← Product Code [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY) · K051687

# ELECSYS PRECICONTROL UNIVERSAL (K051687)

_Roche Diagnostics Corp. · JJY · Jul 25, 2005 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K051687

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY.md)
- **Decision Date:** Jul 25, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Elecsys® PreciControl Universal is used for quality control of the Elecsys® immunoassays on the Elecsys® immunoassay systems. For Prescription Use Only

## Device Story

Elecsys® PreciControl Universal is a lyophilized quality control product consisting of human serum matrix spiked with various antigens (AFP, Cortisol, DHEAS, Estradiol, Ferritin, Folate, FSH, L-Thyroxin, hCG, IgE, Insulin, LH, Progesterone, Prolactin, Testosterone, SHBG, T3, TSH, Vitamin B12). It is used by laboratory professionals to monitor the performance and accuracy of Elecsys® immunoassay systems. The product requires reconstitution with distilled water before use. By comparing the measured values of the control against established target ranges, clinicians verify the reliability of patient sample testing. This process ensures the analytical validity of immunoassay results, supporting accurate clinical decision-making and patient diagnosis.

## Clinical Evidence

No clinical data. The device was evaluated through bench testing for value assignment and stability characteristics.

## Technological Characteristics

Lyophilized human serum matrix with added synthetic and human-derived antigens. Requires reconstitution with 3.0 mL distilled water. Stability varies by analyte and storage condition (refrigerated, frozen, or on-analyzer).

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Elecsys® PreciControl MultiAnalyte ([K033937](/device/K033937.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k051687

B. Purpose for Submission:
Marketing of a laboratory control

C. Measurand:
Control material for AFP, Cortisol, DHEAS, Estradiol, Ferritin, Folate, FSH, FT3, FT4, hCG, IgE, Insulin, LH, Progesterone, Prolactin, SHBG, Testosterone, T3, T4, TSH, T-Uptake, and Vitamin B12

D. Type of Test:
The product is used as a quality control to monitor the precision of laboratory testing procedures.

E. Applicant:
Roche Diagnostics

F. Proprietary and Established Names:
Elecsys® PreciControl Universal

G. Regulatory Information:
1. Regulation section:
21CFR 862.1660 Quality control material (assayed and unassayed).
2. Classification:
Class I
3. Product code:
JJY

{1}

4. Panel:
(75) Chemistry

H. Intended Use:

1. Intended use:
See indications for use below.

2. Indications for use:
Elecsys® PreciControl Universal is used for quality control of the Elecsys® immunoassays on the Elecsys® immunoassay systems.

3. Special conditions for use statement(s):
For Prescription Use Only

4. Special instrument requirements:
Elecsys Immunoassay systems

I. Device Description:
The Elecsys® PreciControl Universal is a lyophilized product consisting of added antigens in human serum matrix. During manufacture, the analytes are spiked into the matrix at the desired concentrations.

J. Substantial Equivalence Information:

1. Predicate device name:
Elecsys PreciControl MultiAnalyte

2. Predicate 510(k) number(s):
k033937

3. Comparison with predicate:

2

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Characteristics | Elecsys PreciControl Universal | Predicate device Elecsys PreciControl MultiAnalyte  |
|  Intended Use | Elecsys® PreciControl Universal is used for quality control of Elecsys® immunoassays on the Elecsys® immunoassay systems. | Elecsys® PreciControl MultiAnalyte is used for quality control of the Elecsys® C-Peptide and Elecsys® Insulin immunoassays on the Elecsys® immunoassay systems.  |
|  Levels | Two | Same  |
|  Format | Lyophilized | Same  |
|  Handling | Reconstitute with exactly 3.0 mL distilled water and allow to stand closed for 30 minutes to reconstitute, and then mix gently. | Reconstitute with exactly 2.0 mL distilled water and allow to stand closed for 15 minutes to reconstitute, and then mix gently.  |
|  Stability | **Unopened:**
Store at 2 - 8°C until expiration date
**Reconstituted (except for Insulin):**
• 20 - 25°C: up to 8 hours
• on the analyzers at 20 - 25°C: up to 8 hours
• at 2-8°C: 3 days
• at -20°C: 1 month (freeze only once)
**Reconstituted Insulin:**
• on the analyzers at 20 - 25°C: up to 5 hours
• 20 - 25°C: up to 5 hours
• at -20°C: 1 month (freeze only once) | **Unopened:**
Store at 2 - 8°C until expiration date
**Reconstituted:**
• on the analyzers at 20 - 25°C: up to 3 hours
• at -20°C: 1 month (freeze only once)
• after thawing: use only once  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Characteristic | Elecsys PreciControl Universal | Predicate device Elecsys PreciControl MultiAnalyte  |
|  Matrix | Human serum with added antigens | Equine serum with added C-Peptide and insulin  |

K. Standard/Guidance Document Referenced (if applicable):

None were referenced in the submission.

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Roche Diagnostics maintains a set of master calibrators for each assay that have values assigned that are traceable to the various method "gold standards." Elecsys® PreciControl products are assayed and compared to these reference preparations and target values and ranges are assigned.

Values are assigned using a total of 12 analyzers except for hCG; hCG values were assigned using 8 analyzers. Two independent series of analyses are performed on each instrument for each analyte. Each sample is tested in duplicate. The target value is calculated as the median of the determined values.

Four studies were performed on reconstituted material to verify the stability claims.

Elecsys PreciControl Universal material and reference material are evaluated in triplicate. The Elecsys PreciControl Universal material is stressed. The reference material is an unstressed set of Elecsys PreciControl Universal. One

{4}

bottle of Elecsys PreciControl Universal material was used per each stability testing. Each result is reported as the average of three samples. The acceptance criterion is recovery of 90%-110% of the concentration based on unstressed material. The results are presented below.

|  Stability | Reconstituted  |   |
| --- | --- | --- |
|  Elecsys PreciControl Universal | Except for Insulin
°on the analyzers at 20 - 25°C: up to 5 hours
°20 - 25°C: up to 8 hours
°at 2-8°C: 3 days
°at -20°C: 1 month (freeze only once) | Insulin
°on the analyzers at 20-25°C: up to 5 hours
°20°-25°C: up to 5 hours
°-20°C: 1 month (freeze only once)  |

Unopened vial stability: 2-8°C until the expiration date.

d. Detection limit:
Not Applicable

e. Analytical specificity:
Not Applicable

f. Assay cut-off:
Not Applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable

3. Clinical studies:

{5}

a. Clinical Sensitivity:
Not Applicable

b. Clinical specificity:
Not Applicable

c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable

4. Clinical cut-off:
Not Applicable

5. Expected values/Reference range:
Not Applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K051687](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K051687)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
