← Product Code [MMI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI) · K230648

# Access hsTnI (K230648)

_Beckman Coulter, Inc. · MMI · Dec 4, 2023 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K230648

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [MMI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI.md)
- **Decision Date:** Dec 4, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1215
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Access hsTnl is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access 2 Immunoassay Systems to aid in the diagnosis of myocardial infarction (MI).

## Device Story

Access hsTnI is a two-site immunoenzymatic 'sandwich' assay performed on the Access 2 Immunoassay System. Input: human serum or plasma sample. Process: sample mixed with monoclonal anti-cTnl antibody conjugated to alkaline phosphatase and surfactant-containing buffer; paramagnetic particles coated with monoclonal anti-cTnl antibody added; cTnl binds to solid phase; unbound materials washed away via magnetic field; chemiluminescent substrate added; light production measured by luminometer. Output: light signal proportional to analyte concentration, automatically calculated via stored calibration. Used in clinical laboratories by trained personnel. Software upgrade modifies washing efficiency (3 probe washes and 1 NaOH exposure) and reaction vessel mixing cycles to improve performance. Results aid clinicians in diagnosing myocardial infarction.

## Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP05-A3 (n=80 replicates per platform). Linearity and detection limits (LoB, LoD, LoQ) evaluated per CLSI EP06-A2 and EP17-A2. Method comparison (n=182 samples total) using Passing-Bablok regression showed high correlation (R=0.997-0.998) with the predicate. Thermal sensitivity studies confirmed performance stability across 18°C-28°C. No new clinical studies were required as the modification did not impact clinical performance claims or cut-offs established in K172787.

## Technological Characteristics

Paramagnetic particle, chemiluminescent sandwich immunoassay. Materials: mouse monoclonal anti-human cTnl antibody, sheep monoclonal anti-human cTnl detection antibody. System: automated immunoassay instrument (Access 2). Connectivity: standalone. Software: updated Assay Protocol File (APF) for wash/mix cycles. Standards: CLSI EP05-A3, EP06, EP17-A2, EP09c.

## Regulatory Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

## Predicate Devices

- Access hsTnI ([K172787](/device/K172787.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

## I Background Information:

A 510(k) Number

K230648

B Applicant

Beckman Coulter, Inc.

C Proprietary and Established Names

Access hsTnI

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  MMI | Class II | 21 CFR 862.1215 - Creatine Phosphokinase/Creatine Kinase Or Isoenzymes Test System | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

Modification of an existing device

B Measurand:

Cardiac Troponin I

C Type of Test:

Quantitative Immunoassay

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K230648 - Page 2 of 7

# III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

Access hsTnI is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access 2 Immunoassay Systems to aid in the diagnosis of myocardial infarction (MI).

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

D Special Instrument Requirements:

Access 2 Immunoassay Systems

# IV Device/System Characteristics:

A Device Description:

Same as described in K172787.

B Principle of Operation:

The principle of operation is unchanged from the previous clearance described in K172787. The device has been modified to resolve assay carryover by improving the washing capabilities of the candidate test system.

V Substantial Equivalence Information:

A Predicate Device Name(s):

Access hsTnI

B Predicate 510(k) Number(s):

K172787

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K230648 | K172787  |
| --- | --- | --- |
|  Device Trade Name | Access hsTnI | Access hsTnI  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Quantitative determination of cardiac | same  |

{2}

|   | troponin I (cTnI) levels in human serum and plasma to aid in the diagnosis of myocardial infarction (MI). |   |
| --- | --- | --- |
|  Assay Principle | Chemiluminescent sandwich assay | Same  |
|  Sample Type | Serum and Lithium Heparin plasma | Same  |
|  Analytical Measuring Range | 2.0 pg/mL to 27,027 pg/mL | Same  |
|  Open Reagent Pack Stability | Stable at 2 to 10°C for 64 days after opening | Same  |
|  General Device Characteristic Differences |  |   |
|  System software | RV mix occurs at earlier cycle. | RV mix only occurs at the end of the 4thcycle.  |
|  Assay Protocol | Three (3) probe washes and one (1) NAOH exposure after reagent probe’s final contact with reaction solutions and RV mix. | One (1) probe wash after probe’s final interaction with reaction solutions and RV mix.  |
|  Dilution factor/Recovery & Extended Recovery Range | 1:5 with new limitation to address potential carryover | 1:10  |

VI Standards/Guidance Documents Referenced:

CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline

CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition

CLSI EP05-A3: Evaluation of Precision of Qualitative Measurement Methods Procedures; Approved Guideline – Third Edition

CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Third Edition

K230648 - Page 3 of 7

{3}

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

The sponsor evaluated precision of the Access hsTnI assay on the Access 2 Immunoassay system in several studies following the recommendations in CLSI EP5-A3.

Four lithium heparin plasma samples with troponin concentrations spanning the assay range were each tested on a single Access 2 in duplicate, in two runs per day, over 20 days for a total of 40 runs and a minimum of 80 replicates per platform using a single reagent pack lot at a single site. All four samples used were unique pools of native patient samples. The two highest concentration samples (1,200 and 15,000 pg/mL targets) were spiked with a natively sourced troponin.

The within-laboratory (total) imprecision includes within-run (repeatability), between-run, and between-day variance components. Variance components analysis was used to estimate within-run, between run, and between-day, %CV and SD. Results for this study are shown below:

|  Access 2 Imprecision Analysis and Results Summary  |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  hsTnI Summary: |   | Within-Run |   | Between-Run |   | Between-Day |   | Within-Laboratory (Total)  |   |
|  Target Sample Concentration (pg/mL) | Sample Concentration Mean (pg/mL) | SD | CV % | SD | CV % | SD | CV % | SD | CV%  |
|  9.0 | 11.49 | 0.249 | 2.17 | 0.1 | 0.79 | 0.4 | 3.9 | 0.52 | 4.54  |
|  18.8 | 23.4 | 0.592 | 2.53 | 0.2 | 0.94 | 0.5 | 2.0 | 0.79 | 3.39  |
|  1,200 | 924.6 | 19.956 | 2.16 | 2.9 | 0.32 | 18.7 | 2.0 | 27.53 | 2.98  |
|  15,000 | 12,844.5 | 411.617 | 3.2 | 270.2 | 2.1 | 202.6 | 1.6 | 532.45 | 4.15  |

The results of the study support that the enhanced washing protocol did not impact the precision of the test system.

2. Linearity:

The sponsor evaluated the linearity of the Access hsTnI assay on the Access 2 Instrument following the recommendations in CLSI EP06-A2 using both serum and lithium heparin plasma sample types covering the full range of the assay. The results of the study demonstrate that the mitigation (the enhanced washing procedure) did not impact the linearity of the test system.

Hook Effect:

Same as described in K172787.

K230648 - Page 4 of 7

{4}

Dilution Recovery:
The sponsor has changed the validated dilution scheme and provided data to support a dilution ratio of 1:5.

3. Analytical Specificity/Interference:
The Analytical Specificity/Interference performance was evaluated in K172787 and the claims are unchanged.

4. Assay Reportable Range:
2.0 - 27,027 pg/mL.

The sponsor performed studies to evaluate intra-assay carryover and included a limitation in the labeling describing the carryover observed. In the study, when a sample with cTnI &gt; 150,000 pg/mL (ng/L) was tested on an Access 2 system, intra-assay carryover was observed if an Access hsTnI was tested after a high cTnI sample. The extent of carryover observed was directly proportional to the cTnI concentration that was present in the high sample. In the studies, the estimated carryover was 3-5 pg/mL (ng/L) from a high sample at 270,000 pg/mL (ng/L) and 5-8 pg/mL (ng/L) from a high sample at 500,000 pg/mL (ng/L).

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability and stability is the same as described in K172787, except the stability of frozen samples has been updated to support the storage of frozen samples at -20°C for up to 180 days.

6. Detection Limit:
Studies were performed to determine the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) for the Access hsTnI assay on the Access 2 instrument using the recommendations in CLSI EP17-A2. Blank samples were used in the LoB studies, lithium heparin samples and serum samples were used in the LoD and LoQ studies. The results provided for LoB, LoD, and LoQ demonstrate that the modification to the assay workflow did not impact the claimed measuring range of 2.0-27,027 pg/mL.

7. Assay Cut-Off:
The Assay Cut-off was evaluated in K172787.

B Comparison Studies:

1. Method Comparison with Predicate Device:
Two method comparison studies were completed to assess the impact of the modification using a protocol based on CLSI EP09c. The first study evaluated 40 Lithium Heparin Plasma and 49 Serum samples across three (3) Access 2 instruments. The second method comparison

K230648 - Page 5 of 7

{5}

study was performed to evaluate 51 Lithium Heparin Plasma and 53 Serum samples across two (2) Access 2 instruments. Each sample was measured using the proposed test system and the predicate test system. Each sample was measured in duplicate. The first replicate result from each sample was utilized to fit a Passing-Bablok linear regression model. Representative results are shown below:

|  Instrument | Sample Type | Sample # | Claimed Range | Range of Samples | Regression Equation | Correlation coefficient/R  |
| --- | --- | --- | --- | --- | --- | --- |
|  Access 2 | Plasma | 41 | 2.0-27,027 | 8.4 to 24112 | y= 0.98514x + -1.293997 | 0.998  |
|   |  Serum | 51 |   | 2.2 to 23977 | y=0.98254 + -0.143634 | 0.997  |

The results of the method comparison study (including analyses around the cutoffs) show that the performance of the assay is unchanged since its clearance in K172787.

2. Thermal Method Comparison:

Two Thermal Sensitivity studies were completed to assess the impact of the design change on the Thermal Sensitivity of the hsTnI immunoassay on the Access 2 Immunoassay system using a protocol based on CLSI EP09c. In the first study, more than 80 samples (Lithium Heparin Plasma and Serum) were analyzed across 2 Access 2 instruments. Each sample was measured at each of three different ambient room temperature conditions and calibration temperature conditions (18°C, 23°C and 28°C). The second study evaluated more than 80 samples (Lithium Heparin Plasma and Serum) across 2 Access 2 instruments. The results were analyzed using Passing-Bablok linear regression comparing each potential run temperature/calibration temperature combination.

The results of the thermal method comparison studies support that the performance of the assay is unchanged since its clearance in K172787.

3. Matrix Comparison:

The sponsor provided data to support the use of lithium heparin samples and serum samples.

C Clinical Studies:

1. Clinical Sensitivity:

The sponsor provided information to support that the modification to the candidate device did not impact the clinical performance claims described in K172787.

K230648 - Page 6 of 7

{6}

2. Clinical Specificity:

See Clinical Sensitivity above.

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable.

D Clinical Cut-Off:

The sponsor provided information to support that the modification to the candidate device did not impact the clinical cutoffs described in K172787.

E Expected Values/Reference Range:

The sponsor provided information to support that the modification to the candidate device did not impact the 99th percentile upper reference limits described in K172787.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K230648 - Page 7 of 7

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K230648](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K230648)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
