The intended use of Ciba Corning Ciba Corning ACS:180 DPD Assay is for the quantitative of deoxypyridinoline (DPD) in urine using the Ciba Corning ACS:180 automated chemiluminescence systems
Device Story
Competitive immunoassay using direct chemiluminescent technology; performed on ACS:180 automated system. Input: urine sample. Mechanism: DPD in sample competes with pyridinoline bound to paramagnetic particles for monoclonal mouse anti-DPD antibody labeled with acridinium ester. Output: relative light units (RLUs) inversely proportional to DPD concentration. Used in clinical laboratory settings by trained technicians. Results assist clinicians in evaluating bone resorption status.
Clinical Evidence
Bench testing only. Method comparison performed on 752 urine samples (range 6-350 nM) against an alternate ELISA method; correlation coefficient r=0.961; regression equation: y=1.05x + 2.13. Total precision (%CV) ranged from 2.84% to 9.76%. Analytical sensitivity (minimum detectable concentration) is 5.0 nM.
Indicated for the quantitative measurement of deoxypyridinoline (DPD) in urine samples to assess bone resorption.
Regulatory Classification
Identification
A hydroxyproline test system is a device intended to measure the amino acid hydroxyproline in urine. Hydroxyproline measurements are used in the diagnosis and treatment of various collagen (connective tissue) diseases, bone disease such as Paget's disease, and endocrine disorders such as hyperparathyroidism and hyperthyroidism.
K980833 — IMMULITE PYRILINKS-D MODEL NUMBERS LKPD1, LKPD5 · Diagnostic Products Corp. · May 1, 1998
K972580 — DEOXYPYRIDINOLINE ASSAY FOR THE BAYER IMMUNO 1 SYSTEM · Bayer Corp. · Sep 22, 1997
K960171 — ACTIVE CROSSLAPS ELISA · Diagnostic Systems Laboratories, Inc. · Mar 28, 1996
K974260 — PYRILINKS-D ASSAY KIT · Metra Biosystems, Inc. · Jan 26, 1998
Submission Summary (Full Text)
{0}
Company Confidential
K963251
OCT 15 1996
# Summary of Safety and Effectiveness
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
## 1. Submitters Information
Contact person: William J. Pignato
Director of Regulatory Affairs
Address: Ciba Corning Diagnostics Corp.
63 North Street
Medfield, MA 02052
Phone: 508 359-3825
Date Summary Prepared: August 15, 1996
## 2. Device Information
Proprietary Name: Ciba Corning ACS:180 DPD Assay
Common Name: Assay for Deoxypyridinoline
## 3. Predicate Device Information
Name: Pyrilinks®-D
Manufacturer: Metra Biosystems, Inc
## 4. Device Description
The Ciba Corning ACS:180 DPD assay is a competitive immunoassay using direct, competitive chemiluminescent technology. DPD in the patient sample competes with pyridinoline bound to the paramagnetic particles in the Solid Phase for a limited amount of monoclonal mouse anti-DPD antibody in the Lite Reagent. The monoclonal mouse anti-DPD is bound to goat anti-mouse antibody labeled with acridinium ester. An inverse relationship exists between the amount of DPD present in the sample and the amount of relative light units (RLUs) detected by the system.
## 5. Statement of Intended Use
The intended use of Ciba Corning Ciba Corning ACS:180 DPD Assay is for the quantitative of deoxypyridinoline (DPD) in urine using the Ciba Corning ACS:180 automated chemiluminescence systems
Ciba Corning ACS:180 DPD
August 15, '996
Page 2
{1}
Company Confidential
## 6. Summary of Technological Characteristics
The Ciba Corning ACS:180 DPD assay is a competitive immunoassay using direct, competitive chemiluminescent technology
## 7. Performance Characteristics
### Sensitivity
The ACS:180 DPD assay measures DPD concentrations up to 350 nM with a minimum detectable concentration of 5.0 nM. Analytical sensitivity is defined as the concentration of DPD that corresponds to the RLUs that are two standard deviations less than the mean RLUs of 20 replicate determinations of the DPD zero standard.
### Method Comparison
For 752 urine samples in the range of 6 to 350 nM, the relationship between ACS:180 DPD and an alternate ELISA method is described by the equation:
$$
\text{ACS:180 DPD} = 1.05 \times (\text{alternate method}) + 2.13
$$
Correlation coefficient (r) = 0.961
### Precision
Total precision (%CV) ranged from 2.84 to 9.76
Ciba Corning ACS:180 DPD
August 15, '996
Page 3
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.