N-GENEOUS LDL-ST CHOLESTEROL REAGENT, N-GENEOUS LDL-ST CHOLESTEROL CALIBRATOR, GENZYME LDL CHOLESTEROL CONTROL SET
Applicant
Genzyme Corp.
Product Code
MRR · Clinical Chemistry
Decision Date
Oct 18, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1475
Device Class
Class 1
Indications for Use
Reagents: For the direct, quantitative determination of low-density lipoprotein cholesterol (LDL-C) in human serum or plasma. Calibrator: For the calibration of the N-geneous® LDL-ST Cholesterol assay in serum or plasma. Controls: To monitor the performance of the Genzyme Direct LDL Cholesterol, N-geneous® LDL Cholesterol and N-geneous® LDL-ST Cholesterol Reagents.
Device Story
Device is a two-reagent homogeneous assay for direct quantitative determination of LDL-C in human serum or plasma; used on Olympus chemistry analyzers. Principle: first detergent selectively solubilizes non-LDL lipoproteins for enzymatic removal; LDL particles remain intact. Second detergent releases LDL cholesterol; subsequent enzymatic reaction with 4-aminoantipyrine produces color proportional to LDL-C concentration. Used in clinical laboratory settings by trained personnel. Output is quantitative LDL-C measurement; assists clinicians in lipid profile assessment and cardiovascular risk management.
Clinical Evidence
Bench testing only. Comparative study of 77 serum samples (LDL 54.5–222.1 mg/dL) on Olympus AU600 analyzer. Correlation with ß-Quantification (n=62): r=0.969, slope 1.08, intercept -8.7 mg/dL. Correlation with predicate N-geneous® LDL (n=77): r=0.990, slope 1.02, intercept -1.7 mg/dL. Precision studies (n=20 per run) showed within-run CV 0.8–1.4% and between-run CV 2.1–2.7%.
Technological Characteristics
Two-reagent homogeneous enzymatic assay. Uses selective detergents for differential solubilization of lipoproteins. Colorimetric detection via peroxidase reaction with 4-aminoantipyrine. Designed for automated chemistry analyzers (e.g., Olympus AU600). No off-line pretreatment or centrifugation required.
Indications for Use
Indicated for the direct, quantitative determination of LDL-C in human serum or plasma for clinical laboratory use.
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
Predicate Devices
N-geneous® LDL Cholesterol Reagents
Reference Devices
ß-Quantification methods
Related Devices
K971573 — N-GENEOUS LDL CHOLESTEROL REAGENT, N-GENEOUS LDL CHOLESTEROL CALIBRATOR, GENZYME LDL CHOLESTEROL CONTROL SET · Genzyme Corp. · Jun 18, 1997
K981303 — DIRECT LDL · Abbott Laboratories · Aug 3, 1998
K020724 — DIMENSION AUTOMATED LDL CHOLESTEROL FLEX REAGENTCARTRIDGE METHOD, MODEL DF131 · Dade Behring, Inc. · May 9, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K992766
## 510(k) PREMARKET NOTIFICATION
| Genzyme Corporation |
|---------------------|
| One Kendall Square |
| Cambridge, MA 02139 |
#### 510(k) Summary Of Safety and Effectiveness Information Upon Which An Equivalence Determination Could be Made
| Trade or Proprietary Name: | Genzyme N-geneous® LDL-ST Cholesterol Reagent<br>Genzyme N-geneous® LDL-ST Cholesterol Calibrator<br>Genzyme LDL Control Set |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Homogeneous assay for low density lipoprotein cholesterol |
| Classification Name: | Low density lipoprotein cholesterol test<br>Calibrator, Primary<br>Low density lipoprotein control |
| Manufacturer: | Genzyme Diagnostics<br>One Kendall Square<br>Cambridge, MA 02139-1562 |
| Contact Person: | Robert Yocher, Vice President, Regulatory Affairs (617) 374-7275<br>Barbara Pizza, Manager, Regulatory Affairs (617) 252-7953. |
The use of the Genzyme N-geneous® LDL-ST Cholesterol Reagents in the clinical laboratory setting is substantially equivalent to the N-geneous® LDL Cholesterol Reagents and the B-Quantification methods.
The Genzyme N-geneous® LDL-ST Cholesterol Reagent is a two-reagent homogeneous method for the direct quantitative determination of low density lipoprotein cholesterol (LDL-C) in human serum and plasma. This method is applicable to Olympus two-reagent Chemistry Analyzers and does not require any off-line pretreatment or centrifugation steps.
The principle of the test is based upon a unique detergent which selectively solubilizes only the non-LDL lipoproteins, allowing them to be removed by cholesterol enzymes prior to the LDL cholesterol reaction. The LDL particles remain intact. The hydrogen peroxide produced by the reaction of the enzymes with the released cholesterol is consumed by a peroxidase reaction with 4-aminoantipyrine, yielding a colorless product. A second detergent, capable of releasing LDL cholesterol molecules, is added. The enzyme reaction with LDL cholesterol in the presence of the coupler produces color which is proportional to the amount of LDL cholesterol in the sample.
Comparative performance studies were conducted using the N-geneous® LDL-ST Cholesterol Reagents and two reference methods: N-geneous® LDL Cholesterol Reagent and ß-Quantification.
{1}------------------------------------------------
# 510(k) PREMARKET NOTIFICATION
Seventy-seven serum samples, with LDL values between 54.5 and 222.1 mg/dL, were tested at Genzyme Corporation using the N-geneous® LDL-ST on the Olympus AU600 Analyzer and the Ngeneous® LDL methods. Of these 77 samples, 62 had sufficient volume to run ß-Quantification method.
| N-geneous® LDL-ST on the<br>Olympus AU600 | vs.<br>β-Quantification<br>(n = 62) | vs.<br>N-geneous® LDL on<br>Hitachi 911<br>(n = 77) |
|-------------------------------------------|-------------------------------------|-----------------------------------------------------|
| Slope | 1.08 | 1.02 |
| Intercept (mg/dL) | -8.7 | -1.7 |
| Correlation Coefficient (r) | 0.969 | 0.990 |
| Mean (mg/dL) | LDL-ST: 119.6<br>β-Q: 119.0 | LDL-ST: 121.6<br>LDL: 121.3 |
| Standard Deviation (mg/dL) | LDL-ST: 28.9<br>β-Q: 26.0 | LDL-ST: 30.7<br>LDL: 29.9 |
| Mean Difference (mg/dL) | 0.6 | 0.3 |
| Mean Percent Difference (%) | -0.1 | 0.1 |
## ß-Q = ß-Quantification
Precision studies were conducted using the N-geneous® LDL-ST Cholesterol Reagents on the Olympus Analyzers. Both within-run and between-run studies were performed using Genzyme Desirable and Risk controls:
| Within-Run Precision | Low | High |
|----------------------|---------------------|----------------------|
| | (<130 mg/dL) | (≥160 mg/dL) |
| n | 20 | 20 |
| Sample Range (mg/dL) | 92.0 - 97.0 | 229.6 - 237.3 |
| Mean (mg/dL) | 93.8 | 233.5 |
| SD (mg/dL) | 1.3 | 1.9 |
| %CV | 1.4 | 0.8 |
| Between run | Low<br>(<130 mg/dL) | High<br>(≥160 mg/dL) |
| n | 20 | 20 |
| Sample Range (mg/dL) | 83 - 90 | 178 - 195 |
| Mean (mg/dL) | 86.9 | 188.4 |
| SD (mg/dL) | 1.8 | 5.0 |
| %CV | 2.1 | 2.7 |
These data demonstrate that the performance of the N-geneous® LDL-ST Cholesterol Reagents in the clinical laboratory is substantially equivalent to the performance of the N-geneous® LDL Cholesterol Reagent and the ß-Quantification methods.
{2}------------------------------------------------
In lieu of a 510(k) statement under 513(i) of the Act, this information is provided as a 510(k) summary for disclosure to any other persons/companies without the specific written authorization from Genzyme Corporation.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the eagle.
OCT 1 8 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Barbara Pizza Manager, Regulatory Affairs Genzyme Corporation One Kendall Square Cambridge, Massachusetts 02139-1562
Re: K992766
> Trade Name: N-geneous® LDL-ST Cholesterol Reagent Regulatory Class: I reserved P Product Code: MRR N-geneous® LDL-ST Cholesterol Calibrator Regulatory Class: II Product Code: JIS Genzyme LDL Cholesterol Control Set Regulatory Class: I reserved Product Code: JJX Dated: August 16, 1999 Received: August 17, 1999
Dear Ms. Pizza:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrlv/dsma/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 3.0 INTENDED USE
Page _1_ of __ 1
510(k) Number (if known): K993746
Device Name: _N-geneous® LDL-ST Cholesterol Reagent
N-geneous® LDL-ST Cholesterol Calibrator
Genzyme LDL Cholesterol Control Set
#### Indications for Use:
Reagents:
For the direct, quantitative determination of low-density lipoprotein cholesterol (LDL-C) in human serum or plasma.
Calibrator:
For the calibration of the N-geneous® LDL-ST Cholesterol assay in serum or plasma.
Controls:
To monitor the performance of the Genzyme Direct LDL Cholesterol, N-geneous® LDL Cholesterol and N-geneous® LDL-ST Cholesterol Reagents.
Tom Coker
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K992761
RxV
CTC
7
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.