← Product Code [MMI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI) · K121214

# ACCESS ACCUTNI+3 REAGENT AND ACCESS ACCUTN1+3 CALIBRATORS FOR USE ON THE ACCESS 2 IMMUNOASSAY SYSTEM (K121214)

_Beckman Coulter, Inc. · MMI · Jun 14, 2013 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K121214

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [MMI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI.md)
- **Decision Date:** Jun 14, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1215
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Access AccuTnI+3 Reagent is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access 2 Immunoassay System to aid in the diagnosis of myocardial infarction. The Access AccuTnI+3 Calibrators are intended to calibrate the Access AccuTnI+3 Reagent for the quantitative determination of cardiac troponin I (cTnl) levels in human serum and plasma using the Access 2 Immunoassay System to aid in the diagnosis of myocardial infarction. The Access 2 Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

## Device Story

The Access AccuTnI+3 system is an automated, paramagnetic particle, chemiluminescent immunoassay used in clinical laboratories to measure cardiac troponin I (cTnI) in human serum and plasma. The system utilizes mouse monoclonal anti-human cTnI antibodies on magnetic particles and an alkaline phosphatase conjugate to form a sandwich immunoassay. The Access 2 Immunoassay System processes samples, performs the assay, and provides quantitative results to clinicians. The device includes a thermal algorithm to monitor and manage internal instrument temperature, ensuring result accuracy within specified ambient ranges. Clinicians use these results to aid in the diagnosis of myocardial infarction. The system benefits patients by providing rapid, quantitative cardiac biomarker data to support timely clinical decision-making.

## Clinical Evidence

Prospective multicenter study of 1,929 patients presenting with chest pain/ischemic symptoms. Adjudicated diagnosis based on 2007 Universal Definition of MI. Sensitivity/specificity evaluated at 0.03 ng/mL (feasibility cutoff), 0.02 ng/mL (99th percentile URL), and 0.04 ng/mL. At 0.03 ng/mL, baseline sensitivity was 87.4% (95% CI: 82.6-91.2) and specificity 89.3% (95% CI: 87.7-90.7). At 0.02 ng/mL, baseline sensitivity was 92.1% (95% CI: 88.1-95.1) and specificity 84.3% (95% CI: 82.5-86.0).

## Technological Characteristics

Paramagnetic particle, chemiluminescent immunoassay. Reagents: mouse monoclonal anti-human cTnI antibodies, alkaline phosphatase conjugate, TRIS/ACES buffered saline, BSA matrix, ProClin 300. System: automated discrete photometric chemistry analyzer. Connectivity: microcomputer-controlled, random/continuous access. Software: Version 3.3 includes thermal algorithm for temperature management.

## Regulatory Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

## Predicate Devices

- ADVIA Centaur TnI-Ultra Assay and ADVIA Centaur TnI-Ultra Calibrators ([K053020](/device/K053020.md))
- Access 2 Immunoassay System ([K922823](/device/K922823.md)/A007)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k121214

B. Purpose for Submission:
New Device

C. Measurand:
Cardiac troponin I (cTnI)

D. Type of Test:
Quantitative paramagnetic-particle chemiluminescent immunoassay

E. Applicant:
Beckman Coulter, Inc.

F. Proprietary and Established Names:
Access AccuTnI+3 Reagent and Access AccuTnI+3 Calibrators for use on the Access 2 Immunoassay System

G. Regulatory Information:

1. Regulation section:
21 CFR 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system
21 CFR 862.1150 Calibrator
21 CFR 862.2160 Discrete photometric chemistry analyzer for clinical use

2. Classification:
Class II (21 CFR 862.1215 and 21 CFR 862.1150),
Class I (21 CFR 862.2160)

3. Product code:
MMI, Immunoassay method, troponin subunit
JIT, Calibrator, secondary
JJE, Analyzer, chemistry (photometric, discrete), for clinical use

4. Panel:
Clinical Chemistry (75)

H. Intended Use:

1. Intended use(s):
See Indication(s) for use

{1}

2. Indication(s) for use:
The Access AccuTnI+3 Reagent is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access 2 Immunoassay Systems to aid in the diagnosis of myocardial infarction.

The Access AccuTnI+3 Calibrators are intended to calibrate the Access AccuTnI+3 Reagent for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access 2 Immunoassay Systems to aid in the diagnosis of myocardial infarction.

The Access 2 Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

3. Special conditions for use statement(s):
For prescription use, for in vitro diagnostic use, not to be used for risk stratification

4. Special instrument requirements:
For use on the Access 2 Immunoassay System

I. Device Description:
The Access AccuTnI+3 Reagent packs contain:
- Paramagnetic particles coated with mouse monoclonal anti-human cardiac troponin I suspended in TRIS buffered saline, with surfactant, bovine serum albumin (BSA) matrix, &lt;0.1% sodium azide, and 0.1% ProClin 300
- 0.1N NaOH
- TRIS buffered saline, surfactant, &lt;0.1% sodium azide and 0.1% ProClin 300
- Mouse monoclonal anti-human cTnI alkaline phosphatase conjugate diluted in ACES buffered saline, with surfactant, BSA matrix, protein (bovine, goat, mouse), &lt;0.1% sodium azide, and 0.25% ProClin 300

The Access AccuTnI+3 Calibrators contain multi-point calibrators for use with the Access AccuTnI+3 Reagent. Individual vials contain zero or approximately 0.3, 1.2, 5.0, 25 and 100 ng/mL (ng/L) of recombinant cardiac troponin I complex, respectively, in a buffered BSA matrix, with surfactant, &lt;0.1% sodium azide, and 0.1% ProClin 300. The calibrators are sold separately.

The Access 2 Immunoassay System is a microcomputer controlled, random-access and continuous-access instrument. The instrument performs enzyme immunoassays utilizing paramagnetic particle solid phase and chemiluminescent detection. A luminometer measures the amount of light generated by the reaction. The Access 2 is designed to be used with numerous different immunoassays. The system software was designed such that immunoassays can be added to the system without changing the system software. A separate assay-specific protocol file, the APF, is loaded into

2

{2}

the system. The APF contains all assay specific parameters used to process a particular assay.

To correct thermal sensitivity from ambient temperature fluctuations that could affect the accuracy of troponin test results, the sponsor developed a software algorithm that normalizes troponin results. This solution was implemented through a combination of system and operating software changes.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

ADVIA Centaur TnI-Ultra Assay

ADVIA Centaur TnI-Ultra Calibrator

Access 2 Immunoassay System

2. Predicate k number(s):

k053020 for the ADVIA Centaur TnI-Ultra Assay and ADVIA Centaur TnI-Ultra Calibrator; k922823/A007 for the Access 2 Immunoassay System

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Access AccuTnI+3 Reagent for use on the Access 2 Immunoassay System | Predicate Device (k053020)  |
|  Intended Use | in vitro diagnostic method for the quantitative measurement of cardiac TnI in serum and plasma | Same  |
|  Assay Principle | Chemiluminescent sandwich immunoassay | Same  |
|  Test System | Automated immunoassay instrument | Same  |
|  Primary Reagent Materials | Solid phase magnetic particles, anti cTnI antibodies | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Access AccuTnI+3 Reagent for use on the Access 2 Immunoassay System | Predicate Device (k053020)  |
|  Indications For Use | Not for risk stratification use | For risk stratification use  |
|  Sample Types | Serum and heparinized plasma | Serum, heparinized plasma and EDTA plasma  |
|  Instrument | Access 2 Immunoassay System with thermal algorithm capability | ADVIA Centaur System  |
|  Specific Reagent | Mouse monoclonal anti-human | Polyclonal goat anti-  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Access AccuTnI+3 Reagent for use on the Access 2 Immunoassay System | Predicate Device (k053020)  |
|  Materials | cTnI alkaline phosphatase conjugate, magnetic particles coated with mouse monoclonal anti-human cTnI | cTnI antibody labeled with acridinium ester, 2 biotinylated mouse monoclonal anti-cTnI antibodies, magnetic particles conjugated with streptavidin  |
|  Acute Myocardial Infarction (AMI) Cut-Off | 0.03 ng/mL validated based on clinical trial outcome | 0.9 ng/mL per WHO-defined cutoff  |
|  Upper Reference Limit | 99th percentile of 0.02 ng/mL | 99th percentile of 0.04 ng/mL  |
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Access AccuTnI+3 Calibrators | Predicate Device (k053020)  |
|  Intended Use | Intended to calibrate the Access AccuTnI+3 Reagent | Intended to calibrate the ADVIA Centaur TnI-Ultra assay  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Access AccuTnI+3 Calibrators | Predicate Device (k053020)  |
|  Calibrator Materials | Recombinant troponin complex in buffered BSA | Bovine cTnI in goat serum  |
|  Calibrator Number And Type | Six Liquid: approximately 0, 0.3, 1.2, 5.0, 25 and 100 ng/mL (with no master curve) | Two lyophilized: high and low (use with Master Curve)  |

# K. Standard/Guidance Document Referenced (if applicable):

- Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline (EP5-A2)
- Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (EP6-A)
- Interference Testing in Clinical Chemistry; Approved Guideline (EP7-A2),
- Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline (EP17-A2)
- Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (EP25-A)
- Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline (C28-A3c)

{4}

L. Test Principle:

The Access AccuTnI+3 Reagent is a two-site immunoenzymatic (“sandwich”) assay. Monoclonal anti-cTnI antibody conjugated to alkaline phosphatase is added to a reaction vessel along with a surfactant-containing buffer and sample. After a short incubation, paramagnetic particles coated with monoclonal anti-cTnI antibody are added. The human cTnI binds to the anti-cTnI antibody on the solid phase, while the anti-cTnI antibody - alkaline phosphatase conjugate reacts with different antigenic sites on the cTnI molecules. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos® 530 is added to the vessel and light generated by the reaction is measured. The light production is directly proportional to the concentration of cTnI in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

The sponsor evaluated precision in several studies based on CLSI EP5-A2.

The first 3 studies were conducted at one of 3 ambient temperature conditions. The studies performed at 18°C and 28°C were performed internally. A low spiked patient pool (lp), 3 commercial controls (c), and one high patient pool (hp), were assayed in replicates of 2, 2 shifts per day, 2 runs per shift, for a total of 21 shifts (42 runs) over 13 days. These studies were performed on 1 instrument and 1 reagent lot.

The third study was performed externally under existing laboratory conditions (measured temperatures ranged from 20°C to 24°C). Six commercial controls (ranging from 0.02 ng/mL to 15 ng/mL) were run in duplicate, 2 runs per day for 20 days for a total of 40 runs. This study was performed on 1 reagent lot and 1 instrument.

The fourth study was a confirmation study that was run across multiple ambient temperatures ranging from 18°C to 28°C. The sponsor used a low spiked patient pool, 3 commercial controls, and one high patient pool. These samples were tested in duplicate in 2 runs per shift, 2 shifts per day, over 14 days. One reagent lot and on one instrument was used in this study.

The following data are representative of the results from the precision studies and are presented in the labeling:

5

{5}

|  Mean | Within-Run (%CV) | Between-Run (%CV) | Total Imprecision (%CV)  |
| --- | --- | --- | --- |
|  0.07 (lp) | 5 | 5 | 7  |
|  0.91 (c) | 2 | 5 | 5  |
|  1.96 (c) | 2 | 5 | 5  |
|  13.97 (c) | 2 | 5 | 5  |
|  56.36 (hp) | 5 | 4 | 6  |

The sponsor also determined the precision of the device using 3 low-level natural patient sample pools, 2 instrument, 2 reagent lots, multiple ambient temperatures ranging from  $18^{\circ}\mathrm{C}$  to  $28^{\circ}\mathrm{C}$  and multiple calibration cycles over 14 days. The samples were assayed in replicates of 2 per run, 2 runs per shift and 2 shifts per day. The results of the study are summarized below:

|  Mean | Total imprecision (SD, ng/mL) | Total imprecision (%CV)  |
| --- | --- | --- |
|  0.04 | 0.003 | 8  |
|  0.43 | 0.022 | 5  |
|  1.01 | 0.043 | 4  |

Field precision studies: The sponsor performed a precision studies in-house to simulate the use of the device by the end user. This study was designed to incorporate different sources of variability that may be found in the field including: the simultaneous running of other assays; variability in test request type (e.g., panel, reflex, stat, and individual testing); variability of ambient temperatures; and variability of instrument age. The sponsor used quality control material for the study, 3 instruments and 1 lot of reagent. The sponsor cycled the temperature of the laboratory from 18 to  $28^{\circ}\mathrm{C}$  over the course of 6 hours over 3 days of testing. The samples were assayed in a minimum of 44 replicates on each instrument across the 3 days of testing. The results of this study are summarized below:

|  Instrument | Sample | Mean ng/mL | Within-instrument % CV  |
| --- | --- | --- | --- |
|  1 | 1 | 0.056 | 4.1  |
|   |  2 | 0.641 | 4.6  |
|   |  3 | 2.040 | 3.7  |
|   |  4 | 0.383 | 5.4  |
|   |  5 | 8.248 | 5.4  |
|  2 | 1 | 0.055 | 8.7  |
|   |  2 | 0.636 | 4.9  |
|   |  3 | 2.054 | 3.5  |
|   |  4 | 0.389 | 6.1  |
|   |  5 | 8.348 | 3.8  |

{6}

7

|  3 | 1 | 0.053 | 5.0  |
| --- | --- | --- | --- |
|   |  2 | 0.594 | 4.6  |
|   |  3 | 1.874 | 3.2  |
|   |  4 | 0.366 | 4.9  |
|   |  5 | 7.802 | 4.5  |

External Field precision study: The sponsor performed this study to evaluate the precision of the device at external sites. Commercially available quality control materials were run in duplicate per run with 3 lots of reagents on 5 instruments at 5 external sites over 10 days (sample 1 fell below the LoQ of the assay and is not listed). Three sites ran a total of 20 runs, 1 site ran a total of 18 runs and 1 site ran a total of 40 runs. One instrument was used at each site. The results of the study are summarized below:

Reagent lot 1

|  Sample | Means (ng/mL) observed at the 5 sites | Within-lab %CVs observed at the 5 sites  |
| --- | --- | --- |
|  2 | 0.034 to 0.042 | 3.5 to 4  |
|  3 | 0.622 to 0.685 | 2.4 to 4.4  |
|  4 | 0.928 to 1.04 | 2 to 5.2  |
|  5 | 2.586 to 2.813 | 2.5 to 4.3  |
|  6 | 14.19 to 16.19 | 1.9 to 7.4  |

Reagent lot 2

|  Sample | Means (ng/mL) observed at the 5 sites | Within-lab %CVs observed at the 5 sites  |
| --- | --- | --- |
|  2 | 0.032 to 0.040 | 4.6 to 11  |
|  3 | 0.602 to 0.642 | 2.4 to 5.0  |
|  4 | 0.883 to 1.055 | 2.8 to 6.5  |
|  5 | 2.492 to 2.68 | 2.8 to 5.4  |
|  6 | 13.48 to 15.19 | 1.8 to 6.9  |

Reagent lot 3

|  Sample | Means (ng/mL) observed at the 5 sites | Within-lab %CVs observed at the 5 sites  |
| --- | --- | --- |
|  2 | 0.030 to 0.036 | 4.3 to 15  |
|  3 | 0.0594 to 0.645 | 2.7 to 5.7  |
|  4 | 0.880 to 0.996 | 1.9 to 6.0  |
|  5 | 2.455 to 2.65 | 2.4 to 5.0  |
|  6 | 13.82 to 14.85 | 2.1 to 8.2  |

b. Linearity/assay reportable range:
Linearity was evaluated in accordance with the CLSI EP6-A guideline utilizing 2 reagent lots and 2 instruments at 3 different temperature conditions (from 18 to 28 °C) using lithium heparin plasma samples spiked with native

{7}

cardiac troponin I analyte. The sponsor evaluated linearity using 9 pools of varying analyte concentration that spanned the reportable range of the assay. A total of 8 replicates of each high and each low sample were analyzed and 4 replicates of each intermediate mixture of high and low were analyzed with each reagent lot and instrument at each temperature. A total of 12 independent analyses were performed. The polynomial fit was significant for all analyses but the maximum deviation from nonlinearity was 14%. The sponsor claims acceptable linearity from 0.003 to 106.04 ng/mL.

High Dose Hook Effect: The sponsor demonstrated that there was no high dose hook effect from the concentration of the S5 calibrator (≈100 ng/mL) to 2,529 ng/mL.

The reportable range of the device is 0.02 ng/mL to 100 ng/mL.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): The Access AccuTnI+3 Calibrators are standardized to an internal standard. They are identical to the previously cleared (k010429) Access AccuTnI Calibrators.

Calibrators are value assigned using primary reference curves and verified using quality control material and patient samples that must meet specifications. The value assignment process was reviewed and found to be acceptable.

Calibrator stability: The calibrators are stable for 12 months when stored unopened at -20°C. Once opened, the calibrators are stable for 60 days when stores at 2 to 8°C. The calibrator stability protocols were reviewed and found to be acceptable.

d. Detection limit: The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) studies were performed following the recommendations in EP17-A2. Testing was performed over 3 days using 2 reagent lots, 3 instruments, 3 calibration cycles with 3 different temperature conditions (from 18 to 28°C).

To estimate the LoB, 4 blank samples were measured in replicates of 10 on each day of testing on each instrument/reagent lot combination at the different temperature conditions. The LoB was established to be 0.003 ng/mL. The sponsor claims that the LoB is &lt;0.01 ng/mL.

To estimate the LoD, 6 native lithium heparin plasma samples containing low levels of troponin I analyte were measured in 2 runs of 10 replicates each during each day of testing on each instrument/reagent lot combination at the different temperature conditions. The sponsor claims an LoD of 0.01 ng/mL.

8

{8}

To estimate the LoQ, 11 native lithium heparin plasma samples containing low levels of troponin I analyte were measured on each instrument/reagent lot combination at the different temperature conditions.

The sponsor claims an LoQ of 0.02 ng/mL with a performance goal of &lt; 20% CV (within-lab). The lowest concentration with a % CV (within-lab) &lt; 10% was estimated to be 0.04 ng/mL.

The sponsor provided data demonstrating that the LoQ of the device using serum samples at 3 different temperature conditions (from 18 to 28 °C) is identical to the LoQ of the device using lithium heparin plasma samples.

The reportable range of the device is 0.02 ng/mL to 100 ng/mL.

e. Analytical specificity:

Two levels of potential interfering substances were added to lithium heparin plasma pools containing either 0.05 or 0.5 ng/mL troponin at 3 different temperature conditions (from 18 to 28 °C). For the controls, the corresponding solvent was added to the lithium heparin plasma pools containing either 0.05 or 0.5 ng/mL troponin. Control troponin samples (control sample) and samples spiked with the potential interferents (test sample) were tested in replicates of 5 (for the 0.5 ng/mL samples) or 10 (for the 0.05 ng/mL samples) and compared. The sponsor used 1 reagent lot and 6 instruments for this study. The sponsor concluded that the following substances at the listed concentrations did not interfere with the performance of the device.

|  Substance Added | Concentration | Difference observed  |
| --- | --- | --- |
|  Acetaminophen | 3 and 20 mg/dL | <10%  |
|  Acetylsalicylic Acid | 50 and 65 mg/dL | <10%  |
|  Allopurinol | 2 and 40 mg/dL | <10%  |
|  Ambroxol | 8.6 and 40 mg/dL | <10%  |
|  Ampicillin | 1.8 and 5 mg/dL | <10%  |
|  Ascorbic Acid | 4 and 6 mg/dL | <10%  |
|  Atenolol | 0.2 and 1 mg/dL | <10%  |
|  Bilirubin conjugated | 5 and 40 mg/dL | <10%  |
|  Bilirubin unconjugated | 4 and 40 mg/dL | <10%  |
|  Biotin | 10 and 290 ng/mL | <10%  |
|  Caffeine | 2 and 10 mg/dL | <10%  |
|  Captopril | 0.5 and 5 mg/dL | <10%  |
|  Cinnarizine | 4.8 and 40 mg/dL | <10%  |
|  Cocaine | 1 and 2 mg/dL | <10%  |
|  Diclofenac | 2 and 5 mg/dL | <10%  |
|  Digoxin | 2 and 200 ng/mL | <10%  |
|  Dopamine * | 30 and 65 mg/dL | <10% at 30 mg/dL,  |

{9}

|   |  | 13.8% at 65 mg/dL  |
| --- | --- | --- |
|  Erythromycin | 2 and 20 mg/dL | <10%  |
|  Fibrinogen | 100 and 1000 mg/dL | <10%  |
|  Furosemide | 3 and 40 mg/dL | <10%  |
|  Hemoglobin | 200 and 500 mg/dL | <10%  |
|  Human Serum Albumin | 5000 and 6000 mg/dL | <10%  |
|  Ibuprofen | 40 and 50 mg/dL | <10%  |
|  Low MW Heparin | 8 U/mL and 28.8 U/mL | <10%  |
|  Methyldopa | 0.75 and 2.5 mg/dL | <10%  |
|  Nifedipine | 20 and 60 μg/dL | <10%  |
|  Nitrofurantoin | 0.2 and 6.4 mg/dL | <10%  |
|  Nystatin | 0.7 and 2.15 mg/dL | <10%  |
|  Oxytetracycline | 0.5 and 24 mg/dL | <10%  |
|  Phenytoin | 5 and 10 mg/dL | <10%  |
|  Propranolol | 1 and 500 μg/mL | <10%  |
|  Quinidine | 0.6 and 2 mg/dL | <10%  |
|  Simvastatin | 4 and 20 μg/mL | <10%  |
|  Theophylline | 2 and 25 mg/dL | <10%  |
|  Triglycerides | 1000 and 3000 mg/dL | <10%  |
|  Trimethoprim | 1.8 and 7.5 mg/dL | <10%  |
|  Verapamil | 0.1 and 16 mg/dL | <10%  |
|  Warfarin | 3 and 30 μg/mL | <10%  |

*For dopamine, the sponsor claims no interference up to 30 mg/dL

To evaluate cross reactivity, the substances shown in the following table were added to lithium heparin plasma pools containing 2 levels of troponin ( $\leq 0.02$  ng/mL and approximately  $0.5\mathrm{ng / mL}$ ). Control and test samples were tested on 6 instruments and 2 reagent lots at 3 temperature conditions (from 18 to 28 °C) in replicates of 10 for the low troponin samples ( $&lt; 0.02\mathrm{ng / mL}$ ) and replicates of 5 for the high troponin samples ( $0.05\mathrm{ng / mL}$ ). For each possible cross-reactant tested, the troponin I concentration ( $\mathrm{ng / mL}$ ) obtained for the spiked sample was compared to the troponin I concentration obtained with the control sample and applied to the following formula:  $\%$  cross reactivity = [(mean dose of spiked - mean dose of control)/amount of cross reactant spiked] X 100. The sponsor concluded that the following proteins at the concentration listed did not cross react (defined as  $&lt; 1\%$  cross reactivity) with the device.

|  Substance | Concentration tested ng/mL  |
| --- | --- |
|  Actin | 1000  |
|  Cardiac troponin C | 1000  |
|  Recombinant human CK-MB | 1000  |
|  Myoglobin | 1000  |
|  Myosin | 1000  |
|  Recombinant human cTnT | 250  |

{10}

|  Skeletal troponin I | 1000  |
| --- | --- |
|  Tropomyosin | 1000  |

HAMA/Heterophile antibodies: The sponsor provided results demonstrating that their formulation reduces the effects of HAMA interferents. They include the following in the Limitations of the Procedure section of the Instructions for Use:

"For assays employing antibodies, the possibility exists for interference by heterophile antibodies in the patient sample. Patients who have been regularly exposed to animals or have received immunotherapy or diagnostic procedures utilizing immunoglobulins or immunoglobulin fragments may produce antibodies, e.g. HAMA, that interfere with immunoassays. Additionally, other heterophile antibodies such as human anti-goat antibodies may be present in patient samples. Such interfering antibodies may cause erroneous results. Carefully evaluate the results of patients suspected of having these antibodies."

f. Assay cut-off: See section 4 "Clinical cut-off"

2. Comparison studies:

a. Method comparison: Not applicable

b. Matrix comparison: The sponsor conducted a clinical matrix comparison study to compare matched lithium heparin plasma samples and serum samples randomly selected from among the entire pivotal trial cohort. Results of Passing-Bablok regression analysis of singlicate results are provided below (the sponsor excluded 26 samples with troponin levels below the LoQ from these analyses):

Passing Bablok Regression by Concentration Ranges

|  Range (ng/mL) | N | Slope (95% CI) | Intercept (95% CI) | Correlation  |
| --- | --- | --- | --- | --- |
|  0.02 to 20 | 96 | 1 (0.98 to 1.00) | 0.00 (0 to 0) | r = 1.00  |
|  0.02 to 5 | 93 | 1 (0.98 to 1.00) | 0.00 (0 to 0) | r= 0.99  |
|  0.02 to 0.04 | 75 | 1 (1.00 to 1.00) | 0.00 (0 to 0) | r= 0.99  |

Clinical concordance analysis using 123 matched samples from the pivotal trial cohort was performed between the 2 sample types at the  $0.03\mathrm{ng / mL}$  cutoff and is provided below: The sponsor demonstrated  $96\%$  agreement between the lithium heparin plasma samples and serum samples.

{11}

Concordance at the 0.03 ng/mL cut-off

|   | Serum < 0.03 | Serum ≥ 0.03 | Total  |
| --- | --- | --- | --- |
|  Plasma < 0.03 | 38 | 3 | 41  |
|  Plasma ≥ 0.03 | 2 | 80 | 82  |
|  Total | 40 | 83 | 123  |

An additional matrix comparison study was performed at 3 different temperature conditions (ranging from 18°C to 28°C) on 2 instruments using 1 reagent lot. The study included 97 matched serum and lithium heparin plasma samples (not spiked or diluted) which were run at each temperature condition. The following are representative results (of singlicate measurements) of Passing-Bablok regression analysis and are presented in the labeling:

|  Range (ng/mL) | n | Slope (95% CI) | r value | Intercept (95% CI)  |
| --- | --- | --- | --- | --- |
|  0.02 - 71.76 | 97 | 1.03 (1.00 - 1.05) | 0.99 | 0.00 (-0.01 – 0.00)  |

# 3. Clinical studies:

# a. Clinical Sensitivity:

A clinical study was performed to evaluate the clinical performance of the device at the different cut-offs. A multicenter prospective study enrolled 1929 patients from Emergency Departments presenting with chest pain or equivalent ischemic symptoms suggestive of Acute Coronary Syndromes. Final diagnoses were adjudicated by an independent panel of expert physicians using criteria consistent with the 2007 Universal Definition of Myocardial Infarction. Serial samples were collected from patients within 9 hours of presentation to the ER. The sample collection times were at baseline, 1 to 3 hours, 3 to 6 hours and 6 to 9 hours after presentation to the ER. Investigators and adjudicators were blinded to the proposed device's results. Adjudicators were also blinded to site diagnoses. All results presented below were based on the adjudicated diagnoses. Testing was performed using lithium heparin plasma samples. The results are summarized below:

Clinical Performance at  $0.03\mathrm{ng / mL}$  (this cut-off was determined in a feasibility study by ROC analysis)

Based on sample collection timepoint:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 87.4 (221/253) | 82.6-91.2 | 89.3 (1495/1675) | 87.7-90.7  |
|  ≥1-3 h | 96.0 (119/124) | 90.8-98.7 | 89.4 (907/1014) | 87.4-91.3  |
|  ≥3-6 h | 94.9 (149/157) | 90.2-97.8 | 86.7 (816/941) | 84.4-88.8  |
|  ≥6-9 h | 90.7 (39/43) | 77.9-97.4 | 87.0 (214/246) | 82.1-90.9  |

{12}

|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 55.1 (221/401) | 50.1-60.1 | 97.9 (1495/1527) | 97.1-98.6  |
|  ≥1-3 h | 52.6 (119/226) | 45.9-59.3 | 99.5 (907/912) | 98.7-99.8  |
|  ≥3-6 h | 54.4 (149/274) | 48.3-60.4 | 99.0 (816/824) | 98.1-99.6  |
|  ≥6-9 h | 54.9 (39/71) | 42.7-66.8 | 98.2 (214/218) | 95.4-99.5  |

Based on hours since symptom onset:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 91.5 (151/165) | 86.2-95.3 | 89.1 (885/993) | 87.0-91.0  |
|  ≥ 8 h | 94.3 (150/159) | 89.5-97.4 | 86.8 (966/1113) | 84.7-88.7  |
|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 58.3 (151/259) | 52.0-64.4 | 98.4 (885/899) | 97.4-99.2  |
|  ≥ 8 h | 50.5 (150/297) | 44.7-56.3 | 99.1 (966/975) | 98.3-99.6  |

Clinical Performance at  $0.02\mathrm{ng / mL}$  ( $99^{\mathrm{th}}$  percentile upper reference limit)

Based on sample collection timepoint:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 92.1 (233/253) | 88.1-95.1 | 84.3 (1412/1675) | 82.5-86.0  |
|  ≥1-3 h | 98.4 (122/124) | 94.3-99.8 | 85.5 (867/1014) | 83.2-87.6  |
|  ≥3-6 h | 98.1 (154/157) | 94.5-99.6 | 81.0 (762/941) | 78.3-83.4  |
|  ≥6-9 h | 93.0 (40/43) | 80.9-98.5 | 76.4 (188/246) | 70.6-81.6  |
|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 47.0 (233/496) | 42.5-51.5 | 98.6 (1412/1432) | 97.9-99.1  |
|  ≥1-3 h | 45.4 (122/269) | 39.3-51.5 | 99.8 (867/869) | 99.2-100  |
|  ≥3-6 h | 46.2 (154/333) | 40.8-51.8 | 99.6 (762/765) | 98.9-99.9  |
|  ≥6-9 h | 40.8 (40/98) | 31.0-51.2 | 98.4 (188/191) | 95.5-99.7  |

Based on hours since symptom onset:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 96.4 (159/165) | 92.3-98.7 | 83.2 (826/993) | 80.7-85.5  |
|  ≥ 8 h | 96.9 (154/159) | 92.8-99.0 | 81.7 (909/1113) | 79.3-83.9  |

{13}

|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 48.8 (159/326) | 43.2-54.3 | 99.3 (826/832) | 98.4-99.7  |
|  ≥ 8 h | 43.0 (154/358) | 37.8-48.3 | 99.5 (909/914) | 98.7-99.8  |

Clinical Performance at  $0.04\mathrm{ng / mL}$  (lowest measured concentration with a  $\% \mathrm{CV}$  (within-lab)  $&lt; 10\%$

Based on sample collection timepoint:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 77.1 (195/253) | 71.4-82.1 | 93.9 (1572/1675) | 92.6-95.0  |
|  ≥1-3 h | 89.5 (111/124) | 82.7-94.3 | 93.5 (948/1014) | 91.8-94.9  |
|  ≥3-6 h | 87.3 (137/157) | 81.0-92.0 | 93.0 (875/941) | 91.2-94.5  |
|  ≥6-9 h | 90.7 (39/43) | 77.9-97.4 | 93.1 (229/246) | 89.2-95.9  |
|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  Baseline | 65.4 (195/298) | 59.7-70.8 | 96.4 (1572/1630) | 95.4-97.3  |
|  ≥1-3 h | 62.7 (111/177) | 55.1-69.9 | 98.6 (948/961) | 97.7-99.3  |
|  ≥3-6 h | 67.5 (137/203) | 60.6-73.9 | 97.8 (875/895) | 96.6-98.6  |
|  ≥6-9 h | 69.6 (39/56) | 55.9-81.2 | 98.3 (229/233) | 95.7-99.5  |

Based on hours since symptom onset:

|  Interval | Sensitivity |   | Specificity  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 85.5 (141/165) | 79.1-90.5 | 93.6 (929/993) | 91.8-95.0  |
|  ≥ 8 h | 88.1 (140/159) | 82.0-92.7 | 92.7 (1032/1113) | 91.0-94.2  |
|  Interval | Positive Predictive Value |   | Negative Predictive Value  |   |
| --- | --- | --- | --- | --- |
|   |  % | 95% CI | % | 95% CI  |
|  < 8 h | 68.8 (141/205) | 62.0-75.1 | 97.5 (929/953) | 96.3-98.4  |
|  ≥ 8 h | 63.3 (140/221) | 56.6-69.7 | 98.2 (1032/1051) | 97.2-98.9  |

b. Clinical specificity:

See clinical specificity information above in 3a

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

# 4. Clinical cut-off:

This assay has 2 claimed clinical cut-offs. One cutoff (0.03) was determined in a feasibility study by ROC analysis. The second cut-off (0.02) is the  $99^{\text{th}}$  percentile upper reference limit. The sponsor also provided clinical performance

{14}

information in the package insert at 0.04 ng/mL which is the lowest concentration with a %CV (within-lab) &lt;10%.

5. Expected values/Reference range:
The sponsor conducted a multicenter prospective study to establish the 99th percentile upper reference limit in a population of apparently healthy adults with no known diseases of the cardiovascular system or other serious acute or chronic diseases or infections. Lithium heparin plasma samples were evaluated. 527 subjects were enrolled at seven geographically diverse locations. Both male and female subjects were included in the reference range study to determine the 99th percentile upper reference limit.

The 99th percentile upper reference limit was demonstrated to be 0.02 ng/mL (95% CI: 0.01 to 0.05).

N. Instrument Name:
Access 2 Immunoassay System

O. Systems Descriptions:
1. Modes of Operation:
Micro computer controlled analyzer with random and continuous access.

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
☑ Yes ☐ X or No

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
☐ Yes ☑ X or No

2. Software:
FDA reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
☑ Yes ☐ X or No

3. Specimen Identification:
Bar code or sample information can also be entered manually.

4. Specimen Sampling and Handling:
Instructions on sample handling are provided in the reagent labeling.

5. Calibration:
An active calibration curve is required for all tests. For the Access AccuTnI+3 Reagent, calibration is required every 56 days.

15

{15}

6. Quality Control:
The sponsor recommends that at least two levels of an appropriate quality control material be tested a minimum of once a day. The sponsor also states that quality control testing should be performed in accordance with laboratory accreditation requirements, applicable laws and good laboratory practices.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Ambient temperature method comparison studies: The sponsor provided data from 4 studies to demonstrate the impact of ambient temperature differences on the test result. Three of these studies focused on troponin concentrations around the cut-offs (0.02 to 0.4 ng/mL for study 1; 0.02 to 0.6 ng/mL for studies 2 and 3). Study 4 evaluated the full reportable range (FR) of the device (0.03 to 94 ng/mL) and regression analysis of low dose (LD) samples with troponin concentrations around the cut-off (0.03 to 0.4 ng/mL) was also performed. In total, 751 lithium heparin plasma samples were tested at different temperature conditions (ranging from 18 to 28°C) on multiple instruments using multiple reagent lots. Samples were divided into aliquots, and tested at different temperature conditions. The sponsor performed Passing-Bablok regression analyses of singlicate test results. The results of the 19 different comparisons provided by the sponsor from the 4 different studies are summarized below:

|  Study | Low temp | High temp | n | Slope (95% CI) | Intercept (95% CI) | r²  |
| --- | --- | --- | --- | --- | --- | --- |
|  1 | 18°C | 26°C | 57 | 1.04 (1.00, 1.08) | 0.01 (0.00, 0.01) | 0.97  |
|   |  18°C | 22°C | 59 | 1.00 (0.97, 1.02) | 0.00 (0.00, 0.00) | 0.99  |
|   |  22°C | 26°C | 57 | 1.05 (1.02, 1.08) | 0.00 (0.00, 0.01) | 0.98  |
|   |  20°C | 28°C | 61 | 0.97 (0.92, 1.03) | 0.00 (0.00, 0.01) | 0.96  |
|   |  20°C | 24°C | 61 | 0.98 (0.95, 1.01) | 0.00 (0.00, 0.00) | 0.99  |
|   |  24°C | 28°C | 61 | 1.00 (0.95, 1.04) | 0.00 (0.00, 0.01) | 0.98  |
|  2 | 18°C | 26°C | 76 | 0.93 (0.90, 0.97) | 0.00 (0.00, 0.01) | 0.97  |
|   |  18°C | 22°C | 75 | 0.99 (0.97, 1.01) | 0.00 (0.00, 0.00) | 0.99  |
|   |  22°C | 26°C | 75 | 0.94 (0.91, 0.97) | 0.01 (0.00, 0.01) | 0.99  |
|   |  20°C | 28°C | 78 | 0.95 (0.90, 0.98) | 0.01 (0.00, 0.01) | 0.99  |
|   |  20°C | 24°C | 78 | 0.99 (0.96, 1.01) | 0.00 (0.00, 0.01) | 0.99  |
|   |  24°C | 28°C | 78 | 0.96 (0.93, 0.99) | 0.00 (0.00, 0.00) | 0.99  |
|  3 | 18°C | 28°C | 379 | 0.94 (0.93, 0.95) | 0.00 (0.00, 0.00) | 0.99  |
|  4 LD | 18°C | 28°C | 42 | 1.00 (0.96, 1.05) | 0.00 (0.00, 0.00) | 0.99  |
|   |  18°C | 23°C | 42 | 0.98 (0.94, 1.02) | 0.00 (0.00, 0.01) | 0.99  |
|   |  23°C | 28°C | 42 | 1.03 (1.00, 1.06) | 0.00 (0.00, 0.00) | 0.99  |
|  4 FR | 18°C | 28°C | 102 | 0.98 (0.97, 0.99) | 0.00 (0.00, 0.00) | 1.00  |
|   |  18°C | 23°C | 102 | 1.02 (1.00, 1.03) | 0.00 (-0.01, 0.00) | 1.00  |
|   |  23°C | 28°C | 102 | 0.96 (0.95, 0.98) | 0.01 (0.00, 0.01) | 0.99  |

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{16}

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

17

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K121214](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/MMI/K121214)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
