SAF-T CLOSED BLOOD COLLECTION SYSTEM WITH SAF-HOLDER DEVICE, WITH 21G, 23G, 25G SAF-T WING

K081229 · Smiths Medical Asd, Inc. · JKA · Jul 21, 2008 · Clinical Chemistry

Device Facts

Record IDK081229
Device NameSAF-T CLOSED BLOOD COLLECTION SYSTEM WITH SAF-HOLDER DEVICE, WITH 21G, 23G, 25G SAF-T WING
ApplicantSmiths Medical Asd, Inc.
Product CodeJKA · Clinical Chemistry
Decision DateJul 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

The Saf-T Closed Blood Collection System® device is attached to a peripheral IV catheter at the time of IV catheter placement for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes. The Saf-T Closed Blood Collection System® device is intended for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes.

Device Story

Venous blood collection device; facilitates direct blood draw into vacuum tubes or syringe-based transfer. Components include tubing, clamps, and vacuum tube holder. Used by clinicians during peripheral IV catheter placement or via direct access. Minimizes Luer manipulations to reduce sample contamination risk. Longer holder design improves vacuum tube alignment. Device provides a closed system for blood specimen collection; aids in clinical diagnostic workflows by simplifying blood sampling from existing IV lines or direct venipuncture.

Clinical Evidence

Bench testing only; no clinical data required.

Technological Characteristics

Vascular access via IV catheter or safety needle; includes tubing, clamps, and vacuum tube holder. Mechanical device; no energy source or software components.

Indications for Use

Indicated for patients requiring venous blood collection via peripheral IV catheter or direct draw; facilitates blood transfer to vacuum tubes; intended for prescription use.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SECTION 5, 510(k) Summary bringing technology to Smiths Medical ASD Inc. Critical Care Division 10 Bowman Drive Keene, NH 03431 0724, USA T: +1 603 352 3812 F: +1 603 355 8157 www.smiths-medical.com 'JUL 2 1 2008 ## Company Information: Smiths Medical ASD, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812, prompt 4, ext 2493 Contact: Brian D. Farias Regulatory Affairs Manager Summary Prepared: April 25, 2008 ### Product Name: Trade Name: Saf-T Closed Blood Collection System® Common Name: Venous Blood Collection Device Classification Name: 862.1675 Blood Specimen Collection Device ### Predicate Device(s): K.072783 (Smiths Medical ASD, Inc.) Saf-T Closed Blood Collection System® Device K922445 (Smiths Medical ASD, Inc.- formerly Concord/Portex) Venipuncture Needle Pro® Device {1}------------------------------------------------ #### Device Description: The Saf-T Closed Blood Collection System® device is a venous blood drawing device that is currently used as syringe draw and transfer device to fill vacuum tubes. This submission expands the indications to allow a direct blood draw into a vacuum tube. A syringe draw and transfer remains as an option - the Clinician can choose the method depending upon patient status. The use of this device involves minimal Luer manipulations, which minimizes the risk of sample contamination. This submission also increases the holder length to be equal to similar devices, the longer length will aid in vacuum tube alignment into the holder. ### REF Code 99200 # Device Name: Saf-T Closed Blood Collection System® with Saf-T Holder® Device and Male Luer Connector Indications for Use: The Saf-T Closed Blood Collection System® device is attached to a peripheral IV catheter at the time of IV catheter placement for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes. ### REF Code 992XX # Device Name: Saf-T Closed Blood Collection System® with Saf-T Holder® and Saf-T Wing® Device Indications for Use: The Saf-T Closed Blood Collection System® device is intended for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes. ### Technological Characteristics: The proposed and predicate devices have the same characteristics, i.e. vascular access via a newly placed IV catheter line or direct access via a safety needle; with tubing and clamps to control blood flow, and a vacuum tube holder for vacuum tube placement. {2}------------------------------------------------ # Non-Clinical Data: Bench testing confirms that the proposed device and the predicate device have similar performance specifications. ## Clinical Data: Not Required # Conclusion: The bench testing conducted demonstrates that the proposed device is safe and effective and is substantially equivalent to the predicate device. 12 Very truly yours, SMITHS MEDICAL ASD, INC. BA Brian D. Farias Regulatory Affairs Managcr {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Brian Farias Regulatory Affairs Manager Smiths Medical ASD, Incorporated 10 Bowman Drive Keene, New Hampshire 03431 JUL 2 1 2008 Re: K081229 Trade/Device Name: Saf-T Closed Blood Collection System® Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: II Product Code: JKA Dated: April 25, 2008 Received: April 30, 2008 Dear Mr. Farias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Farias Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # KV,81229 ### SECTION 4, Indications for Use Statement # Indications for Use 510(k) Number (if known): #### REF Code 99200 Device Name: Saf-T Closed Blood Collection System® with Saf-T Holder® Device and Male Luer Connector Indications for Use: The Saf-T Closed Blood Collection System® device is attached to a peripheral IV catheter at the time of IV catheter placement for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes. ### REF Code 992XX Device Name: Saf-T Closed Blood Collection System® with Saf-T Holder® and Saf-T Wing® Device ### Indications for Use: The Saf-T Closed Blood Collection System® device is intended for use as a direct blood draw device into a vacuum tube or to allow a syringe blood draw and transfer to fill vacuum tubes. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of Klaell for Chulin (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K081227 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...