CHEK-STIX U.T.I. SELF-TEST

K960888 · Bayer Corp. · JMT · Jul 5, 1996 · Clinical Chemistry

Device Facts

Record IDK960888
Device NameCHEK-STIX U.T.I. SELF-TEST
ApplicantBayer Corp.
Product CodeJMT · Clinical Chemistry
Decision DateJul 5, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1510
Device ClassClass 1

Indications for Use

CHEK-STIX U.T.I. Test Strips for the Detection of Urinary Tract Infection is an O.T.C. home use test for nitrite and leukocytes (WBC’s) in urine. It is intended to be used by lay persons to detect urinary tract infections.

Device Story

Plastic test strip with two reagent pads; detects nitrite and leukocyte esterase in urine specimens. User dips strip in urine; reagent pads undergo color change (pink/purple) upon reaction. User compares strip color to provided bottle label chart. Used in home setting by lay persons. Output provides qualitative detection of urinary tract infection markers; assists in identifying potential infection; facilitates timely clinical consultation.

Clinical Evidence

Clinical study comparing lay person performance to healthcare professionals. Sample size not specified. Primary endpoint: agreement between user groups. Results: 97% agreement for nitrite test; 85% agreement for WBC test.

Technological Characteristics

Dry reagent strip; colorimetric reaction (diazo) for nitrite and leukocyte esterase. Form factor: plastic strip with two reagent pads. No electronic components or software.

Indications for Use

Indicated for lay persons for the detection of urinary tract infections via testing for nitrite and leukocytes in urine.

Regulatory Classification

Identification

A nitrite (nonquantitative) test system is a device intended to identify nitrite in urine. Nitrite identification is used in the diagnosis and treatment of uninary tract infection of bacterial origin.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Bayer Corporation, Business Group Diagnostics CHEK-STIX U.T.I. Test Strips 510(k) Safety and Effectiveness Summary Page 1 of 2 JUL - 5 1996 # 510(k) SAFETY AND EFFECTIVENESS SUMMARY Prepared: June 19, 1996 *K960888* Submitter: Bayer Corporation, Business Group Diagnostics Address: 1884 Miles Avenue, P. O. Box 70 Elkhart, IN 46515 (219) 262-6929 Contact: Rosanne M. Savol, R.A.C. Manager, Regulatory Affairs Device: Trade/Proprietary Name: CHEK-STIX U.T.I. Test Strips for the Detection of Urinary Tract Infection Common/Usual Name: Test for nitrite in urine Test for leukocytes (WBC’s) in urine Document Control Number: K960888 Classification Information: Division of Clinical Laboratory Devices Panel - Clinical Chemistry and Toxicology Classification Codes -- CH(75) JMT [Diazo (Colorimetric), Nitrite (Urinary, Non-Quant.)] CH(75) LJX (Urine Leukocyte Test) Predicate Device: MULTISTIX® 2 Reagent Strips for Urinalysis Manufactured by Bayer Corporation Device Description: CHEK-STIX U.T.I. Test Strips for the Detection of Urinary Tract Infection is a plastic strip to which is affixed two reagent pads to test for nitrite and leukocytes (WBC’s) in a urine specimen. The product is provided as a carton containing a bottle of five strips and a package insert. Intended Use: CHEK-STIX U.T.I. Test Strips for the Detection of Urinary Tract Infection is an O.T.C. home use test for nitrite and leukocytes (WBC’s) in urine. It is intended to be used by lay persons to detect urinary tract infections. Revised 6/3/96 {1} Bayer Corporation, Business Group Diagnostics CHEK-STIX U.T.I. Test Strips 510(k) Safety and Effectiveness Summary Page 2 of 2 ## Technological Characteristics: CHEK-STIX U.T.I. Test Strips are dry reagent strips which, when in contact with a urine specimen, reacts with nitrite and leukocyte esterase in the specimen to give a pink and a purple color. The color of the strip is compared to a color chart on the bottle label to obtain the result. CHEK-STIX U.T.I. has the same test pads as are on strips used by doctors and clinical labs. ## Assessment of Performance: A study was conducted in clinical settings to demonstrate that lay persons can obtain results with CHEK-STIX U.T.I. comparable to results obtained by health care professionals. Agreement between the two groups was 97% for the nitrite test and 85% for the WBC test. ## Conclusion: The results of the patient-use evaluation demonstrate that lay persons can use CHEK-STIX U.T.I. Test Strips and obtain results that are in agreement with results obtained by health care professionals. The instructions are simple to understand and afford the home user a way to detect urinary tract infections. Revised 6/3/96
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...