The Blood Transfer Accessory is a sterile non invasive device for use in venous blood collection procedures in combination with blood collection devices with a male luer feature.
Device Story
Blood Transfer Accessory (BTA) is a single-use, sterile, non-invasive device for venous blood collection. It consists of a polypropylene holder with a female luer feature, a stainless-steel needle, and a rubber sleeve. The device connects to blood collection devices with a male luer feature. During use, a sample collection tube is inserted into the holder; the needle pierces the tube to facilitate blood transfer. The holder design protects the user from needle sticks and direct blood contact. After collection, the tube is removed. The device is intended for use by healthcare personnel or users in clinical settings. It functions as a mechanical interface to safely transfer blood from a venous access point to a collection tube.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemocompatibility) and mechanical performance testing (flow rate, leakage, insertion/evacuation force, needle/sleeve holding strength, and deformation resistance).
Technological Characteristics
Materials: Polypropylene holder, Acrylonitrile butadiene styrene (ABS) adapter, stainless-steel needle, rubber sleeve. Sterilization: Ethylene Oxide (EtO) to SAL 10^-6 per ISO 11135:2014. Connectivity: Mechanical interface (female luer). Standards: ISO 80369-7:2016, ISO 80369-20:2015, ISO 10993 series.
Indications for Use
Indicated for use in venous blood collection procedures in combination with blood collection devices with a male luer feature.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
K190485 — Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) · Kawasumi Laboratories, Inc. · Jul 8, 2019
K160532 — Greiner Holdex · Greiner Bio-One Na, Inc. · Nov 3, 2016
K223810 — Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) · Sb-Kawasumi Laboratories, Inc. · Feb 24, 2023
K242320 — BD Vacutainer® One Use Holder · Becton, Dickinson and Company · Nov 1, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 30, 2019
NSP Tech Pte Ltd HC Tan 10 Admiralty Street, Northlink Building, #02-06 757695 Singapore
Re: K181743
Trade/Device Name: Blood Transfer Accessory (BTA) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: April 16, 2019 Received: June 28, 2019
Dear HC Tan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name Blood Transfer Accessory (BTA)
### Indications for Use (Describe)
The Blood Transfer Accessory is a sterile non invasive device for use in venous blood collection procedures in combination with blood collection devices with a male luer feature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
#### SUBMITTER: 1.
### Applicant Name:
NSP Tech Pte Ltd 10 Admiralty Street, Northlink Building, #02-06 Singapore 757695
## Contact Person:
Name: HC Tan Phone: (65)98180450 Fax:(65)67476533 Email: hockchoon.tan@nsptech.com.sg
| Establishment registration number | :3008337059 |
|-----------------------------------|-----------------|
| Date prepared | : July 29, 2019 |
# 2. DEVICE
| Trade Name | : | Blood Transfer Accessory (BTA) |
|-------------------|---|----------------------------------|
| Common Name | : | Blood Specimen Collection Device |
| Regulatory Class | : | Class II |
| Regulation Number | : | 21 CFR 862.1675 |
| Regulation Name | : | Blood Specimen Collection Device |
| Product Code | : | JKA |
| Panel | : | Clinical Chemistry |
#### PREDICATE DEVICE 3.
Vacutainer® Brand Multiple Sample Luer Adapter
510(k) Number: K991088 Regulatory Class: Class II Regulation Number: 21 CFR 862.1675 JKA Product Code:
#### DEVICE DESCRIPTION: 4.
The Blood Transfer Accessory (BTA) is a single use device comprising of a holder made from polypropylene plastic with a female luer feature. A stainless-steel needle covered with a rubber sleeve is attached to the inside of the holder.
The Blood Transfer Accessory (BTA) has a Holder that encases the Needle preventing any needle stick issue with user. The Holder of the Blood Transfer Accessory (BTA) also house the sample collection tube and protects the user or healthcare personnel from direct contact with
{4}------------------------------------------------
blood specimen. The Adapter connected to the blood collection device for the blood transfer. The sample collection tube is inserted from the far end of the holder where the opening is located. This transfer is conducted until the required amount of blood is completed. The sample collection tube is then evacuated from the Blood Transfer Accessory (BTA) by pulling away.
#### INDICATION FOR USE STATEMENT 5.
The Blood Transfer Accessory is a sterile non invasive device for use in venous blood collection procedures in combination with blood collection devices with a male luer feature.
#### SUBSTANTIAL EQUIVALENCE COMPARISON 6.
### TABLE OF COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
| | NSP Tech Pte Ltd<br>Blood Collection<br>Accessory(BUCA) | Vacutainer® Brand<br>Multiple Sample Luer<br>Adapter | Discussion/Comment |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Number | K181743 | K991088 | same |
| Classification | Class II | Class II | same |
| Product Code | JKA | JKA | same |
| Indication For Use | The Blood Transfer<br>Accessory is a<br>sterile non invasive<br>device for use in<br>venous blood<br>collection<br>procedures in<br>combination with<br>blood collection<br>devices with a male<br>luer feature. | The Vacutainer Brand<br>Luer Adapter is a sterile,<br>non-invasive device used<br>to connect venous access<br>devices such as needles,<br>blood collection sets, and<br>infusion sets to blood<br>collection tubes. They<br>are also used in<br>connection with non-<br>needle devices for<br>collection of blood from<br>catheters. The<br>Vacutainer Brand Luer<br>Adapter is sold by itself<br>and as a component of<br>other Vacutainer Brand<br>devices. | Similar- The difference<br>is in the name of the<br>device and this does not<br>raise any questions of<br>safety or effectiveness.<br>The predicate also<br>specifies that devices<br>that it connects to. The<br>subject device states it<br>connects to devices<br>with a male luer. |
| Manufacturer | NSP Tech Pte Ltd | Becton Dickinson | same |
| Product<br>Configuration | Holder –<br>Polypropylene<br>Adapter - Acrylonitrile<br>butadiene styrene<br>Sleeve — Not made<br>with natural<br>rubber<br>latex<br>Needle - Stainless<br>Steel | Holder – Polypropylene<br>Adapter - Polycarbonate<br>Sleeve – Not made<br>with natural<br>rubber latex<br>Needle - Stainless Steel | The only major<br>difference between the<br>subject device and the<br>predicate device is in<br>the material used for<br>the part, Adapter. The<br>subject device, Blood<br>Transfer Accessory (BTA)<br>uses the polymer,<br>Acrylonitrile butadiene<br>styrene (ABS) for the<br>Adapter part while the<br>predicate device uses<br>Polycarbonate (PC)<br>polymer.<br>Biocompatibility Testing<br>and Performance<br>Testing was conducted<br>to demonstrate SE. |
| Sterilization | Ethylene Oxide (EtO)<br>Gas sterilization | Ethylene Oxide (EtO) Gas<br>sterilization | same |
| Sterility | Meet the Sterility<br>Assurance Level of<br>10-6 as per the<br>requirement of ISO<br>11135:2014 | Not known | Non-Clinical Performance<br>Testing was conducted to<br>demonstrate SE. |
| Packaging | Sealed Blister<br>Packaging | Sealed Blister Packaging | Non-Clinical<br>Performance Testing<br>was conducted to<br>demonstrate SE. |
| Use | Single-Use only | Single-Use only | same |
| Bio compatibility | ISO 10993 series | ISO 10993 series | same |
| Shelf Life | 1 year | 3 years | Non-Clinical Performance<br>Testing was conducted to<br>demonstrate SE. |
{5}------------------------------------------------
# 7. TECHNOLOGICAL CHARACTERISTICS COMPARISON
The Blood Transfer Accessory (BTA)device has similarities with the predicate device in areas such as Intended Use, Product Configuration, and performance. The only difference between the subject device and the predicate device is in the use of material for the part, Adapter. The Blood Transfer Accessory (BTA) uses the polymer material, Acrylonitrile butadiene styrene (ABS) while the predicate device uses the polymer material, Polycarbonate (PC). Biocompatibility tests and performance tests based on international standards demonstrated the device performs as intended.
{6}------------------------------------------------
In addition the indications for use statement is different in that the subject device connects to male luer blood collection devices and the predicate device specifies the devices it may connect to. Performance testing demonstrated that this does not raise different questions of safety or effectiveness.
## 8. PERFORMANCE DATA/NON-CLINICAL TESTING
The Blood Transfer Accessory (BTA)was tested for validation and verification of functions based on risk analysis and the results passed predetermined acceptance criteria.
## Biocompatibility
Biocompatibility was evaluated per the FDA Guidance titled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" – Guidance for Industry and Food and Drug Administration Staff issued in June 2016 with the following endpoints:
Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Pyrogenicity Hemocompatibility (Coagulation, Platelets and Hemolysis)
## Performance Data
Performance testing was carried out on the Blood Transfer Accessory (BTA) using applicable international standards as reference and predicate device as benchmark for comparison.
- . Luer Lok Adapter
For international standard related test, the Luer Lok adapter is tested with reference to the ISO 80369-7:2016 and ISO 80369-20:2015.
- PerformanceTesting
As for the benchmarking tests related to the predicate device, blood collection performances such as the flow rate and quality of holder, holding strength of sleeve to adapter and needle to adapter are assessed and compared against that of the predicate device. The test methods employed are as summarized below:
- o Flow Rate The time taken to fill up a 3.0 ml blood collection Test tube with liquid of viscosity similar to blood is recorded.
- With the device held in place by a fixture, the o Leakage Testmaximum numbers of round of insertion and evacuation of blood collection tube the device can withstand before leakage start occurring at the device is recorded.
Page 4 of 5
{7}------------------------------------------------
- The device is held in place by a fixture and a force Evacuation o Force Test taken to insert or evacuate a blood collection tube into the holder of the device is recorded.
- Using fixture to hold the adapter in place, the o Holding Strength of needle is pull in a direction away from the adapter. Needle to The force taken to dislodge the needle from the adapter is recorded. Adapter Test -
- With the adapter held in place by a fixture, the o Holding Strength of sleeve is pulled in a direction away from the adapter and the force taken to dislodge the sleeve Sleeve to from the adapter is recorded. Adapter Test -
- o Deformation The device is held in place by a fixture and a 5 kg Test force is applied on the device for a period of 5 minutes. The permanent change in outer diameter of the holder of the device before and after compression by the applied force is measured and recorded. This test assesses the quality of the holder.
#### ● Sterilization
Sterilization is done in accordance to ISO 11135-1:2014.
The Sterility Assurance Level (SAL) is 10-6.
#### SUBSTANTIAL EQUIVALENCE CONCLUSION 9.
The result of non-clinical testing demonstrates that the Blood Transfer Accessory (BTA) is substantially equivalent to the predicate device Vacutainer® Brand Multiple Sample Luer Adapter and has been demonstrated to meet its predetermined specifications.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.