← Product Code [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX) · K984327

# ARCHITECT TOTAL T3 MASTERCHECK, MODEL NUMBER 6C51-05 (K984327)

_Bio-Rad · JJX · Dec 22, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K984327

## Device Facts

- **Applicant:** Bio-Rad
- **Product Code:** [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX.md)
- **Decision Date:** Dec 22, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Architect Total T3 MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range of the Total T3 assay on the Abbott Architect i System.

## Device Story

Architect Total T3 MasterCheck is a multi-level liquid control product containing human serum and Total T3 (except Level 0). It is used in clinical laboratory settings to verify the performance of the Total T3 assay on the Abbott Architect i System. The device serves as a calibration verifier, allowing laboratory technicians to assess the sensitivity, calibration linearity, and reportable range of the assay. By comparing the measured values of the MasterCheck levels against expected ranges, the healthcare provider ensures the accuracy and reliability of patient test results. This verification process helps maintain the quality of diagnostic testing, ultimately benefiting patients by ensuring that thyroid hormone measurements are precise and clinically actionable.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Liquid control material; matrix is human serum; contains sodium azide as a preservative. Levels 1-3 contain Total T3. Storage at 2-8°C. Standalone diagnostic control product.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Serum Multi-Analyte Verification Test Set ([K950469](/device/K950469.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

22

K98432-7

Image /page/0/Picture/2 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is white text on a black background. The text is in a bold, sans-serif font. The logo is simple and recognizable.

Laboratories

Diagnostics Group 500 Jeronimo Road Irvine, California 92618-20 Telephone: (949) 598-12

## 510(k) Summary

Submitter Bio-Rad Laboratories 9500 Jeronimo Road Irvine, CA (949)598-1285 Fax (949)598-1555

Contact Person Elizabeth Platt

Date of Summary Preparation December 02, 1998

Device (Trade & Common Name) Architect Total T3 MasterCheck

Classification Name Class I. 75JJX CFR 862.1660: Single (Specified) Analyte Controls (Assayed and Unassayed);

Devices to Which Substantial Equivalence is Claimed Document Serum Multi-Analyte Verification Test Set Casco Standards Yarmouth, ME K950469

## Statement of Intended Use

Architect Total T3 MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range of the Total T3 assay on the Abbott Architect i System.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in a black rounded rectangle. The text "BIO-RAD" is in white, bold, sans-serif font.

Bio-Rad Laboratories Diagnostics Group 9500 Jeronimo Road Irvine, California 92618-2017 Telephone: (949) 598-1200

Description of the Device

Architect Total T3 MasterCheck Level 0 contains human serum.

Architect Total T3 MasterCheck Levels 1, 2, and 3 contain Total T3 prepared in human serum.

Preservative: Sodium Azide.

## Statement of How Technological Characteristics Compare to Substantial Equivalence Device

A table is provided below comparing the similarities between the Bio-Rad Architect Total T3 MasterCheck and the devices to which substantial equivalence is claimed.

|                    | Architect Total T3 MasterCheck                                                                                                               | Casco Standards Document<br>Serum Multi-Analyte<br>Verification Test Set                                                                                                                                           |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use    | Verification of sensitivity,<br>calibration linearity, and<br>reportable range of the Total<br>T3 assay on the Abbott<br>Architect i System. | <i>In vitro</i> diagnostic use in the<br>quantitative determination of<br>linearity, calibration<br>verification and verification<br>of reportable range using<br>automated, semi-automated<br>and manual methods. |
| Form               | Liquid                                                                                                                                       | Liquid                                                                                                                                                                                                             |
| Matrix             | Human Serum                                                                                                                                  | Human Serum                                                                                                                                                                                                        |
| Storage            | 2-8°C                                                                                                                                        | -10 to -20°C                                                                                                                                                                                                       |
| Analytes           | Total T3                                                                                                                                     | Multiple                                                                                                                                                                                                           |
| Open Vial<br>Claim | 3 Days at 2-8°C.                                                                                                                             | 30 Days at 2-8°C.                                                                                                                                                                                                  |
| Differences        | Calibration verifier for the<br>Architect Total T3 assay                                                                                     | Calibration verifier for<br>multiple analytes.                                                                                                                                                                     |

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or other bird-like figure with its wings spread, positioned to the right of the text. The text, arranged in a circular fashion around the bird, reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

DEC 22 1990

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Elizabeth Platt Staff Regulatory Affairs Representative BIO-RAD LABORATORIES 9500 Jeronimo Road Irvine, CA 92618-2017

Re: K984327 Trade Name: Architect Total T3 MasterCheck Requlatory Class: I Product Code: JJX Dated: December 2, 1998 Received: December 3, 1998

Dear Ms. Platt:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

t

510(k) Number: _______________________________________________________________________________________________________________________________________________________________ Device Name: Architect Total T3 MasterCheck

Indications for Use:

Architect Total T3 MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range of the Total T3 assay on the Abbott Architect i System.

## (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Concurrence of CDRH, Office of Device Evaluation)

![Signature](signature.jpg)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K984327

| Prescription Use | <img alt="Check Mark" src="check_mark.jpg"/> | OR | Over-The Counter Use |  |
|------------------|----------------------------------------------|----|----------------------|--|
|------------------|----------------------------------------------|----|----------------------|--|

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K984327](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K984327)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
