RAPIDPOINT 500 SYSTEM MEASUREMENT CARTRIDGE WITH CALIBRATORS
Device Facts
| Record ID | K122398 |
|---|---|
| Device Name | RAPIDPOINT 500 SYSTEM MEASUREMENT CARTRIDGE WITH CALIBRATORS |
| Applicant | Siemens Healthcare Diagnostics, Inc. |
| Product Code | JIX · Clinical Chemistry |
| Decision Date | Oct 12, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Indications for Use
The RAPIDPoint® 500 System measurement cartridge with calibrators is an in vitro diagnostic product for the calibration of the analytes on the RAPIDPoint® 500 System. The analytes measured are partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), pH, Sodium (Na *), Potassium (K'), ionized Calcium (Ca*), Chloride (Cl), Glucose (Glu), total hemoglobin (tHb), Neonatal Bilirubin (nBili) and Lactate (Lac).
Device Story
RAPIDPoint 500 System uses measurement and wash/waste cartridges containing aqueous salt solution calibrators; calibrates pH, pO2, pCO2, Na+, K+, Ca++, Cl-, Glucose, tHb, and Lactate. Neonatal Bilirubin calibration is derived from tHb calibration. System operates in point-of-care settings; used by clinicians to ensure accuracy of blood analysis. Calibrator values assigned via flame photometry and mass spectrometry. Device ensures analytical accuracy for clinical decision-making regarding patient metabolic and respiratory status.
Clinical Evidence
Bench testing only. Stability protocols, value assignment procedures, and traceability to NIST reference materials were validated. No clinical data presented.
Technological Characteristics
Aqueous salt solution calibrators; liquid format. Traceable to NIST SRM reference materials (e.g., SRM186, 918, 919, 915, 917) and standard methods (e.g., flame photometry, hexokinase, coulometric, lactate dehydrogenase). Storage at 2-8°C or 2-30°C depending on component. Standalone cartridge system.
Indications for Use
Indicated for in vitro diagnostic calibration of blood gas, electrolyte, metabolite, and hemoglobin analytes on the RAPIDPoint 500 System. No specific patient population or contraindications provided.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K061730 — OSMETECH OPTI R CRITICAL CARE ANALYZER, MODEL GD7044 · Osmetech, Inc. · Aug 4, 2006
- K110648 — NOVA STAT PROFILE PHOX ULTRA ANALYZER SYSTEM · Nova Biomedical Corporation · Aug 29, 2011
- K061974 — GEM PREMIER 4000 WITH IQM, MODEL 4000; GEM CVP 1 AND 2 WITH CO-OX; GEM CVP 3 AND 4 HEMATOCRIT · Instrumentation Laboratory CO · Sep 15, 2006