K071433 · Specialty Medical Supplies · JJY · Jun 28, 2007 · Clinical Chemistry
Device Facts
Record ID
K071433
Device Name
SMS GLUCOSE CONTROL, MODEL 2120059
Applicant
Specialty Medical Supplies
Product Code
JJY · Clinical Chemistry
Decision Date
Jun 28, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The SMS Glucose Control is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of TheraSense FreeStyle Blood Glucose Monitor.
Device Story
SMS Glucose Control is an aqueous, viscosity-adjusted liquid solution containing a known quantity of glucose. Packaged in plastic dropper-tipped bottles, it includes a red dye to assist users in confirming application to test strips. Used by healthcare professionals and patients with diabetes in clinical or home settings to verify the performance of the TheraSense FreeStyle Blood Glucose Monitoring System. The user applies the control solution to a test strip; the meter reads the result, which is then compared against the expected range provided by the manufacturer. This process ensures the meter and test strips are functioning correctly, providing confidence in subsequent patient blood glucose readings.
Clinical Evidence
No clinical data. Bench testing only. Stability determined via real-time studies (24-month shelf-life, 90-day open-vial stability). Traceability established via in-house glucose preparation; values assigned using three lots of test strips with 10 replicates per lot.
Technological Characteristics
Buffered aqueous solution containing D-glucose, viscosity modifier, preservatives, and red dye. Non-sterile, non-hazardous, no human/animal-derived materials. Plastic dropper-tipped bottle. Standard: CLSI EP5-A for precision evaluation.
Indications for Use
Indicated for healthcare professionals and home users with diabetes mellitus to verify the performance of the TheraSense FreeStyle Blood Glucose Monitoring System.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
TheraSense FreeStyle Control (k031260)
Liberty Glucose Control (k060481)
Related Devices
K070649 — SMS GLUCOSE CONTROL, MODEL 2120061 · Spectrum Medical Supply, Inc. · Apr 12, 2007
K070506 — SMS MULTI GLUCOSE CONTROL, MODEL 2120049 · Medical Specialties Supply Co., Inc. · Apr 18, 2007
K103450 — FDI GLUCOSE CONTROL SOLUTION FOR FREESTYLE · Fujirebio Diagnostics,Inc. · Jan 25, 2011
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k071433
B. Purpose for Submission:
New device
C. Measurand:
Quality control material for glucose
D. Type of Test:
Not applicable
E. Applicant:
Specialty Medical Supplies (SMS)
F. Proprietary and Established Names:
SMS Glucose Control
G. Regulatory Information:
1. Regulation section:
21CFR Section 862.1660 – Quality control material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJX, single (specified) analyte controls (assayed and unassayed)
4. Panel:
Clinical Chemistry (75)
{1}
H. Intended Use:
1. Intended use:
See intended use section below.
2. Indication(s) for use:
For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the TheraSense FreeStyle Blood Glucose System.
3. Special conditions for use statement(s):
Over-the-Counter use
4. Special instrument requirements:
TheraSense FreeStyle Blood Glucose Meter.
I. Device Description:
The product is a one-level, aqueous liquid glucose control solution for use with the TheraSense FreeStyle Blood Glucose System. The glucose control consists of a buffered aqueous solution of D-glucose, a viscosity modifier, preservatives, red dye, and other non-reactive ingredients. The device is non-sterile, non-hazardous and contains no human or animal derived materials.
The product is packaged in a plastic dropper tipped bottle for easy application of the control solution to the test strip and contains sufficient volume (3.6 mL) to run 75 tests. A red coloration is included to aid the user to visually confirm application of the control.
J. Substantial Equivalence Information:
1. Predicate device name(s):
TheraSense FreeStyle Control
Liberty Glucose Control
2. Predicate 510(k) number(s):
k031260
k060481
{2}
3
3. Comparison with predicate:
The subject device and both predicate devices contain D-Glucose and no human or animal derived materials.
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicates |
| Target Users | Professional and home use | same |
| Matrix | Buffered Aqueous solution | same |
| Packaging | Plastic bottle with dropper tip | same |
| Analytes | Glucose | same |
| Number of levels | 1 | same |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A, Evaluation of the Precision Performance of Clinical Chemistry Devices.
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The D-Glucose used in this control is traceable to an in-house glucose preparation. Values are assigned by repeat analysis using three different lots of test strips, 10 replicates per strip lot. For each lot, the mean, standard deviation, and % CV were calculated. A representative range of values for this product is 85 – 129 mg/dL.
Stability characteristics of the SMS Glucose Control were determined using real-time studies. The unopened shelf-life is 24 months and the open vial stability is 90 days at the recommended storage of 36 °F to 86 °F.
d. Detection limit:
{3}
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{4}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.