation section:
21CFR Sec.-862.1345-Glucose test system
21CFR Sec.-862.2100 - Calculator/data processing module for clinical use

2. Classification:
Class 2

3. Product code:
NBW - System, Test, Blood Glucose, Over the Counter
JQP-Calculator/Data Processing Module, For Clinical Use

4. Panel:
Chemistry (75)

H. Intended Use:

1. Intended use(s):

{1}

See indication(s) for use below.

2. **Indication(s) for use:**
The IDEAL LIFE Pod™ receives data wirelessly from IDEAL LIFE devices to transmit over the Internet or common telephone lines from the user's home. The Pod is an optional accessory to IDEAL LIFE devices, including the GlucoManager™, the BP-Manager™ and/or the Body Manager™. The Pod is intended to aid people at home and health care professionals to review and evaluate historical blood glucose, weight and blood pressure test results, to support effective health care management.

The IDEAL LIFE Pod makes no interpretation, evaluation, medical judgment or recommendations for treatment. This device is not intended as a substitute for medical care.

3. **Special conditions for use statement(s):**
For over the counter use

4. **Special instrument requirements:**
IDEAL LIFE GlucoManager (k080283) glucose meter

I. **Device Description:**
The IDEAL LIFE Pod is used to upload the data stored in the memory of various IDEAL LIFE devices to the IDEAL LIFE website via a common telephone line or via the Internet. Two modes of transmission of data are possible: (1) via the telephone line which connects to an RJ11 jack on the IDEAL LIFE Pod, or (2) via the Internet through the Ethernet cable attached to the RJ45 jack on the IDEAL LIFE Pod.

J. **Substantial Equivalence Information:**
1. **Predicate device name(s):**
IN TOUCH Diabetes Management Software System
Ideal Life POD (Model ILP 0001A Version 1)
2. **Predicate 510(k) number(s):**
k984527 and k060504
3. **Comparison with predicate:**

|   | Predicate 1
K984527, IN TOUCH
Diabetes Management
Software System | New Device
IDEAL LIFE POD
(Model ILP 0001A
Version 2) | EQUIVALENT
OR
Different
Impact on S&E  |
| --- | --- | --- | --- |
|  Indication for
Use | The IN TOUCH Diabetes
Management Software System
is intended for use in home
and clinical settings to aid
people with diabetes and
health care professionals in the | The IDEAL LIFE Pod™
receives data wirelessly from
IDEAL LIFE devices to
transmit over the Internet or
common telephone lines from
the user's home. The Pod is an | Same as Predicate
1,  |

2 of 10

{2}

|   | review, analysis, and evaluation of historical blood glucose test results to support effective diabetes management. It is intended for use as an accessory to LifeScan Brand blood glucose monitoring systems with data management capabilities. | optional accessory to IDEAL LIFE devices, including the GlucoManager™, the BP-Manager™ and/or the Body Manager™. The Pod is intended to aid people at home and health care professionals to review and evaluate historical blood glucose, weight and blood pressure test results, to support effective health care management.

The IDEAL LIFE Pod makes no interpretation, evaluation, medical judgment or recommendations for treatment. This device is not intended as a substitute for medical care. |   |
| --- | --- | --- | --- |
|   | **Predicate 1**
**K984527, IN TOUCH**
**Diabetes Management Software System** | **New Device**
**IDEAL LIFE POD**
**(Model ILP 0001A Version 2)** | **EQUIVALENT OR DIFFERENT**  |
|  Intended Use | Blood Glucose Analyzer | Blood Glucose Analyzer and Physiological Signal Transmitter | Equivalent  |
|  Users | General adult population | General adult population | Equivalent  |
|  Prescription Status;
Environment of Use | Non-prescription
Use in a clinical or non-clinical environment | Non-prescription
Use in a clinical or non-clinical environment | Equivalent  |
|  Part of a System | Use with LifeScan Brand blood glucose monitors | Use with IDEAL LIFE BP-Manager™, Gluco-Manager™, Body Manager™ | Equivalent  |
|  Purpose of the System | Transmission of information, access to Internet-based database and personal data management system, ability to view data and graphical analysis of the data, enter personal information and reminders (diet and exercise, medications, notes), enter alert levels. | Same | Equivalent  |
|   | **Predicate 1** | **New Device** | **EQUIVALENT**  |

3 of 10

{3}

|   | K984527, IN TOUCH Diabetes Management Software System | IDEAL LIFE POD (Model ILP 0001A Version 2) | OR DIFFERENT  |
| --- | --- | --- | --- |
|  Display | N/A | Same as previous Pod | different from Predicate 1  |
|  Automatic | Automatic after connection | Same | Equivalent  |
|  Communication method from device | Hard wire connection of blood glucose monitor directly to personal computer | Same | different from Predicate 1  |
|  Power Source | N/A | Same |   |
|  Operating Temp/Humidity |  | Same |   |
|  Storage Temp/Humidity |  | Same |   |
|  Console Wt. |  | Same |   |
|  Size |  | Same |   |
|  Connectors |  | Same
RJ45 jack for Ethernet cable connection for data transmission | Different – RJ11 new adapter  |
|   | Predicate 2
K060504 IDEAL LIFE POD (Model ILP 0001A Version 1) | New Device
IDEAL LIFE POD (Model ILP 0001A Version 2) | EQUIVALENT
OR DIFFERENT
Impact on S&E  |
|  Indication for Use | The IDEAL LIFE POD™ is an optional accessory to the IDEAL LIFE BP-MANAGER™ and is a transmitter that is intended to transmit data obtained by the IDEAL LIFE BP-MANAGER™ to the internet via common telephone lines from the user's home setting. The IDEAL LIFE POD™ is an optional accessory designed to assist in the management of user information. This device makes no interpretation, evaluation, medical judgments or recommendations for treatment. Clinical judgment and experience are required to check or interpret information transmitted. This device is | The IDEAL LIFE Pod™ receives data wirelessly from IDEAL LIFE devices to transmit over the Internet or common telephone lines from the user's home. The Pod is an optional accessory to IDEAL LIFE devices, including the GlucoManager™, the BP-Manager™ and/or the Body Manager™. The Pod is intended to aid people at home and health care professionals to review and evaluate historical blood glucose, weight and blood pressure test results, to support effective health care management.

The IDEAL LIFE Pod makes no interpretation, evaluation, medical judgment or recommendations for treatment. | Same as Predicate 1,
Different (Expanded) from Predicate 2)

Does this impact S&E? No, because testing has demonstrated that new device meets Intended Use.  |

4 of 10

{4}

|   | not intended as a substitute for medical care. | This device is not intended as a substitute for medical care. |   |
| --- | --- | --- | --- |
|   | Predicate 2 K060504 IDEAL LIFE POD (Model ILP 0001A Version 1) | New Device IDEAL LIFE POD (Model ILP 0001A Version 2) | EQUIVALENT OR DIFFERENT  |
|  Intended Use | Physiological Signal Transmitter | Blood Glucose Analyzer and Physiological Signal Transmitter | Equivalent  |
|  Users | General adult population | General adult population | Equivalent  |
|  Prescription Status; Environment of Use | Non-prescription Use in a non-clinical environment | Non-prescription Use in a clinical or non-clinical environment | Equivalent  |
|  Part of a System | Use with IDEAL LIFE BP-Manager™ | Use with IDEAL LIFE BP-Manager™, Gluco-Manager™, Body Manager™ | Equivalent  |
|  Purpose of the System | Same | Same | Equivalent  |
|   | Predicate 2 K060504, IDEAL LIFE POD (Model ILP 0001A Version 1) | New Device IDEAL LIFE POD (Model ILP 0001A Version 2) | EQUIVALENT OR DIFFERENT  |
|  Display | 3-LED's | Same as previous Pod | Equivalent to Predicate 2, different from Predicate 1  |
|  Automatic | Automatic after connection | Same | Equivalent  |
|  Communication method from device | Wireless RF protocol | Same | Equivalent to Predicate 2, different from Predicate 1  |
|  Power Source | AC Adapter or AA batteries | Same  |   |
|  Operating Temp/Humidity | 10°C to 40°C/below 85% RH | Same  |   |
|  Storage Temp/Humidity | -5°C to 50°C/below 90% RH | Same  |   |
|  Console Wt. | 142 g. without batteries | Same  |   |
|  Size | 120mm W x 105mm D x 30mm H | Same  |   |
|  Connectors | AC adapter jack for electrical power RJ11 jack (telephone line transmission) | Same Same RJ45 jack for Ethernet cable connection for data transmission  |   |
|   |  |  | Different – RJ11 new adapter  |

5 of 10

{5}

K. Standard/Guidance Document Referenced (if applicable):
- ISO 15197; In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, 2003
- ANSI/AAMI/ ISO 14971, Medical Devices – Application of risk management to medical devices, 2007.
- ANSI C63.4; Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 kHz to 40 GHz, 2003.

L. Test Principle:
The IDEAL LIFE Pod accessory to a compatible glucose meter the IDEAL LIFE Gluco-Manager Model GMM 0001, which uses specific test principles, the subject of k080283.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility: see above associated device
b. Linearity/assay reportable range: see above associated device
c. Traceability, Stability, Expected values (controls, calibrators, or methods): see above associated device
d. Detection limit: see above associated device
e. Analytical specificity: see above associated device
f. Assay cut-off: see above associated device

2. Comparison studies:
a. Method comparison with predicate device: see above associated device
b. Matrix comparison: Not Applicable

3. Clinical studies:
a. Clinical Sensitivity: Not Applicable
b. Clinical specificity: Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable): User testing of the POD was conducted to determine:
1. If the POD and Gluco-Manager are easy to use and function as planned when the user is provided only the instructions and materials that would be available if a person were to purchase the devices; and

6 of 10

{6}

2. To verify that test data transmitted with (a) the POD connected via the RJ-11 telephone jack, and (b) the POD connected via the RJ-45 Ethernet cable are user friendly and provide the expected Gluco-Manager output on the Internet.

The goal was to select subjects who have never used these devices, who have varying educational levels, and who range in age from 18 to 70+ years. Twenty (20) subjects were enrolled in this study.

Inclusion Criteria:
1. Male and female patients aged 18 to 70+ years.
2. Subjects must be able to read English and must be physically able to conduct the test by themselves.
3. Subjects must have some familiarity with computers, as the use of the two devices with a computer and Internet-based system is central to the study.
4. Subjects must agree to sign the Informed Consent document.
5. Subjects will not be paid for their participation in this study.

Exclusion Criteria:
1. Subjects who have used either of these IDEAL LIFE devices before will be excluded from the trial.

Evaluation
Each Study Subject was provided with the following:
- One POD
- One Gluco-Manager
- One Ethernet cable
- One telephone cable
- Pod and Gluco-Manager Instruction Manuals
- Written Instructions
- Survey parts 1 and 2

Study Subjects were instructed to follow the Instruction Manual(s) to set up the Gluco-Manager and Pod. The Instruction Manuals provide a toll-free number to register the Pod and to receive a user name and password. If the Study Subject asked questions of IDEAL LIFE staff at the time that the toll-free telephone number is called for product registration, questions will be allowed as this will be standard procedure during actual device use.

Seven data sets are preloaded in the Gluco-Manager and the subjects transmitted preloaded data using the Modem/RJ-11 jack setup and then view their data on the Internet. After this process was completed, Part 2 of

7 of 10

{7}

the evaluation was completed to evaluate the usability of the devices and the accuracy of the data transmitted. At the end of the user questions, the subjects were directed to repeat this process using the Ethernet connection. Subjects transmitted preloaded data using the Ethernet/RJ-45 jack setup and then viewed their data on the Internet. After this process was completed, Part 3 of the evaluation was completed to evaluate the usability of the devices and the accuracy of the data transmitted.

Descriptive statistics were used to analyze the mean and standard deviation of VAS scores, as well as the coefficient of variation (SD/Mean). Comments of study subjects were analyzed to determine if changes to the Instruction Manual or the device are required or would be beneficial. A log of telephone calls and questions was maintained for the same purpose. Adverse events, if any, were processed according to IDEAL LIFE MDR Reporting Procedure 902P.

Applicant's Analysis and Conclusions

The Pod is a "Plug-N-Play" device that is, by design, simple to set up and to use. Validation and Verification testing conducted by IDEAL LIFE engineers during development and prior to finalizing the software and hardware design, including this Usability Test and one smaller previous test. The user test was designed to test the usability of the Pod with the Gluco-Manager for naive users, to reflect initial operation by a new purchaser of the devices and identify potential design flaws in actual use situations.

Average scores for all questions were below "2" on a scale of 1 to 10, 1 being the most favorable response. All open comments were positive.

In conclusion, the positive response to this user test and a previous usability test indicates that there are no areas of concern about the usability of the device by the intended users.

4. Clinical cut-off: Not Applicable
5. Expected values/Reference range: See above associated device

N. Instrument Name:

Ideal Life POD (Model ILP 0001)

O. System Descriptions:

8 of 10

{8}

9 of 10

1. **Modes of Operation:**

The device is for data transmission the communication of the Pod with compatible IDEAL LIFE devices is shown using wireless Bluetooth technology. After the data from these devices is transmitted to the Pod, the Pod acts like a modem to communicate through the user's telephone line cord (using the LINE connector, white in the photo below) or over the Internet (using the Ethernet connector via an RJ45 connector, red in the photo below).

|  Model | IDEAL LIFE Model ILP 0001, Version 2  |
| --- | --- |
|  Display | 3 LED's  |
|  Power Source | AC adapter  |
|  Operating Temperature/Humidity | 10°C to 40°C / <85% relative humidity  |
|  Storage Temperature/Humidity | -5°C to 50°C / <90% relative humidity  |
|  Console Weight | Approximately 142 g without batteries  |
|  Outer Dimensions | Approximately 120mm W × 105mm D × 30mm H  |
|  Connectors | AC adapter jack for electrical power; USB port  |
|   |  RJ11 jack marked "line" for telephone line transmission of data to IDEAL LIFE  |
|   |  RJ11 jack marked "phone" for connection of user's telephone  |
|   |  RJ45 jack for Ethernet cable connection for transmission of data over the Internet  |
|  Frequency Range | 2400 to 2483.5 MHz  |
|  Channel Band Width (99%) | 840 kHz  |
|  Conducted Output Power | 1.28 dBm  |

2. **Software:**

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☑ or No ☐

3. **Specimen Identification:**

Glucose meter controlled, by time and date stamp

4. **Specimen Sampling and Handling:**

Not Applicable

{9}

5. Calibration: Not Applicable
6. Quality Control: Not Applicable

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The applicant performed the following two sets of test scripts:

**Dial up connection**
This test script was to ensure that “Gluco-Manager” (GMM) readings are correct and can be displayed on the Ideal life Website. POD-2N dial up connection was used to transmit readings to the Ideal Life DB. This Test Case was a combination of Hardware Test for the device “Gluco-Manager” (GMM) and System Test for the Web-Based Application.
All executed Test Cases have the status - PASS

**Internet connection**
This test script was to ensure that “Gluco-Manager” (GMM) Readings are correct and can be displayed on the Ideal life Website. POD-2N internet connection was used to transmit readings to the Ideal Life DB. The Test Case was a combination of Hardware Test for the device “Gluco-Manager” (GMM) and System Test for the Web-Based Application.
All executed Test Cases have the status - PASS

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

10 of 10

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K080538](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K080538)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
