PRECICHEK NS-101 POCT PROFESSIONAL BLOOD GLUCOSE MONITORING SYSTEM
K113314 · Hmd Biomedical, Inc. · LFR · Nov 9, 2012 · Clinical Chemistry
Device Facts
Record ID
K113314
Device Name
PRECICHEK NS-101 POCT PROFESSIONAL BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Hmd Biomedical, Inc.
Product Code
LFR · Clinical Chemistry
Decision Date
Nov 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by health care professionals in a clinical setting: The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices. The PRECICHEK KP Blood Glucose Test Strip is for use with the PRECICHEK NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips to check that the meter and test strips are working together properly and that the test is performing correctly.
Device Story
System measures glucose in capillary, venous, or arterial whole blood; utilizes electrochemical biosensor technology (GDH-FAD) and capillary action; requires 0.5 μL sample volume. Meter processes electrical current generated by glucose-reagent reaction; displays quantitative glucose concentration. Used in clinical settings by healthcare professionals for multi-patient monitoring; requires disinfection between uses. Includes meter, test strips, control solutions, and check strip. Meter features barcode scanner for lot identification; calibration via code entry. Provides rapid results (5 seconds) to aid diabetes management; supports clinical decision-making regarding glycemic control.
Clinical Evidence
Bench testing only. Compliance with ISO 15197:2003, IEC 60601-1, IEC 60601-1-2, and CLSI standards (EP05-A2, EP06-A, EP07-A2, EP09-A2) confirmed performance criteria met.
Indicated for quantitative glucose measurement in fresh capillary (fingertip), venous, and arterial whole blood by healthcare professionals in clinical settings to monitor diabetes control. Contraindicated for neonatal use, diabetes diagnosis/screening, and patients who are dehydrated, hypotensive, in shock, critically ill, or in a hyperosmolar state.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
SureStepFlexx Professional Blood Glucose Management System (K023194)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k113314
B. Purpose for Submission:
New device
C. Measurand:
Capillary, Venous and Arterial whole blood glucose
D. Type of Test:
Quantitative Amperometric assay (Glucose Dehydrogenase (GDH-FAD))
E. Applicant:
HMD Biomedical, Inc.
F. Proprietary and Established Names:
Precichek NS-101 POCT Professional Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1345, Glucose, Dehydrogenase, Glucose
21 CFR § 862.1345, Glucose Test System
21 CFR 862.1660, Quality Control Material (assayed and unassayed)
2. Classification:
Class II and Class I (reserved)
3. Product code:
{1}
LFR, JJX
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indications for Use below
2. Indication(s) for use:
Precichek NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, venous whole blood and arterial whole blood samples. The Prechichek NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by health care professionals in a clinical setting.
The Precichek NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a healthcare professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control.
The Precichek NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The Precichek KP Blood Glucose Test Strip is for use with the Precichek NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose in fresh capillary whole blood from the fingertips, venous and arterial whole blood samples.
The Precichek Glucose Control Solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the Precichek NS-101 POCT Professional Blood Glucose monitoring system and the Precichek KP Blood Glucose Test Strips to check that meter and test strips are working together properly and that the test is performing correctly.
3. Special conditions for use statement(s):
- For in vitro diagnostic use only
- For Professional Use only
- Not intended for use on neonates
- Not for diagnosis of or screening for diabetes mellitus
- Not to be used for patients who are dehydrated, hypotensive, in shock, critically ill or in a hyperosmolar state
{2}
- Multiple-patient use meters must be disinfected between use following labeling recommendations
- Multiple patient use systems should only use single use, auto disabling lancin
4. Special instrument requirements:
PreciChek NS-101 POCT Professional Blood Glucose Monitoring System
# I. Device Description:
The PreciChek NS-101 POCT Professional Blood Glucose Monitoring System contains a blood glucose meter (PreciChek NS-101 POCT Professional Blood Glucose Meter), PreciChek KP Blood Glucose Test Strips, PreciChek Glucose Control Solution (Level I and Level II), and Check Strip. The Check Strip, can be used to check that the meter is operating properly. Meter contains a USB cable which is used to charge the battery of the meter but does not transfer results to a computer. Owner's booklet is provided in the kit. Single use Auto-disabling lancets are sold separately.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Sure Step Flexx (LifeScan, Inc.)
Sure Step Flexx Glucose Control Solution (LifeScan Inc.)
2. Predicate 510(k) number(s):
k023194
k032985
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device PreciChek NS-101 POCT | Predicate SureStep Flexx (k023914) |
| Intended Use | To quantitatively measure the concentration of glucose (sugar) in capillary whole blood samples drawn from the fingertip, venous, and arterial whole blood by healthcare professionals to monitor patient's blood glucose in clinical settings. The system is used outside the body (in vitro diagnostic use) | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device Precichek NS-101 POCT | Predicate SureStep Flexx (k023914) |
| Code System | Bar code | Same |
| Blood sample source | Fresh whole blood (capillary, venous, arterial) | Same |
| Calibration | Plasma Equivalent | Same |
| Operating Procedure | Menu Control-Operational ID, Patient ID, Strip Lot | Same |
| Characteristics | For Professional Use Only | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Methodology | Glucose Dehydrogenase(GDH-FAD) | Glucose Oxidase |
| Principle | Electrochemical reaction | Photochemical reaction |
| Reaction time | 5 seconds | 30 seconds |
| Blood Glucose Range | 20 - 600 mg/dl | 0-500 mg/dl |
| Operational environment | 10-40°C, 50-104°F 20-80% RH | 18-30°C, 64-86°F 30-70% RH |
| Blood Sample Volume | 0.5 μL | 5μL-30μL |
| Power source | 3.7V Li battery, rechargeable via USB | 3AA 1.5v alkaline batteries |
# K. Standard/Guidance Document Referenced (if applicable):
ISO 15197: In vitro diagnostic test systems-Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
ISO 14971: Medical Devices-Applications of risk management to medical devices
IEC 60601-1: Medical Electrical Equipment-Part1: General Requirements for Safety
# L. Test Principle:
This test is based on electrochemical biosensor technology and the principle of capillary action. The electrical current generated by the reaction of glucose with the reagent of the strip is measured by the meter and is displayed as the corresponding blood glucose level. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
a. Precision/Reproducibility:
Within-day precision was performed using venous whole blood samples at five glucose concentrations, three test strip lots, and 10 Precichek NS-101 POCT meters. Each level was evaluated 10 times for a total of 100 tests per each glucose level. Day-to-day precision was evaluated using three glucose control solutions with concentration levels of low (45mg/dL), normal (100mg/dL) and high (300 mg/dL). The day-to-day precision was evaluated over a ten-day period using three test strip lots and 10 Precichek NS-101 POCT meters. A summary of the test results is presented below.
Within-day Precision (whole blood)
KPH 14081 (Lot 1)
| Concentration (mg/dL) of glucose | YSI | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| 10-50 | 40.6 | 100 | 40.7 | 2.0 | 4.8 |
| 51-110 | 93.3 | 100 | 100.1 | 4.1 | 4.1 |
| 111-150 | 120 | 100 | 124.8 | 3.1 | 2.5 |
| 151-250 | 213 | 100 | 215.8 | 7.8 | 3.6 |
| 251-400 | 336 | 100 | 340.0 | 12.1 | 3.5 |
| 401-600 | 532 | 100 | 537.4 | 14.0 | 2.6 |
KPH15112 (Lot 2)
| Concentration (mg/dL) of glucose | YSI | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| 10-50 | 40.6 | 100 | 40.5 | 1.7 | 4.2 |
| 51-110 | 93.3 | 100 | 96.9 | 3.6 | 3.7 |
| 111-150 | 120 | 100 | 122.8 | 3.6 | 3.0 |
| 151-250 | 213 | 100 | 206.9 | 7.8 | 3.8 |
| 251-400 | 336 | 100 | 322.9 | 9.5 | 3.0 |
| 401-600 | 532 | 100 | 522.8 | 16.3 | 3.1 |
KPH 14082 (Lot 3)
| Concentration (mg/dL) of glucose | YSI | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| 10-50 | 40.6 | 100 | 40.0 | 1.8 | 4.6 |
| 51-110 | 93.3 | 100 | 94.9 | 4.5 | 4.7 |
| 111-150 | 120 | 100 | 125.4 | 3.5 | 2.8 |
| 151-250 | 213 | 100 | 206.7 | 9.0 | 4.4 |
{5}
6
| 251-400 | 336 | 100 | 316.4 | 14.3 | 4.5 |
| --- | --- | --- | --- | --- | --- |
| 401-600 | 532 | 100 | 516.4 | 16.7 | 3.2 |
All 3 lots Combined
| Concentration (mg/dL) of glucose | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- |
| 10-50 | 100 | 40.4 | 1.9 | 4.6 |
| 51-110 | 100 | 97.3 | 4.5 | 4.6 |
| 111-150 | 100 | 124.3 | 3.5 | 2.8 |
| 151-250 | 100 | 209.8 | 9.0 | 4.3 |
| 251-400 | 100 | 326.5 | 14.3 | 4.4 |
| 401-600 | 100 | 525.5 | 16.7 | 3.2 |
Day-to-Day Precision (control materials)
KPH 14081 (Lot 1)
| Control level | YSI (mg/dL) | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level I (45 mg/dL) | 50.0 | 100 | 60.3 | 2.7 | 4.5 |
| Level II (100 mg/dL) | 100 | 100 | 124.4 | 5.9 | 4.7 |
| Level III (300 mg/dL) | 300 | 100 | 433.6 | 15.2 | 3.5 |
KPH15112 (Lot 2)
| Control level | YSI (mg/dL) | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level I (45 mg/dL) | 50.0 | 100 | 61.5 | 2.7 | 4.3 |
| Level II (100mg/dL) | 100 | 100 | 122.9 | 5.2 | 4.2 |
| Level III (300 mg/dL) | 300 | 100 | 438.0 | 15.4 | 3.5 |
KPH 14082 (Lot 3)
| Control level (mg/dL) | YSI (mg/dL) | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 (45 mg/dL) | 50.0 | 100 | 60.1 | 2.9 | 4.8 |
| Level II (100 mg/dL) | 100 | 100 | 124.2 | 5.3 | 4.2 |
| Level III (300 mg/dL) | 300 | 100 | 440.8 | 16.4 | 3.7 |
{6}
All 3 Lots Combined
| Control level (mg/dL) | YSI (mg/dL) | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 (45 mg/dL) | 50.0 | 100 | 60.6 | 2.8 | 4.7 |
| Level II (100 mg/dL) | 100.0 | 100 | 123.8 | 5.5 | 4.4 |
| Level III (300mg/dL) | 300.0 | 100 | 437.5 | 15.9 | 3.6 |
b. Linearity/assay reportable range:
Linearity testing was evaluated using 8 levels of venous whole blood samples (18, 41, 103, 128, 200, 313 and 654 mg/dL) as measured by YSI. Measurements were taken 10 times for each concentration and the values of the Precichek NS-101 POCT meter were compared to those obtained from the YSI-2300. 3 lots of test strips were used in the linearity testing.
Linearity of Precichek NS-101 vs YSI 2300 for each test strip lot
| Strip lot | Linear Regression | 95% CI slope | 95% CI intercept | R² | N |
| --- | --- | --- | --- | --- | --- |
| 1 | y= 0.98x + 0.26 | ± 0.01 | ± 3.46 | 0.995 | 160 |
| 2 | y= 0.98x + 0.94 | ± 0.01 | ± 3.21 | 0.996 | 160 |
| 3 | y= 0.98x + 0.79 | ± 0.01 | ± 3.47 | 0.995 | 160 |
| combined | y= 0.98x + 0.66 | ± 0.01 | ± 1.94 | 0.995 | 480 |
The claimed measuring range for this device is 20- 600 mg/dL
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The Precichek NS-101 POCT BGMS is traceable to the NIST SRM 917 reference material.
Stability: Real time stability studies were conducted to assess the shelf-life and open vial stability of the control solutions and test strips. Real time stability studies are still on-going. For controls, the sponsor claims an unopened shelf life of 18 months and stability of 90 days after opening, when stored at 50-80 °F (10-30 °C). For test strips, the sponsor claims that the unopened test strips have a 24 month shelf life and are stable for 90 days after first opening, when stored at 10-40°C (50-104°F). The sponsor's protocol and acceptance criteria were reviewed and considered acceptable.
Control Value Assignment: Previously cleared under k032985. The value assignment of the Precichek glucose control solutions were determined by in-house procedure. The control solutions were prepared by gravimetric addition of glucose to achieve target values of 45 mg/dL for Level I and 300 mg/dL for Level II. Values
{7}
were confirmed by YSI 2300 method. Verification of the control solutions were tested with 30 test strips and 5 Precichek NS-101 POCT Professional Blood Glucose Meters per level of control. The target ranges were set at the following:
| Glucose Control solution | Target concentration | Acceptable Range |
| --- | --- | --- |
| Level 1 | 45 mg/dL | 35 – 65 mg/dL |
| Level 2 | 300 mg/dL | 336-545 mg/dL |
d. Detection limit:
The measuring range of the system is $20 - 600\mathrm{mg / dL}$ . This range was verified by the linearity study above (section M.1.b)
e. Analytical specificity:
Whole blood samples were spiked to nominal glucose concentrations of 90-120 mg/dL and 300-400 mg/dL. The interfering substance was prepared at 5 concentrations to include normal (therapeutic) and high (toxic). Each level was analyzed 6 times The sponsor defines no significant interference as $\leq 10\%$ difference.
| Endogenous and Exogenous Substances | Highest Concentration tested showing no interference (mg/dL) |
| --- | --- |
| Acetaminophen | 25 |
| Bilirubin | 20 |
| Dopamine | 13 |
| Ibuprofen | 40 |
| Methyl dopa | 10 |
| Tolarzamide | 10 |
| Cholesterol | 500 |
| Creatinine | 30 |
| Ephedrine | 10 |
| L-dopa | 2 |
| Salicylate | 40 |
| Tolbutamide | 12 |
| Tetracyclin | 4 |
| Ascorbic Acid | 2.25 |
| Galactose | 1000 |
| Lactose | 50 |
| Maltose | 800 |
| D-Xylose | 50 |
{8}
| Urea | 500 |
| --- | --- |
| Glutathione | 55 |
| Triglycerides | 1000 |
| Uric Acid | 15 |
The labeling states "A sample with large amounts of the substances such as Triglycerides $(>1,000\mathrm{mg / dL})$ , ascorbic acid $(>2.25\mathrm{mg / dL})$ , uric acid $(>15\mathrm{mg / dL})$ and xylose $(>50\mathrm{mg / dL})$ may cause a higher result.
f. Assay cut-off:
Not applicable
# 2. Comparison studies:
a. Method comparison with predicate device:
The sponsor performed accuracy studies to demonstrate the performance of the Precichek NS-101 POC meter compared to the YSI 2300 reference method. The system accuracy evaluation was performed with 105 fresh capillary blood samples (5 spiked and 5 altered), with glucose concentrations distributed over the range of $20 - 600\mathrm{mg / dL}$ . Samples were analyzed using 3 Precichek NS-101 POCT meters and 3 lots of test strips. To obtain the blood glucose concentrations $< 50\mathrm{mg / dL}$ and $>500\mathrm{mg / dL}$ , samples were allowed to glycolize or were spiked to achieve desired concentrations. The results relative to YSI are summarized in the tables below.
# Capillary Whole Blood:
System accuracy results for glucose concentrations $< 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 1 | 9/16 (56.3%) | 14/16 (87.5%) | 16/16 (100.0%) |
| 2 | 11/16 (68.8%) | 14/16 (87.5%) | 16/16 (100.0%) |
| 3 | 9/16 (56.3%) | 15/16 (93.8%) | 16/16 (100.0%) |
System accuracy results for glucose concentrations $\geq 75\mathrm{mg / dL}$
| Strip Lot | Within ±5% | Within ±10% | Within ±15% | Within ± 20% |
| --- | --- | --- | --- | --- |
| 1 | 40/89 (44.9%) | 70/89 (78.7%) | 82/89 (92.1%) | 86/89 (96.6%) |
| 2 | 46/89 (51.7%) | 69/89 (77.5%) | 81/89 (91.0%0 | 86/89 (96.6%) |
| 3 | 46/89 51.7%0 | 73/89 (82.0%) | 86/89 (96.6%) | 87/89 (97.8%) |
Regression Analysis vs. YSI (glucose range 14.0 -617 mg/dL)
| Lot Number | Slope and y-intercept | R² | N |
| --- | --- | --- | --- |
| 1 | Y = 1.00x + 3.70 | 0.98 | 105 |
| 2 | Y = 0.95x + 9.33 | 0.98 | 105 |
| 3 | Y = 0.95x + 6.81 | 0.98 | 105 |
{9}
# b. Matrix comparison:
The sponsor performed matrix comparison studies to demonstrate the performance of the Precichek NS-101 POC meter compared to the YSI 2300 reference method. The evaluation was performed 112 EDTA venous blood samples(5 spiked and 5 altered) and 100 EDTA arterial samples (5 spiked, 5 altered) with glucose concentrations distributed over the range of $20 - 600\mathrm{mg / dL}$ . Samples were analyzed using 3 Precichek NS-101 POCT meters and 3 lots of test strips. To obtain the blood glucose concentrations $< 50\mathrm{mg / dL}$ and $>500\mathrm{mg / dL}$ , samples were allowed to glycolize or were spiked to achieve desired concentrations. The results relative to YSI are summarized in the tables below.
# Venous Whole Blood:
System accuracy results for glucose concentrations $< 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 1 | 9/17 (52.9%) | 14/17 (82.4%) | 17/17 (100.0%) |
| 2 | 6/17 (35.3%) | 13/17 (76.5%) | 17/17 (100.0%) |
| 3 | 8/17 (47.1%) | 15/17 (88.2%) | 17/17 (100.0%) |
System accuracy results for glucose concentrations $\geq 75\mathrm{mg / dL}$
| Strip Lot | Within ±5% | Within ±10% | Within ±15% | Within ± 20% |
| --- | --- | --- | --- | --- |
| 1 | 34/95 (35.8%) | 66/95 (69.5%) | 88/95 (92.6%) | 93/95 (97.8%) |
| 2 | 39/95 (41.1%) | 71/95 (74.7%) | 88/95 (92.6%) | 93/95 (97.8%) |
| 3 | 50/95 (52.6%) | 80/95 (84.2%) | 89/95 (93.7%) | 93/95 (97.8%) |
Regression Analysis vs. YSI (glucose range 14.0 -617 mg/dL)
| Lot Number | Slope and y-intercept | R2 | N |
| --- | --- | --- | --- |
| 1 | Y = 0.95x + 0.19 | 0.97 | 112 |
| 2 | Y = 0.95x + 7.44 | 0.97 | 112 |
| 3 | Y = 0.98x + 6.93 | 0.98 | 112 |
# Arterial Whole Blood:
System accuracy results for glucose concentrations $< 75\mathrm{mg / dL}$
| Strip Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| 1 | 9/17 (52.9%) | 14/17 (82.4%) | 17/17 (100%) |
| 2 | 5/17 (29.4%) | 13/17 (76.5%) | 17/17 (100%) |
| 3 | 8/17 (47.1%) | 15/17 (88.2%) | 17/17 (100%) |
System accuracy results for glucose concentrations $\geq 75\mathrm{mg / dL}$
| Strip Lot | Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- | --- |
| 1 | 33/83 (40.0%) | 64/83 (77.1%) | 79/83 (95.2%) | 81/83 (97.6%) |
{10}
11
| 2 | 36/83 (43.4%) | 60/83 (72.3%) | 73/83 (88.0%) | 79/83 (95.2%) |
| --- | --- | --- | --- | --- |
| 3 | 34/83 (41.0%) | 64/83 (77.1%) | 77/83 (92.8%) | 81/83 (97.6%) |
Regression Analysis vs. YSI (glucose range 14.0 -617 mg/dL)
| Lot Number | Slope and y-intercept | R² | N |
| --- | --- | --- | --- |
| 1 | Y = 1.02x + 2.77 | 0.98 | 100 |
| 2 | Y = 1.04x + 1.11 | 0.98 | 100 |
| 3 | Y = 1.04x + 0.00 | 0.98 | 100 |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
None
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
| Time of day | Expected Range, Non-diabetes |
| --- | --- |
| Fasting and before meals | Less than 100 mg/dL |
| 2 hours after meals | Less than 140 mg/dL |
Sponsor referenced: American Diabetes, Association Position Statement, Diabetes Care Vol. 34 (Supp.1) p.S13 (2011)
N. Instrument Name:
Precichek NS-101 POCT Professional Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
{11}
Each test strip is single use and must be replaced with a new strip for additional readings. Each test strip requires a sample volume of $0.5\mu \mathrm{L}$.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? :
Yes _______ or No ☐ X
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes _______ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No _______
The applicant has provided documentation that indicates the device was designed and developed under good software life-cycle processes.
3. Specimen Identification:
The device has a scanner lens. When you press and hold the “Scan” button, the red LED light will be turned on to penetrate the scanner lens and read the bar coded labeled sample.
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood samples drawn from the fingertip, venous and arterial whole blood EDTA samples. No special handling or storage is needed.
5. Calibration:
The device is calibrated by a coding process. The code number is the last two digits of the strip lot number. The user should input the last two digits of the strip lot number into the memory before use. While performing the blood glucose test, the strip lot number printed on the vial label is scanned to match the one in the memory. The meter then applies a formula to calculate the glucose value.
6. Quality Control:
Two levels of glucose control solutions (Level 1, 2) are provided within the kit from the
12
{12}
sponsor. The labeling provides instructions on when to test control materials. An acceptable range for each control level is printed on the test strip vial label. If the control values fall outside these ranges, the user is referred to customer support for problems and more information.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Altitude Studies: Venous whole blood samples, with glucose concentrations, according to YSI, from 30-594 mg/dL were analyzed using 3 Precichek NS-101 POCT test strip lots and 10 Precichek NS-101 POCT meters. Results were compared to YSI values and resulted in no significant bias (≤ 10%) for all samples tested. The data submitted support use of the device up to 8800 feet (2,750 meters).
2. Temperature and Humidity Studies: The sponsor performed temperature and humidity studies using venous whole blood samples at 6 glucose ranges (30-50, 51-110, 151-250, 251-400 and 401-600 mg/dL), 5 Precichek NS-101 POCT meters and 2 lots of Precichek NS-101 test strip lots. Results demonstrated that the system can be used in conditions of 50° - 104°F (10° - 40°C) with relative humidity 20-80% and stored under conditions of 50° - 104°F (10° - 40°C) and relative humidity of 20-80%. No significant effect (≤ 10% relative to YSI) was observed at the temperature and humidity combinations tested. The following combinations were tested: low temperature/low humidity, low temperature/high humidity, high temperature/low humidity and high temperature/high humidity. The results support the claims in the labeling that the system can be used in conditions of 50° - 104°F (10° - 40°C) with relative humidity of 20-80%.
3. Sample Volume Studies: Sample volume studies were performed using venous whole blood samples with concentrations of 49.3, 97, 123, 210, 336, and 488 mg/dL at 10 sample volumes (0.3, 0.35, 0.4, 0.45, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0). Ten Precichek POCT meters and 2 lots of test strips were used in the study and results were compared to YSI. Results support the claimed minimum sample volume of 0.5μL.
4. Infection Control Studies: The device is intended for multi-patient use. Disinfection efficacy studies were performed on the materials comprising the meter by outside commercial testing demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant (Clorox Germicidal Wipes) (EPA Registration #67619-12). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials for the meter after 18,250 cleanings and 18,250 disinfection cycles with the Clorox Germicidal Wipes. The robustness studies were designed to simulate 5 years of multi-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
5. Readability Study: Not applicable. This device is for "Professional Use" only.
6. Hematocrit Study: The effect of hematocrit was evaluated using whole blood samples spiked with seven hematocrit levels (25, 30, 35, 40, 45, 55, 65%) at glucose
13
{13}
concentrations distributed within the measuring range (20-50, 51-110, 111-150, 151-250, 251-400, and 401-600 mg/dL) of the device. Each concentration was tested in replicates of 6. The study incorporated 6 glucose meters. Blood analysis was also performed on the YSI 2300 glucose analyzer. Results of samples with hct levels between 30-55% showed bias versus the reference method of ≤ ± 15%)
7. EMC Testing: The sponsor provided the appropriate documentation certifying that electromagnetic testing (EMC) had been performed and the Precichek NS-101 POCT Professional Blood Glucose Monitoring System was found compliant.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.