ABC CALIBRATION VERIFIERS

K973791 · Sc Calibrators & Controls, LLC · JJY · Oct 14, 1997 · Clinical Chemistry

Device Facts

Record IDK973791
Device NameABC CALIBRATION VERIFIERS
ApplicantSc Calibrators & Controls, LLC
Product CodeJJY · Clinical Chemistry
Decision DateOct 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1
Attributes3rd-Party Reviewed

Indications for Use

ABC Calibration Verifiers is intended to be used to verify the calibration, linear operating range and reportable range (linearity) of methods used to determine the concentration of therapeutic drugs.

Device Story

ABC Calibration Verifiers is a five-level, human serum-based in vitro diagnostic set containing 12 analytes (antibiotics and cardiac drugs). Used in clinical laboratories to verify calibration and linearity of TDM methods. Levels 1-5 span the analytical range; level 1 is zero, level 5 is near the upper limit, and levels 2-4 are targeted at clinical interest concentrations. The device is assayed like a patient sample on automated or manual TDM systems. Healthcare providers compare the observed results against expected values to confirm instrument performance and reportable range. This verification ensures accuracy of therapeutic drug concentration measurements, supporting clinical decision-making for drug dosing.

Clinical Evidence

Bench testing only. Performance validated on two automated TDM systems (Abbott TDx Assay System and Behring EMIT system) and HPLC methods for selected analytes. Testing confirmed the device's ability to verify calibration and linear operating range across the intended analytical range.

Technological Characteristics

Five-level set; human serum matrix; non-azide preservative. Analytes: 12 therapeutic drugs. Calibration: Gravimetric and traceable. Stability: 60 days at 2-8°C (Vancomycin 30 days). Compatible with automated and manual immunoassay methods.

Indications for Use

Indicated for use in clinical laboratories to verify calibration, linear operating range, and reportable range of therapeutic drug monitoring (TDM) methods and instruments. Required for modified moderately complex, highly complex, and in-house methods; recommended for use following reagent lot changes, major instrument maintenance, or when control values exhibit trends/shifts.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SC Calibrators & Controls LLC, 1701 Berkeley Street, Santa Monica CA 90404 Tel: (310) 828-7423 Fax: (310) 453-1666 Image /page/0/Picture/2 description: The image shows a logo with three overlapping letters 'C' stacked vertically within a black rectangle. A partial circle is positioned to the left of the letters, partially obscuring the top 'C'. The letters and the partial circle are white, contrasting with the black background of the rectangle. ## 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: Roy F. Schall, Jr., Ph.D. Contact Name: August 31, 1997 Date: ABCtm Calibration Verifiers Product Name: Common Names:Calibration Verifiers, Linearity Verifiers Document TDM | Linearity Set, Casco Standards, Inc. Predicate Devices: LiniCAL Protein I Calibration Verifiers, International Enzymes, Inc. ABC Calibration Verifiers is an in vitro diagnostic Description of the Device: medical device intended for use with automated and manual methods monitoring selected therapeutic drugs (TDM). ABC Calibration Verifiers is a five level set of verifiers, with each level containing 12 analytes, is made with human serum and is used to confirm the proper calibration, linear operating range and reportable range (linearity) of TDM methods. The "ABC" designation derives from Antibiotic and Cardiac drugs. ABC Calibration Verifiers contains concentrations of analytes extending over a wide analytical range. Level 1 is a zero level, and level 5 has concentrations near the upper limit of instruments. Levels 2, 3, and 4 are targeted to be near analyte concentrations of clinical interest. When assayed like patient samples, the verifiers assist in determination of calibration and linear operating range (linearity) of methods for the analytes included. (A second product, ABC Linearity Verifiers, is identical in every respect to this product except that the concentrations of levels 1.2.3, and 4 are related by linear dilution. Both products will perform linear range and calibration verification, and their composition, stabilities and performance are identical except for concentration of middle levels.) Intended Use: ABC Calibration Verifiers is intended to be used to verify the calibration, linear operating range and reportable range (linearity) of methods used to determine the concentration of therapeutic drugs. {1}------------------------------------------------ # Technical Characteristics Compared to Predicate Devices: ## Comparison of Features of ABC Calibration Verifiers and the two predicate devices: DOCUMENT TDM I Linearlty Test Set and LiniCAL Protein | Calibration Verifiers. | Attribute | ABC Calibration Verifiers | DOCUMENT TDM I | LiniCAL Protein I | |--------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------| | No Analytes/bottle | 12 | 19 | 6 | | No of Levels/set | 5 | 8 | 5 | | Vol per level (mL) | 5 | 5 | 1 | | Type of Analytes | Therapeutic Drugs | Therapeutic Drugs | Serum Proteins | | Method(s) | Automated, Manual<br>Immunoassay, other | Automated<br>Immunoassay | Beckman Array<br>Nephelometer | | Base Matrix | Human Serum | Human Serum | Human Serum | | Preservative | Non-Azide | Azide | Azide | | Calibration | Gravimetric + Some<br>Traceable | Assayed<br>(Two Methods) | Assayed + Traceable | | Use. | Verification of Calibration,<br>Linear Operating Range (OR),<br>Reportable Range<br>All Methods | Verif. of Calib,<br>Linear OR,<br>Reportable<br>Range for<br>Two Methods | Verif. of Calib,<br>Linear OR<br>Beckman Array | | Stability | 60 days at 2-8° C;<br>To Expir. Date if frozen<br>(Vancomycin 30 days 2-8°C) | To Expiration<br>at 2-8° C | To Expiration Date<br>unopened at 2-8°C<br>14 days opened | Technical Characteristics by Assessment of Performance: The performance of ABC Verifiers has been tested on two commonly used automated therapeutic drug monitoring systems (Abbott TDx Assay System and Behring (formerly Syva) EMIT system reagents) and by HPLC methods for selected analytes to validate their use. All methods tested demonstrated the desired functionality of the product. Conclusions: Based upon the purpose of the device, the descriptions and labeling of the predicate devices, and upon the safety and efficacy using multiple instruments and methods, and stability data generated for the ABC Calibration Verifiers, the {2}------------------------------------------------ product is substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized image of a bird with three human profiles incorporated into its design, suggesting a connection between health, humanity, and service. OCT 1 4 1997 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SC Calibrators & Controls LLC Ms. Carole Stamp c/o TUV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891 Re : K973791 Trade Name: ABC™ Calibration Verifiers™ Regulatory Class: I Product Code: JJY Dated: September 19, 1997 Received: October 6, 1997 Dear Ms. Stamp: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ರ್ಗ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Stamp Under the Clinical Laboracory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655 . This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 12.0 The Office of Device Evaluation (ODE) has developed the attached optional form to assist them with instituting the requirement for all original 510(k)s received by ODE on or after 1/2/96. The requirement is for all 510(k) submissions to have clearly defined "Indications for Use". These indications will be attached by ODE to any substantial equivalence (SE) letter to define what the device is cleared for. No 510(k) submitted on or after 1/2/96 will be cleared for marketing by ODE without the inclusion of the indications for use information, which will be attached to an SE letter. ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: ABC ™ Calibration Verifiers Indications For Use: ## Indications For Use ABC Calibration Verifiers should be used any time it is necessary to confirm the proper calibration and linear operating range of TDM methods and instruments. The Clinical Laboratory Improveent Act and its requlations (CLIA) require verification of linearity and calibration twice a year for wodified moderately complex methods, highly complex methods and in-house methods. Verification should be performed more often if any of the following occurs: introduction of procedures for which control values change with a new readent lot, major preventative maintenance or replacement of critical parts on the instrument, control values exhibit an unusual trend or shift or are outside the acceptable limits of the laboratory, and if the laboratory's schedule of verification is more frequent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | <div> <img alt="Signature" src="signature.png"/> </div> (Division Sign-Off) | |---------------|-----------------------------------------------------------------------------| | | Division of Clinical Laboratory Devices | | 510(k) Number | K973791 | | Description Use | OR | Over-The-Counter Use | |-----------------|----|----------------------| |-----------------|----|----------------------| (Per 21 CFR 801.109) (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...