SYNCHRON SYSTEMS CAL 5 PLUS

K971283 · Beckman Instruments, Inc. · JIX · Jun 3, 1997 · Clinical Chemistry

Device Facts

Record IDK971283
Device NameSYNCHRON SYSTEMS CAL 5 PLUS
ApplicantBeckman Instruments, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateJun 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

The SYNCHRON® Systems CAL 5 Plus, when used in conjunction with SYNCHRON® Systems and reagents, is intended for use in the calibration of Antistreptolysin-O (ASO-) on the SYNCHRON CX® Systems and for Antistreptolysin-O (ASO-), C-reactive protein (CRP), and Rheumatoid factor (RF) chemistries on the SYNCHRON LX™ Systems.

Device Story

SYNCHRON Systems CAL 5 Plus is a frozen liquid human serum matrix calibrator. It is used in clinical laboratory settings to establish reference points for Antistreptolysin-O (ASO), C-reactive protein (CRP), and Rheumatoid factor (RF) assays. The device is intended for use with Beckman SYNCHRON CX and LX clinical chemistry analyzers. The calibrator is processed by the analyzer to assign values to patient samples, ensuring accurate measurement of analytes via spectrophotometry. Healthcare providers use the resulting calibrated measurements to aid in clinical decision-making regarding patient disease states. The device benefits patients by ensuring the accuracy and reliability of diagnostic test results.

Clinical Evidence

Bench testing only. Stability studies were conducted at 32°C, 37°C, and 41°C to confirm shelf-life. Value assignment and verification processes were performed to ensure accuracy on SYNCHRON LX and CX systems. No clinical data provided.

Technological Characteristics

Frozen liquid human serum matrix. Calibrator for clinical chemistry analyzers. Operates via spectrophotometry on SYNCHRON CX/LX systems. No electronic components or software algorithms; purely a chemical reference standard.

Indications for Use

Indicated for the calibration of Antistreptolysin-O (ASO-), C-reactive protein (CRP), and Rheumatoid factor (RF) assays on SYNCHRON CX and LX clinical chemistry systems. No specific patient population or contraindications defined.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{1} K971283 JUN - 3 1997 # Summary of Safety & Effectiveness ## SYNCHRON® Systems CAL 5 Plus ### 1.0 Submitted By: Sheri Hall Product Submissions Manager Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 993-8961 FAX: (714) 961-4457 ### 2.0 Date Submitted: 28 March 1997 ### 3.0 Device Name(s): #### 3.1 Proprietary Names SYNCHRON® Systems CAL 5 Plus #### 3.2 Classification Name (21 CFR § 862.1150) Calibrator ### 4.0 Predicate Device(s): | SYNCHRON Systems CAL 5 Plus | Predicate | Predicate Company | Docket Number | | --- | --- | --- | --- | | SYNCHRON Systems CAL 5 Plus | Beckman CAL 5 Plus | Beckman Instruments, Inc. | K926236/A | {2} Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON® Systems CAL 5 Plus Summary of Safety & Effectiveness ## 5.0 Description: This SYNCHRON Systems CAL 5 Plus is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON CX® and LX™ Clinical Systems. The SYNCHRON Systems CAL 5 Plus is for use in the calibration of SYNCHRON Systems Antistreptolysin-O (ASO-), C-reactive protein (CRP), and Rheumatoid factor (RF) chemistries. ## 6.0 Intended Use: The SYNCHRON® Systems CAL 5 Plus, when used in conjunction with SYNCHRON® Systems and reagents, is intended for use in the calibration of Antistreptolysin-O (ASO-) on the SYNCHRON CX® Systems and for Antistreptolysin-O (ASO-), C-reactive protein (CRP), and Rheumatoid factor (RF) chemistries on the SYNCHRON LX™ Systems. ## 7.0 Comparison to Predicate(s): The SYNCHRON Systems CAL 5 Plus is a frozen liquid human serum matrix identical to the Beckman CAL 5 Plus product. The only difference between the Beckman CAL 5 Plus and the SYNCHRON Systems CAL 5 Plus is the instrument platform on which they are used. The Beckman Immunochemistry Systems are fully automated, specific protein analyzers, that measure by nephelometry; while the SYNCHRON Systems are fully automated, clinical chemistry analyzers that measure by spectrophotometry. ## 8.0 Summary of Performance Data: The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence of the SYNCHRON Systems CAL 5 Plus to the Beckman CAL 5 Plus and is stable for ASO, CRP, and RF. The value assignment process for each analyte is correlated to a known standard via the anchor method. The SYNCHRON Systems CAL 5 Plus value assignment and verification processes yield acceptable calibrator assigned values for calibration on the SYNCHRON LX System and SYNCHRON CX Systems. SYNCHRON Systems CAL 5 Plus Stability Testing Summary | Stress Temperature | Duration of Incubation | Predicted Stability | Beckman Stability Claim* | | --- | --- | --- | --- | | 32°C | 40 Days | 32 Months | 24 Months | | 37°C | 24 Days | 33 Months | 24 Months | | 41°C | 15.5 Days | 32 Months | 24 Months | * Expiration dating placed on the package based on date of manufacture This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92. file: CAL5510K.DOC {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN - 3 1997 Ms. Sheri Hall Product Submissions Manager Beckman Instruments, Inc. 200 S. Kraemer Boulevard, W-337 Brea, California 92822-8000 Re: K971283 Trade Name: SYNCHRON® Systems CAL 5 Plus Regulatory Class: II Product Code: JIX Dated: April 1, 1997 Received: April 7, 1997 Dear Ms. Hall: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. {4} Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} page 1 of 1 510(k) Number (if known): Device Name: SYNCHRON® Systems CAL 5 Plus Indications for Use: The SYNCHRON® Systems CAL 5 Plus, when used in conjunction with SYNCHRON® Systems and reagents, is intended for use in the calibration of Antistreptolysin-O (ASO-) on the SYNCHRON CX® Systems and for Antistreptolysin-O (ASO-), C-reactive protein (CRP), and Rheumatoid factor (RF) chemistries on the SYNCHRON LX™ Systems. 21 CFR § 862.1150 Calibrator (a) Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (b) Classification. Class II. ![img-0.jpeg](img-0.jpeg) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (per 21 CFR 801.109) OR Over-the-Counter Use ☐ Optional Format 1-2-96
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...