RANDOX IMMUNOASSAY SPECIALITY CONTROL (II), LEVEL 1, LEVEL 2 AND LEVEL 3
Applicant
Randox Laboratories, Ltd.
Product Code
JJY · Clinical Chemistry
Decision Date
Nov 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The Randox Immunoassay Speciality Control (II), Level 1, Level 2, Level 3 are intended for in vitro diagnostic use as assayed quality control material for Calcitonin. Gastrin, and Procalcitonin to monitor the precision of the laboratory testing systems listed in the package insert. This device is for prescription use only.
Device Story
Randox Immunoassay Speciality Control (II) consists of three levels of lyophilized human serum-based control material; contains added Calcitonin, Gastrin, and Procalcitonin. Used in clinical laboratories to monitor precision of testing systems for these specific analytes. Requires reconstitution with distilled water before use. Laboratory technicians analyze the control alongside patient samples; results are compared against assigned values to verify system performance. Provides quality assurance for diagnostic testing; helps ensure accuracy and reliability of clinical results for patients undergoing testing for these hormones/markers.
Clinical Evidence
No clinical data. Bench testing only. Stability protocols (shelf-life and open-vial) were reviewed and found acceptable. Value assignment was determined using multiple analyzer platforms.
Technological Characteristics
Lyophilized human serum matrix; contains synthetic Calcitonin, human Gastrin, and E. coli-derived Procalcitonin. Supplied in 1ml vials. Storage at 2-8°C. No electronic components or software.
Indications for Use
Indicated for use as an assayed quality control material to monitor the precision of laboratory testing systems for Calcitonin, Gastrin, and Procalcitonin. Intended for prescription use in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Biorad Lyphochek® Immunoassay Plus Control Levels 1, 2 and 3 (k981532)
Related Devices
K130337 — RANDOX SPECIALITY CONTROL (I) LEVELS 1, 2 AND 3 · Randox Laboratories Limited · Jun 19, 2013
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k121603
B. Purpose for Submission:
New device
C. Measurand:
Multi-analyte control materials (calcitonin, gastrin, procalcitonin)
D. Type of Test:
Not applicable
E. Applicant:
Randox Laboratories Limited
F. Proprietary and Established Names:
Randox Immunoassay Speciality Control (II) Levels 1, 2 and 3
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1660, Quality control material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJY
4. Panel:
Clinical Chemistry
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The Randox Immunoassay Speciality Control (II), Level 1, Level 2, Level 3 are intended for in vitro diagnostic use as assayed quality control material for Calcitonin, Gastrin, and Procalcitonin to monitor the precision of the laboratory testing systems listed in the package insert. This in vitro diagnostic device is intended for prescription use only.
3. Special conditions for use statement(s):
For in vitro diagnostic use
For prescription use only
4. Special instrument requirements:
Values are listed in the package insert for several analyzers
I. Device Description:
Randox Immunoassay Speciality Control (II) is manufactured at three levels, Level 1, Level 2 and Level 3. Each control is prepared from human serum with added constituents of human origin, chemicals, stabilizers and preservatives. They are supplied in lyophilized form in 5x1ml vials and require reconstitution with 1ml of distilled water.
Human source material from which this product has been derived and has been tested at the donor level for the Human Immunodeficiency Virus (HIV1 & HIV2) antibody, Hepatitis B surface antigen (HbsAg) and the Hepatitis C virus (HCV) antibody and were found to be non-reactive based on FDA approved methods.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Biorad Lyphochek® Immunoassay Plus Control Levels 1, 2 and 3
2. Predicate 510(k) number(s):
k981532
{2}
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Randox Speciality Immunoassay Control (II) Levels 1, 2 and 3 | BIO-RAD LyphocheckR Immunoassay Plus Control Levels 1, 2 & 3 k981532 |
| Intended Use/Indications for use | Same | For use as an assayed quality control serum to monitor the precision of laboratory testing procedures listed in the package insert |
| Format | Lyophilized | Same |
| Matrix | Human Serum | Same |
| Storage (Unopened) | +2 to +8 °C Until expiration date | Same |
| Open Vial Claim | Store refrigerated +2 to +8°C. In reconstituted serum Procalcitonin is stable for 1 day, Gastrin and Calcitonin are stable for 8 hours at +2 to +8°C if kept capped in original container and free from contamination. The control is stable if frozen once for 28 days at -20°C. | Calcitonin and Gastrin should be assayed immediately after reconstitution. After reconstituting and freezing the control, all analytes are stable for 20 days when stored tightly capped at -10°C to -20°C with the exception of Calcitonin as there is no frozen stability claim supplied. |
| Shipping Temperature | +2 to +8°C | +2 to +8°C |
| Analytes | Procalcitonin, Gastrin and Calcitonin. | 92 analytes, including Gastrin and Calcitonin |
| Volume | 1 mL | 5 mL |
{3}
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
The analytes contained in the Randox Speciality Immunoassay Control (II) Levels 1, 2 and 3 were obtained from commercially available sources (see table below).
| ANALYTE | SUPPLIER | PRODUCT NUMBER | ORIGIN | SOURCE |
| --- | --- | --- | --- | --- |
| Calcitonin | Sigma | T-3525 | Synthetic Analytical Grade Chemical | Commercial source, added volumetrically |
| Gastrin | Sigma | G-9020 | Human Gastrin | Commercial source, added volumetrically |
| Procalcitonin | Randox | RCP9522 | Extracted and purified from E.coli | Commercial source, added volumetrically |
Stability
Shelf-life and open-vial stability protocols and acceptance criteria were reviewed and found acceptable. Closed vial (shelf-life) stability at the recommended storage temperature of (2 to 8 °C) was demonstrated based on
{4}
accelerated stability. Closed vials are stable until the date printed on individual vials. Real-time closed vial stability studies are on-going. In reconstituted serum Procalcitonin is stable for 1 day, Gastrin and Calcitonin are stable for 8 hours at +2 to +8°C if kept capped in original container and free from contamination. The control is stable if frozen once for 28 days at -20°C. All storage recommendations are provided in the labeling.
## Value Assignment
Value assignment was determined for each analyte contained in the Randox Speciality Immunoassay Control (II) Levels 1, 2 and 3 using multiple analyzer platforms.
Value assignment data were collated and an appropriate target value was assigned to each analyte based on the average of the observed values. Ranges were then assigned as +/- 25% of the target. The labeling states that obtained values should fall within the specified range provided on lot-specific value sheets and that laboratories should establish appropriate acceptance criteria when using this product for its intended use.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
{5}
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Target and range values for representative analyzers are found in the package insert for each specific lot.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.