← Product Code [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX) · K974825

# BOEHRINGER MANNHEIM C.F.A.S. - HDL/LDL-C-PLUS (K974825)

_Boehringer Mannheim Corp. · JIX · Jan 30, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K974825

## Device Facts

- **Applicant:** Boehringer Mannheim Corp.
- **Product Code:** [JIX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX.md)
- **Decision Date:** Jan 30, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Boehringer Mannheim c.f.a.s. - HDL/LDL-C plus is intended to be used in the calibration of test systems for the quantitative determination of high-density lipoprotein Cholesterol (HDL-C) and low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.

## Device Story

The Boehringer Mannheim c.f.a.s. - HDL/LDL-C plus is a calibrator consisting of lyophilized human serum supplemented with HDL-Cholesterol and LDL-Cholesterol. It is used in clinical laboratory settings to establish calibration curves for automated test systems measuring HDL-C and LDL-C levels in patient serum or plasma samples. By providing a known reference standard, the device enables the quantitative determination of these cholesterol fractions, assisting healthcare providers in assessing lipid profiles and managing cardiovascular health.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Lyophilized human serum matrix supplemented with HDL-Cholesterol and LDL-Cholesterol. Designed for use as a multi-analyte calibrator in automated clinical chemistry test systems.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Equal Diagnostics LDL Direct Liquid Select™ Cholesterol Calibrator
- Equal Diagnostics HDL Direct Liquid Select™ Cholesterol Calibrator

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN 30 1998

## K974825

## 510(k) Summary

| Introduction                                           | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence.                                                                                                                                         |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) Submitter<br>name, address,<br>contact              | Boehringer Mannheim Corporation<br>9115 Hague Rd<br>Indianapolis, IN 46250<br>(317) 845-2386                                                                                                                                                                                                                               |
|                                                        | Contact person: Edward R. Kimmelman                                                                                                                                                                                                                                                                                        |
|                                                        | Date prepared: Dec. 23, 1997                                                                                                                                                                                                                                                                                               |
|                                                        | 2) Device name                                                                                                                                                                                                                                                                                                             |
| Common name: c.f.a.s.- HDL/LDL-C                       |                                                                                                                                                                                                                                                                                                                            |
| Classification name: Calibrator, Multi-analyte mixture |                                                                                                                                                                                                                                                                                                                            |
| 3) Predicate<br>device                                 | We claim substantial equivalence to the Equal Diagnostics LDL Direct<br>Liquid Select™ Cholesterol Calibrator and the Equal Diagnostics HDL Direct<br>Liquid Select™ Cholesterol Calibrator. The intended use of the above<br>calibrators is the establishment of calibration curves for their respective test<br>systems. |
| 4) Device<br>description                               | The Boehringer Mannheim c.f.a.s. - HDL/LDL-C plus calibrator consists of<br>lyophilized human serum with added HDL-Cholesterol and LDL-Cholesterol.                                                                                                                                                                        |
| 5) Intended use                                        | The Boehringer Mannheim c.f.a.s. - HDL/LDL-C plus is intended to be used<br>in the calibration of test systems for the quantitative determination of high-<br>density lipoprotein Cholesterol (HDL-C) and low-density lipoprotein<br>Cholesterol (LDL-C) in serum and plasma.                                              |

Continued on next page

{1}------------------------------------------------

## 510(k) Summary, Continued

6) Comparison to the predicate device

The Boehringer Mannheim Direct LDL-Cholesterol test is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Equal Diagnostics HDL Direct Liquid Select™ Cholesterol Calibrator and the Equal Diagnostics LDL Direct Liquid Select™ Cholesterol Calibrator.

The intended use of this BM calibrator and the predicate devices is the same in that they are intended to be used for the calibration of test systems for the measurement of their labeled analytes.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird in flight.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 30 1998

Edward R. Kimmelman . Program Director, Requlatory Affairs and Compliance Boehringer Mannheim Corporation 9115 Haque Road Indianapolis, Indiana 46250

Re : K974825 Boehringer Mannheim Calibrator for Automated Systems -HDL/LDL-C-PLUS Regulatory Class: II Product Code: JIX Dated: December 16, 1997 Received: December 24, 1997

Dear Mr. Kimmelman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.com armann

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with ----the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note:

this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic -----Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to ---premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

K974825

Device Name: Boehringer Mannheim Calibrator for Automated Systems -HDL/LDL-C plus

Indications for Use: For the calibration of test systems for the measurement of HDL and LDL Cholesterol in human serum or plasma.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED) .............................................................................................................................................................................. Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional format 1-2-96)

(DiDivision Sign-Off)
DivDivision of Clinical Laboratory Devices
510510(k) Number. 12974825

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K974825](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JIX/K974825)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
