K070593 · Home Diagnostics, Inc. · NBW · May 7, 2007 · Clinical Chemistry
Device Facts
Record ID
K070593
Device Name
TRACKRECORD DATA MANAGEMENT SOFTWARE
Applicant
Home Diagnostics, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
May 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
TrackRecord Data Management Software is intended for use in the home or in clinical settings, for single or multi-patient use, to assist people with diabetes as well as their healthcare professionals in the review, analysis and evaluation of historical blood glucose test results to support effective diabetes management.
Device Story
TrackRecord Data Management Software functions as a data management tool for diabetes care. It accepts historical blood glucose test results as input; processes and organizes this data to facilitate review and evaluation; and produces reports or displays for patients and healthcare professionals. Used in home or clinical settings by patients or clinicians. Output assists in diabetes management decision-making by providing a structured view of historical glucose trends.
Clinical Evidence
No clinical diagnostic data; bench testing only. A Human Factors Study with 101 subjects (home users and healthcare professionals) evaluated design, ease-of-use, and instruction comprehension. Subjects performed 21 feature/function tests across five categories (Installation, UI, Setup, Downloading, Patient Management). Performance was rated by observers and reviewed by an expert panel. Results confirmed users could successfully install the software, upload data, and generate accurate reports.
Technological Characteristics
Software-based data management system for blood glucose test results. Operates as a standalone application for review and analysis of historical data. No specific hardware materials or energy sources defined as it is a software-only device.
Indications for Use
Indicated for people with diabetes and their healthcare professionals to review, analyze, and evaluate historical blood glucose test results for diabetes management. Suitable for home or clinical settings, single or multi-patient use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Submission Summary (Full Text)
{0}
1 of 10
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k070593
B. Purpose for Submission:
New submission for an accessory data management software application for glucose meters
C. Measurand:
Glucose
D. Type of Test:
An accessory device, the TrackRecord DMS, for diabetes management, is a stand-alone optional data management software accessory supplied in CD format for use with Home Diagnostics brand Smart System blood glucose meters such as the TrueTrack and Prestige IQ Smart System meters. The software is installed onto a personal computer (PC) and the blood glucose meter is connected to the PC via a data cable (k040670). When used with TrueTrack or Prestige IQ meters, TrackRecord DMS permits the management of blood glucose data transferred from the meter memory to the computer for enhanced data management capability.
E. Applicant:
Home Diagnostics, Inc.
F. Proprietary and Established Names:
TrackRecord Data Management Software
G. Regulatory Information:
1. Regulation section:
21CFR Sec. - 862.1345-Glucose test system.
21CFR Sec.-862.2100 - Calculator/data processing module for clinical use.
2. Classification:
Class II and I, respectively
3. Product code:
NBW - System, Test, Blood Glucose, Over The Counter
JQP-Calculator/Data Processing Module, For Clinical Use
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use.
2. Indication(s) for use:
TrackRecord Data Management Software (DMS) is intended for use in the home or in clinical settings, for single or multi-patient use, to assist people with diabetes as well as their healthcare professionals in the review, analysis and evaluation of historical blood glucose test results to support effective diabetes management.
3. Special conditions for use statement(s):
Not Applicable
{1}
4. Special instrument requirements:
Home Diagnostics brand Smart System blood glucose meters such as the TrueTrack and Prestige IQ Smart System meters.
TrackRecord DMS is compatible with Windows 98 Second Edition (SE), Windows 2000 and Windows XP Home and Professional. The software operates per 600 MHz Intel® Pentium® III or equivalent, with a minimum 128 MB of RAM, and uses 100 - 200 MB during installation and 100 MB after installation. A 9-pin / 25-pin COM or USB port is required with a serial or USB data cable. A CD-ROM is required.
I. Device Description:
Software application features include data management for single at-home patient use and multiple-patient use for the healthcare professional in the clinical setting. Users can input patient healthcare and ID information, and select date range, glucose units, target glucose ranges, and meal time blocks.
Five reports are available:
1) Summary Report that combines the reports listed below.
2) Electronic Log Book that displays glucose results in a similar manner to a paper log book
3) Conformance Report to graphically represent the percentages of glucose results outside the target range and below the hypoglycemic threshold previously set up by the user.
4) Trend Report to show a graphical representation of daily blood glucose results.
5) Extended Log Book Report is a more comprehensive report than the Log Book Report displaying patient information, glucose target and hypoglycemic ranges, reference method, date of the report, and date range of 2 glucose results. The report calculates the average result for the time slot, the number of results above the target range for that time slot, the number of results below the target range for that time slot, the number of hypoglycemic results for that time slot, and the percentages for above target, below target, within target, and hypoglycemic.
J. Substantial Equivalence Information:
1. Predicate device name(s):
LifeScan IN TOUCH Diabetes Management System
2. Predicate 510(k) number(s):
k984527
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| | IN TOUCH Diabetes Management Software | TrackRecord Data Management System |
| Installation of program | Installed using CD | Installed using CD |
{2}
| Similarities | | |
| --- | --- | --- |
| | IN TOUCH Diabetes Management Software | TrackRecord Data Management System |
| Package Contents | Choice of Software and Electronic Instructions for Use/Registration – or– Software, Electronic Instructions for Use/Registration and USB/9 pin | Choice of Software and Electronic Instructions for Use/Registration – or– Software, Electronic Instructions for Use/Registration and USB/9 pin |
| Cable availability | Serial Cable, Cable available separately | Serial Cable, Cable available separately |
| Capable of uploading data from various devices | Software driver must be uploaded on the device or installed on PC. | Software driver must be uploaded on the device or installed on PC. |
| Computer System Requirements | 9-pin or 25 pin COM or USB port | 9-pin or 25 pin COM or USB port |
| | CD-ROM drive | CD-ROM drive |
| | Windows Operating System 98 SE | Windows Operating System 98 SE |
| | Windows 2000 Professional, and | Windows 2000 Professional, and |
| | Windows XP Home and Professional. | Windows XP Home and Professional. |
| | Minimum of 128 MB of RAM. | Minimum of 128 MB of RAM. |
| | 100-200 MB minimum free hard disc space during installation, 100 MB after installation. | 100-200 MB minimum free hard disc space during installation, 100 MB after installation. |
| | Video Monitor and adapter with at least 800x600 pixel screen resolution and 256 colors | Video Monitor and adapter with at least 800x600 pixel screen resolution and 256 colors |
| Software use indications | Home or clinic settings | Single or Multiple user settings. |
| Manual Entry | Ability to add records manually. | Ability to add records manually. |
| Copy database to separate file. | Yes | Yes |
| Deleting Results | Only Manual entry results may be deleted. | Only Manual entry results may be deleted. |
| Deleting Patients and all accompanying records. | Yes | Yes |
3 of 10
{3}
| Similarities | | |
| --- | --- | --- |
| | IN TOUCH Diabetes Management Software | TrackRecord Data Management System |
| Copy saved database back to active DMS database. | Yes | Yes |
| Ability to default to manufacturer settings (mealtime slots, target glucose ranges, etc.) | Yes | Yes |
| Language capabilities | English, Spanish | English, Spanish |
| Ability to uninstall DMS program | Yes | Yes |
| Ability to access DMS program via icon or explorer. | Yes | Yes |
| Viewing the Owner’s Manual | Link provided via icon. | Accessed via Help on toolbar or F1 on computer |
| Search for specific patient in multiple (clinic) user function. | Yes | Yes |
| Default glucose target ranges available | Yes | Yes |
| Ability to personalize target ranges. | Yes | Yes |
| Customizable schedule | Yes | Yes |
| Search patient capability | Yes | Yes |
| Result type display | Choice of whole blood or plasma result types. | Choice of whole blood or plasma result types. |
| Units of measure display | Choice of mmol/L or mg/dL | Choice of mmol/L or mg/dL |
| Ability to enter hypoglycemic range | Yes | Yes |
| Ability to set default favorite report. | Yes | Yes |
| Report Types | Logbook, Glucose Trend, Pie Chart (Conformance), Summary | Logbook, Glucose Trend, Pie Chart (Conformance), Summary |
| Ability to synchronize meter clock to PC upon download | Yes | Yes |
| Ability to display12 or 24 hour clock format | Yes | Yes |
| Ability to change date format | Yes (mm/dd/yy or dd/mm/yy) | Yes ( mm-dd-yy, mmm-dd-yyyy or yyyy-mm-dd) |
4 of 10
{4}
| Similarities | | |
| --- | --- | --- |
| | IN TOUCH Diabetes Management Software | TrackRecord Data Management System |
| Auto-detect COM port | Yes | Yes |
| Downloaded results cannot be edited or deleted | Yes | Yes |
| Technical support | Yes | Yes |
| Differences | | |
| --- | --- | --- |
| | IN TOUCH Diabetes | TrackRecord Data Management |
| | Management Software | System |
| Units of measure automatically set by country in setup installation. | Yes | No |
| Ability to link to older database versions. | Yes | No older versions available at this time. |
| Required information on patient entry. | No | Yes, Patient ID (multiple users only), patient name (first and last), street address, and city. |
| Insurance information | Only one insurance provider may be entered. | Two insurance providers may be entered. |
| Doctor information | One doctor may be entered. | Two doctors may be entered. |
| Diabetes Educator information | One diabetes educator may be entered. | Two diabetes educators may be entered. |
| Diabetes control | Insulin list, medication list, diet/exercise options | Limited to insulin list and medication list |
| Ability to limit result selection by last transfer | Yes, with the exception of the IN TOUCH Ultra | No |
| Indication for primary test method. | Not available. | Fingerstick or alternate site. |
| Specifying/Entering medications/insulin | Yes, up to 3 different medications/insulin types | Unlimited |
| Ability to enter insulin regiment | Yes | No |
| Ability to specify complications associated with diabetes by patient | Yes | No |
| Ability to set default target range by diabetes type (Type I, Type II Gestational, etc.), | Yes | No |
5 of 10
{5}
| Differences | | |
| --- | --- | --- |
| | IN TOUCH Diabetes | TrackRecord Data Management |
| | Management Software | System |
| Report types | Standard Day, Average Readings, Insulin, Exception, Data List, Histogram, Health Checks | Others available, not these specific types |
| Ability to view results and sort without generating report | No | On screen results may be sorted by individual meter (serial number), date, time, result, time-slot, control results, and manual entry result |
| Ability to view control results | No | Yes |
| Ability to set meter clock to a specific date and time | Yes | No |
| Ability to show cholesterol results/select units of measure | Yes | No |
| Ability to modify meter average results | Yes, 7, 14, 30,60, or 90 days | No |
| Change meter audio cues | Yes | No |
| Ability to clear meter results in memory | Yes | No |
| Ability to input additional information on patient and downloaded results | Yes – Insulin doses, carb data, exercise data, health notes, comments | No |
| Ability to email report from PC directly from program | Yes | No |
| Ability to enter test site for manual result entry | Yes | No |
| Ability to input additional information on patient | Yes - medication dosing, exercise data, food data, health records, health notes/data, | No |
| Ability to input additional information on manual result | Yes – test site, comments, | No |
K. Standard/Guidance Document Referenced (if applicable):
CLSI (NCCLS) GP14-A, Labeling of Home-Use In Vitro Testing Products.
FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
{6}
L. Test Principle:
The TrackRecord DMS is an accessory to the TrueTrack and Prestige IQ Smart System meters, see k030703 and k010403 respectively for meter test principle.
M. Performance Characteristics (if/when applicable):
The TrackRecord DMS is an accessory device to Home Diagnostics brand Smart System blood glucose meters such as the TrueTrack (k030703) and Prestige IQ Smart System (k010403) meters
1. Analytical performance:
a. Precision/Reproducibility: see above associated devices
b. Linearity/assay reportable range: see above associated devices
c. Traceability, Stability, Expected values (controls, calibrators, or methods): see above associated devices
d. Detection limit: see above associated devices
e. Analytical specificity: see above associated devices
f. Assay cut-off: see above associated devices
2. Comparison studies:
a. Method comparison with predicate device: see above associated devices
b. Matrix comparison: see above associated devices
3. Clinical studies:
a. Clinical Sensitivity: see above associated devices
b. Clinical specificity: see above associated devices
c. Other clinical supportive data (when a. and b. are not applicable): A Human Factors Study using 101 subjects was conducted to evaluate the TrackRecord DMS with the following objectives:
1. To evaluate design and ease-of-use
2. To document participant feedback on the features / functions
3. To document Observer feedback on participant performance
4. To validate the user instructions are accurate and easy to follow
5. To evaluate participant ratings with respect to features / functions
6. To report any discrepancies (bugs/error messages) found during the evaluation
These objectives address ease-of-use, instructions for use and product labeling information. This study was comprehensive and included application tests (feature/function tests) and questionnaire items. The subjects (Users) had
7 of 10
{7}
never used TrackRecord DMS and had no prior exposure to TrackRecord DMS or its corresponding Instructions for Use.
Subjects were required to have a minimum of basic computer skills and feel comfortable using a Windows-based Operating System. The study subjects consisted of a mixture of healthcare professionals (multiple patient usage) and persons representing home users (single patient usage).
Conclusions were based on the subjects' ability to demonstrate by usage and evaluation (rating) of features/functions (recorded as responses to 21 feature/function tests), their understanding of the applications of the DMS software, their understanding of how to use the software (install, upload/enter data and patient information, and generate/print reports), their comprehension of the contents of the instructions for use, as well as user assessment of ease-of-use based on a separate questionnaire. The 21 feature/function tests per user were divided into 5 main categories:
1) Installation
2) User Interface / Owner's Manual
3) Set up
4) Downloading Results
5) Patient Management / Reports
Performance of each User was recorded and evaluated by HDI study administrators (Observers), who are familiar with proper use of TrackRecord DMS. Observers rated each subject's performance using four general questions as follows:
1) Subject was successful in installing the DMS program?
2) Report printouts were accurate and complete?
3) Subject performed required functions satisfactorily?
4) Subject accessed "Help" and Technical Support contact information?
Data forms completed by both Users (study subjects) and Observers were provided. Evaluation of the data was performed by an Expert Review Panel, comprised of professionals with expertise in the areas of medical device design, customer and technical support, product quality and quality assurance, and clinical application of the product class.
4. Clinical cut-off: see above associated devices
5. Expected values/Reference range: see above associated devices
8 of 10
{8}
9 of 10
N. Instrument Name:
Home Diagnostics TrackRecord DMS - accessory to Home Diagnostics brand Smart System blood glucose meters such as the TrueTrack (k030703) and Prestige IQ Smart System (k010403) meters.
O. System Descriptions:
1. Modes of Operation:
Windows Operating System 98 SE Windows 2000 Professional, and Windows XP Home and Professional. Minimum of 128 MB of RAM. 100-200 MB minimum free hard disc space during installation, 100 MB after installation. Video Monitor and adapter with at least 800x600 pixel screen resolution and 256 colors. The application is a Microsoft Access developed database.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
The TrackRecord DMS has single or multiple patient capabilities. The application determines which patient owns the glucose meter so the results can be properly assigned to that patient. The meter serial number is the link to the patient record. A patient can have more than one meter serial number, but one meter serial number cannot belong to more than one patient.
The TrueTrack meter can upload from zero to 365 glucose results, plus up to one glucose control result.
The Prestige IQ meter can upload from zero to 365 glucose results, plus up to 12 glucose control results and up to 6 Standard Strip results.
4. Specimen Sampling and Handling:
Capillary whole blood
5. Calibration:
See k030703 and k010403
{9}
6. Quality Control:
See k030703 and k010403
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not Applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10 of 10
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.