The PIVO™ Pro Needle-free Blood Collection Device attaches to a peripheral IV catheter system for use to obtain venous blood specimens into a vacuum tube or syringe from adult and pediatric patients, including those with difficult intravenous access who may have small, fragile, and/or non-palpable veins.
Device Story
The PIVO™ Pro is a sterile, single-use, needle-free blood collection device designed to attach to peripheral IV catheters and extension sets (including NearPort™ IV Access). It consists of an inner flow tube with a slider, a proximal flexible tube with a female Luer, an outer housing, and a distal winged clip. The clinician attaches the winged clip to the IV system and the female Luer to an evacuated tube holder or syringe. The inner flow tube is advanced through the IV catheter into the vein to collect blood, then retracted and removed. Used in clinical settings, it facilitates blood draws in patients with difficult venous access, potentially reducing the need for additional venipunctures and improving patient comfort.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including biocompatibility (ISO 10993-1), packaging (ASTM D4169, ISO 11607-1), sterilization validation (ISO 11137-1), and internal BD studies (catheter compatibility, flow rate, hemolysis, leak testing, and usability testing).
Indicated for adult and pediatric patients, including those with difficult intravenous access (DIVA) having small, fragile, or non-palpable veins, to obtain venous blood specimens via peripheral IV catheter.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 28, 2023
Becton Dickinson Infusion Therapy Systems, Inc. Amy Moore Senior Regulatory Affairs Specialist 9450 S State St Sandy, Utah 84070
Re: K230865
Trade/Device Name: PIVOTM Pro Needle-free Blood Collection Device Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: August 25, 2023 Received: August 28, 2023
Dear Amy Moore:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230865
Device Name
PIVO™ Pro Needle-free Blood Collection Device
#### Indications for Use (Describe)
The PIVO™ Pro Needle-free Blood Collection Device attaches to a peripheral IV catheter system for use to obtain venous blood specimens into a vacuum tube or syringe from adult and pediatric patients, including those with difficult intravenous access who may have small, fragile, and/or non-palpable veins.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the BD logo. The logo consists of an orange circle with a white starburst inside on the left. To the right of the circle are the letters "BD" in blue.
# K230865 510(k) Summary (21 CFR §807.92) PIVO™ Pro Needle-free Blood Collection Device
| Submitter<br>Information | Submitter Name: | Becton Dickinson Infusion Therapy Systems Inc. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| | Submitter Address: | 9450 South State Street, Sandy, Utah 84070 |
| | Contact Person: | Amy Moore, Sr. Regulatory Affairs Specialist |
| | Email Address: | amy.moore@bd.com |
| | Phone Number: | (801) 522-5000 |
| | Date of Preparation: | September 28, 2023 |
| Subject Device | Trade Name: | PIVO™ Pro Needle-free Blood Collection Device |
| | Common Name: | Blood Collection Device |
| | 510(k) Reference: | K230865 |
| | Regulation Number: | 21 CFR §862.1675 |
| | Regulation Name: | Blood Specimen Collection Device |
| | Regulatory Class: | II |
| | Product Code: | JKA |
| | Classification Panel: | General Hospital |
| Predicate<br>Device | Trade Name: | PIVO™ Needle-free Blood Collection Device |
| | Common Name: | Blood Collection Device |
| | 510(k) Reference: | K193569 |
| | Regulation Number: | 21 CFR 862.1675 |
| | Regulation Name: | Blood Specimen Collection Device |
| | Regulatory Class: | II |
| | Product Code: | JKA |
| | Classification Panel: | General Hospital |
| Reason for<br>Submission | The purpose of this submission is to notify the FDA of the introduction of PIVO™ Pro<br>Needle-free Blood Collection Device, allowing for compatibility with a peripheral IV<br>catheter with NearPort™ IV Access and addition of a patient population statement to<br>the Indications for Use. | |
| Device<br>Description | The PIVO™ Pro Needle-free Blood Collection Device is a sterile, single use needle-<br>free collection device that attaches to a peripheral intravascular catheter (IV) system.<br>The device is comprised of an inner flow tube with a slider, proximal flexible tube | |
| with female Luer, outer housing, and winged clip on the distal end. The winged clip<br>attaches to the IV system. The female Luer attaches to an evacuated tube holder or<br>syringe. The inner flow tube is then advanced to collect a blood sample. Once<br>complete, the inner flow tube is retracted, and the device is removed from the IV.<br><br>The device is available in three sizes: 20GA, 22GA, and 24GA IV compatible. The<br>device is compatible with the corresponding IV gauge and larger IV catheters.<br>The PIVOTM Pro Needle-free Blood Collection Device is compatible with peripheral<br>IV catheters and extension sets with NearPort™ IV Access. | | |
| Indications for<br>Use<br>(21 CFR §<br>807.92(a)(5)) | The PIVOTM Pro Needle-free Blood Collection Device attaches to a peripheral IV<br>catheter system for use to obtain venous blood specimens into a vacuum tube or<br>syringe from adult and pediatric patients, including those with difficult intravenous<br>access who may have small, fragile, and/or non-palpable veins. | |
| Technological<br>Characteristics | Technological characteristics of the subject device are substantially equivalent to the<br>predicate device. The subject PIVOTM Pro Needle-free Blood Collection Device<br>achieves its intended use based on the same technology and principles of operation as<br>the predicate device. | |
{5}------------------------------------------------
A comparison of the subject and predicate device technological characteristics is provided in the table below.
| Attribute | SUBJECT (K230865)<br>PIVO™ Pro Needle-free<br>Blood Collection Device | PREDICATE (K193569)<br>PIVO™ Needle-free Blood<br>Collection Device | Comparison & Discussion |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 862.1675<br>Class II<br>JKA – Blood Specimen<br>Collection Device | 21 CFR 862.1675<br>Class II<br>JKA – Blood Specimen<br>Collection Device | Same |
| Indications for<br>Use | The PIVO™ Pro Needle-free<br>Blood Collection Device<br>attaches to a peripheral IV<br>catheter system for use to<br>obtain venous blood<br>specimens into a vacuum tube<br>or syringe from adult and<br>pediatric patients, including<br>those with difficult<br>intravenous access who may<br>have small, fragile, and/or<br>non-palpable veins. | The PIVO™ device is<br>attached to a peripheral IV<br>catheter for use as a direct<br>blood draw device into a<br>vacuum tube or a syringe. | Minor grammatical wording<br>changes with no change to<br>meaning.<br>Addition of a patient<br>population statement, with<br>specific reference to pediatric<br>patients and patients with<br>DIVA (difficult intravenous<br>access) which are subsets of<br>'General Use' and do not<br>introduce new risks not<br>normally associated with the<br>general use of the subject<br>device. Therefore,<br>modifications, do not raise |
| Attribute | SUBJECT (K230865)<br>PIVOTM Pro Needle-free<br>Blood Collection Device | PREDICATE (K193569)<br>PIVOTM Needle-free Blood | Comparison & Discussion |
| | | Collection Device | new or different questions of<br>safety or effectiveness. |
| Intended Use | Venous blood draw | Venous blood draw | Same |
| Patient<br>Interface | Separately placed<br>commercially available<br>peripheral IV catheter | Separately placed<br>commercially available<br>peripheral IV catheter | Same |
| PIV Attachment | Clip-to-Connect | Clip-to-Connect | Same |
| Blood Collection<br>Attachment | Female Luer to Blood<br>Transfer Device or Syringe | Female Luer to Blood<br>Transfer Device or Syringe | Same |
| Blood Control<br>Mechanism | Cap on female luer and clamp<br>on flexible tubing | Cap on female luer and clamp<br>on flexible tubing | Same |
| Tubing | Transparent<br>Flexible | Transparent<br>Flexible | Same |
| | Primary Components Material Composition | | |
| ISO 10993-1<br>Biocompatibility<br>Contact Type<br>and Duration | Body Contact: Externally<br>communicating device<br>Contact: Circulating blood<br>Contact Duration: Limited (A)<br>(<24 hrs) | Body Contact: Externally<br>communicating device<br>Contact: Circulating blood<br>Contact Duration: Limited (A)<br>(<24 hrs) | Same |
| Housing | Polycarbonate | Polycarbonate | Same |
| Inner Flow<br>Tube | Polyimide | Polyimide | Same |
| Proximal<br>Tubing | Vestamid | Vestamid | Same |
| Luer | Nylon | Nylon | Same |
| Slider (Pusher) | Nylon | Nylon | Same |
| Winged Clip | Polycarbonate | Polycarbonate | Same |
| Clamp | Nylon | HDPE | Same |
| Color | Pink<br>Blue<br>Yellow | Pink<br>Blue<br>Yellow | Same |
| Compatible PIV<br>Sizes | The device is available in<br>three sizes: 20 GA, 22 GA,<br>and 24 GA IV compatible.<br>The device is compatible with<br>the corresponding IV gauge<br>and larger IV catheters. | 14 GA - 24 GA | Clarification only. Both the<br>predicate and subject devices<br>are available in three sizes:<br>20GA, 22GA and 24 GA IV<br>compatible, as indicated on<br>the product label. The subject<br>and predicate devices are<br>compatible with the corresponding IV gauge and larger IV catheters. |
| Attribute | SUBJECT (K230865)<br>PIVO™ Pro Needle-free<br>Blood Collection Device | PREDICATE (K193569)<br>PIVO™ Needle-free Blood<br>Collection Device | Comparison & Discussion |
| | | | corresponding IV gauge and<br>larger IV catheters. For<br>example, a 20GA IV<br>compatible subject or<br>predicate device can be used<br>with a 20GA or larger IV<br>catheter (e.g., 18GA). |
| | | | The clarification does not<br>change the use of the device<br>or device specifications and,<br>therefore, does not raise new<br>or different questions of<br>safety or effectiveness. |
| Proximal<br>Tubing | Transparent<br>Flexible | Transparent<br>Flexible | Same |
| Inner Flow<br>Tube (Distal<br>Tubing) Length | 20 GA = 155.28 mm<br>22 GA = 155.28 mm<br>24 GA = 132.68 mm | 20 GA = 141.2 mm<br>22 GA = 141.2 mm<br>24 GA = 141.2 mm | The results of design<br>validation and verification<br>testing demonstrate that the<br>modifications to the subject<br>device perform as intended<br>and do not raise new or<br>different questions of safety<br>or effectiveness. |
| Outer Diameter<br>(OD) of Inner<br>Flow Tube<br>(Distal Tubing) | 20 GA = 0.709 mm max<br>22 GA = 0.543 mm max<br>24 GA = 0.400 mm max | 20 GA = 0.709 mm max<br>22 GA = 0.543 mm max<br>24 GA = 0.400 mm max | Same |
| Wall Thickness<br>of Inner Flow<br>Tube (Distal<br>Tubing) | 20 GA = 0.709 mm<br>22 GA = 0.0635 mm<br>24 GA = 0.0575 mm | 20 GA = 0.709 mm<br>22 GA = 0.0635 mm<br>24 GA = 0.0575 mm | Same |
| Sample<br>collection | Device attaches to female luer<br>of PIV system, tube inserted<br>into PIV, blood is drawn<br>through tube into a blood<br>transfer device | Device attaches to female luer<br>of PIV system, tube inserted<br>into PIV, blood is drawn<br>through tube into a blood<br>transfer device | Same |
| Packaging<br>Material | Tyvek/PET<br>OR<br>Nylon/Nylon | Tyvek/PET<br>OR<br>Nylon/Nylon | Same |
| Complete<br>Retraction | Yes | Yes | Same |
| Attribute | SUBJECT (K230865)<br>PIVOTM Pro Needle-free<br>Blood Collection Device | PREDICATE (K193569)<br>PIVOTM Needle-free Blood<br>Collection Device | Comparison & Discussion |
| Sterilization<br>Method | Gamma | Gamma | Same |
| Single Use Only | Yes | Yes | Same |
{6}------------------------------------------------
Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, Utah 84070 USA
### bd.com
{7}------------------------------------------------
Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, Utah 84070 USA
### bd.com
{8}------------------------------------------------
Performance tests completed on the subject devices were limited to those tests required Summary of Performance to support a determination of substantial equivalence to the predicate device. A risk Tests analysis was conducted to assess the impact of the proposed modifications to the predicate device. When technological characteristics between the subject and predicate devices were found to be identical, results of performance testing conducted on the predicate devices were applied to the subject device. The performance tests listed below were conducted to ensure that the subject device meets predetermined design requirements:
- 1) Compliance Testing
- Biocompatibility (ISO 10993-1) ●
- Body contact: Externally communicating device -
- Contact: Circulating blood -
- -Contact duration: Limited (A) (<24 hrs)
- Packaging Testing (ASTM D4169) ●
- Sterilization Validation (ISO 11137-1) ●
### 2) BD Internal Studies
- Catheter Gauge Compatibility ●
- Extension Length
- Catheter Perforation ●
- Insertion ●
- Alignment
- Atraumatic Tip ●
- Flow Rate
- Hemolysis ●
- Pusher Pull Force
- Blood Leak ●
- Pressure Leak ●
- . Vacuum Leak
3) Packaging verification testing per ISO 11607-1 Packaging for terminally sterilized medical devices. Part 1: Requirements for materials, sterile barrier systems and packaging systems
{9}------------------------------------------------
| | 4) Usability testing per Applying Human Factors and Usability Engineering to<br>Medical Devices Guidance for Industry and Food and Drug Administration Staff,<br>issued February 2016 |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Per design control requirements specified in 21 CFR 820.30, the subject device met all<br>predetermined acceptance criteria for the above-listed performance tests, demonstrating<br>substantial equivalence to the predicate devices. |
| | Clinical studies are not required to demonstrate substantial equivalence to the predicate<br>device. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and results of<br>performance testing, the subject PIVOT™ Pro Needle-free Blood Collection Device has<br>been demonstrated to be substantially equivalent to the predicate PIVOT™ Needle-free<br>Blood Collection Device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.