← Product Code [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX) · K100810

# ELECSYS  C-PEPTIDE CALCHECK 5 (K100810)

_Roche Diagnostics · JJX · Apr 26, 2010 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K100810

## Device Facts

- **Applicant:** Roche Diagnostics
- **Product Code:** [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX.md)
- **Decision Date:** Apr 26, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Elecsys C-Peptide CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the serum and plasma assay range established by the Elecsys C-Peptide reagent on the indicated Elecsys and cobas e immunoassay analyzers.

## Device Story

Elecsys C-Peptide CalCheck 5 is a set of five lyophilized levels of synthetic human C-peptide in HEPES buffered equine serum; used for calibration verification and assay range verification of Elecsys C-Peptide reagent. Intended for professional use on Modular Analytics E170/cobas e 601 and Elecsys 2010/cobas e 411 analyzers. Value assignment is performed by testing each level in duplicate on multiple analyzers against a master calibration curve derived from WHO reference reagent 84/510. The device provides laboratories with a tool to verify the accuracy of the assay range, ensuring reliable clinical decision-making for C-peptide measurements.

## Clinical Evidence

No clinical data. Bench testing only. Stability testing (real-time and accelerated) confirmed 15-month closed-vial stability at 2-8°C and 4-hour stability at 25°C after reconstitution. Value assignment validated using multiple analyzer platforms.

## Technological Characteristics

Lyophilized synthetic human C-peptide in HEPES buffered equine serum. Five-level concentration set (0, 5, 20, 30, 40 ng/mL). Designed for use with Elecsys/cobas e immunoassay analyzers. Traceable to WHO 1st IRR 84/510.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Elecsys C-Peptide CalCheck (k040157)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k100810

B. Purpose for Submission:
New device

C. Measurand:
Calibration verification and assay range verification for Elecsys C-Peptide Reagent.

D. Type of Test:
Not applicable.

E. Applicant:
Roche Diagnostics

F. Proprietary and Established Names:
Elecsys C-Peptide CalCheck 5

G. Regulatory Information:

1. Regulation section:
21 CFR 862.1660, Quality Control Material (Assayed and Unassayed)

2. Classification:
Class I, reserved

3. Product code:
JJX – Single (Specified) Analyte Controls (Assayed and Unassayed)

4. Panel:
Clinical Chemistry (75)

{1}

2

H. Intended Use:

1. Intended use(s):

See indications for use below.

2. Indication(s) for use:

The Elecsys C-Peptide CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the serum and plasma assay range established by the Elecsys C-Peptide reagent on the indicated Elecsys and cobas e immunoassay analyzers.

3. Special conditions for use statement(s):

For professional use only.

4. Special instrument requirements:

For use with Elecsys C-Peptide reagent on the Modular Analytics E170/cobas e 601 and Elecsys 2010/cobas e 411 analyzers.

I. Device Description:

The Elecsys C-Peptide CalCheck 5 set contains five lyophilized levels of synthetic human C-peptide in HEPES buffered equine serum. During manufacture, human C-peptide is spiked into the matrix at the target concentrations listed below. CalCheck Level 1 contains no analyte.

|  Elecsys C-Peptide CalCheck Level | Target Value (ng/mL)  |
| --- | --- |
|  Level 1 | 0  |
|  Level 2 | 5.0  |
|  Level 3 | 20  |
|  Level 4 | 30  |
|  Level 5 | 40  |

All human source materials were prepared exclusively from the blood of donors tested individually and found nonreactive for HBsAg, antibodies to HCV and HIV. The testing methods were FDA approved or cleared in compliance with the European Directive 98/79/EC.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Elecsys C-Peptide CalCheck

{2}

2. Predicate 510(k) number(s): k040157
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Indications for Use | Same | An assayed control for use in the verification of the calibration established by the Elecsys C-Peptide reagent  |
|  Analyte Reagent | Same | Elecsys C-Peptide  |
|  Matrix | Same | Equine serum  |
|  Format | Same | Lyophilized  |
|  Stability | Same | Unopened: Store at 2-8°C until expiration date. Reconstituted: Four hours at 20-25°C.  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Levels | Five | Three  |

K. Standard/Guidance Document Referenced (if applicable):

Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material.

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

{3}

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

## Traceability and Value Assignment

A master calibration curve is generated on an Elecsys Modular Analytics E170 /cobas e 601 analyzer using nine target values from the World Health Organization reference reagent, 1st IRR, code 84/510 (NIBSC) standard. The master calibration curve is used to calculate ng/mL for the respective CalChecks.

For each Elecsys C-peptide CalCheck 5 lot manufactured, value assignment is performed as follows. Each CalCheck is tested in duplicate using at least three Elecsys Modular Analytics E170 /cobas e 601 analyzers. Additionally, each CalCheck is assayed once on eight Elecsys 2010 / cobas e 411 analyzers. The target value for each CalCheck is the median of the observed values. The labeling states that laboratories should establish appropriate acceptance criteria when using this product for its intended use.

## Stability

Real time and accelerated stability testing protocols and acceptance criteria were described and found to be adequate. The current closed vial stability claim is 15 months at 2° to 8°C. The reconstituted vials are stable for four hours at 25°C.

d. Detection limit:

Not applicable

e. Analytical specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Not applicable

b. Matrix comparison:

{4}

Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The expected values are provided in the labeling for each specific lot.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K100810](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K100810)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
