LINICAL HSCRP CALIBRATION VERIFIERS LEVELS A-E FOR OLYMPUS AU SYSTEMS

K022491 · Cliniqa Corporation · JJX · Aug 27, 2002 · Clinical Chemistry

Device Facts

Record IDK022491
Device NameLINICAL HSCRP CALIBRATION VERIFIERS LEVELS A-E FOR OLYMPUS AU SYSTEMS
ApplicantCliniqa Corporation
Product CodeJJX · Clinical Chemistry
Decision DateAug 27, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

LiniCAL™ hsCRP Calibration Verifiers Levels A – E for Olympus AU Systems™ is intended for use as an assayed quality control material for analysis.

Device Story

LiniCAL™ hsCRP Calibration Verifiers are assayed quality control materials used to verify the calibration and linearity of high-sensitivity C-reactive protein (hsCRP) assays on Olympus AU Systems. The product consists of five levels (A-E) of liquid-stable material. Laboratory technicians use these verifiers to monitor assay performance, ensure accuracy of patient test results, and maintain quality control standards within the clinical laboratory environment. The verifiers provide a known reference point to confirm that the analytical system is operating within established specifications.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Liquid-stable assayed quality control material; five levels (A-E); designed for use with Olympus AU Systems for hsCRP analysis.

Indications for Use

Indicated for use as an assayed quality control material for analysis in clinical laboratory settings using Olympus AU Systems.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Submission Summary (Full Text)

{1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with outstretched arms, overlaid with three horizontal bars that resemble a stylized wave or heartbeat. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the figure. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## AUG 2 7 2002 Ms. Carol Ruggiero Director of Regulatory Affairs CLINIOA Corporation 1432 South Mission Road Fallbrook, CA 92028 Re: k022491 Trade/Device Name: LiniCAL™ hsCRP Calibration Verifiers Levels A – E for Olympus AU Systems™ Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJX Dated: July 25, 2002 Received: July 29, 2002 Dear Ms. Ruggiero: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 0224 510(k) Number (if known): Device Name: LiniCAL™ hsCRP Calibration Verifiers Levels A - E for Olympus AU Systems ™ Indications For Use: LiniCAL™ hsCRP Calibration Verifiers Levels A – E for Olympus AU Systems™ is intended for use as an assayed quality control material for analysis. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Toronica D. Celcie for Dan Cooper Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...