SEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, SEJOY hCG One Step Pregnancy Test Midstream

K212447 · Hangzhou Sejoy Electronics & Instruments Co., Ltd. · LCX · Feb 18, 2022 · Clinical Chemistry

Device Facts

Record IDK212447
Device NameSEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, SEJOY hCG One Step Pregnancy Test Midstream
ApplicantHangzhou Sejoy Electronics & Instruments Co., Ltd.
Product CodeLCX · Clinical Chemistry
Decision DateFeb 18, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The SEJOY hCG One Step Pregnancy Test Strip is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic glycoprotein (hCG) in urine to aid in the early detection of pregnancy. The SEJOY hCG One Step Pregnancy Test Cassette is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The SEJOY hCG One Step Pregnancy Test Midstream is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy.

Device Story

Rapid, one-step lateral flow sandwich immunoassay for qualitative hCG detection in urine; aids early pregnancy detection. Formats: strip, cassette, midstream. Principle: monoclonal antibodies directed against alpha and beta hCG subunits; hCG in sample binds to form colored line at test zone ('T'); control zone ('C') confirms proper performance. User-operated (OTC): dip strip, add urine to cassette well, or urinate on/dip midstream absorbent tip. Visual result read at 3-5 minutes. Two lines (T and C) indicate positive; one line (C only) indicates negative; no C line indicates invalid. Benefits: provides rapid, accessible pregnancy screening for lay users.

Clinical Evidence

Bench testing included precision/reproducibility studies (n=270 per concentration/format), interference testing (exogenous/endogenous substances, pH, specific gravity, hook effect, beta-core fragment), and method comparison against predicate (n=150). Lay user study (n=440) demonstrated 100% agreement between lay user and professional results. No clinical sensitivity/specificity studies required for this qualitative immunoassay.

Technological Characteristics

Lateral flow sandwich immunoassay; monoclonal antibodies against alpha/beta hCG subunits. Cut-off: 25 mIU/mL. Formats: strip, cassette, midstream. Traceable to WHO 5th International Standard (NIBSC 07/364). No electronic components, software, or energy sources. Interference testing per CLSI EP07.

Indications for Use

Indicated for qualitative detection of hCG in urine to aid in early pregnancy detection. Intended for over-the-counter (OTC) use by lay users.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K212447 B Applicant Hangzhou Sejoy Electronics & Instruments Co., Ltd. C Proprietary and Established Names SEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, SEJOY hCG One Step Pregnancy Test Midstream D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LCX | Class II | 21 CFR 862.1155 - Human Chorionic Gonadotropin (HCG) Test System | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: New Device B Measurand: Human Chorionic Gonadotropin (hCG) C Type of Test: Qualitative chromatographic immunoassay ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} B Indication(s) for Use: The SEJOY hCG One Step Pregnancy Test Strip is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The SEJOY hCG One Step Pregnancy Test Cassette is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. The SEJOY hCG One Step Pregnancy Test Midstream is a rapid, one-step lateral flow immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. C Special Conditions for Use Statement(s): OTC - Over The Counter D Special Instrument Requirements: None. IV Device/System Characteristics: 1. Device Description: The SEJOY hCG One Step Pregnancy test will be sold in three formats: strip, cassette, and midstream. Each format is qualitative lateral flow immunoassay for the detection of hCG in human urine samples, as an aid in the detection of pregnancy. The SEJOY hCG One Step Pregnancy Test Strip and SEJOY hCG One Step Pregnancy Test Midstream formats consist of one test device sealed in a desiccated aluminum pouch and a package insert. The SEJOY hCG One Step Pregnancy Test Cassette format consists of one test device, a disposable plastic dropper, and a package insert. 2. Principle of Operation: The SEJOY hCG One Step Pregnancy Test (strip/cassette/midstream) is a lateral flow sandwich immunoassay employing monoclonal antibodies that are specifically directed against the alpha and beta subunits of hCG. To use the test, the user either dips the device in the urine specimen (strip format), adds the urine specimen to the specimen well (cassette format), or urinates on the absorbent tip or immerses the absorbent tip in urine (midstream format). If hCG is present in the sample, it will reach the test zone ("T") of the membrane and form a colored line. When the test is performed properly, a colored line will always appear in the control zone ("C"). The test result is shown in the result window and read visually between 3 and 5 minutes of urine application. Two distinct colored lines, one in the test region and another in the control line region indicate a positive test result (pregnant). Absence of a colored line in the test region and only a colored line in the control line indicates a negative test result (not pregnant). Absence of a colored line in the control line even in the presence of a colored line in the test line region indicates an invalid test result. K212447 - Page 2 of 9 {3} V Substantial Equivalence Information: A Predicate Device Name(s): CLUNGENE HCG Pregnancy Rapid Test Cassette, CLUNGENE HCG Pregnancy Rapid Test Strip, CLUNGENE HCG Pregnancy Rapid Test Midstream B Predicate 510(k) Number(s): K193132 C Comparison with Predicate(s): | Device & Predicate Device(s): | K212447 | K193132 | | --- | --- | --- | | Device Trade Name | SEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, SEJOY hCG One Step Pregnancy Test Midstream | CLUNGENE HCG Pregnancy Rapid Test Cassette, CLUNGENE HCG Pregnancy Rapid Test Strip, CLUNGENE HCG Pregnancy Rapid Test Midstream | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | Aid in the early detection of pregnancy | Same | | Type of use | Over the counter use | Same | | Methodology | Chromatographic immunoassay | Same | | Specimen type | Urine | Same | | Assay cut-off | 25 mIU/mL | Same | | Device format | Strip, Cassette, Midstream | Same | | General Device Characteristic Differences | | | | Reading Time | 3-5 minutes | 3 minutes | VI Standards/Guidance Documents Referenced: - CLSI EP07 3rd Edition, Interference Testing in Clinical Chemistry. K212447 - Page 3 of 9 {4} VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: A precision study was performed using negative urine samples spiked with hCG to obtain samples with hCG concentrations of 0, 12.5, 18.75, 22.5, 25, 50, 100 and 200 mIU/mL. Tests were performed by three different operators for each sample concentration in 2 runs per day for 15 days using 3 lots of each device format (strip, cassette, midstream). The device cutoff is 25 mIU/mL hCG. The results are summarized in the tables below: Strip Format | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Combined lots | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | + | - | + | - | + | - | + | - | | | | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | 0 | | 12.5 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | 0 | | 18.75 | 74 | 16 | 73 | 17 | 72 | 18 | 219 | 51 | 81.1 | 18.9 | | 22.5 | 51 | 39 | 48 | 42 | 40 | 50 | 139 | 131 | 51.5 | 48.5 | | 25 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | | 50 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | | 100 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | | 200 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | Cassette Format | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Combined lots | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | + | - | + | - | + | - | + | - | | | | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100.0 | 0.0 | | 12.5 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100.0 | 0.0 | | 18.75 | 70 | 20 | 75 | 15 | 73 | 17 | 218 | 52 | 80.7 | 19.3 | | 22.5 | 50 | 40 | 44 | 46 | 44 | 46 | 138 | 132 | 51.1 | 48.9 | | 25 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0.0 | 100.0 | | 50 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0.0 | 100.0 | | 100 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0.0 | 100.0 | | 200 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0.0 | 100.0 | Midstream Format | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Combined lots | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | + | - | + | - | + | - | + | - | | | | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | 0 | | 12.5 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | 0 | | 18.75 | 76 | 14 | 74 | 16 | 72 | 18 | 222 | 48 | 82.2 | 17.8 | | 22.5 | 46 | 44 | 46 | 44 | 49 | 41 | 141 | 129 | 52.2 | 47.8 | | 25 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | | 50 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | K212447 - Page 4 of 9 {5} K212447 - Page 5 of 9 | hCG Concentration (mIU/mL) | Lot 1 | | Lot 2 | | Lot 3 | | Combined lots | | % Negative | % Positive | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | + | - | + | - | + | - | + | - | | | | 100 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | | 200 | 0 | 90 | 0 | 90 | 0 | 90 | 0 | 270 | 0 | 100 | 2. Linearity: Linearity is not applicable since this is a qualitative test. 3. Analytical Specificity/Interference: Interference from exogenous and endogenous substances To evaluate the potential for interference by certain exogenous and endogenous compounds, each interferent was prepared by diluting stock interferent material to the desired concentration. Negative and positive female urine samples containing 0 and 25 mIU/mL hCG, respectively, were spiked with the compounds to obtain the desired test concentration. The samples were tested using 3 lots of each device format (strip, cassette, midstream). The results demonstrated no interference from exogenous compounds at the following concentrations for both negative and positive hCG urine samples. | Analyte | Concentration | | --- | --- | | Acetaminophen | 20 mg/dL | | Acetoacetic Acid | 2000 mg/dL | | Ascorbic Acid | 20 mg/dL | | β-hydroxybutyrate | 2000 mg/dL | | Caffeine | 20 mg/dL | | Ephedrine | 20 mg/dL | | Gentisic Acid | 20 mg/dL | | Phenylpropanolamine | 20 mg/dL | | Salicylic Acid | 20 mg/dL | | Phenothiazine | 20 mg/dL | | EDTA | 80 mg/dL | | Acetylsalicylic Acid | 20 mg/dL | | Benzoylecgonine | 10 mg/dL | | Cannabinol | 10 mg/dL | | Codeine | 6 μg/dL | | Ethanol | 1.00% | | Methanol | 10% | | Albumin | 2000 mg/dL | | Glucose | 2000 mg/dL | | Bilirubin | 2 mg/mL | | Atropine | 20 mg/dL | | Estriol-17-beta | 1400 μg/dL | {6} | Analyte | Concentration | | --- | --- | | Hemoglobin | 500 mg/dL | | Pregnanediol | 1500 μg/dL | | Thiophene | 20 mg/dL | | Ampicillin | 20 mg/dL | | Tetracycline | 20 mg/dL | | Ketone | 20 mg/dL | ## Cross-reactivity of structurally-related compounds To evaluate specificity and cross-reactivity, negative (0 mIU/mL hCG) and positive female urine samples (25 mIU/mL hCG) were spiked with various concentrations of glycoprotein hormones, including LH (500 mIU/mL), FSH (1000 mIU/mL) and TSH (1 mIU/mL). The samples were tested using 3 lots of each device format (strip, cassette, midstream). The results showed that there is no interference at 500 mIU/mL LH, 1000 mIU/mL FSH, and 1 mIU/mL TSH for both negative and positive urine samples. ## Effect of urine pH To evaluate potential interference from changes in pH, urine samples containing 0 mIU/mL, 10 mIU/mL, and 25 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The samples were tested on all three device formats (strip, cassette, midstream). The results demonstrated that urine samples within the pH range of 4.0 to 9.0 do not interfere with either positive or negative results from the device. ## Effect of urine Specific Gravity To evaluate potential interference from changes in specific gravity, urine samples containing 0, 12.5 and 25 mIU/mL hCG were adjusted to specific gravities of 1.000, 1.012, 1.020, 1.030, 1.037 and tested using the candidate device. The samples were tested on all three device formats (strip, cassette, midstream). The results demonstrated that samples within the specific gravity range of 1.000 to 1.037 do not interfere with either positive or negative results from the device. ## High dose hook effect study Negative urine samples were spiked with varying hCG concentrations (25, 100, 1,000, 10,000, 100,000, 150,000, 500,000, 1,000,000, and 2,000,000 mIU/mL) using 3 lots of each device format (strip, cassette, midstream). The results showed no hook effect up to 2,000,000 mIU/mL hCG. ## Effect of hCG β-core fragment To evaluate potential interference from hCG β-core fragment, negative and positive urine samples containing 0 and 25 mIU/mL hCG were spiked with 2,000,000 pmol/L hCG β-core fragment. The samples were tested using 3 lots of each device format (strip, cassette, midstream). The results demonstrated that there is no interference at 2,000,000 pmol/L hCG-β-core fragment in either negative or positive urine samples. ## 4. Assay Reportable Range: Not applicable. K212447 - Page 6 of 9 {7} 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The SEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, and SEJOY hCG One Step Pregnancy Test Midstream are traceable to the World Health Organization (WHO) 5th International Standard (IS) for hCG (NIBSC code 07/364). 6. Detection Limit: Refer to the Precision and Reproducibility section VII.A.1 above. 7. Assay Cut-Off: The cut-off of the SEJOY hCG One Step Pregnancy Test Strip, SEJOY hCG One Step Pregnancy Test Cassette, SEJOY hCG One Step Pregnancy Test Midstream is 25mIU/mL hCG. Refer to the Precision and Reproducibility section VII.A.1 above. B Comparison Studies: 1. Method Comparison with Predicate Device: a) The performance of the three device formats (strip, cassette, midstream) of the SEJOY hCG One Step Pregnancy test was compared to the predicate device. Testing was performed by laboratory professionals at three point-of-care (POC) sites with urine samples from 150 women between the ages of 18 to 60. About half of the women were suspected to be pregnant. All samples were masked and randomized prior to testing using each device format (strip, cassette, midstream). The data show that the agreement of the SEJOY hCG One Step Pregnancy Test (strip, cassette, and midstream) with the predicate device was 100%. Strip format | Product | Predicate device | | | | --- | --- | --- | --- | | | | Negative | Positive | | SEJOY hCG One Step Pregnancy Test Strip | Negative | 72 | 0 | | | Positive | 0 | 78 | Cassette format | Product | Predicate device | | | | --- | --- | --- | --- | | | | Negative | Positive | | SEJOY hCG One Step Pregnancy Test Cassette | Negative | 72 | 0 | | | Positive | 0 | 78 | Midstream format (dip method) | Product | Predicate device | | | | --- | --- | --- | --- | | | | Negative | Positive | | SEJOY hCG One Step Pregnancy Test Midstream | Negative | 72 | 0 | | | Positive | 0 | 78 | b) The performance of the SEJOY hCG One Step Pregnancy Test Midstream, using the dip method and simulated stream methods, was compared to the predicate device. Testing was performed by laboratory professionals at three POC sites with urine samples from 135 women between the ages of 18 to 60. About half of the women were suspected to be pregnant. All samples were masked and randomized prior to testing using the dip method K212447 - Page 7 of 9 {8} and simulated stream method for midstream format. The data show that the agreement of hCG Test Midstream with the predicate device was 100%. Midstream format (dip method): | Product | Predicate device | | | | --- | --- | --- | --- | | | | Negative | Positive | | SEJOY hCG One Step Pregnancy Test Midstream | Negative | 72 | 0 | | | Positive | 0 | 63 | Midstream format (simulated stream method): | Product | Predicate device | | | | --- | --- | --- | --- | | | | Negative | Positive | | SEJOY hCG One Step Pregnancy Test Midstream | Negative | 72 | 0 | | | Positive | 0 | 63 | 2. Matrix Comparison: Not Applicable. The device is intended for urine samples only. C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Lay user study A total of 440 lay users with varying ages and educational backgrounds were recruited at three sites. At each testing site, 110 lay users tested their own urine sample using one test method (strip, cassette, midstream (stream method), or midstream (dip method)) according to the package insert, provided a sample for professional testing, and took a questionnaire regarding their experience with the candidate device. The questionnaire results indicated that lay users found the test easy to use, the results clear and easy to read, and the instructions for use easy to understand. The data show that the agreement between lay user results and professional results was 100%. Strip format | SEJOY hCG One Step Pregnancy Test Strip | Professional | | | | --- | --- | --- | --- | | | | Positive | Negative | | Lay person | Positive | 46 | 0 | | | Negative | 0 | 64 | K212447 - Page 8 of 9 {9} K212447 - Page 9 of 9 Cassette format | SEJOY hCG One Step Pregnancy Test Cassette | Professional | | | | --- | --- | --- | --- | | | | Positive | Negative | | Lay person | Positive | 44 | 0 | | | Negative | 0 | 66 | Midstream format (dip method): | SEJOY hCG One Step Pregnancy Test Midstream | Professional | | | | --- | --- | --- | --- | | | | Positive | Negative | | Lay person | Positive | 42 | 0 | | | Negative | 0 | 68 | Midstream format (simulated stream method): | SEJOY hCG One Step Pregnancy Test Midstream | Professional | | | | --- | --- | --- | --- | | | | Positive | Negative | | Lay person | Positive | 45 | 0 | | | Negative | 0 | 65 | D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Not applicable. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...