← Product Code [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW) · K063026

# OK METER BLOOD GLUCOSE MONITORING SYSTEM (K063026)

_Ok Biotech Co., Ltd. · NBW · Feb 9, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063026

## Device Facts

- **Applicant:** Ok Biotech Co., Ltd.
- **Product Code:** [NBW](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW.md)
- **Decision Date:** Feb 9, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The OK Meter Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Testing is done outside the body (in vitro diagnostic use). It is indicated for both lay uses by people with diabetes, and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates.

## Device Story

OK Meter Blood Glucose Monitoring System; electrochemical biosensor; capillary action. Input: 0.7 μL fresh capillary whole blood applied to test strip. Operation: glucose oxidase enzyme reaction on strip electrodes; current measured by meter proportional to glucose concentration. Output: blood glucose result displayed in 6 seconds. Used by patients (home) or healthcare professionals (clinic). User enters batch-specific code from strip vial into meter for calibration. Output used to monitor glucose levels in Diabetes Mellitus; aids clinical decision-making for glycemic control.

## Clinical Evidence

Pre-clinical and clinical data were submitted in accordance with FDA Guidance for In Vitro Diagnostic Test Systems. No specific performance metrics (e.g., sensitivity, specificity) are provided in the summary document.

## Technological Characteristics

Electrochemical biosensor; glucose oxidase enzyme; amperometric detection. Sample volume: 0.7 μL. Measurement range: 20-600 mg/dL. Calibration: batch-specific code entry. Connectivity: none. Sterilization: not specified. Software: embedded firmware for signal conversion and display.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Roche Diagnostics Corp. Accu-Chek Active Test System (k012324)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k063026

B. Purpose for Submission:
Clearance of a new device.

C. Measurand:
Glucose

D. Type of Test:
Blood Glucose Concentration through a Quantitative Amperometric Assay (Glucose Oxidase)

E. Applicant:
OK Biotech Co., Ltd.

F. Proprietary and Established Names:
OK Meter Blood Glucose Monitoring System

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1345, Glucose Test System
21 CFR § 862.1660, Single (specified) analyte controls (assayed and unassayed)
2. Classification:
Class II
Class I (reserved)
3. Product code:
NBW, CGA, JJX
4. Panel:
75 (Clinical Chemistry)

H. Intended Use:
1. Intended use(s):
See Indications for use.
2. Indication(s) for use:
The OK Meter Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Testing is done outside the body (in vitro diagnostic use). It is indicated for both lay uses by people with diabetes, and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates.
3. Special conditions for use statement(s):
For Over-the-Counter use.
4. Special instrument requirements:
OK Meter Blood Glucose Monitoring System

I. Device Description:
The OK Meter Blood Glucose Monitoring System is based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. The OK Meter Blood

{1}

Glucose Monitoring System uses a whole blood volume of 0.7 μL. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 6 seconds. The control solutions available are used to test the performance of the device.

## J. Substantial Equivalence Information:

1. Predicate device name(s):
Roche Diagnostics Corp. Accu-Chek Active Test System
2. Predicate 510(k) number(s):
k012324
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Detection Method | Amperometry | Amperometry  |
|  Enzyme | Glucose Oxidase | Glucose Oxidase  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Test Range | 20 – 600 mg/dL | 10 – 600 mg/dL  |
|  Volume Required | 0.7 μL | 1-2 μL  |
|  Hematocrit Range | 20-60 mg/dL | 20-70 mg/dL  |
|  Test Time | 6 seconds | 5 seconds  |

## K. Standard/Guidance Document Referenced (if applicable):

CLSI EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices
CLSI EP6-P: Evaluation of the Linearity of Quantitative Measurement Procedures

## L. Test Principle:

Glucose measurement is based on electrical potential caused by the reaction of glucose with the reagents contained on the strip’s electrodes. The glucose in the sample is oxidized by the enzyme glucose oxidase, and the current resulting from this enzymatic reaction is measured and converted to glucose concentration by the meter. The magnitude of the current is proportional to the concentration of glucose in the sample.

## M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

The sponsor evaluated the within run precision of the device using 20 replicate measurements of 5 concentrations of glucose-adjusted venous whole blood on 5 meters each tested with 3 strip lots. Results are summarized below.

{2}

The sponsor also evaluated the between run precision of the device using replicate measurements of glucose controls. Twenty replicates at 5 different concentrations (30-48, 104-148, 172-243, 293-386, 413-585) and 2 levels of controls (a low  $\sim 120$  and high  $\sim 270$ ) were each tested with 3 strip lots over 10 days. Results met the sponsor's acceptance criteria of a standard deviation  $\leq 4.5 \mathrm{mg} / \mathrm{dL}$  when the glucose concentration was  $&lt; 75 \mathrm{mg} / \mathrm{dL}$  and less than a  $6 \%$  CV for glucose concentrations  $\geq 75 \mathrm{mg} / \mathrm{dL}$ .

# b. Linearity/assay reportable range:

To establish the linearity of the OK Meter system through the range of 15 to 598 mg/dL glucose adjusted whole blood samples were compared to YSI 2300. Linear regression yields the following statistics:

|  N | 20  |
| --- | --- |
|  Slope | 0.991  |
|  y-intercept | 5.54  |
|  r2 | 0.9993  |

The sponsor claims  $20\mathrm{mg / dL}$  as the lowest detectable limit in the labeling.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): The device is traceable to a laboratory analyzer which is calibrated to a glucose standard (NIST SRM 965a). Stability characteristics of the two control solutions were determined using real-time stability studies to determine the storage stability at room temperature to be 18 months.

# d. Detection limit:

The measuring range of the OK Meter Blood Glucose Monitoring System is 20 - 600 mg/dL. This range was verified by the linearity study (above).

# e. Analytical specificity:

The sponsor tested the following substances for interference using two levels of glucose concentrations ( $\sim 140\mathrm{mg / dL}$  and  $\sim 340\mathrm{mg / dL}$ ). Interference was defined as  $&gt;15\%$  bias compared to YSI. The following results were determined with regard to interfering substances:

|  Substance | Therapeutic/Normal Concentration | Highest Concentration with <15% bias compared to YSI  |
| --- | --- | --- |
|  Acetaminophen | 1-2 mg/mL | 15 mg/dL  |
|  Dexamethasone | 1-2 mg/mL | 15 mg/dL  |
|  Dexamethasone | 1-2 mg/mL | 15 mg/dL  |

{3}

4

|  Substance | Therapeutic/Normal Concentration | Highest Concentration with <15% bias compared to YSI  |
| --- | --- | --- |
|  Dopamine | <87 pg/mL | 300 mg/dL  |
|  L-Dopa | Not applicable | 300 mg/dL  |
|  Tolbutamide | 5.3-10 mg/dL | 220 mg/dL  |
|  Ascorbic acid | 0.8-1.2 mg/dL | 5 mg/dL  |
|  Bilirubin | 1.3 mg/dL | 20 mg/dL  |
|  Galactose | Not applicable | 10 mg/dL  |
|  Maltose | Not applicable | 20 mg/dL  |
|  Triglyceride | 36-165 mg/dL | 300 mg/dL  |
|  Uric acid | 7 mg/dL | 13 mg/dL  |

An altitude study was performed with 7 whole blood samples ranging from 15-600 mg/dL. The spiked whole blood samples were measured by the OK Meter 6 times each. The samples were compared to YSI and found to have deviation less than 12 mg/dL for samples &lt;70 mg/dL and less than a 6% CV for samples ≥70. The sponsor claims that altitude up to 4691 feet has no effect on blood glucose measurements when using this test system.

To test the claimed hematocrit range, 20 replicates were tested at each hematocrit level and each glucose concentration. Samples were adjusted to hematocrit levels of 20%, 30%, 45%, 50%, and 65% and at glucose concentrations of 50, 80, 150, 300, 400, 500 mg/dL. Acceptable results were defined as 95% of samples falling into zone A of an error grid analysis and less than 5% falling into zone B compared when compared to a 45% hematocrit level. The performance was acceptable across the claimed hematocrit range of 20 – 60 %.

f. Assay cut-off:
Not Applicable.

2. Comparison studies:

a. Method comparison with predicate device:

The sponsor demonstrated that the OK Meter Blood Glucose Monitoring System for finger stick is equivalent to a standard method (YSI-2300) by having 150 patient samples with a hematocrit range of 35-55% test their own blood. A technician collected blood for YSI measurement and tested using the subject meter at the same time. Samples ranged from 71 to 414 mg/dL (according to YSI). The study results are summarized below:

|   | YSI vs. Patient Finger  |
| --- | --- |
|  N | 150  |
|  Slope | 0.9713  |
|  Intercept | 2.707  |
|  r² | 0.9623  |

The sponsor also evaluated the results obtained by patients versus those obtained by

{4}

trained technicians. The sponsor found that differences between 150 samples obtained by both a patient and a technician were within ± 20%.

Meter versus YSI at each site met the ISO 15197 requirement of ninety-five percent (95 %) of the individual glucose results falling within ±15 mg/dL of the results of the manufacturer's measurement procedure at glucose concentrations for samples &lt;75 mg/dL and within ± 20 % at glucose concentrations ≥75 mg/dL. These results are summarized in the table below.

|  Site | Finger  |
| --- | --- |
|  N | 150  |
|  Percentage That Met ISO Requirement | 100% (150/150)  |

b. Matrix comparison:
Not applicable. Capillary whole blood is the only indicated matrix.

3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided a readability study that indicated that the user manual, strip labeling, and control solutions are at a 7th grade reading level.

4. Clinical cut-off:
Not Applicable.

5. Expected values/Reference range:
The sponsor included the following Expected Values for normal glucose levels in their strip labeling:

|  Status | Range (mg/dL) | Range (mmol/L)  |
| --- | --- | --- |
|  Before breakfast | 70-105 | 3.9-5.8  |
|  Before lunch or dinner | 70-110 | 3.9-6.1  |
|  1 hour after meals | <160 | <8.9  |
|  2 hours after meals | <120 | <6.7  |
|  Between 2 and 4 AM | >70 | >3.9  |

Source: Krall, L.P. and Beaser, R.S.: Joslin Diabetes Manual. Philadelphia: Lea and Febiger (1989), 138.

{5}

6

N. Instrument Name:
OK Meter Blood Glucose Monitoring System

O. System Descriptions:

1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.

2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐

3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.

4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger only. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues.

5. Calibration:
A code number is provided with each batch of test strips to calibrate the meter for that batch. The code number must be entered into the meter to match the number found on the test strip vial. No further calibrations are required of the user.

6. Quality Control:
The sponsor is providing a single level glucose control solution with this device, though two levels are available for purchase as stated in the labeling. When a test strip is inserted into the meter, a control can be run. An acceptable range for each control level is printed on the test strip vial label. The user is referred to a troubleshooting section of the owner’s manual to identify possible reasons control results fall outside these ranges.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None.

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063026](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/NBW/K063026)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
