RAICHEM HDL-CHOLESTEROL DIRECT REAGENT, HDL-CHOLESTEROL CALIBRATOR
Applicant
Raichem
Product Code
JIX · Clinical Chemistry
Decision Date
Sep 21, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Raichem HDL-Cholesterol Direct Reagent is intended for the quantitative determination of HDL-Cholesterol in serum or plasma. The Raichem HDL-Cholesterol Calibrator is intended to be used for the calibration of the Raichem HDL-Cholesterol Direct assay.
Device Story
Raichem HDL-Cholesterol Direct Reagent is an in vitro diagnostic assay for quantitative HDL-cholesterol measurement in human serum or plasma. The assay utilizes a two-step reaction sequence: first, chylomicrons and VLDL are selectively reacted with cholesterol esterase and eliminated; second, HDL-cholesterol reacts with cholesterol esterase to produce color. The resulting increase in absorbance is directly proportional to HDL-cholesterol concentration. The system includes a serum-based calibrator with values assigned via ultracentrifugation, traceable to NIST SRM 1951a. Used in clinical laboratory settings by trained personnel, the assay provides quantitative results that assist healthcare providers in assessing lipid profiles and cardiovascular risk. The device is intended for prescription use.
Clinical Evidence
Bench testing only. Precision studies (48 runs over 24 days) showed total CVs of 2.6-2.7%. Linearity established from 1 to 150 mg/dL. Method comparison against CDC designated comparison method (n=47) yielded a correlation coefficient of 0.987 and regression y = 0.9488x + 2.0. Comparison against previous Raichem reagent (n=48) yielded r=0.988 and y = 1.011x + 0.192. Lower detection limit is 1 mg/dL.
Technological Characteristics
In vitro diagnostic reagent system. Uses surfactants and MgCl2 in a buffered environment for selective HDL-cholesterol reaction. Includes serum-based calibrator traceable to NIST SRM 1951a. Designed for use on automated clinical chemistry analyzers (e.g., Hitachi 717).
Indications for Use
Indicated for the quantitative determination of HDL-Cholesterol in human serum or plasma for clinical diagnostic purposes.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Raichem HDL-Cholesterol Reagent (K830207A)
Boehringer Mannheim Direct HDL-Cholesterol (K963213)
K972041 — EZ HDL CHOLESTEROL REAGENT(PROCEDURE 354)/EZ HDL CALIBRATOR (354-5) · Sigma Diagnostics, Inc. · Jul 22, 1997
K153435 — Direct HDL Cholesterol (HDL) · Randox Laboratories, Ltd. · Jan 8, 2016
K971526 — ACE HDL-C REAGENT · Schiapparelli Biosystems, Inc. · May 21, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
## 1. Submitter's Name/Contact Person
Melissa Saner/Technical Marketing Manager
## Address
Raichem Division of Hemagen Diagnostics, Inc. # 2022395 Formerly: Reagents Applications, Inc. 8225 Mercury Court San Diego, CA 92129 Phone: 619-569-8009 extension 111 Fax: 619-569-6208 Email: msaner@raichem.com
#### Date Prepared
August 4, 1998
## 2. Device Name
| Trade Name: | Raichem HDL-Cholesterol Direct Reagent |
|------------------------|----------------------------------------|
| Common Name: | High density (HDL) cholesterol test |
| Classification Name: | Lipoprotein test system |
| Device Classification: | I |
| Regulation Number: | 21 CFR 862.1475 |
| Panel: | Chemistry (75) |
| Product Code: | LBS |
| Trade Name: | HDL-Cholesterol Calibrator |
| Common Name: | HDL-Cholesterol Calibrator |
| Classification Name: | Calibrator, Primary |
| Device Classification: | II |
| Regulation Number: | 21 CFR 862.1150 |
| Panel: | Chemistry (75) |
| Product Code: | JIS |
{1}------------------------------------------------
## 3. Predicate Device
CDC Designated Comparison Method: Cholesterol Reference Method Laboratory Network
Raichem HDL-Cholesterol Reagent: 510(k) Docket No. K830207A Boehringer Mannheim Direct HDL-Cholesterol: 510(k) Docket No. K963213 Boehringer Mannheim C.f.a.s. HDL-C (Calibrator); 510(k) Docket No. K963213 Raichem Serum Calibrator (Ortho Automated Reference Serum Assayed -Marketed by Chiron Diagnostics): No 510(k) Docket Number Assigned (Pre Medical Device Amendments of 1976) (Medical Device Listing Document No. A 268140)
-
## 4. Description of Device
The Raichem HDL-Cholesterol Direct test is based upon two sets of reaction sequences. In the first set of reactions Chylomicrons. VLDL selectively react with cholesterol esterase and are then eliminated from the reaction scheme. The addition of the second reagent causes the reaction of HDL-Cholesterol with cholesterol esterase and the subsequent production of color. The resultant increase in absorbance is directly proportional to the concentration of HDL-Cholesterol in the sample. The Raichem HDL-Cholesterol Calibrator is a serum calibrator used for the calibration of the Raichem HDL-Cholesterol Direct assay.
## 5. Intended Use of Device
The Raichem HDL-Cholesterol Direct Reagent is intended for the quantitative determination of HDL-Cholesterol in serum or plasma. The Raichem HDL-Cholesterol Calibrator is intended to be used for the calibration of the Raichem HDL-Cholesterol Direct assay.
## 6. Substantial Equivalence
## A. Technological Characteristics
#### Proposed Device
- The Raichem HDL-Cholesterol Reagent involves the direct selective reaction of HDL-1. Cholesterol utilizing surfactants and MgCl2 in a specific buffered environment.
- 2. The Raichem HDL-Cholesterol Calibrator is a serum calibrator used for the calibration of the Raichem HDL-Cholesterol Direct assay.
#### Predicate Device
- 1. The Designated Comparison Method is performed by a laboratory that is part of the Cholesterol Reference Method Laboratory Network. This method is standardized to the CDC reference method for HDL-Cholesterol.
- 2. The Raichem HDL-Cholesterol Reagent involves the precipitation of VLDL and LDL utilizing dextran sulfate (500,000 MW) and magnesium sulfate.
- ന് The Boehringer Mannheim Direct HDL-Cholesterol is a homogenous assay involving the selective reaction of HDL-Cholesterol utilizing PEG and sulfated cyclodextrin.
- 4. The Raichem Serum Calibrator (Ortho Automated Reference Serum) is a device purchased from Chiron Diagnostics by Raichem. This device is a serum calibrator used for the calibration of clinical chemistry assays.
{2}------------------------------------------------
Page 3 510(k) Summary HDL-Cholesterol Direct
## B. Performance Data
#### Precision
Precision studies were performed in 48 runs over a period of 24 days following the NCCLS EP5-T2 Tentative Guideline.
| | Total | | Within Run | | |
|--------------|-------|------|------------|------|--|
| Mean (mg/dL) | SD | CV % | SD | CV % | |
| 33.1 | 0.89 | 2.7 | 0.26 | 0.8 | |
| 51.21 | 1.35 | 2.6 | 0.32 | 0.6 | |
## Standardization of the Calibrator
The value for the calibrator is established by ultracentrifugation. It is traceable to the National Institute of Standards Standard Reference Material (1951a). The procedures used to assign the value to this calibrator material meet the requirements of the "HDL Cholesterol Method Evaluation Protocol for Manufacturers" of the U.S. National Reference System for Cholesterol, CRMLN (Cholesterol Reference Method Laboratory Network), August 1995.
## Lower Detection Limit
The lower detection limit for the proposed device was determined to be 1 mg/dL.
#### Linearity
1 to 150 mg/dL
#### Comparison Testing
- Comparison of Raichem HDL-Cholesterol Direct Reagent (y) with the CDC HDL-1. Cholesterol designated comparison method (x). Testing was performed in accordance with NCCLS EP9-A Approved Guideline. The Raichem HDL-Cholesterol Direct assay was performed using the Raichem Serum Calibrator (Ortho Automated Reference Serum) on the Hitachi 717.
| Number of sample pairs (n): | 47 |
|-----------------------------|---------------------|
| Range of results (mg/dL): | 24 – 70 |
| Correlation Coefficient: | 0.987 |
| Regression Equation: | $y = 0.9488x + 2.0$ |
- 2. Comparison of Raichem HDL-Cholesterol Direct Reagent (y) with Raichem HDL-Cholesterol Reagent (500,000 MW dextran sulfate) (x), following the NCCLS EP9-A Approved Guideline.
| Number of sample pairs (n): | 48 |
|-----------------------------|----------------------|
| Range of results (mg/dL): | 23.0 - 77.1 |
| Correlation Coefficient: | 0.988 |
| Regression Equation: | $y = 1.011x + 0.192$ |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized representation of an eagle or bird-like figure with three curved lines forming its body and wings. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
SEP 2 1 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Melissa Saner Technical Marketing Manager Raichem 8225 Mercury Court San Diego, California 92111-1203
K982767 Re : HDL-Cholesterol Direct Reagent and Calibrator Regulatory Class: I & II Product Code: JIX, LBS Dated: August 4, 1998 Received: August 6, 1998
Dear Ms. Saner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact
the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Steven Bitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
HDL-Cholesterol Calibrator Device Name:
## Indication(s) For Use
This calibrator is intended to be used for the calibration of the Raichem HDL-Cholesterol Direct assay.
HDL-Cholesterol Direct Reagent Device Name:
Indication(s) For Use
This reagent is intended to be used for the quantitative determination of HDL-Cholesterol in human serum or plasma.
(PLEASE DO NOT WRITE BELOW THIS LINE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter-Use_
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K982767
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.