← Product Code [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW) · K063409

# LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067 (K063409)

_Agamatrix · NBW · Jan 12, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063409

## Device Facts

- **Applicant:** Agamatrix
- **Product Code:** [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW.md)
- **Decision Date:** Jan 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

AgaMatrix Liberty™ Blood Glucose Monitoring System: AgaMatrix Liberty™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip, palm, or forearm. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Blood Glucose Meter: AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Blood Glucose Test Strips: AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix Liberty™ Control Solutions (Normal level and High level): AgaMatrix Liberty™ Control Solutions are intended for use with the AgaMatrix Liberty™ Meter and AgaMatrix Liberty™ Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

## Device Story

Portable, battery-operated blood glucose meter; uses test strips to measure glucose in fresh capillary whole blood samples (fingertip, palm, or forearm). Operated by patients at home or healthcare professionals in clinical settings. Device provides quantitative glucose concentration readout to assist in diabetes management. System includes meter, test strips, lancing device, and control solutions. No complex AI/ML processing described; standard electrochemical glucose monitoring principle.

## Clinical Evidence

Clinical study of 49 subjects (Type II diabetes majority) compared forearm vs. fingertip glucose readings. Linear regression analysis showed slope 0.98 (lay user) and 1.01 (HCP), with 94-96% of results within ±20% of reference. Median glucose 152 mg/dL (range 84-320 mg/dL).

## Technological Characteristics

Portable, battery-operated electrochemical glucose meter. Materials and design identical to predicate. Standards: ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, IEC 61000-4-3.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Liberty™ Blood Glucose Monitoring System (k052762)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k063409

B. Purpose for Submission:
Addition of an alternate testing site

C. Measurand:
Whole blood glucose

D. Type of Test:
Quantitative

E. Applicant:
AgaMatrix Inc.

F. Proprietary and Established Names:
Liberty™ Blood Glucose Monitoring System

G. Regulatory Information:

1. Regulation section:
21 CFR §862.1345, Blood Glucose Test System
21 CFR §862.1660, Quality control material

2. Classification:
Class II

3. Product code:
NBW, system, test, blood glucose, over the counter
CGA, glucose oxidase, glucose
JJX, single (specified) analyte controls

4. Panel:
Clinical Chemistry (75)

H. Intended Use:

1. Intended use(s):
See Indications for use below.

2. Indication(s) for use:
AgaMatrix Liberty™ Blood Glucose Monitoring System:
AgaMatrix Liberty™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip, palm, or forearm. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by

{1}

persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

## AgaMatrix Liberty™ Blood Glucose Meter:

AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

## AgaMatrix Liberty™ Blood Glucose Test Strips:

AgaMatrix Liberty™ Blood Glucose is intended for use with AgaMatrix Liberty™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood. Testing is done outside the body (in vitro diagnostic use). It is intended for use at home (over the counter (OTC)) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

## AgaMatrix Liberty™ Control Solutions (Normal level and High level):

AgaMatrix Liberty™ Control Solutions are intended for use with the AgaMatrix Liberty™ Meter and AgaMatrix Liberty™ Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

3. Special conditions for use statement(s): Provides plasma equivalent results.

4. Special instrument requirements: Liberty™ Blood Glucose Monitoring System

## I. Device Description:

The Liberty™ Blood Glucose Monitoring System consists of: the Liberty™ Blood Glucose Meter, Test Strips, and Control Solutions.

## J. Substantial Equivalence Information:

1. Predicate device name(s): Liberty™ Blood Glucose Monitoring System
2. Predicate 510(k) number(s): k052762
3. Comparison with predicate: The only modification claimed by the sponsor is to add an alternate site claim for forearm sampling. The meter and its software, the test strips, and the controls are unchanged.

## K. Standard/Guidance Document Referenced (if applicable):

See k052762 for more information.

2

{2}

L. Test Principle:

The Liberty blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents on the electrode of the test strip. The blood sample is drawn into the end of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and hexaammineruthenium (III) chloride, generating a current that is proportional to the glucose concentration in the sample. The result is shown on the meter display.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
For a detailed description see k052762.
b. Linearity/assay reportable range:
For a detailed description see k052762.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
For a detailed description see k052762.
d. Detection limit:
For a detailed description see k052762. The operating range of the meter is 20 – 600 mg/dL.
e. Analytical specificity:
For a detailed description see k052762.
f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:
49 individuals participated in a study to see if glucose readings from the forearm were comparable to glucose reading from fingertips. The subjects ranged in age, education, and were about equally divided between males and females; the majority of the participants had Type II diabetes. The study was performed in a clinic setting. Forearm blood was tested once by the participant then, in a second lanced site, by a healthcare provider (HCP); fingertip blood was tested by the HCP.

Subjects claimed to be in a glucose steady state having not eaten, taken diabetes medication, or exercised vigorously for two hours prior to the study; steady state was not confirmed independently.

The median fingertip Liberty blood glucose result for the 49 study subjects was 152 mg/dL, with results ranging from 84 to 320 mg/dL. The median hematocrit of the subject blood samples was 41.3% with results ranging from 31.5 to 46.6%.

Values obtained from the forearm testing were compared to fingerstick results; linear regression analysis yielded the following results:

{3}

Regression Analysis of Liberty BGM:
Forearm Testing Compared to Fingerstick Testing

|  AST Site | Slope | r | w/in ± 20%  |
| --- | --- | --- | --- |
|  Lay User | y = 0.98x + 10.8 | 0.93 | 96 %  |
|  HCP * | y = 1.01x + 5.8 | 0.97 | 94 %  |

* As per ISO 15197 Section 7.3.3.1, one outlier was excluded from the regression calculations but not from the determination of system accuracy. In this patient, the HCP fingerstick value was 84 mg/dL while the HCP forearm value was 133 mg/dL. The forearm reading obtained by the lay user was 120 mg/dL.

b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
The normal fasting adult glucose range for a non-diabetic is 70 – 105 mg/dL. One to two hours after a meal, normal blood glucose levels should be less than 140 mg/dL. A medical professional should determine the range that is appropriate for diabetes patients.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063409](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063409)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
