RANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3
K130236 · Radiometer Medical Aps · JJS · Mar 15, 2013 · Clinical Chemistry
Device Facts
Record ID
K130236
Device Name
RANGE + QUALICHECK LEVEL 1, RANGE + QUALICHECK LEVEL 2, RANGE + QUALICHECK LEVEL 3
Applicant
Radiometer Medical Aps
Product Code
JJS · Clinical Chemistry
Decision Date
Mar 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil
Device Story
Range+ QUALICHECK is an assayed quality control system consisting of three levels of aqueous solutions containing biological buffers, salts, glucose, lactate, dyes, and preservatives. It is used to verify the accuracy and precision of Radiometer ABL700, ABL800/8x7, ABL77, ABL80, and ABL90 series analyzers. The solution is equilibrated with carbon dioxide and oxygen. The healthcare provider uses the control solution by conditioning ampoules at 18-32°C, then running the sample on the analyzer. The analyzer measures the parameters and compares results against the provided control ranges. This process ensures the analyzer is functioning correctly, allowing the clinician to rely on patient test results for clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Stability and shelf-life were validated using three lots of each level, measured on ABL735 analyzers over 36 months. Value assignment was performed using 144 measurements across multiple validated ABL7xx series devices to establish mean ±2SD target ranges. Traceability to NIST and IUPAC standards was documented.
Technological Characteristics
Aqueous solution containing biological buffer, salts, glucose, lactate, dyes, and preservative. Equilibrated with CO2 and O2. Provided in 2mL ampoules. Three-level system. Traceable to NIST SRM and IUPAC pH scale. Storage: 2-8°C. In-use: 18-32°C.
Indications for Use
Indicated for use as an assayed quality control system for evaluating the accuracy and precision of blood gas, oximetry, electrolyte, metabolite, and hemoglobin parameters on specified Radiometer ABL analyzers.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
QUALICHECK5+ (k980135)
Related Devices
K130415 — HIGH METABOLITE QUALICHECK · Radiometer Medical Aps · Mar 26, 2013
K961355 — MULTICHECK · Radiometer America, Inc. · Jun 14, 1996
K961354 — QUALICHECK 4 METABOLITE(S3630,S3640,S3650,S3660) · Radiometer America, Inc. · Jun 14, 1996
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130236
B. Purpose for Submission:
New Device
C. Measurand:
Quality control material for blood gases
D. Type of Test:
Not applicable
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
Range+ QUALICHECK, Level 1
Range+ QUALICHECK, Level 2
Range+ QUALICHECK, Level 3
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660, Quality Control Material (Assayed and Unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJS, Controls for Blood-Gases
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
Refer to Indications for use.
2. Indication(s) for use:
For In Vitro Diagnostic Use. This Range+ QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges.
Analytes are: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not applicable
I. Device Description:
The Range+ QUALICHECK is an assayed quality control system, which can be used for quality control of the ABL700 Series, ABL800/8x7, ABL77, ABL80 and ABL90 analyzers from Radiometer. The Range+ QUALICHECK is a three level quality control system. Each level consists of 30 2mL ampoules of solution per box. The quality control solution is an aqueous solution containing a biological buffer, salts, glucose, lactate, dyes and a preservative, and it is equilibrated with carbon dioxide and oxygen. Range + QUALICHECK measures the following parameters: pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil.
J. Substantial Equivalence Information:
1. Predicate device name(s):
QUALICHECK5+
2. Predicate 510(k) number(s):
k980135
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Indications for Use | For In Vitro Diagnostic Use. It is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. | Same |
| Matrix | Aqueous solution | Same |
| Blood gas measurement | pH, pO2, pCO2 | Same |
| Oximetry Measurement | ctHb, sO2, FO2Hb, FCOHb, FMetHb | Same |
| Electrolyte Measurement | cK+, cNa+, cCa2+, cCl- | Same |
| Metabolite Measurement | cGlucose, cLactate | Same |
| Hemoglobin Measurement | ctBil | Same |
| Compatible analyzers | ·ABL700/705/710/715/720/725/730/735 ·ABL805/810/815/820/825/830/835 ·ABL80 - OSM SW ·ABL90 FLEX | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Storage | 2°C to 8°C until expiration date 5 hours between 18-32°C | 2°C to 25°C until expiration date, including up to a total of 15 days at up to 32°C |
| Oximetry Measurement | FHbf in Level 1 | FHbf in all four levels |
| Levels | Three levels | Four levels |
K. Standard/Guidance Document Referenced (if applicable):
None
{3}
# L. Test Principle:
Not applicable
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The materials are traceable to the following standard materials:
| Parameter | Unit | Traceable to |
| --- | --- | --- |
| pH | - | The IUPAC pH scale. Primary pH standards are certified by the Danish Primary Laboratory for Electrochemistry (DPLEC) at the Danish Institute of Fundamental Metrology (DFM) under DANAK accreditation no. 255. The system is validated by comparison with SRM produced by National Institute of Standards and Technology (NIST) |
| pCO2 | mmHg | NIST Standard Reference Material (SRM) gas 1674b and SRM 2625a |
| pO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cK+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 999b |
| cNa+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 919b |
| cCa2+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) 915 |
| cCl- | mmol/L (37°C) | NIST Standard Reference Material (SRM) 999b |
| cGlucose | mmol/L (37°C) | NIST Standard Reference Material (SRM) 917b |
| cLactate | mmol/L (37°C) | L-Lactic Acid Lithium Salt. SIGMA L2250 |
| cMalt | mmol/L (37°C) | L-Lactic Acid Lithium Salt. SIGMA L2250 |
| pH | - | NIST Standard Reference Material (SRM) gas 1674b and SRM 2625a |
| pCO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| pO2 | mmHg | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cK+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cNa+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cCa2+ | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cCl- | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
| cGlucose | mmol/L (37°C) | NIST Standard Reference Material (SRM) gas 2658a and NIST 2659a |
{4}
| ctHb | g/dL | NIST SRM (absorbance, wavelength). Hemoglobin-cyanide standard. J.T. Baker (Product No. 3061) |
| --- | --- | --- |
| sO₂ | % | NIST SRM (absorbance, wavelength). NIST SRM gas, whole blood, pH=7.4, ctHb=15 g%, sO₂=100% |
## Stability:
Shelf-life characteristics for the Range+ QUALICHECK control solutions were determined using three different lots of each level at 32°C for 10 days followed by 8°C for 36 months. The products were placed at 32°C for 10 days to assess the potential short term temperature deviations during transportation and handling. Shelf-life and in-use stability was tested by measurements on 2 ABL735 analyzers, 6 replicates at each time point (0, 12, 25, and 36 months). Protocols and acceptance criteria have been reviewed and found to be adequate.
The control solutions are stable for 2 years at 2-8°C.
In-use Stability: Ampoules should be conditioned for at least 5 hours at a constant temperature between 18°C-32°C before use. The contents should be used immediately after opening.
## Value Assignment
The sponsor provided assigned values and control ranges for each level. To determine the assigned values and control ranges for Range+ QUALICHECK, six (6) trays of 1000 ampoules are sampled randomly from the Range+ QUALICHECK batch. Twelve ampoules are sampled from each of the trays and 72 ampoules are sampled from the reference batch. The samples are conditioned and shaken at 25°C in a water bath for 6 hours. Measurements are performed on a minimum of 3 validated ABL7xx series devices with data collection. Measurement of each parameter is performed alternately on the reference ampoule and the sample ampoule and repeated 12 times on each ABL7xx for a total of 144 measurements. Target ranges are calculated based on the mean ±2SD.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
{5}
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Representative target values for the control solutions have been provided in the labeling.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.