← Product Code [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY) · K032074

# LINICAL ESOTERICS CALIBRATION VERIFIERS LEVELS A-E FOR BECKMAN COULTER SYNCHRON ANALYZERS (K032074)

_Cliniqa Corporation · JJY · Oct 21, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K032074

## Device Facts

- **Applicant:** Cliniqa Corporation
- **Product Code:** [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY.md)
- **Decision Date:** Oct 21, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

CLINIQA LiniCAL™ Esoterics Calibration Verifiers are intended for use in the clinical laboratory to verify calibration and/or assess linearity of the Beckman Coulter Synchron® Analyzers. Five assayed levels of AMM (ammonia), ETOH (alcohol), MA (microalbumin), M-TP (microprotein) and SALY (salicylate) are provided to allow monitoring of the reportable range.

## Device Story

LiniCAL™ Esoterics Calibration Verifiers are quality control materials used in clinical laboratories to verify calibration and assess linearity of Beckman Coulter Synchron® Analyzers. The device consists of five assayed levels (A-E) of specific analytes: ammonia, alcohol, microalbumin, microprotein, and salicylate. Laboratory personnel use these materials to monitor the reportable range of the analyzer. By comparing the measured values of the verifiers against the provided assayed values, clinicians can confirm the accuracy and linearity of the instrument's calibration, ensuring reliable patient test results.

## Clinical Evidence

No clinical data. Bench testing only. Stability established via real-time studies comparing vials stored at 2-8°C against -70°C controls at 3-month intervals, ensuring analytes remain within 10% of original values.

## Technological Characteristics

Liquid, assayed quality control material containing ammonia, alcohol, microalbumin, microprotein, and salicylate. Provided in five levels (A-E) to cover the reportable range of Beckman Coulter Synchron® Analyzers.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k032074

B. Analyte:
Ammonia, Alcohol (ethanol), albumin (microalbumin), protein (microprotein), and Salicylate

C. Type of Test:
Calibration Verification Material

D. Applicant:
Cliniqa Corporation

E. Proprietary and Established Names:
LiniCAL Esoterics Calibration Verifiers Levels A-E for Beckman Coulter Synchron® Analyzers

F. Regulatory Information:
1. Regulation section:
862.1660 for Quality Control Material (Assayed and Unassayed)
2. Classification:
I
3. Product Code:
JJY
4. Panel:
75

G. Intended Use:
1. Indication(s) for use:
CLINIQA LiniCAL™ Esoterics Calibration Verifiers are intended for use in the clinical laboratory to verify calibration and/or assess linearity of the Beckman Coulter Synchron® Analyzers. Five assayed levels of AMM (ammonia), ETOH (alcohol), MA (microalbumin), M-TP (microprotein) and SALY (salicylate) are provided to allow monitoring of the reportable range.
2. Special condition for use statement(s):
Not applicable.
3. Special instrument Requirements:
Beckman Coulter Synchron® Analyzers

H. Device Description:
CLINIQA LiniCAL™ Esoterics Calibration Verifiers are human serum protein based, containing assayed constituents of chemically defined origin, including Immunoglobulin Fraction (IF), human albumin, ACS Reagent Grade of alcohol, ammonium chloride, and the sodium salt, Salicylic acid. Preservatives, stabilizers, and sodium azide have been added to maintain product integrity. They are manufactured without glycerol and glycol. The product is provided in liquid form and is ready to use.

{1}

Page 2 of 3

Constituent concentrations in Level A are at the low end to allow assessment of the lower limit of the reportable range. Constituent concentrations in Level E are at the high end and are designed to challenge the upper limit of the reportable range. Due to variation of analytical methods, Level E may exceed the limit of linearity for some test systems. Levels B, C, and D provide intermediate constituent concentrations over the reportable range.

I. Substantial Equivalence Information:

1. Predicate device name(s):
LiniCAL Protein 1 Calibration Verifiers Levels A-E for Beckman Coulter Immage

2. Predicate K number(s):
K013332

3. Comparison with predicate:
Both are serum based products into which protein has been added, and they are manufactured using the same processes. They differ in terms of the constituents and their target concentrations.

J. Standard/Guidance Document Referenced (if applicable):
The sponsor did not reference any standards.

K. Test Principle:
Not applicable.

L. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:
Not applicable.

b. Linearity/assay reportable range:
Not applicable.

c. Traceability (controls, calibrators, or method):
The sponsor has not provided any information regarding the traceability of the values assigned to the product. Values are assigned by the following process:

Assays used to establish the values will be run by at least three laboratory groups, and will incorporate at least one independent control. A statistical method will be used to evaluate data for possible Outliers. After Outlier analysis, the data will be averaged to obtain a representative expected value for each constituent.

Assignment of values shall be performed using the Beckman Coulter Synchron® Analyzer reagents and calibrators, and independent controls available at the time of assay.

Stability: Real time stability studies of the first three lots of manufactured product will take place to establish expiration dating. Stability data is recorded at 3 month intervals over the life of the product to at least 110% of the expiration date. Vials of the

{2}

Page 3 of 3

calibration verifiers are stored at 2-8 Degrees C. and are compared at each time point to vials stored at -70 degrees C. Each analyte in the product is tested at all five levels and should remain within 10% of the original value.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable.

f. Assay cut-off:
Not applicable.

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a and b are not applicable):

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:

Representative Assigned Values From One Lot of Product

|  Analyte | Method | Units | Level A
Lot 123456 | Level B
Lot 123457 | Level C
Lot 123458 | Level D
Lot 123459 | Level E
Lot 123460  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  AMM | SYNCHRON | μmol/L | 21.5 | 241 | 451 | 683 | 891  |
|  ETOH | SYNCHRON | mg/dL | 12.0 | 147 | 269 | 413 | 531  |
|  MA | SYNCHRON | mg/dL | 1.05 | 5.65 | 12.7 | 19.8 | 25.9  |
|  M-TP | SYNCHRON | mg/dL | 8.0 | 49.5 | 68.5 | 101 | 123  |
|  SALY | SYNCHRON | mg/dL | 6.05 | 25.5 | 46.1 | 68.2 | 86.1  |

M. Conclusion:
Substantially equivalent to predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K032074](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K032074)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
