BD Vacutainer® Trace Element K2EDTA Tubes, BD Vacutainer® Trace Element Serum Tubes
K213953 · Becton, Dickinson and Company · JKA · Jul 17, 2023 · Clinical Chemistry
Device Facts
Record ID
K213953
Device Name
BD Vacutainer® Trace Element K2EDTA Tubes, BD Vacutainer® Trace Element Serum Tubes
Applicant
Becton, Dickinson and Company
Product Code
JKA · Clinical Chemistry
Decision Date
Jul 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
BD Vacutainer® Trace Element Serum Tubes and BD Vacutainer® Trace Element K₂EDTA Tubes are plastic, evacuated, sterile, single use, in vitro diagnostic medical devices. They are intended to be used in settings where venous blood specimens are collected by trained healthcare professionals for the collection, transportation, and processing of blood in a closed tube. Blood collected in the BD Vacutainer® Trace Element Serum Tubes and BD Vacutainer® Trace Element K₂EDTA Tubes are used for trace element testing (e.g., Arsenic, Cadmium, Chromium, Copper, Lead, Manganese, Mercury, Selenium, and Zinc).
Device Story
Evacuated plastic blood collection tubes; used for venous blood collection; intended for trace element testing. Serum tubes contain spray-dried silica clot activator; K2EDTA tubes contain spray-dried dipotassium EDTA anticoagulant. Tubes feature BD Hemogard™ closures (rubber stopper with protective plastic shield) to reduce blood exposure. Operated by trained healthcare professionals in clinical settings. Blood collected in closed system; processed for laboratory analysis. Output is a blood sample suitable for trace element testing; aids clinicians in diagnosing or monitoring trace element levels. Benefits include reduced risk of tube breakage compared to glass and enhanced user safety during handling and disposal.
Clinical Evidence
Clinical testing compared subject devices to a legally marketed comparator. Evaluated within-tube stability, shelf-life performance, and repeatability/reproducibility. Results confirmed functional clinical equivalence for identifying clinically relevant levels of Arsenic, Cadmium, Chromium, Copper, Lead, Manganese, Mercury, Selenium, and Zinc. All pre-determined acceptance criteria were met.
Indicated for use by trained healthcare professionals in clinical settings for the collection, transport, and processing of venous blood specimens for trace element testing (Arsenic, Cadmium, Chromium, Copper, Lead, Manganese, Mercury, Selenium, and Zinc).
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
BD Vacutainer® Trace Element Serum Plus Tube and BD Vacutainer® Trace Element K2EDTA Plus Tube (K041071)
Related Devices
K981013 — VACUTAINER PLUS TUBE WITH EDTA ANTICOAGULANT ANDD VACUTAINER PLUS SERUM TUBE · Becton Dickinson Vacutainer Systems · Jun 17, 1998
K041071 — BD VACUTAINER TRACE ELEMENT SERUM PLUS TUBE AND BD VACUTAINER TRACE ELEMENT K2EDTA PLUS TUBE, MODELS 368380/368381 · Becton, Dickinson & CO · May 14, 2004
K213670 — BD Vacutainer K2EDTA Blood Collection Tubes, BD Vacutainer K3EDTA Blood Collection Tubes · Becton, Dickinson and Company · Aug 25, 2023
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K213953
B Applicant
Becton Dickinson and Company
C Proprietary and Established Names
BD Vacutainer® Trace Element K₂EDTA Tubes, BD Vacutainer® Trace Element Serum Tubes
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JKA | Class II | 21 CFR 862.1675 - Blood Specimen Collection Device | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Modifications to existing blood collection devices.
B Measurand:
Not applicable. Blood collection tube.
C Type of Test:
Not applicable.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
BD Vacutainer® Trace Element Serum Tubes and BD Vacutainer® Trace Element K₂EDTA Tubes are plastic, evacuated, sterile, single use, in vitro diagnostic medical devices. They are intended to be used in settings where venous blood specimens are collected by trained healthcare professionals for the collection, transportation, and processing of blood in a closed tube. Blood collected in the BD Vacutainer® Trace Element Serum Tubes and BD Vacutainer® Trace Element K₂EDTA Tubes are used for trace element testing (e.g., Arsenic, Cadmium, Chromium, Copper, Lead, Manganese, Mercury, Selenium, and Zinc).
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
BD Vacutainer® Trace Element Tubes have not been evaluated for other testing applications which use serum or EDTA anticoagulated blood.
BD Vacutainer® Trace Element tubes are controlled only for the trace elements listed within their respective upper limits. Other trace elements may be present, however, these levels have not been evaluated and are not controlled. Users should assess whether the use of these tubes is appropriate for their intended testing purposes.
D Special Instrument Requirements:
Not applicable.
IV Device/System Characteristics:
A Device Description:
The BD Vacutainer® Trace Element K₂EDTA Tube is a sterile, plastic, evacuated plastic blood collection tube. It consists of closure assembly, a plastic tube, and K₂EDTA coating and is available in two draw sizes, 6 mL (13 x 100 mm) and 3 mL (13 x 75 mm). The plastic tube is manufactured from PET (polyethylene terephthalate) plastic. The K₂EDTA anticoagulant coating is spray-dried on the walls of the tube.
The BD Vacutainer® Trace Element Serum Tube is a sterile, plastic, evacuated plastic blood collection tube consisting of a closure assembly, a plastic tube and contains a silica clot activator additive which is spray-dried on the interior walls of the tube. It is available in one draw size, 6 mL (13 x 100 mm). The plastic tube is manufactured from PET.
Both types of tubes are closed with a BD Hemogard™ Closure, which consists of a rubber stopper and protective plastic shield to help reduce user exposure to blood. The stopper/closures
K213953 - Page 2 of 10
{2}
are color coded in royal blue to reflect that the tubes are intended to collect samples for trace element.
## B Principle of Operation:
The BD Vacutainer® Trace Element K₂EDTA and BD Vacutainer® Trace Element Serum tubes use controlled vacuum to pull a specific volume of blood into the sterile interior of the tube. Within the serum tube, silica clot activator serves to enhance clot activation of the blood, thus rendering serum samples for trace element testing. The other tube contains K₂EDTA, which prevents blood coagulation, thus rendering whole blood or plasma samples for trace element testing.
The tubes are compatible with BD Vacutainer® Blood Collection Needles, Blood Collection Sets, Transfer Devices, Holders and Adapters. Once the vein of the patient has been penetrated using a standard needle, the collection tube is centered in the holder and pushed onto the needle, puncturing the stopper of the tube. Immediately after the blood has been drawn, the tube is gently inverted 8-10 times for K₂EDTA and 8 times for serum to mix the blood with the additive. For the serum tubes, blood should be allowed to clot for a minimum of 60 minutes before centrifugation, The recommended centrifugation conditions are 1300 RCF (relative centrifuge force or g) for 10 minutes.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
BD Vacutainer® Trace Element Serum Plus Tube and BD Vacutainer® Trace Element K2EDTA Plus Tube
### B Predicate 510(k) Number(s):
K041071
### C Comparison with Predicate(s):
Due to an administrative error, the shelf life of the candidate device in the table below was listed incorrectly and is now updated:
| Device & Predicate Device(s): | K213953 | K041071 |
| --- | --- | --- |
| Device Trade Name | BD Vacutainer® Trace Element K₂EDTA Tubes, BD Vacutainer® Trace Element Serum Tubes | BD Vacutainer® Trace Element Serum Plus Tubes and BD Vacutainer® Trace Element K₂EDTA Tubes |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | Plastic evacuated blood collection tubes that provide a means of collecting, | Same |
K213953 - Page 3 of 10
{3}
K213953 - Page 4 of 10
| Device & Predicate Device(s): | K213953 | K041071 |
| --- | --- | --- |
| | transporting, separating, and processing venous whole blood in a closed tube for trace element testing. | |
| Evacuated Blood Collection Tube | Yes | Same |
| Additive Type | BD Vacutainer® Trace Element Serum Tube: Serum Clot Activator
BD Vacutainer® Trace Element K_{2}EDTA Tube: K_{2}EDTA | Same |
| Additive Application/Quantity | Spray dried | Same |
| Tube Material | PET (polyethylene terephthalate) plastic | Same |
| Tube Closure | BD Hemogard™/Royal Blue | Same |
| Sterilization Method | Gamma irradiation | Same |
| Sterility Assurance Level | 10^{-6} | Same |
| **General Device Characteristic Differences** | | |
| Tube Dimensions and Draw Volume | BD Vacutainer® Trace Element Serum Tube: 13 x 100 mm, 6 mL
BD Vacutainer® Trace Element K_{2}EDTA Tube: 13 x 100 mm, 6 mL and 13 x 75 mm, 3 mL | BD Vacutainer® Trace Element Serum Plus Tube: 13 x 100 mm, 6 mL
BD Vacutainer® Trace Element K_{2}EDTA Tube: 13 x 100mm, 6 mL |
| Shelf Life | 13 x 100 mm, 6 mL BD Vacutainer® Trace Element Serum Tube and BD Vacutainer® Trace Element K_{2}EDTA Tube: 12 months
13 x 75 mm, 3 mL BD Vacutainer® Trace Element K_{2}EDTA Tube:10 months | 12 months |
| Elements Tested | Arsenic, Cadmium, Chromium, Copper, Lead, Manganese, Mercury, Selenium, and Zinc | Arsenic, Cadmium, Calcium, Chromium, Copper, Iron, Lead, Magnesium, Manganese, Mercury, Selenium, and Zinc |
{4}
VI Standards/Guidance Documents Referenced:
CLSI GP34-A: 2010 Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline
ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems
ANSI/AAMI/ISO 11137-1:2006, A1: 2013, A2 2018 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
ANSI/AAMI/ISO 11137-2: 2013 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
ANSI/AAMI/ISO 11137-3:2017 Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control
ANSI/AAMI/ISO 11737-1:2018 Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products
ANSI/AAMI/ISO 11737-2:2019 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
ANSI AAMI ST67:2019 Sterilization of health care products - Requirements and guidance for selecting a sterility assurance level (SAL) for products labeled "sterile"
EN ISO 14971:2012 Medical Devices – Application of risk management to medical devices
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A study was conducted to evaluate BD Vacutainer® Trace Element K₂EDTA Tubes (6 mL) and BD Vacutainer® Trace Element Serum Tubes (6 mL) for repeatability (within tube) and reproducibility (lot to lot and tube to tube) performance using lead and zinc as representative trace elements for testing. Two sample matrices, including serum (for zinc testing) and anticoagulated whole blood (for lead testing), were evaluated on two instrument platforms for Zinc and one platform for Lead to confirm tube performance. Three lots of each of the evaluation tubes (BD Vacutainer® Trace Element K₂EDTA Tubes and BD Vacutainer® Trace Element Serum Tubes) were evaluated.
The first study (whole blood lead testing) was performed using samples collected from 35 subjects. The second study (serum zinc testing) was performed using samples collected from 43 subjects.
K213953 - Page 5 of 10
{5}
The samples were tested for trace elements using atomizing methods, i.e., inductively coupled plasma mass spectrometry (ICP-MS).
Variance Components and Confidence Intervals (CI): ICP-MS Platform 1
| Analyte/Unit | Tube Type | Mean | Variance Component | CV (%) | 95% CI |
| --- | --- | --- | --- | --- | --- |
| Lead (μg/L) | BD Trace Element K₂EDTA | 5.4056 | Between Lots | 0 | 0, 0 |
| | | | Between Tubes | 3.6 | 3.1, 4.1 |
| | | | Within Tube | 1.2 | 1.1, 1.4 |
| | | | Total | 3.8 | 3.0, 5.4 |
| Zinc (μg/L) | BD Trace Element Serum | 74.5167 | Between Lots | 3.3 | 1.9, 4.3 |
| | | | Between Tubes | 4.3 | 3.6, 4.9 |
| | | | Within Tube | 1.5 | 1.4, 1.6 |
| | | | Total | 5.6 | 4.6, 7.2 |
Variance Components and Confidence Intervals (CI): ICP-MS Platform 2
| Analyte/Unit | Tube Type | Mean | Variance Component | CV (%) | 95% CI |
| --- | --- | --- | --- | --- | --- |
| Zinc (μg/L) | BD Trace Element Serum | 751.5803 | Between Lots | 0 | 0, 3.5 |
| | | | Between Tubes | 5.6 | 3.8, 6.9 |
| | | | Within Tube | 7.3 | 6.6, 8 |
| | | | Total | 9.1 | 7.7, 11.3 |
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Benchtop studies were conducted to evaluate interference from stopper materials over the sample storage time and over the shelf life of the tube. Study protocols, acceptance criteria and results for these studies were provided and found to be acceptable.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
a) Shelf-life stability:
Real Time stability testing of the BD Vacutainer® Trace Element K₂EDTA Tubes and BD Vacutainer® Trace Element Serum Tubes showed that the 6 mL BD Vacutainer® Trace Element K₂EDTA Tubes and 6 mL BD Vacutainer® Trace Element Serum Tubes are stable for 12 months (6 mL tubes) and the 3 mL BD Vacutainer® Trace Element
K213953 - Page 6 of 10
{6}
$\mathrm{K}_2\mathrm{EDTA}$ Tubes are stable for 10 months, when stored at $4^{\circ}\mathrm{C}$ to $25^{\circ}\mathrm{C}$. Stability study protocol and acceptance criteria has been reviewed and found to be acceptable.
## b) Analyte stability:
Multiple analyte stability studies were conducted to assess the analyte within tube stability for As, Cd, Cr, Cu, Hg, Pb, Mn, Se and Zn in BD Vacutainer® Trace Element $\mathrm{K}_2\mathrm{EDTA}$ Tubes and BD Vacutainer® Trace Element Serum Tubes. Stability studies were assessed at initial time (Time 0) and 24 hrs at room temperature storage (for As, Cd, Cr, Cu, Hg, Pb, Mn, Se) and 4 hrs at room temperature storage (Zn).
The results of the study support the sponsor’s claim, the within-tube stability was demonstrated in the BD Vacutainer® Trace Element $\mathrm{K}_2\mathrm{EDTA}$ Tubes and BD Vacutainer® Trace Element Serum Tubes for up to 24 hrs at room temperature storage for As, Cd, Cr, Cu, Hg, Pb, Mn, Se and up to 4 hrs at room temperature storage for Zn.
## c) Clinical Transportation Study:
A study was conducted to assess the analyte within tube stability for As, Cd, Cr, Cu, Hg, Pb, Mn, Se and Zn when exposed to simulated transportation conditions and worst-case room temperature. Conditions evaluated included:
- Exposure to vibration and impact simulations that replicate the vibratory and impact loadings which a small parcel (e.g., small box) is expected to experience when shipped via a parcel carrier.
- Exposure to the upper limit of the labeled temperature of $25^{\circ}\mathrm{C}$.
The results of the study support that the tubes can be used to transport venous blood sample when subjected to the drop, vibration, temperature and humidity conditions used in this test method. Study protocols and acceptance criteria have been reviewed, and performance was considered acceptable.
## d) Additional Bench Testing on the Candidate Device:
Due to an administrative error, the packaging performance study in the section below was previously redacted and is now included:
Benchtop studies were conducted to assess draw volume, X-value, $2^{\mathrm{nd}}$ stopper pullout, stopper/shield separation, stopper leakage, tube leakage, breakage resistance during drop testing, breakage resistance during centrifugation testing, packaging performance and the trace metal content prior to blood draw. The study protocols were reviewed, and performance was considered acceptable.
## 6. Detection Limit:
Not applicable.
## 7. Assay Cut-Off:
Not applicable.
K213953 - Page 7 of 10
{7}
B Comparison Studies:
1. Method Comparison with Predicate Device:
A study was conducted to evaluate the Clinical Equivalence of BD Vacutainer® Trace Element Blood Collection Tubes (BD Vacutainer® Trace Element Serum and K₂EDTA) in comparison with Greiner Bio-One Vacuette® Trace Element Blood Collection Tubes (Greiner Bio-One Vacuette® Trace Element Serum and Sodium Heparin) for testing trace elements.
A total of 159 subjects were enrolled in the study across several study parts. All samples were collected by routine venipuncture into the corresponding candidate tube and also in no additive tubes that were used for the preparation of the contrived samples to obtain results for high levels of trace elements that spanned the analytical measurement range.
Immediately after collection and prior to centrifugation, tubes were inspected for complete fill, clot formation and stopper leakage. After centrifugation, additional visual assessments were performed. Tubes were then handled and processed according to manufacturer's recommended handling conditions for tube inversions and centrifugation time and force.
The following tube comparisons were performed for clinical equivalence:
- 6 mL BD Trace Element K₂EDTA vs 6 mL Greiner Trace Element NaHep
- 6 mL BD Trace Element Serum vs 6 mL Greiner Trace Element Serum
- 3 mL BD Trace Element K₂EDTA vs 6 mL Greiner Trace Element NaHep
- 3 mL BD Trace Element K₂EDTA vs 6 mL BD Trace Element K₂EDTA
Samples collected in the BD Trace Element K₂EDTA were analyzed on one ICP-MS instrument platform for testing As, Cd, Cr, Pb, Mn, Hg and Se in anticoagulated whole blood and samples collected in the BD Trace Element Serum were tested for Cr, Cu, Mn, Se and Zn in serum. Biases between tube types at pre-determined medical decision points were estimated with 95% intervals. The mean biases and 95% limit were compared to the clinically acceptable limit (CAL). Data was analyzed using Passing Bablok regression.
Summary of the Passing-Bablok regressions analyses is provided in the table below:
| Analyte | Comparison | N | Range Tested (μg/mL) | Slope | Intercept | r |
| --- | --- | --- | --- | --- | --- | --- |
| Arsenic | BD Trace Element K₂EDTA 3 mL vs Greiner Trace Element NaHep | 121 | 0.340-676.800 | 0.99 | -0.02 | 0.986 |
| | BD Trace Element K₂EDTA 6 mL vs Greiner Trace Element NaHep | 121 | 0.340-676.800 | 1.01 | -0.02 | 0.988 |
| Cadmium | BD Trace Element K₂EDTA 3 mL vs Greiner Trace Element NaHep | 121 | 0.140-33.580 | 0.97 | 0.01 | 0.997 |
| | BD Trace Element K₂EDTA 6 mL vs Greiner Trace Element NaHep | 121 | 0.140-33.580 | 1 | 0 | 0.997 |
K213953 - Page 8 of 10
{8}
| Analyte | Comparison | N | Range Tested (μg/mL) | Slope | Intercept | r |
| --- | --- | --- | --- | --- | --- | --- |
| Chromium | BD Trace Element K_{2}EDTA 3 mL vs Greiner Trace Element NaHep | 122 | 0.040-5.210 | 1 | 0.1 | 0.949 |
| | BD Trace Element K_{2}EDTA 6 mL vs Greiner Trace Element NaHep | 121 | 0.040-5.210 | 1.03 | 0.1 | 0.929 |
| | BD Trace Element Serum vs Greiner Trace Element Serum | 110 | 0.060-8.560 | 1 | 0 | 0.987 |
| Copper | BD Trace Element Serum vs Greiner Trace Element Serum | 108 | 17.132-415.634 | 1 | 0.51 | 0.996 |
| Lead | BD Trace Element K_{2}EDTA 3 mL vs Greiner Trace Element NaHep | 121 | 0.180-88.766 | 0.97 | 0.02 | 0.997 |
| | BD Trace Element K_{2}EDTA 6 mL vs Greiner Trace Element NaHep | 121 | 0.180-88.766 | 0.99 | -0.01 | 0.997 |
| Manganese | BD Trace Element K_{2}EDTA 3 mL vs Greiner Trace Element NaHep | 120 | 3.690-93.656 | 1 | 0.19 | 0.979 |
| | BD Trace Element K_{2}EDTA 6 mL vs Greiner Trace Element NaHep | 120 | 3.690-93.656 | 1.03 | -0.39 | 0.981 |
| | BD Trace Element Serum vs Greiner Trace Element Serum | 106 | 0.128-169.554 | 0.99 | 0.07 | 0.995 |
| Mercury | BD Trace Element K_{2}EDTA 3 mL vs Greiner Trace Element NaHep | 120 | 0.250-73.637 | 1.04 | -0.18 | 0.979 |
| | BD Trace Element K_{2}EDTA 6 mL vs Greiner Trace Element NaHep | 119 | 0.250-73.637 | 1.06 | -0.15 | 0.978 |
| Selenium | BD Trace Element K_{2}EDTA 3 mL vs Greiner Trace Element NaHep | 110 | 134.053-711.770 | 0.99 | 2.74 | 0.976 |
| | BD Trace Element K_{2}EDTA 6 mL vs Greiner Trace Element NaHep | 110 | 91.547-711.770 | 0.99 | 2.67 | 0.976 |
| | BD Trace Element Serum vs Greiner Trace Element Serum | 108 | 53.862-569.510 | 1.01 | -0.73 | 0.977 |
| Zinc | BD Trace Element Serum vs Greiner Trace Element Serum | 108 | 34.560-259.571 | 1.01 | -1.25 | 0.994 |
2. Matrix Comparison:
Not applicable.
K213953 - Page 9 of 10
{9}
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K213953 - Page 10 of 10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.