← Product Code [MRG](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MRG) · K123660

# 25-OH  VITAMIN D ELISA (K123660)

_Euroimmun US · MRG · Jul 17, 2013 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MRG/K123660

## Device Facts

- **Applicant:** Euroimmun US
- **Product Code:** [MRG](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MRG.md)
- **Decision Date:** Jul 17, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1825
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

EUROIMMUN’s 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Li-heparin). Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.

## Device Story

The EUROIMMUN 25-OH Vitamin D ELISA is an in vitro diagnostic test kit used in clinical laboratories to measure 25-OH vitamin D levels in human serum or plasma. The device utilizes a competitive immunoassay principle: patient samples and biotin-labeled 25-OH vitamin D compete for binding sites on sheep monoclonal anti-25-OH vitamin D antibodies coated on a microplate. After washing, peroxidase-labeled streptavidin is added, followed by a TMB substrate to produce a color reaction. The resulting color intensity is inversely proportional to the concentration of 25-OH vitamin D in the sample. A microwell plate reader measures optical density at 450/620 nm. Clinicians use the quantitative results, alongside other clinical and laboratory data, to assess vitamin D sufficiency in adult patients. The device aids in identifying vitamin D deficiency or insufficiency, supporting clinical decision-making regarding patient supplementation or treatment.

## Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP05-A2 (intra-assay CV 4.2-12.4%, inter-assay CV 6.7-16.2%). Linearity confirmed 2.0-129 ng/mL (R²=0.999). Method comparison with predicate (n=240) yielded y=1.08x-0.78, R=0.9858. Analytical specificity, interference, and matrix equivalence (EDTA/Li-heparin) studies performed. Reference range established using 206 healthy subjects.

## Technological Characteristics

Competitive ELISA; 96-well microplate format. Reagents: monoclonal sheep anti-25-OH Vitamin D IgG, biotin-labeled 25-OH Vitamin D, peroxidase-labeled streptavidin, TMB substrate. Measuring range: 4.0–120 ng/mL. Instrument: ELISA plate reader (450/620nm). Storage: 2-8°C. Traceability: UV spectrophotometric analysis of 25-OH vitamin D3.

## Regulatory Identification

A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.

## Special Controls

*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.

## Predicate Devices

- IDS Immunodiagnostic Systems Ltd. OCTEIA 25-Hydroxy Vitamin D ([K021163](/device/K021163.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k123660

B. Purpose for Submission:
New Assay

C. Measurand:
25-hydroxyvitamin D and other hydroxylated metabolites

D. Type of Test:
Quantitative, competitive enzyme immunoassay

E. Applicant:
EUROIMMUN US

F. Proprietary and Established Names:
25-OH Vitamin D ELISA

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  MRG | Class II | 21 CFR 862.1825
Vitamin D Test System | Clinical Chemistry (75)  |

H. Intended Use:
1. Intended use(s):
See Indications for Use

2. Indication(s) for use:
EUROIMMUN’s 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and

{1}

plasma (EDTA, Li-heparin). Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

Microwell plate reader capable of measuring OD at 450nm and at 620nm for dual wavelength readings.

I. Device Description:

EUROIMMUN’s 25-OH Vitamin D ELISA consists of a microwell ELISA plate coated with anti-25-OH vitamin D antibodies (sheep), 6 calibrators, 2 controls, Biotin concentration solution, sample buffer, enzyme conjugate, wash buffer concentrate, TMB chromogen/substrate solution and stop solution.

Six levels of calibrators (level 1 to 6) containing various concentrations of 25OHD come with the kit. Each bottle contains a horse serum based with active ingredients of 0.09% ProClin 950 and 0.09% sodium azide.. Calibrators are 25-OH vitamin D3 spiked and are supplied in 1 mL bottle for each level and are liquid material ready to use.

Two levels of controls (level 1 and 2) containing a low and a high concentration of 25OHD come with the kit. Each bottle contains a horse serum based with active ingredients of 0.09% ProClin 950 and 0.09% sodium azide.. Controls are supplied in 1 mL bottle for each level and are liquid material ready to use.

J. Substantial Equivalence Information:

1. Predicate device name(s):

IDS 25-Hydroxy Vitamin D EIA

2. Predicate 510(k) number(s):

k021163

3. Comparison with predicate:

2

{2}

|  Item | Candidate device- 25-OH Vitamin D ELISA k123660 | Predicate device- IDS 25-Hydroxy Vitamin D EIA k021163  |
| --- | --- | --- |
|  Intended Use/Indication for use | For the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations. | same  |
|  Sample type | Serum and plasma | same  |
|  Test format | 96-well microplate assay | same  |
|  Assay components | Micro-titer plate; Calibrators; Biotin labeled 25-OH Vitamin D; Avidin-Conjugate; Assay buffer; Wash buffer; Substrate; Stop solution | same  |
|  Instrument | ELISA plate reader | same  |
|  Measuring wavelength | 450/620nm | same  |
|  Approximate assay time | 3.5 hours | same  |
|  Antigen used in calibrators | 25-OH Vitamin D3 | same  |
|  Assay principle | Competitive immunoassay | same  |
|  Reagent storage temperature | 2-8 °C | same  |
|  Interpretation of results | Standard curve | same  |
|  Traceability | Standardized using UV quantification of 25-OH vitamin | same  |
|  Specificity | 25-OH Vitamin D and other hydroxylated vitamin D metabolites | same  |

{3}

|  Antibody Calibrators | Monoclonal sheep anti-25-OH-Vitamin D IgG antibody | Polyclonal sheep anti-25-OH-Vitamin D IgG antibody  |
| --- | --- | --- |
|  Assay range | 4.0 – 120 ng/mL | 2.4 – 144 ng/mL  |
|  Sample Volume | 20 μl | 25 μl  |

K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods
CLSI EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
CLSI EP7-A2 Inference Testing in Clinical Chemistry
CLSI EP9-A Method Comparison and Bias Estimation Using Patient Samples
CLSI EP17-A Protocols for Determination of Limits of Detection and Limits of Quantification
CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
EN 13612:2002: Performance evaluation of in vitro diagnostic medical devices
EN 13640:2002: Stability testing of in vitro diagnostic reagents

L. Test Principle:
The method for quantitative determination of 25-OH Vitamin D is a direct, competitive immunosorbent assay (ELISA). In the first analysis step, the calibrators and patient samples are diluted with biotin-labeled 25-OH vitamin D and added to microplate wells coated with monoclonal sheep anti-25-OH Vitamin D antibodies. During the incubation an unknown amount of 25-OH vitamin D in the patient sample and a known amount of biotin labeled 25-OH vitamin D compete for the antibody binding sites in the microplate wells plate. Unbound 25-OH vitamin D is removed by washing. For the detection of bound biotin-labeled 25-OH vitamin D, a second incubation is performed using peroxidase-labeled streptavidin. In a third incubation using the peroxidase substrate tetramethylbenzidine (TMB) the bound peroxidase promotes a color reaction. The color intensity is inversely proportional to the 25-OH vitamin D concentration in the sample. Results for the samples can be calculated directly using a standard curve.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
The precision/reproducibility of the test was investigated following CLSI standard EP05-A2. Intra-assay and inter-assay coefficients of variation (CV) were determined using 8 natural human serum samples with different concentrations of 25OHD. Two of the 8 serum samples were spiked to achieve higher levels of 25OHD. The intra-assay and inter-assay CVs are based on 40 determinations, with inter-assay performed in 10 different runs over 5 different days (with 4 replicates per run). The following precision results were obtained:

{4}

|  Intra-assay precision  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  n = 40 | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | Sample 7 | Sample 8  |
|  Mean value (ng/mL) | 4.1 | 16.8 | 24.6 | 28.8 | 42.9 | 46.3 | 68.7 | 93.3  |
|  Standard deviation | 0.5 | 0.9 | 1.7 | 2.1 | 1.8 | 2.8 | 3.5 | 6.3  |
|  Coeff. of variation (CV,%) | 12.4 | 5.5 | 6.9 | 7.4 | 4.2 | 6.0 | 5.1 | 6.7  |
|  Inter-assay precision  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  n = 40 | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | Sample 7 | Sample 8  |
|  Mean value (ng/mL) | 5.8 | 16.6 | 22.3 | 34.8 | 43.5 | 55.3 | 67.8 | 94.4  |
|  Standard deviation | 0.9 | 1.3 | 1.8 | 3.1 | 3.0 | 3.7 | 5.8 | 7.8  |
|  Coeff. of variation (CV,%) | 16.2 | 7.8 | 8.1 | 8.8 | 7.0 | 6.7 | 8.6 | 8.3  |

The lot-to-lot reproducibility of the test was investigated following CLSI standard EP05-A2. Inter-lot coefficients of variation (CV) were determined using 8 natural human serum samples with different concentrations of 25OHD. Two of the 8 serum samples were spiked to achieve higher levels of 25OHD. The inter-lot CVs are based on 32 determinations performed in 8 different runs on 4 different lots. The following reproducibility results were obtained:

{5}

|  Lot to lot reproducibility  |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  n = 32 | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | Sample 7 | Sample 8  |
|  Mean value (ng/mL) | 7.3 | 18.5 | 24.8 | 37.4 | 47.6 | 58.0 | 74.3 | 97.3  |
|  Standard deviation | 0.89 | 1.89 | 1.81 | 2.29 | 4.25 | 4.07 | 6.59 | 9.26  |
|  Coeff. of variation (CV,%) | 12.2 | 10.2 | 7.3 | 6.1 | 8.9 | 7.0 | 8.9 | 9.5  |

b. Linearity/assay reportable range:

The linearity of the test was investigated following CLSI standard EP6-A. Sample preparations were prepared by mixing of natural low patient sample (2 ng/mL) and high 129 ng/mL patient blood samples to create 9 different inter-mixture concentrations. The sample preparations covered the concentration range of 2 to 129 ng/mL. All samples were tested in double determinations per sample preparation, the mean of the two determinations for each sample was calculated and polynomial regression was performed of observed results vs. expected results. The linear regression generated was:  $\mathrm{Y} = 0.98\mathrm{X} - 0.57$ ,  $\mathrm{r}^2 = 0.999$

Based on the results of the linearity study sponsor claimed that the candidate assay is linear from 2.0 to  $129\mathrm{ng / mL}$

The 25-OHD assay has a measuring range of 4.0 to  $120\mathrm{ng / mL}$

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

Calibrators and controls are traceable to concentrations determined by UV spectrophotometric analysis. An in-house stock solution is prepared gravimetrically by reconstituting a vial of 25-OH vitamin D3 with ethanol. The antigen concentration of the ethanolic stock is spectrophotometrically calculated using the OD coefficient of 18.2 at  $264\mathrm{nm}$  to calculate the concentration from the absorbance value. The ethanolic stock of antigen is used to build an intermediate stock volumetrically by dilution into horse serum based on the spectrophotometric reading of the ethanolic stock solution. The intermediate stock is used in the manufacturing of lot specific calibrators and controls volumetrically.

Value assignments

Target ranges for control materials are: C1 (low):  $15\mathrm{ng / mL}$ $(10 - 25\mathrm{ng / mL})$  and C2

{6}

(high): 40 ng/mL (25-60 ng/mL). Target values are confirmed using two commercially available devices and a HPLC method. Testing of the controls and assignment of values involves multiple replicates with multiple operators to generate a total of 40 values. The mean obtained values are assigned as control values. The controls are labeled with the assigned values as mean C1±50% and C2±30%.

Calibrator value assignments are based on an internal procedure. The calibrator levels for each batch are confirmed by running a minimum of eight quality control sera covering the whole concentration range of the calibration curve measured in multiple replicates with multiple operators over days. The initial value assignment for calibrators was performed using two commercially available devices and a HPLC method. The final calibrators values are then verified &amp; assigned by adjusting their initial values to meet the specified ranges when tested against the commercially available assays. Once confirmation of the calibrator values is established, the new calibrators are tested again in bulk and final. The target ranges of the calibrators are as follows:

Calibrator 1 = 0 ng/mL, Calibrator 2 = 4 ng/mL, Calibrator 3 = 10 ng/mL, Calibrator 4 = 25 ng/mL, Calibrator 5 = 60 ng/mL, Calibrator 6 = 120 ng/mL,

## Stability

Stability studies were conducted using accelerated stability study. The predicted shelf-life, based on results of an accelerated testing at 37°C, is at least 12 months at 2-8°C for the control materials and the calibrators. Real-time stability testing is ongoing to support the predicted shelf-life. Open-vial stability is 3 months when stored at 2-8°C. Biotin was found to be stable for at least 2 weeks when diluted to working strength. Stability study protocol has been reviewed and found to be adequate.

## d. Detection limit:

Limit of blank (LoB), limit of detection (LoD) and limit of quantification (LoQ) were investigated following CLSI standard EP17-A. LoB, LoD, and LoQ studies were performed with three lots of reagents. LoB was determined using the mean of 60 replicates for calibrator 1 (0 ng/mL) and the mean of 20 replicates for calibrator 2 (4 ng/mL). LoB was found to be 0.54 ng/mL. LoD was determined as the LoB + 1.645 times the standard deviation of 200 determinations from 5 samples in the low range of 2 to 10 ng/mL, measured in 5 independent runs with 8 replicates per run. The mean LoD was determined to be 2.1 ng/mL. LoQ is defined as the lowest concentration at which the curve line crosses the 20% CV line and was determined from a plot of the mean concentrations (X-axis) vs. % CVs (Y-axis). LoQ was determined to be 4.0 ng/mL.

Sponsor claims that the candidate assay has a measuring range of 4.0 to 120 ng/mL.

{7}

# e. Analytical specificity:

# Cross reactivity study:

Cross reactivity was investigated to evaluate the potential cross-reactants. A 25-OH Vitamin D free sample was aliquoted and spiked with potential cross reacting Vitamin D metabolites at the concentrations listed below. The following cross-reactions were calculated based on the following equation and results summarized in the table below:

Cross reactivity  $(\%) = \frac{\mathrm{x}(\mathrm{ng} / \mathrm{mL}) \times 100}{\mathrm{c}(\mathrm{ng} / \mathrm{mL})}$

with

$\mathbf{x} =$  observed value

$\mathbf{c} =$  spiked concentration of potential cross reactant

|  Potential cross reacting substance | Concentration spiked | Concentration observed | Cross reactivity  |
| --- | --- | --- | --- |
|   |  ng/mL | ng/mL | %  |
|  25-OH Vitamin D3 | 10.0 | 10.0 | 100  |
|  25-OH Vitamin D2 | 25.0 | 24.3 | 100  |
|  24,25-OH Vitamin D3 | 100 | 0.3 | 0.3  |
|  Cholecalciferol (Vitamin D3) | 10,000 | 3.4 | 0.03  |
|  Ergocalciferol (Vitamin D2) | 10,000 | 5.1 | 0.05  |
|  1,25-OH Vitamin D3 | 10.0 | 4.3 | 45  |
|  1,25-OH Vitamin D2 | 10.0 | 19.8 | 212  |
|  3-epi-25-OH Vitamin D3 | 10.0 | 1.7 | 17  |

# Interference study:

To investigate the influence from hemoglobin, triglycerides and bilirubin, 3 different sera at different 25-OH vitamin D concentrations (9 to  $76\mathrm{ng / mL}$ ) were spiked with potential interfering substances and were incubated with the test system according to the package insert. Non-significant interference is defined as  $&lt; 10\%$  bias between the spiked and unspiked sample. No significant interference was observed for concentrations of up to  $750\mathrm{mg / dL}$  for hemoglobin,  $2000\mathrm{mg / dL}$  for triglyceride,  $40\mathrm{mg / dL}$  for bilirubin,  $400\mathrm{mg / dL}$  for cholesterol,  $1000\mathrm{mg / dL}$  for biotin, and  $10.0\mathrm{mg / mL}$  ascorbic acid.

# f. Assay cut-off:

N/A

{8}

# 2. Comparison studies:

# a. Method comparison with predicate device:

The method comparison against the predicate device was performed following CLSI standard EP09-A2. 240 human serum samples were obtained from different sources for the method comparison study. The samples, ranging from 4.1 to  $119.1\mathrm{ng / mL}$ , spanned the claimed measuring range of the candidate assay. To ensure that the tested concentrations of 25-OH Vitamin D are distributed across the reportable dynamic range, 36 samples  $(15\%)$  in the set were spiked to represent difficult to achieve concentrations between  $46.9 - 119.1\mathrm{ng / mL}$ .

Results of linear regression analysis are shown in the table below.

|  n | 240  |
| --- | --- |
|  Concentration range (predicate) | 4.1 – 110.4 ng/mL  |
|  Concentration range (candidate) | 4.1 – 119.1 ng/mL  |
|  Regression equation (y = candidate, x = predicate) | y = 1.08 x – 0.78  |
|  95% C.I. of Intercept | -0.06 – 1.63  |
|  95% C.I. of Slope | 1.06 – 1.11  |
|  Correlation coefficient 95% C.I. of R | 0.9858 0.9817 – 0.9890  |

# b. Matrix comparison:

The usability of plasma was investigated using sample pairs each of serum and corresponding plasma (EDTA, and Li-heparin) from donors. To ensure that the tested concentrations of 25- OH Vitamin D are distributed across the reportable dynamic range, 3 sample pairs in the set were spiked with 25-OH Vitamin D stock solution of 25-OH Vitamin D3. Passing-Bablok regression was calculated for the comparison of serum (X) to plasma (Y) and results are summarized below:

|   | EDTA plasma | Li-heparin plasma  |
| --- | --- | --- |
|  n | 38 | 38  |
|  Concentration range (serum) | 8.6 – 105.1 ng/ml | 8.6 – 105.1 ng/ml  |
|  Concentration range (plasma) | 8.8 – 104.1 ng/ml | 8.5 – 104.1 ng/ml  |
|  Regression equation with (95% C.I. of intercept) (95% C.I. of slope) | y = 0.99 x + 0.29 (-0.37 – 1.18) (0.93 – 1.02) | y = 0.97 x + 0.55 (-0.65 – 1.33) (0.93 – 1.04)  |
|  Coefficient of determination R2 | 0.996 | 0.993  |

{9}

10

3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
N/A

4. Clinical cut-off:
N/A

5. Expected values/Reference range:

The levels of 25-OH vitamin D were analyzed in a panel of 206 serum samples from healthy subjects (70 men and 139 women with an average age of 63 years; age range: 21 – 99 years) from a commercial source from the US. Samples are known to be from healthy blood donors from the mid-West region of the United States and drawn in October. The results are shown in the table below.

|  n = 206 | 25-OH Vitamin-D ELISA  |
| --- | --- |
|   |  ng/mL  |
|  Minimum | < 4  |
|  Maximum | 64.8  |
|  Mean | 20.8  |
|  Median | 19.4  |
|  2.5% percentile | 5.4  |
|  97.5% percentile | 47.0  |

The sponsor states that the data is provided for guidance only. It is important for each laboratory to establish its own reference ranges, representative of its typical population. Also, published studies representing the local population can be taken into consideration.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MRG/K123660](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MRG/K123660)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
