← Product Code [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX) · K963181

# ELECSYS CALCHECK T UPDATE (K963181)

_Boehringer Mannheim Corp. · JJX · Sep 10, 1996 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K963181

## Device Facts

- **Applicant:** Boehringer Mannheim Corp.
- **Product Code:** [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX.md)
- **Decision Date:** Sep 10, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Elecsys CalCheck T Uptake is intended for use in the quantitative verification of the calibration curve established by the Elecsys T Uptake reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.

## Device Story

Elecsys CalCheck T Uptake; three-level single-analyte set; lyophilized human-based material. Used for calibration verification of Elecsys T Uptake reagents/calibrators on Elecsys 1010/2010 immunoassay analyzers. User reconstitutes material; assays in triplicate; compares results to target values. Verifies calibration curve accuracy on automated immunoassay systems; supports clinical decision-making by ensuring analytical performance of T Uptake testing.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Three-level lyophilized human-based material; single analyte set; designed for use on Elecsys 1010/2010 immunoassay analyzers.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Tosoh Medic AIA-Pack HCG Calibration Verification Test Set ([K924862](/device/K924862.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SEP 10 1996
K963181

# 510(k) Summary

## Elecsys® CalCheck™ T Uptake

|  Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.  |
| --- | --- |
|  1) Submitter name, address, contact | Boehringer Mannheim Corporation
9115 Hague Rd.
Indianapolis, IN 46250
(317) 845-2000

Contact Person: John D. Stevens

Date Prepared: August 12, 1996  |
|  2) Device name | Proprietary name: Elecsys CalCheck T Uptake

Common name: Calibration Verification Material

Classification name: Single (specified) analyte controls (assayed + unassayed)  |
|  3) Predicate device | We claim substantial equivalence to the Tosoh Medic AIA-Pack HCG Calibration Verification Test Set, K924862. The intended use for both products is to verify the calibration curve of automated immunoassay analyzers  |
|  4) Device Description | The Elecsys CalCheck T Uptake is a three level single analyte set of lyophilized, human based materials. After reconstitution they are assayed in triplicate and the results are compared to the target values.  |
|  5) Intended use | Elecsys CalCheck T Uptake is intended for use in the quantitative verification of the calibration curve established by the Elecsys T Uptake reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.  |

{1}

K963181

# 510(k) Summary, Elecsys® CalCheck™ T Uptake, Continued

## 6) Comparison to predicate device

The Boehringer Mannheim Elecsys CalCheck T Uptake is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Tosoh Medics AIA-Pack HCG Calibration Verification Test Set, K924862.

The intended use of both the Elecsys CalCheck T Uptake and the predicate device are the same in that they are intended to be used for the calibration verification for their labeled analytes on automated immunoassay analyzers.

12

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K963181](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K963181)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com