EASYQC - CHEMISTRY WITH ELECTROLYTES; CHEMISTRY; ELECTROLYTES KIT; CHEMISTRY CALIBRATOR, CREA-U CLAIBRATOR

K080404 · Medica Corp. · JIX · Mar 13, 2009 · Clinical Chemistry

Device Facts

Record IDK080404
Device NameEASYQC - CHEMISTRY WITH ELECTROLYTES; CHEMISTRY; ELECTROLYTES KIT; CHEMISTRY CALIBRATOR, CREA-U CLAIBRATOR
ApplicantMedica Corp.
Product CodeJIX · Clinical Chemistry
Decision DateMar 13, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

MEDICA’s EasyQC – Chemistry, Levels A and B, is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s Easy ElectroLytes kit is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s EasyQC - Chemistry and Electrolytes, Levels A and B is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s EasyCal - Multiconstituent Calibrator is intended for use in the MEDICA EasyRA Chemistry Analyzer to establish points of reference in the measurement of a number of analytes in human serum. MEDICA CREA-U Cal is intended for use in the MEDICA EasyRA Chemistry Analyzer to establish points of reference in the measurement of Creatinine in urine.

Device Story

The device consists of a suite of quality control materials (EasyQC Chemistry, EasyElectrolytes, EasyQC Chemistry & Electrolytes) and calibrators (EasyCal, CREA-U Cal) designed for the MEDICA EasyRA Chemistry Analyzer. These materials are used by laboratory professionals to verify the precision and accuracy of clinical chemistry assays. Quality control materials allow for the detection of systemic analytical deviations, while calibrators establish reference points for measuring analytes in human serum or urine. By ensuring the analyzer is properly calibrated and performing within expected precision limits, these products support the reliability of clinical diagnostic results, ultimately aiding healthcare providers in making accurate clinical decisions based on patient test data.

Clinical Evidence

No clinical data. Bench testing only. Value assignment for calibrators performed via master lot comparison using NIST or gravimetric standards. Control values determined via replicate assays across multiple laboratories using manufacturer-provided reagents and instruments.

Technological Characteristics

The device comprises liquid-based quality control materials and calibrators formulated for clinical chemistry analysis. These are designed for use with the MEDICA EasyRA Chemistry Analyzer. The products function as reference standards or control samples to monitor analytical performance. No specific materials of construction, energy sources, or software algorithms are described for these chemical reagents.

Indications for Use

Indicated for use as quality control materials to estimate test precision and detect systemic analytical deviations, or as calibrators to establish reference points for analyte measurement in human serum or urine, specifically for use with the MEDICA EasyRA Chemistry Analyzer.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k080404 B. Purpose for Submission: New device C. Measurand: Calibrator and quality control materials for general chemistry analytes (see device description below). D. Type of Test: N/A E. Applicant: Medica Corporation F. Proprietary and Established Names: EasyQC- Chemistry EasyQC- Electrolytes kit EasyQC- Chemistry with Electrolytes EasyCal- Chemistry Calibrator EasyCal- CREA-U Calibrator G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJY | I | 21 CFR 862.1660-Quality control material (assayed and unassayed) | Chemistry | {1} | JJR | I | 21 CFR 862.1660-Quality control material (assayed and unassayed) | Chemistry | | --- | --- | --- | --- | | JIX | II | 21 CFR 862.1150-Calibrator | Chemistry | | JIT | II | 21 CFR 862.1150-Calibrator | Chemistry | H. Intended Use: 1. Intended use(s): See indications for use statement below. 2. Indication(s) for use: MEDICA’s EasyQC – Chemistry, Levels A and B, is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s Easy ElectroLytes kit is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s EasyQC - Chemistry and Electrolytes, Levels A and B is a quality control material for clinical chemistry intended for use in the MEDICA EasyRA Chemistry Analyzer to (estimate) test precision and to detect any potential systemic analytical deviations. MEDICA’s EasyCal - Multiconstituent Calibrator is intended for use in the MEDICA EasyRA Chemistry Analyzer to establish points of reference in the measurement of a number of analytes in human serum. MEDICA CREA-U Cal is intended for use in the MEDICA EasyRA Chemistry Analyzer to establish points of reference in the measurement of Creatinine in urine. 3. Special conditions for use statement(s): For in vitro diagnostic use; for prescription use 4. Special instrument requirements: MEDICA EasyRA Chemistry Analyzer {2} I. Device Description: The EasyQC Chemistry kit is a dual level lyophilized powder that is prepared from human serum. The human source material, from which this product was manufactured, has been tested at the single donor unit basis and found non-reactive for HBsAg, anti-HIV ½, anti-HCV and HIV-1 Ag. The kit contains the following analytes: - Albumin - Alk. Phosphatase - ALT - Amylase - AST - Bili-D - Bili-T - Calcium - CO2 - Cholesterol - Creatine Kinase - Creatinine - Iron - GGT - Glucose - HDL Cholesterol - LDH - Magnesium - Phosphorus - Total Protein - Triglycerides - Urea-N (BUN) - Uric Acid The EasyQC Chemistry and Electrolytes kit is a dual level lyophilized powder that is prepared from human serum. The human source material from which this product was manufactured has been tested at the single donor unit basis and found non-reactive for HBsAg, anti-HIV ½, anti-HCV and HIV-1 Ag. The kit contains the same analytes from the EasyQC Chemistry kit shown above, in addition to the following electrolytes: - Sodium - Potassium - Chloride - Lithium. The EasyQC Electrolyte is a tri-level constituted solution containing salts, buffers, bovine albumin and preservatives. It does not contain human serum or human serum products. The analytes in this kit are sodium, potassium, chloride and lithium. The EasyCal Multi-constituent Calibrator is provided in 3 mL vials of lyophilized, human serum powder. The human source material from which this product was manufactured has been tested at the single donor unit basis and found non-reactive for HBsAg, anti-HIV ½, anti-HCV and HIV-1 Ag. The kit contains the following analytes: - Albumin - Alk. Phosphatase - Bili-D - Bili-T - Calcium - CO2 - Cholesterol - Creatinine 3 {3} -Iron -Glucose -Magnesium -Phosphorus -Total Protein -Triglycerides -Urea-N (BUN) -Uric Acid The CREA-U Cal is a single 10 mL vial containing liquid creatinine prepared in 0.1 N Hydrochloric acid (for stability). ## J. Substantial Equivalence Information: 1. Predicate device name(s): DRY-Chemistry Lyophilized Control- k060322 Moni-trol Lyophilized chemistry controls- k960103 EasyLyte QC kit- k990971 Calibrate CH DRY-Chemistry calibrator- k060322 Creatine Urine Standard- k011126 2. Predicate K number(s): k060322 k960103 k990971 k060322 k011126 3. Comparison with predicate: The candidate devices contains the same analytes, has the same stability and traceability as the predicate devices. The only two differences is that the k990971 (EasyLyte QC kit) contains two additional analytes that are not contained in this submission (iCa and pH). 4 {4} K. Standard/Guidance Document Referenced (if applicable): None were reference. L. Test Principle: Not applicable. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable. b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Assignment of values for the analytes within the two calibrators, EasyCal Multiconstituent Calibrator and CREA-U Cal, were obtained from the master calibrator lot using 4 runs performed 2-4 EasyRA analyzers. Each new lot of calibrator material is tested in duplicate against the master calibrator lot in replicates of ten. Included in each run are duplicates of NIST or gravimetric standards (where NIST standards are not available) and two levels of QC material. The assigned values are based on the recoveries (and verification) of the master calibrator lot and the lot being assigned from the multiple runs on multiple device lots over several days. Assignment of values for EasyQC Chemistry, EasyQC Electrolytes and, EasyQC Chemistry and Electrolytes control kits are determined from the mean of replicate assays from at least two laboratories using the reagents, calibrators, and controls available from the instrument manufacturer. The master calibrators are traceable to NIST and other primary standards and are listed in the insert. Stability for each device is shown in the table below: | Stability | | | | --- | --- | --- | | | Use-life store at 2-8C | Shelf-life at 2-8 C | | EasyQC- Electrolytes | 8 weeks | 24 months | {5} | kit | | | | --- | --- | --- | | EasyQC- Chemistry | 7 days (1 day for Co2 and 3 days for Total and Direct bilirubin) | 36 months | | Easy QC-Chemistry and electrolytes | 5 days (1 day for CO2 and 3 days for Total and Direct bilirubin) | 36 months | | EasyCal- Chemistry Calibrator | 5 days (1 day for CO2 and 3 days for total and direct bilirubin) | 36 months | | EasyCal- Creatinine Calibrator | 18 months | 18 months | d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Not applicable. b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. {6} b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Not applicable. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...