ADVIA CENTAUR CYCLOSPORINE MASTER CURVE MATERIAL

K092993 · Siemens Healthcare Diagnostics · JJX · Nov 20, 2009 · Clinical Chemistry

Device Facts

Record IDK092993
Device NameADVIA CENTAUR CYCLOSPORINE MASTER CURVE MATERIAL
ApplicantSiemens Healthcare Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateNov 20, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

The ADVIA Centaur® Cyclosporine Master Curve Material is for in vitro diagnostic use in the verification of calibration and reportable range in the ADVIA Centaur® Cyclosporine (CsA) assay. For prescription use only.

Device Story

ADVIA Centaur® Cyclosporine Master Curve Material consists of 5-level human serum-based solutions containing cyclosporine at specific concentrations (0, 100, 500, 900, 1400 ng/mL). Used in clinical laboratories to verify calibration and reportable range of the ADVIA Centaur® Cyclosporine (CsA) assay. Liquid, ready-to-use format; stored at 2-8°C. Healthcare providers use the material to ensure assay accuracy; results confirm the instrument's performance within the established reportable range, supporting reliable therapeutic drug monitoring for patients on cyclosporine therapy.

Clinical Evidence

No clinical data. Bench testing only. Value assignment performed via comparison studies to a reference lot using the ADVIA Centaur Cyclosporine Assay. Within-run CV for 20 replicates was ≤ 8%. Assigned value acceptance criteria met ± 15% of reference lot. Shelf-life stability established at 2 years; open-vial stability established at 21 weeks at 2-8°C.

Technological Characteristics

Human serum-based matrix; liquid form. Contains cyclosporine, detergents, glycerol, anti-foam, and preservatives. Traceable to internal standard using USP grade cyclosporine. Storage 2-8°C. Standalone material for use with ADVIA Centaur systems.

Indications for Use

Indicated for in vitro diagnostic use to verify calibration and reportable range of the ADVIA Centaur® Cyclosporine (CsA) assay in clinical laboratory settings.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k092993 B. Purpose for Submission: New device C. Measurand: Calibration verification materials D. Type of Test: N/A E. Applicant: Siemens Healthcare Diagnostics, Inc. F. Proprietary and Established Names: ADVIA Centaur Cyclosporine Master Curve Material G. Regulatory Information: 1. Regulation section: 21 CFR 862.1660 Single, specified controls (assayed and unassayed) 2. Classification: Class I, reserved 3. Product code: JJX 4. Panel: Clinical Chemistry (75) {1} H. Intended Use: 1. Intended use(s): See indications for use below 2. Indication(s) for use: The ADVIA Centaur® Cyclosporine Master Curve Material is for *in vitro* diagnostic use in the verification of calibration and reportable range in the ADVIA Centaur® Cyclosporine (CsA) assay. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: ADVIA Centaur® Systems I. Device Description: The ADVIA Centaur® Cyclosporine Master Curve Material are 5 level human serum based solutions containing varying concentrations of cyclosporine in human serum, detergents, glycerol, anti-foam and preservatives. The Cyclosporine Master Curve Materials have expected values (lot specific) of approximately 0, 100, 500, 900 and 1400 ng/mL. The Cyclosporine Master Curve Material (1.0 mL/vial) is liquid and ready to use. Storage is at 2 - 8° C. The labeling contains the following information: “Contains human source material. While each human serum or plasma donor unit used in the manufacture of this product was tested by FDA-approved methods and found nonreactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C (HCV) and antibody to HIV-1/2, all products manufactured using human source material should be handled as potentially infectious. Because no test method can offer complete assurance that hepatitis B or C viruses, HIV, or other infectious agents are absent, these products should be handled according to established good laboratory practices.” J. Substantial Equivalence Information: 1. Predicate device name(s): VALIDATE Thyroid Calibration Verification Test Set {2} 2. Predicate 510(k) number(s): k062501 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | For in vitro diagnostic use in the verification of calibration and reportable range in the ADVIA Centaur Cyclosporine Assay | For in vitro diagnostic use in the quantitative determination of linearity, calibration verification, and verification of reportable range in automated, semi-automated and manual chemistry systems | | Form | liquid | liquid | | Storage | 2 -8° C | -10 to -20 ° C | | Matrix | human serum | human serum | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Analyte list | Cyclosporine only | Multi-analytes including triiodothyronine, thyroxine, TSH and cortisol | K. Standard/Guidance Document Referenced (if applicable): None referenced L. Test Principle: Not applicable M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: {3} Not applicable b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The Master Curve Material is traceable to an internal standard manufactured using USP grade cyclosporine. The ADVIA Centaur® Cyclosporine Master Curve Material is prepared utilizing various amounts of a cyclosporine positive stock solution and negative basepool. The assigned values are established by comparison studies to a reference lot of master curve materials using the ADVIA Centaur Cyclosporine Assay. Value assignment is performed using one ADVIA Centaur instrument with one lot of ADVIA Centaur Cyclosporine reagent assaying the material and reference material in multiple replicates and multiple runs. The within run CV of the 20 replicates met the sponsor's acceptance criteria of ≤ 8%. The acceptance criteria of the test lot assigned value to the reference lot met the acceptance criteria of ± 15%. The ranges for the assigned values were established by calculating ± 25% of the assigned value. The labeling includes a statement that if results do not fall within expected values, technical support should be contacted. The protocol for establishing shelf-life for the Master Curve Material was reviewed and is adequate. The shelf stability is 2 years when properly stored at 2-8 °C. Open stability was established to be up to 21 weeks when stored at 2-8 °C. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: {4} a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Lot specific target values are provided in the labeling. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...