← Product Code [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX) · K983578

# ARCHITECT TESTOSTERONE MASTERCHECK, MODEL 6C28-05 (K983578)

_Bio-Rad · JJX · Nov 12, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K983578

## Device Facts

- **Applicant:** Bio-Rad
- **Product Code:** [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX.md)
- **Decision Date:** Nov 12, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Architect Testosterone MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range on the Abbott Architect i System

## Device Story

Architect Testosterone MasterCheck is a liquid, bovine serum-based control product containing testosterone. It is designed for use in clinical laboratory settings by laboratory professionals. The device serves as a calibration verification tool for the Abbott Architect i System. By processing these samples, the system verifies the sensitivity, calibration linearity, and reportable range of the testosterone assay. The output allows the healthcare provider to confirm that the analyzer is performing within established specifications, ensuring the accuracy of patient testosterone test results.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Liquid bovine serum matrix containing testosterone; includes antimicrobial preservative. Designed for use with Abbott Architect i System. Storage at 2-8°C. Open vial stability: 3 days at 2-8°C.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Serum Multi-Analyte Verification Test Set Casco Standards ([K950469](/device/K950469.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 1 2 1998

Image /page/0/Picture/1 description: The image shows the logo for BIO-RAD. The logo is white text on a black rounded rectangle. The text is in all caps and bolded.

Bio-Rad Laboratories

Diagnostics Group 9500 Jeronimo Road Irvine, California 92618-2017 Telephone: (949) 598-1200

4983578

## 510(k) Summary

Submitter Bio-Rad Laboratories 9500 Jeronimo Road Irvine, CA (949)598-1285 Fax (949)598-1555

Contact Person Elizabeth Platt

Date of Summary Preparation November 2, 1998

Device (Trade & Common Name) Architect Testosterone MasterCheck

Classification Name Class I. 75JJX CFR 862.1660: Single (Specified) Analyte Controls (Assayed and Unassayed)

Devices to Which Substantial Equivalence is Claimed

Document Serum Multi-Analyte Verification Test Set Casco Standards Yarmouth, ME K950469

Statement of Intended Use Architect Testosterone MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range on the Abbott Architect i System

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the Bio-Rad logo. The logo is white text on a black background. The text is in a bold, sans-serif font.

Bio-Rad Laboratories

Diagnostics Group 9500 Jeronimo Road Irvine, California 92618-2017 Telephone: (949) 598-1200

## Description of the Device

Architect Testosterone MasterCheck Level 0 contains bovine serum Architect Testosterone MasterCheck Levels 1, 2 and 3 contain testosterone prepared in bovine serum.

Preservative: Antimicrobial agent.

## Statement of How Technological Characteristics Compare to Substantial Equivalence Device

A table is provided below comparing the similarities between the Bio-Rad Architect Testosterone MasterCheck and the devices to which substantial equivalence is claimed.

|                    | Architect Testosterone<br>MasterCheck                                                                               | Casco Standards Document<br>Serum Multi-Analyte<br>Verification Test Set                                                                                                                                    |
|--------------------|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use    | Verification of sensitivity,<br>calibration linearity, and<br>reportable range on the<br>Abbott Architect i System. | In vitro diagnostic use in the<br>quantitative determination of<br>linearity, calibration<br>verification and verification<br>of reportable range using<br>automated, semi-automated<br>and manual methods. |
| Form               | Liquid                                                                                                              | Liquid                                                                                                                                                                                                      |
| Matrix             | Bovine serum                                                                                                        | Human Serum                                                                                                                                                                                                 |
| Storage            | 2-8°C                                                                                                               | -10 to -20°C                                                                                                                                                                                                |
| Analytes           | Testosterone                                                                                                        | Multiple                                                                                                                                                                                                    |
| Open Vial<br>Claim | 3 Days at 2-8°C.                                                                                                    | 30 Days at 2-8°C.                                                                                                                                                                                           |
| Differences        | Calibration verifier for the<br>Architect testosterone assay                                                        | Calibration verifier for<br>multiple analytes.                                                                                                                                                              |

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three profiles facing right, representing the department's mission to protect the health of all Americans. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Elizabeth Platt Staff Regulatory Affairs Representative Bio-Rad Laboratories, Irvine Facility Diagnostics Group 9500 Jeronimo Road Irvine, California 92618-2017

Re : K983578 Architect Testosterone MasterCheck Trade Name: Product Code: Regulatory Class: F JJX צביר I Dated: October 9, 1998 Received: October 13, 1998

Dear Ms. Platt:

NOV 1 2 1998

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Routman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number: K983578
Device Name: Architect Testosterone MasterCheck

Indications for Use:

Architect Testosterone MasterCheck is intended for use in the verification of sensitivity, calibration linearity, and reportable range on the Abbott Architect i System

## (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Concurrence of CDRH, Office of Device Evaluation)

Prescription Use

OR Over-The Counter Use _

(Division Sign-Off)
Division of Clinical Laboratory vices
510(k) Number k98557
25

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K983578](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K983578)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
