Device Facts
| Record ID | K962406 |
|---|---|
| Device Name | N CRP STANDARD SY |
| Applicant | Behring Diagnostics, Inc. |
| Product Code | JIX · Clinical Chemistry |
| Decision Date | Jul 16, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
N CRP Standard SY is intended to be used for the establishment of reference curves.
Device Story
N CRP Standard SY is a lyophilized human serum-based calibrator containing stabilizers and preservatives. Used with Behring Nephelometer systems (K860894, K892223, K943997) to calibrate N Latex CRP mono tests. Operates by establishing reference curves for immunology assays. Used in clinical laboratory settings by trained personnel. Provides standardized reference values to ensure accurate quantification of C-reactive protein (CRP) levels in patient samples, aiding clinical diagnosis and monitoring of inflammatory conditions.
Clinical Evidence
Bench testing only. Precision studies on Behring Nephelometer yielded %CVs ranging from 1.43% to 4.47%. Stability testing confirmed shelf life of at least 12 months (lyophilized) and 4 weeks (reconstituted).
Technological Characteristics
Lyophilized human serum matrix with stabilizers and preservatives. Single-analyte calibrator. Designed for use with nephelometric immunoassay systems.
Indications for Use
Indicated for use as a calibrator for the establishment of reference curves in immunology assays performed on Behring Nephelometer systems.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- N Protein Standard SY
Reference Devices
Related Devices
- K991210 — K-ASSAY CRP CALIBRATOR SET · Kamiya Biomedical Co. · May 11, 1999
- K962523 — N LATEX CRP MONO REAGENT · Behring Diagnostics, Inc. · Jul 10, 1996
- K964062 — N PROTEIN STRANDARD SL · Behring Diagnostics, Inc. · Nov 21, 1996
- K021882 — CRPEX-HS CRP CALIBRATOR SET, CATALOG NO: 1510 · Good Biotech Corp. · Jul 11, 2002
- K060140 — CRP-ADVANCE MULTI CALIBRATOR SET, MODEL SE-250 · Diagnostic Chemicals , Ltd. · Apr 24, 2006