DIMENSION VISTA RF FLEX REAGENT CARTRIDGE; PROTEIN 2 CALIBRATOR AND CONTROL

K062035 · Dade Behring, Inc. · JIX · Oct 4, 2006 · Clinical Chemistry

Device Facts

Record IDK062035
Device NameDIMENSION VISTA RF FLEX REAGENT CARTRIDGE; PROTEIN 2 CALIBRATOR AND CONTROL
ApplicantDade Behring, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateOct 4, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

Flex® reagent cartridge: The RF method is an in vitro diagnostic reagent for the quantitative determination of rheumatoid factors (RF) in human serum and lithium heparin plasma on the Dimension Vista™ System. Measurements are used as an aid in the diagnosis of rheumatoid arthritis. Protein 2 Calibrator: PROT2 CAL is an in vitro diagnostic product for the calibration of the C-reactive protein (CRP), high sensitivity CRP (hsCRP) and Rheumatoid Factors (RF) methods on the Dimension Vista™ System. Protein 2 Controls L and H: PROT2 CON L and H are assayed intra-laboratory quality controls for the assessment of precision and analytical bias in the determination of C-reactive protein (CRP) and Rheumatoid Factor (RF) on the Dimension Vista™ System.

Device Story

The Dimension Vista™ RF Flex® reagent cartridge is an in vitro diagnostic reagent used on the Dimension Vista™ System. It utilizes polystyrene particles coated with an immunocomplex (human immunoglobulin and anti-human IgG from sheep). When mixed with patient serum or lithium heparin plasma samples containing RF, the particles aggregate. The system passes a light beam through the sample; the intensity of scattered light is measured, which is proportional to the RF concentration. Results are determined by comparison with a known standard. The system is intended for use in clinical laboratories by trained personnel to aid in the diagnosis of rheumatoid arthritis. The associated Protein 2 Calibrator and Controls (L and H) are used to calibrate the system and assess analytical precision and bias.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (repeatability <6%, within-lab ≤6.2%). Linearity confirmed across 10-600 IU/mL range (slope 1.04, r=1.0). Interference testing per CLSI EP7-A2 showed minimal interference (<10%) from common substances. Method comparison with predicate (n=120) yielded slope 0.953, r=0.95. Matrix comparison (serum vs. lithium heparin plasma) showed no significant bias (mean recovery -2.5%). No clinical sensitivity/specificity studies provided.

Technological Characteristics

Immunoturbidimetric assay using polystyrene particles coated with human immunoglobulin and anti-human IgG from sheep. Analyte concentration is proportional to scattered light intensity. System is an automated clinical chemistry analyzer. Reagents are liquid-stable.

Indications for Use

Indicated for quantitative determination of rheumatoid factors (RF) in human serum and lithium heparin plasma to aid in the diagnosis of rheumatoid arthritis. Used on the Dimension Vista™ System.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k062035 B. Purpose for Submission: New Device C. Measurand: Rheumatoid Factor D. Type of Test: Quantitative, nephelometry E. Applicant: Dade Behring Inc. F. Proprietary and Established Names: Dimension Vista™ Rheumatoid Factors Flex ® reagent cartridge Dimension Vista™ Protein 2 Calibrator Dimension Vista™ Protein 2 Control L Dimension Vista™ Protein 2 Control H G. Regulatory Information: 1. Regulation section: 21 CFR § 866.5775 –Rheumatoid Factor Immunological test system 21 CFR § 862.1150 –Calibrator 21 CFR § 862.1660 –Quality Control Material 2. Classification: Class II 3. Product code: DHR - System, Test, Rheumatoid Factor JIX - Calibrator, multi-analyte mixture JJY - Multi-analyte controls, all kinds (assayed and unassayed) 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): Flex® reagent cartridge: The RF method is an in vitro diagnostic reagent for the quantitative determination of rheumatoid factors (RF) in human serum and lithium heparin plasma on the Dimension Vista™ System. Measurements are used as an aid in the diagnosis of rheumatoid arthritis. Protein 2 Calibrator: PROT2 CAL is an in vitro diagnostic product for the calibration of the C-reactive protein (CRP), high sensitivity CRP (hsCRP) and Rheumatoid Factors (RF) methods on the Dimension Vista™ System. Protein 2 Controls L and H: PROT2 CON L and H are assayed intra-laboratory quality controls for the assessment of precision and analytical bias in the determination of C-reactive protein (CRP) and Rheumatoid Factor (RF) on the Dimension Vista™ System. 2. Indication(s) for use: Same as Intended Use. {1} 3. Special conditions for use statement(s): Prescription use only 4. Special instrument requirements: Dimension Vista™ System I. Device Description: Dimension Vista™ Rheumatoid Factors Flex® regent cartridge (RF): RF reagent consists of polystyrene particles coated with an immunocomplex consisting of human-IgG/anti-human IgG from sheep and RF supplement reagent consisting of a phosphate buffer and polyethylene glycol. Dimension Vista™ Protein 2 Calibrator: PROT2 CAL is a liquid, human serum based product containing C-reactive protein and Rheumatoid Factors. Dimension Vista™ Protein 2 Control L and H: PROT2 CON L and H are liquid, multi-analyte, human serum based products containing C-reactive protein and Rheumatoid Factors in two concentration ranges: low and high. J. Substantial Equivalence Information: 1. Predicate device name(s) and 510(k) numbers: ROCHE Tina-Quant RF II assay k032535 N Rheumatology Standard SL k964527 N/T Rheumatology Control SL k962373 2. Comparison with predicate: Dimension Vista™ Rheumatoid Factors Flex® regent cartridge | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | For the quantitative determination of rheumatoid factors (RF) in human serum and lithium heparin plasma | For the quantitative in vitro determination of rheumatoid factors in human serum and plasma | | Indications for Use | Measurements are used as an aid in the diagnosis of rheumatoid arthritis. | Same | | Measurement type | Quantitative | Same | | Stability: Unopened | 90 days 2-8°C | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Method | Immunonephelometry | Immunoturbidimetry | | Capture Antibody | Immune complexes of human immunoglobulin and sheep anti-human IgG antibody | Human IgG | | Instrument system | Dimension Vista™ System | ROCHE/Hitachi 917 analyzer | {2} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Plasma | lithium heparin plasma | lithium heparin plasma and EDTA plasma | | Measuring Range | 10-600 IU/mL | 7-650 IU/mL | | Expected values | <15 IU/mL | <14 IU/mL | | Stability: Open | 21 days 2-8°C | 90 days 2-8°C | Dimension Vista™ Protein 2 Calibrator | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | For the calibration of the specified methods | For establishment of reference curves for specific immunonephelometric determinations | | Composition | Pooled Human sera | Same | | Reagent Preparation | Liquid, Ready-for-Use | Same | | Traceability | 1stBritish Standard 64/002 | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Analytes | C-reactive protein (CRP), high sensitivity CRP (hsCRP) and Rheumatoid Factors (RF) | Rheumatoid factors (RF), anti-Streptolysin O (ASL) and C-reactive protein (CRP) | | Instrument system | Dimension Vista™ System | BN Systems | | Stability: Open | 12 days 2-8°C | 14 days 2-8°C | | Materials Provided | 6 vials, 2 mL per vial | 3 vials, 1 mL per vial | Dimension Vista™ Protein 2 Control L and H | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Assayed intra-laboratory quality controls for the assessment of precision and analytical bias in the determination of specific assays | Assayed controls for accuracy and precision in the quantitative determination of specific analytes | | Composition | Pooled human sera | Same | | Reagent Preparation | Liquid, Ready to use | Same | | Concentration Range | Low and High | Same | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Analytes | C-reactive protein (CRP) and Rheumatoid Factor (RF) | Rheumatoid factors (RF), anti-streptolysin O (ASL) and C-reactive protein (CRP) | | Instrument system | Dimension Vista™ System | BN Systems and the TurbiTime System | | Stability Open | 12 days 2-8°C | 14 days 2-8°C | | Materials Provided | 6 vials, 2 mL per vial | 3 vials, 1 mL per vial | # K. Standard/Guidance Document Referenced (if applicable): CLSI Evaluation of Precision Performance of Clinical Devices Approved Guideline: EP5-A2; CLSI Method Comparison and Bias Estimation Using Patient Samples Approved Guideline EP9-A2; CLSI Interference Testing in Clinical Chemistry Approved Guideline EP7-A2. # L. Test Principle: Polystyrene particles coated with an immunocomplex consisting of human-IgG/anti-human IgG from sheep are aggregated when mixed with samples containing RF. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Precision testing was performed in accordance with CLSI EP5-A2. Four samples derived from serum pools representing RF concentrations that span the measuring range, and the PROT2 CON L and H were analyzed in duplicate, for 20 days, two times per day. The repeatability and within-lab standard deviations (SD) and percent coefficient of variation (%CV) were calculated. Repeatability imprecision was $&lt;6\%$ , and within-lab imprecision was $\leq6.2\%$ . | Material | Mean IU/mL | Repeatability | | Within-Lab | | | --- | --- | --- | --- | --- | --- | | | | SD | %CV | SD | %CV | | PROT2 CON L | 73.7 | 1.5 | 2.0 | 1.6 | 2.2 | | PROT2 CON H | 183.9 | 3.5 | 1.9 | 3.9 | 2.1 | | Serum pool* VL | 14.5 | 0.9 | 5.9 | 0.9 | 6.2 | | Serum pool VL | 30.5 | 0.8 | 2.6 | 1.8 | 5.9 | | Serum pool L | 90.0 | 2.1 | 2.3 | 2.3 | 2.6 | | Serum pool H | 544.3 | 21.7 | 4.0 | 23.9 | 4.4 | *Tested for 5 days # b. Linearity/assay reportable range: Linearity across the assay range of 10-600 IU/mL was confirmed by testing a {4} calibrator with high concentration of RF. The calibrator was serially diluted in 7, two-fold increments (measuring range from 8.36 to 836.86 IU/mL). Each dilution was tested in replicates of three. Data were analyzed using linear regression analysis with the following results: | Slope | Y-Intercept IU/mL | Correlation Coefficient | n | | --- | --- | --- | --- | | 1.040 | -2.348 | 1.0 | 21 | A recovery study using 1st British Standard 64/002 (5 replicates) showed that % recovery ranged from 96.4 to 98.6% with a mean recovery of 97.9%. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibration material values were referenced to the 1st British Standard 64/002. d. Detection Limit Limit of Detection: 10 IU/mL represents the lower limit of the reportable range of the RF assay. Analytical sensitivity was determined by assaying analyte-free System Diluent 20 times and determining the mean value plus two SDs: 0.814 IU/mL. e. Analytical specificity: i. Interference Studies: Interference testing was performed according to CLSI EP7-A2, to determine the effect of various exogenous substances on the Dimension Vista™ RF assay. The following exogenous interferents were tested by spiking into test samples containing two concentrations of RF (one at the clinical decision point 12-15 IU/mL and one higher), and compared to a control sample without interferent: bilirubin (conjugated and unconjugated, 60 mg/dL), hemoglobin (1000 mg/dL), creatinine (30 mg/dL), albumin (6 g/dL), urea (500 mg/dL), cholesterol (500 mg/dL), uric acid (20 mg/dL). Minimal interference (&lt;8%) was observed. Interference by triglycerides (native and clarified) was tested on 5 test samples containing RF concentrations between 12 and 25 IU/mL, and minimal interference (&lt;7%) was observed. Additionally, 41 potentially interfering drugs were also assayed and shown to exhibit minimal interference (&lt;10%). ii. Cross-reactivity with autoimmune antibodies common to other systemic autoimmune diseases was not tested. iii. Antigen Excess: the effect of antigen excess was evaluated using a serum sample above the assay range. No effect was seen up to 2877 IU/mL. Samples above this level are reported as exceeding assay range (&gt;600 IU/mL). f. Assay cut-off: The assay cut-off of 15 IU/mL was determined by measuring 300 normal blood donors from Europe and taking the mean between the 95th and 97.5th percentile value. 2. Comparison studies: a. Method comparison with predicate device: The Dimension Vista™ RF assay was compared to the ROCHE Tina-Quant {5} RF II assay on the Hitachi 917 Analyzer by assaying 120 serum samples with concentrations ranging from 10-520 IU/mL on both systems. Passing Bablock regression analysis of these results yielded the following statistics: | Comparative Method | Slope (95%CI) | Y-Intercept (IU/mL) (95%CI) | Correlation Coefficient | n | | --- | --- | --- | --- | --- | | ROCHE Tina-Quant | 0.953 | -9.71 | 0.95 | 120 | | RF II/Hitachi 917 | (0.898 to 1.024) | (-16.192 to -6.021) | | | b. Matrix comparison: Fourteen matched serum and lithium heparin plasma samples, containing concentrations of RF spanning the measuring range (10-515 IU/mL), were assayed and compared. Regression analysis showed no significant bias between the two matrices and yielded the following information: | % recovery in Li-Hep plasma vs. serum | Slope (95%CI) | Y-Intercept (IU/mL) | Correlation Coefficient | n | | --- | --- | --- | --- | --- | | Mean = -2.5% (Range = -9.7% to 2.3%) | 0.98 (0.94-1.01) | -0.04 | 1.00 | 14 | 3. Clinical studies: a. Clinical Sensitivity: Not provided b. Clinical specificity: Not provided c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: See Assay cut-off 5. Expected values/Reference range: Testing of serum from 300 European blood donors resulted in a 95th percentile of &lt;10 IU/ml and a 97.5th percentile of 19.5 IU/ml. The expected value of the normal healthy adults is &lt;15 IU/ml. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...