The TheraSense, Inc., FreeStyle Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The TheraSense, Inc., FreeStyle Tracker Diabetes Management System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. Additionally, the TheraSense, Inc. FreeStyle Tracker Diabetes Management System is intended for use in home and clinical setting to aid people with diabetes and healthcare professionals in the review, analysis, and evaluation of historical blood glucose test results to support an effective diabetes management program. The TheraSense, Inc. FreeStyle Tracker Diabetes Management System is specifically indicated for use on the finger, forearm, upper arm, thigh, calf and hand.
Device Story
System uses electronic biosensor test strips and handheld meter to measure blood glucose; capillary action draws 0.3 microliters of whole blood into strip chamber; glucose dehydrogenase enzyme reacts with glucose to produce electrical current; meter measures current to calculate glucose level; results displayed in ~15 seconds. Used in home and clinical settings by patients and healthcare professionals. Tracker system integrates with PDA for data management, analysis, and evaluation of historical results. Benefits include effective diabetes management through monitoring and historical data review.
Clinical Evidence
Clinical study compared palm and fingertip glucose values during conditions of rapid glucose change. Results demonstrated equivalence between sites, supporting the recommendation of the palm as an alternative test site.
Technological Characteristics
Electrochemical coulometric sensor; glucose dehydrogenase enzyme; capillary fill (0.3 microliters); result range 20-500 mg/dL; operating temp 50-95°F; humidity 5-90%; battery-powered (AAA or rechargeable PDA battery); data management via Tracker software; plasma equivalent calibration.
Indications for Use
Indicated for quantitative whole blood glucose measurement in patients with diabetes mellitus; for use by healthcare professionals and patients at home. Not for neonates, diagnosis, or screening. Tracker system indicated for use on finger, forearm, upper arm, thigh, calf, and hand.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Freestyle Blood Glucose Monitoring System (K992684)
Freestyle Blood Glucose Monitoring System (K000582)
Freestyle Blood Glucose Monitoring System (K012014)
FreeStyle Tracker Diabetes Management System (K020866)
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number: k031260
B. Analyte: whole blood glucose
C. Type of Test: Quantitative, utilizing Glucose Dehydrogenase technology.
D. Applicant: TheraSense, Inc.
E. Proprietary and Established Names:
FreeStyle Blood Glucose Monitoring System
FreeStyle Tracker Diabetes Management System
F. Regulatory Information:
1. Regulation section:
21 CFR 862.1345 Glucose Test System
21 CFR 862.1660 Quality Control Material
2. Classification:
Class II, Class I
3. Product Code:
NBW, System, Test, Blood Glucose, Over The Counter
LFR, Glucose Dehydrogenase, Glucose
JJX, Single (Specified) Analyte Controls, Class I
4. Panel:
Chemistry (75)
G. Intended Use:
5. Indication(s) for use:
FreeStyle Blood Glucose Monitoring System
The ThereSense, Inc. FreeStyle Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not
{1}
Page 2 of 7
intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
## FreeStyle Tracker Diabetes Management System
The TheraSense, Inc. FreeStyle Tracker Diabetes Management System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
Additionally, the TheraSense, Inc. FreeStyle Tracker Diabetes Management System is intended for use in home and clinical settings to aid people with diabetes and healthcare professionals in the review, analysis, and evaluation of historical blood glucose test results to support an effective diabetes management program.
The TheraSense, Inc. FreeStyle Blood Glucose Monitoring System, and FreeStyle Tracker Diabetes Management System is specifically indicated for use on the finger, forearm, upper arm, thigh, calf and hand.
6. Special condition for use statement(s):
Provides plasma equivalent results.
Not Intended for use on neonates.
7. Special instrument Requirements:
N/A
## H. Device Description:
The FreeStyle Blood Glucose Monitoring System comprises an electronic biosensor glucose reagent test strip, a handheld meter, a quality control solution, a complete Owner's Booklet and Quick Reference Guide. A lancing device, lancets and a logbook for recording test results are also included with the system.
## I. Substantial Equivalence Information:
1. Predicate device name(s):
The predicate devices are the TheraSense, Inc. FreeStyle Blood Glucose Monitoring System and the TheraSense, Inc. FreeStyle Tracker Diabetes Management System
{2}
Page 3 of 7
2. Predicate K number(s):
FreeStyle Blood Glucose Monitoring System (K992684)
(K000582)
(K012014)
FreeStyle Tracker Diabetes Management System (K020866)
3. Comparison with predicate:
All attributes of device to remain the same, only the labeling is to be modified.
Tracker Blood Glucose Monitoring Features Comparison to Predicate
| | Tracker System | FreeStyle System Predicate Device |
| --- | --- | --- |
| Device Name & Manufacturer | TheraSense, Inc. | Same |
| Device Description | Glucose Test System | Same |
| 510(k) Number | K020866 | K992684, K000582, K012014, K994433 |
| Regulatory Class | II; 75 LFR | Same |
| Device Classification | 21 CFR 862.1345 | Same |
| Intended Use | Quantitative glucose measurement in whole blood for self monitoring in diabetics. | Same as K992684, K000582, and K012014 |
| Enzyme | Glucose Dehydrogenase | Same |
| Reagent Form | Test Strip | Same |
| Detection Methodology | Electrochemical | Same |
| Recommended Sample | Capillary whole blood | Same |
| Minimum Sample size | 300 nanoliters (0.3 microliters) | Same |
| Calibration | Test Strip
Plasma equivalent | Same |
| Sample Application | Whole blood applied to a test strip that is external to the meter | Same |
| Result Range | 20 to 500 mg/dL (1.1 – 27.8 mmol/L) | Same |
| Test Time | Average of 15 seconds | Same |
| Assay Method | Coulometric Electrochemical Sensor | Same |
{3}
Page 4 of 7
| | Tracker System | FreeStyle System Predicate Device |
| --- | --- | --- |
| Measurement Units | Either mg/dL or mmol/L | Same |
| Memory | 250 most recent blood glucose and control solution tests | Same |
| Automatic Shut-off | 2 minutes after last user action | Same |
| Power Source | 2 replaceable AAA batteries, or rechargeable PDA battery | Two AAAA replaceable batteries |
| Battery Life | 2-4 weeks | 1,000 tests |
| Operating Temperature | 50 – 95 degrees F | Same |
| Operating Relative Humidity | 5-90 % | Same |
| Hematocrit | 30-60 % | Same |
| Wiping Required? | No | No |
| Timing Required? | No | No |
| Date / Time Stamp? | Yes | Yes |
| Communication | PDA | Serial Port |
| Dimensions | 85.0 x 54.5 x 20.0 mm
Module only | 97 x 52 x 25 mm |
| | 150.0 x 77.0 x 24.0 mm
Module and PDA | n/a |
| Weight | Approx. 40 grams
Module only | Approx. 60 grams |
| | Approx. 1140 grams
Module and PDA | n/a |
| Data Management Capability | Yes, Tracker Data Management Software | Yes, Connect Data Management Software |
J. Standard/Guidance Document Referenced (if applicable):
NCCLS EP-7; AdvaMed Points to Consider for Glucose Monitoring Devices Intended for Alternative Site Testing
K. Test Principle:
The user obtains a blood sample using a conventional lancing technique. The user inserts a test strip into the meter, which turns the meter on. When the strip is touched to the blood drop, the sample chamber on the strip is filled by capillary action in approximately 2 seconds. The blood sample volume required is approximately 0.3 microliters (300 nanoliters), which can be obtained from the finger or other areas of
{4}
Page 5 of 7
the body such as the arm. Test results are displayed in about 15 seconds. The time required to display test results varies depending on the blood glucose concentration (approximately 15 t o 45 seconds).
The glucose in the blood sample reacts with the glucose Dehydrogenase enzyme to yield gluconolactone, and produces a small electrical current. This current is measured by the FreeStyle meter and displayed as a glucose level.
## L. Performance Characteristics (if/when applicable):
The performance of the FreeStyle Test Strips has been tested both in laboratory and clinical studies. The testing range of the FreeStyle System is 20 mg/dL to 500 mg/dL (1.1 to 27.8 mmol/L).
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Within-lot and within-vial precision of FreeStyle test strips was measured with venous blood samples in the laboratory. The pooled precision data for fifty-four test strip lots (n= 5,184) is shown in the tables below:
**Within-Lot Precision**
| Average Glucose Concentration (mg/dL) | 43 | 194 | 380 |
| --- | --- | --- | --- |
| SD (mg/dL) | 2.4 | 6.2 | 13.7 |
| CV (%) | 5.6 | 3.2 | 3.6 |
**Within-Vial Precision**
| Average Glucose Concentration (mg/dL) | 43 | 194 | 380 |
| --- | --- | --- | --- |
| SD (mg/dL) | 1.8 | 3.6 | 7.4 |
| CV (%) | 4.1 | 1.8 | 2.0 |
Variability in blood test from strip to strip was 5.6% or less.
#### b. Linearity/assay reportable range:
The established linearity and reportable range was cleared with submission K020866, which received clearance on 11 June 2002.
#### c. Traceability (controls, calibrators, or method):
Cleared with K020866
#### d. Detection limit
20 – 500 mg/dL
1.1 – 27.8 mmol/L
#### e. Analytical specificity:
Cleared with K020866.
{5}
Page 6 of 7
f. Assay cut-off: N/A
2. Comparison studies:
a. Method comparison with predicate device:
The accuracy of the FreeStyle system was tested by comparing blood glucose results obtained by study subjects with those obtained using a YSI Model 2300 Glucose Analyzer. The results below were obtained from subjects with Type 1 or Type 2 diabetes. The regression statistics are derived from a plot of the FreeStyle arm data versus YSI capillary data.
Slope = 0.931
y-intercept + 8.1 mg/dL
Correlation coefficient (R) 0.969
Number of samples tested in duplicate 197
Range tested 51 to 487 mg/dL
b. Matrix comparison: N/A
3. Clinical studies:
a. Clinical sensitivity: N/A
b. Clinical specificity: N/A
c. Other clinical supportive data (when a and b are not applicable):
A clinical study was conducted to compare glucose readings using blood from the finger and blood from the palm. Forty-four subjects with type 1 diabetes were given a glucose challenge to raise glucose to a hyperglycemic state and then given subcutaneous insulin to reduce glucose to a normal or hypoglycemic state. Three subjects were excluded from the data analysis because they did not achieve the glucose rate of change required by the protocol. The differences between readings from the finger and palm sites for all glucose levels and for low glucose levels are summarized below.
| | Range (mg/dL) | Number of test pairs | Average difference | Standard Deviation |
| --- | --- | --- | --- | --- |
| All test | 43 – 477 | 1309 | -2.1% | 12.7% |
| Low Glucose | 43 – 69 | 72 | 4.9 mg/dL | 9.0 mg/dL |
Physiologic differences in the circulation between the finger or palm and other test sites like the forearm, upper arm, thigh, calf and hand, may result in differences in blood glucose measurements from the other test sites and your finger or palm. Differences in blood glucose measurements between the other test sites and your finger or palm may be observed in finger or palm blood samples sooner than flood samples from the forearm
{6}
Page 7 of 7
and other alternate sites. Vigorous rubbing of the alternate test sites before lancing will help to minimize the difference between finger and alternate site test results. Refer to the Owners Booklet for complete information about testing your blood glucose.
4. Clinical cut-off: N/A
5. Expected values/Reference range:
## Normal Glucose Values
The normal fasting glucose range for a non-diabetic adult is 70 to 110 mg/dL (3.9 to 6.1 mmol/L) 3. Burtis CA Ashwood ER eds: Teitz Textbook of Clinical Chemistry. 2nd Edition W.B. Saunders. Philadelphia. 1994. P. 2190 One to two hours after meal, normal glucose values should be less than 120 mg/dL (6.7 mmol/L) 4. Krall LP and Beaser T: Joslin Diabetes Manual. Lea and Febiger. Philadelphia 1989 p. 138. Consult your physician or healthcare professional for the target glucose values that are right for you.
## Low Glucose Values
The FreeStyle Meter displays results between 20 and 500 mg/dL (1.1 and 27.8 mmol/L). If your result is lower than 20, "Low" (LO) will appear on the meter display. This indicates severe low blood sugar (hypoglycemia). You should immediately treat low blood sugar as recommended by your healthcare professional.
## High Glucose Values
If your test result is above 500 mg/dL (27.8 mmol/L) "High"(HI) will appear on the meter display screen. This indicates severe high blood sugar (hyperglycemia). You should immediately treat low blood sugar as recommended by your healthcare professional.
## Unexpected Results
Low or high blood sugar readings can indicate a potentially serious medical condition. If your blood sugar is unusually low or high, or if you do not feel the way your results indicate, repeat the test with a new test strip. If your reading is not consistent with your symptoms of if your blood glucose result is less than 60 mg/dL (3.3 mmol/L) or higher than 240 mg/dL (13.3 mmol/L), you should contact your healthcare professional and follow his or her treatment advice.
## M. Conclusion:
Based upon review of the information and labeling provided, this device is Substantially Equivalent to 21 CFR 862.1345, 75 NBW, System, Test, Blood Glucose, Over the Counter and 75 LFR Glucose Dehydrogenase, Glucose.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.