VALIDATE CHEM 3 CALIBRATION VERIFICATION TEST SET, MODEL 10003
K012119 · Maine Standards Co. · JJY · Aug 20, 2001 · Clinical Chemistry
Device Facts
Record ID
K012119
Device Name
VALIDATE CHEM 3 CALIBRATION VERIFICATION TEST SET, MODEL 10003
Applicant
Maine Standards Co.
Product Code
JJY · Clinical Chemistry
Decision Date
Aug 20, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
VALIDATE Chem 3 Calibration Verification Test Set is intended for in vitro diagnostic use in quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the following analytes: total protein, albumin, cholesterol and iron.
Device Story
VALIDATE Chem 3 Calibration Verification Test Set is a liquid, human serum-based, multi-level (6 levels including zero) test set. Used in clinical laboratories by trained professionals to verify calibration, validate reportable ranges, and determine linearity of automated, semi-automated, or manual chemistry systems. The device provides a known reference to compare against the system's actual performance for analytes: total protein, albumin, cholesterol, and iron. Healthcare providers use the resulting correlation data to ensure the accuracy and reliability of patient test results within the chemistry system's reportable range, thereby supporting clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated using pre-production lots on a Roche Diagnostics Hitachi 911 instrument. Linear regression analysis performed in triplicate for each analyte (TP, ALB, CHOL, FE) against predicate devices. Results showed high correlation coefficients (r = 0.99920 to 0.99998) and comparable regression equations, confirming functional equivalence.
Technological Characteristics
Liquid, human serum-based calibration verification test set. Contains 6 levels (including zero) in 5 mL bottles. Analytes: total protein, albumin, cholesterol, iron. Storage: 2-8°C. Ready-to-use format.
Indications for Use
Indicated for use by trained laboratory professionals to verify calibration, validate reportable ranges, and determine linearity of chemistry systems for total protein, albumin, cholesterol, and iron.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
DOCUMENT Total Protein/Albumin/Cholesterol CAL-VER (K883861)
{0}------------------------------------------------
# MAINE STANDARDS COMPANY, LLC
# AUG 2 0 2001
#### 510(k) SUMMARY
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: KOI@ II 9
| Submitter: | Maine Standards Company |
|------------|---------------------------------------|
| Address: | 765 Roosevelt Trail Windham, ME 04062 |
| Telephone: | 207-892-1300 |
| Fax: | 207-892-2266 |
| Contact: | Christine Beach, Mgr. RA/QA |
Summary prepared on: June 29, 2001
| Proprietary Name: | VALIDATE Chem 3 Calibration Verification Test Set |
|----------------------|---------------------------------------------------|
| Common Name: | Calibration Verification |
| Classification Name: | Calibrator, Multi-Analyte |
#### Predicate Devices:
- DOCUMENT Total Protein/Albumin/Cholesterol CAL-VER, K883861, 1. manufactured by CASCO NERL Diagnostics.
- 2. DOCUMENT Iron/Magnesium/Triglyceride CAL-VER, K893142, manufactured by CASCO NERL Diagnostics.
Device description: VALIDATE Chem 3 Calibration Verification Test Set is a liquid, human serum based calibration verification test set containing multiple levels used establish the relationship between theoretical operation and actual performance of each of the included analytes Each set contains one bottle each of six (6) ievels, including zero. Each bottle contains 5 milliliters.
Intended use: VALIDATE Chem 3 Calibration Test Set is intended for in vitro diagnostic use in quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the following analytes: total protein, albumin, cholesterol and iron.
> Office location · 765 Roosevelt Trail · Windham, Maine 04062 Mailing address · P.O. Box 8360 · Portland, ME 04104-8360 Tel. 207-892•1300 • Fax 207•892•2266 • Toll Free 800•377•9684
{1}------------------------------------------------
# Comparison of VALIDATE Chem 3 Calibration Verification Test Set to the predicate devices:
Table 1 compares characteristics of the VALIDATE Chem 3 Calibration Verification Test Set with those of the DOCUMENT Iron / Magnesium / Triglyceride CAL�VER, DOCUMENT Total Protein / Albumin / Cholesterol CAL•VER.
| | VALIDATE CHEM 3<br>Calibration<br>Verification Test Set | DOCUMENT Total<br>Protein / Albumin /<br>Cholesterol CAL•VER | DOCUMENT<br>Iron, Magnesium,<br>Triglyceride<br>CAL•VER |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Catalog # | 10003 | M-102 | M-103 |
| Intended Use | For in vitro diagnostic<br>use in quantitatively<br>verifying calibration,<br>validating reportable<br>ranges, and<br>determining linearity in<br>automated, semi-<br>automated and manual<br>chemistry systems. | For in vitro diagnostic<br>use in the quantitative<br>determination of<br>linearity in manual,<br>automated and semi-<br>automated chemistry<br>systems. | For in vitro diagnostic<br>use in the quantitative<br>determination of<br>linearity in manual,<br>automated and semi-<br>automated chemistry<br>systems. |
| Analytes | TP, ALB, CHOL, FE | TP, ALB, CHOL | FE |
| Matrix | protein | protein | aqueous |
| Number of<br>Levels | 6 including Zero | 5 | 5 |
| Preparation | Liquid,<br>ready to use | Liquid,<br>ready to use | Liquid,<br>ready to use |
| Packaging | 5.0 mL each level | 8.0 mL each level | 10.0 mL each level |
| Stability | Until Expiration | Until Expiration | Until Expiration |
| Storage | 2-8°C | 2-8°C | 2-8°C |
#### Comparison of Products TABLE 1.
The performance of VALIDATE Chem 3 Calibration Verification Test Set solutions on the Roche Diagnostics Hitachi 911 instrument system as compared to DOCUMENT Iron / Magnesium / Triglyceride CAL . VER, DOCUMENT Total Protein / Albumin / Cholesterol CAL•VER has been shown to be substantially equivalent using pre-production lots of VALIDATE Chem 3 Calibration Verification Test Set. The results of correlation comparisons between the VALIDATE Chem 3 Calibration Verification Test Set and the predicate devices are presented in Table 2.
{2}------------------------------------------------
la119
TABLE 2. Linear Regression Statistical Comparison of VALIDATE Chem 3 Calibration Verification Test Set to the predicate devices.
| | VALIDATE<br>Chem 3<br>Calibration Verification<br>Test Set | | DOCUMENT<br>Total Protein / Albumin /<br>Cholesterol<br>CAL·VER | | DOCUMENT<br>Iron, Magnesium,<br>Triglyceride<br>CAL·VER | |
|---------|------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------|
| Analyte | Correlation<br>Coefficient<br>(r) | Regression<br>Equation<br>Y=intercept +<br>slope(X) | Correlation<br>Coefficient<br>(r) | Regression<br>Equation<br>Y=intercept +<br>slope(X) | Correlation<br>Coefficient<br>(r) | Regression<br>Equation<br>Y=intercept +<br>slope(X) |
| TP | 0.99984 | .109 + .969 | 0.99997 | .125 + .982 | N/A | N/A |
| ALB | 0.99920 | .022 + 1.025 | 0.99856 | .502 + .874 | N/A | N/A |
| CHOL | 0.99998 | 2.117 + .988 | 0.99997 | -3.359 + 1.016 | N/A | N/A |
| FE | 0.99950 | -7.942 + .996 | N/A | N/A | 0.99992 | -3.464 + 1.021 |
#### Summary:
ਨੇ
Linear regression analysis was carried out on recovered values for each analyte. Each analyte was tested in triplicate. The VALIDATE Chem 3 Calibration Verification Test Set has been shown to be functionally equivalent for calibration verification and linearity assessment to DOCUMENT Iron/Magnesium/Triglyceride CAL•VER, DOCUMENT Total Protein/Albumin/Cholesterol CAL•VER.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is facing to the right. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG 2 0 2001
Ms. Christine V. Beach Manager, RA/QA Maine Standards Company, LLC 765 Roosevelt Trail Windham, ME 04062
Re: 510(k) Number: K012119 Trade/Device Name: VALIDATE Chem 3 Calibration Verification Test Set Regulation Number: 862.1660 Regulatory Class: Class I Product Code: JJY Dated: July 3, 2001 Received: July 6, 2001
Dear Ms. Beach:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical-Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number: Kola119
Device Name: VALIDATE Chem 3 Calibration Verification Test Set
### Indications for Use:
The VALIDATE Chem 3 Calibration Verification Test Set is used by trained laboratory professionals for quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the following analytes: total protein, albumin, cholesterol, and iron.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) X OR Over-The-Counter Use
Kesia Alexander for Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number: K012119
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.